28.03.2022 Views

ABB AV2123000 Positioner User Manual

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

SECTION 4 - CALIBRATION

INTRODUCTION

This section contains the following information:

• Calibration.

• Gain adjustment.

• Speed control.

• Troubleshooting calibration.

NOTES:

1. All procedures in this section assume direct acting operation.

2. The troubleshooting calibration procedure in this section is

not required for normal operation. The procedure is used only

when operating difficulties occur. The corrective action column of

Table 6-1 indicates when to use the troubleshooting calibration.

3. Calibration can also involve cam shaping to achieve the

required control characteristics. Refer to Appendix C for detailed

information about the cam.

CALIBRATION

Tools required:

3⁄

32

-inch Allen wrench.

Mechanical linkage is performed during installation (Section 3)

that normally consists of adjusting the connection between the

positioner and actuator so that the mechanical zero on the positioning

cam corresponds to the mechanical zero of the actuator.

The calibration procedure detailed here involves fine adjustments

to the zero and span such that the input signal will cause the

desired travel.

Calibration consists of applying an input signal pressure to the

positioner to set the zero and span. These adjustments are made

with the supply air applied to the positioner.

Type AV positioners have position transmitter options that are

available for feedback purposes. Refer to Appendix A for more

information about the position transmitter options.

Zero Adjustment

The positioner zero adjustment sets the tension on the range

spring and adjusts the input signal level that causes the actuator

to move off of the minimum travel stop. The zero percent radial

line of the cam should align with the center of the cam roller (Fig.

4-2).

INTRODUCTION

4 - 1

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!