28.03.2022 Views

ABB AV2123000 Positioner User Manual

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

CALIBRATION

TROUBLESHOOTING CALIBRATION ADJUSTMENTS

The I/P converter of the Type AV2 positioner comes from the factory

completely calibrated. Use the following procedures only

when operating difficulties occur. The corrective action column

of Table 6-1 indicates when to use the troubleshooting calibration.

The purpose of this procedure is to adjust the I/P converter so that

a four to 20-milliamp input signal will produce a 20.7 to 103.4-kilopascal

(3.0 to 15.0-pounds per square inch gage) signal. Use the

following procedure to adjust the I/P converter. The supply air

must be connected to the positioner for this procedure.

Tools Required: • Digital voltmeter.

• Calibrated 0 to 30-psig instrument gage.

• Small screwdriver.

• Current source.

CAUTION

Do not exceed the maximum supply pressure of 1034 kilopascals

(150 pounds per square inch gage). Exceeding this

pressure could cause equipment damage.

1. Connect the calibrated zero to 30-psig instrument gage or

other accurate pressure measuring device to the I port to monitor

the signal pressure during calibration.

2. Turn on the supply air to the positioner.

3. Provide a four-mA signal to the I/P converter and observe the

output pressure. If the pressure equals 20.7 kPa (3.0 psig) go to

Step 5. If the pressure does not equal 20.7 kPa (3.0 psig) go to

Step 4.

4. Adjust the zero screw (→0←) on the bottom of the I/P converter

(Fig. 4-7) until the output pressure equals 20.7 kPa (3.0

psig). Clockwise rotation increases pressure.

5. Apply a 20-mA signal to the I/P converter and observe the output

pressure. If the pressure equals 103.4 kPa (15.0 psig) go to

Step 7. If the pressure does not equal 103.4 kPa (15.0 psig) go to

Step 6.

6. Adjust the span screw (⏐↔⏐ ) on the bottom of the I/P converter

(Fig. 4-7) until the output pressure equals 103.4 kPa (15.0

psig). Clockwise rotation increases pressure.

7. Repeat Steps 3 through 6 until the 20.7 kPa (3.0 psig) and

103.4 kPa (15.0 psig) outputs are attained with four and 20-mA

signal inputs.

TROUBLESHOOTING CALIBRATION ADJUSTMENTS

4 - 8

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!