63(&,$/ 0$,17(1$1&(<br />
&ROOHW IDLOV WR RSHQ<br />
With reference to Fig. 3:<br />
If collet (B) fails to open after lever (A) has been turned through 90°, or if<br />
the tool is insufficiently clamped when the lever is in the "locked" position,<br />
adjust as follows:<br />
- turn lever (A) through 90° to bring the collet to its maximum extension<br />
position;<br />
- immobilize handpiece spindle (C) with a cross tool;<br />
- insert a 1.5mm Allen key in the collet (B), locked to the shaft base (C) by<br />
an Allen screw;<br />
- loosen this screw with the Allen Key by a half turn in a counterclockwise<br />
direction. This will release the collet (B);<br />
- to adjust clamping force, turn the collet with your fingers: clockwise to<br />
increase or counterclockwise to reduce;<br />
- use the Allen key to tighten the screw which locks the collet;<br />
- remove the cross tool to release the spindle.<br />
5HSODFLQJ WKH KDQGSLHFH<br />
The handpiece can be removed, exclusively when the motor is stopped,<br />
by turning it counterclockwise until it is detached from the motor.<br />
ÃÃÃÃ ÃÃÃ