You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
Reference handbook
DL50/DL53/DL55/DL58
Titrators
1. Introduction
2. Setup
3. Methods
4. Sample data memory
5. Analysis
6. Auxiliary functions
7. DL58
8. Symbols and explanations
9. Error messages and malfunctions
10. Installation and maintenance
11. Accessories
12. Technical data
13. Index
14. Data sheet
15. ISO certificate, System validation
Introduction
1. Introduction
In contrast to the Quick Guide, this Reference Handbook provides a complete description of
the operating principles of the DL50, DL53, DL55 and DL58 Titrators. Its layout has been kept
modular to ensure future supplementation or replacement of individual sections or pages; new
texts carry the date of issue in the footer on every page.
Safety notes
Measures for your personal protection and for operational safety are described in Section 2
of the Quick Guide. We strongly advise you to read these. The symbols listed there also appear
for texts in this Reference Handbook which require safety measures.
Keywords used
appears before all texts which refer to an error in the measured value acquisition,
data storage
NOTICE
etc.
Note
appears before all texts which contain additional explanations.
Index
The index in Section 13 comprises key words from the Quick Guide and the Reference
Handbook.
Software version
The Reference Handbook applies to software version 2.x for all four titrators.
Note:
All information concerning automation in this titrator's Reference Handbook relates
to the earlier METTLER TOLEDO Sample changer ST20A. Its name continues to
appear in some places. Operation of other METTLER TOLEDO sample changers
via your titrator is described in the corresonding Operating Instructions.
The following pages provide an overview of the menus and commands which are available
under the menu and auxiliary function keys.
FCC rules
This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to both
Part 15 of the FCC Rules and the radio interference regulations of the Canadian Department of Communications.
These limits are designed to provide reasonable protection against harmful interference when the equipment is
operated in a commercial environment. This equipment generates, uses and can radiate radio frequency energy
and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio
communications. Operation of this equipment in a residential area is like to cause harmful interference in which
case the user will be required to correct the interference at his own expense.
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 1-1
Menus: Overview
Setup
Method
Resources
Titrants
Sensors
Temperature sensors
Modify
SETUP
∇
Methods
Method ID
Standard methods
User methods
METTLER methods
METHOD
Print OK
Titrants
SETUP
Standard methods
METHOD
NaOH 0.1 mol/L
HCl 0.1 mol/L
HClO 4 0.1 mol/L
Esc Delete Add Modify OK
∇
Equivalence point titr'n
End point titration (EP)
Titer by EQP titration
Esc Print Modify
∇
Titrant parameters
SETUP
Method:
METHOD
Name
NaOH
Concentration [mol/L] 0.1
Titer 1
Esc Modify OK
∇
Title
Sample
Stir
Esc Modify OK
∇
Titrant names
SETUP
HCl
∆
HClO 4
AgNO 3
EDTA
∇
Esc OK
Title
METHOD
Method ID 00001
Title Equivalence point titr'n
Date/time 00-00-0000 00:00
Esc OK
Titrant parameters
SETUP
Method: 00001
METHOD
Name HClO 4
Concentration [mol/L] 0.1
Titer 1
Esc Modify OK
∇
Title
Sample
Stir
Esc Modify OK
∇
Methods
Method ID 00001
Standard methods
User methods
Delete Print Modify
METHOD
∇
1-2 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Menus: Overview
Sample
Run
No. Status Wt/vol.
Meth. ID
Samples to be analyzed
RUN
Add
Number of samples 1
Method ID 00001
User
Start
Add
SAMPLE
Defined are
RUN
New sample series
Esc OK
Stirrer 1: Stand 1
Sensor 1: DM141
Drive 2: 0.1 mol/L AgNO 3
Esc Print OK
Sample entry
SAMPLE
Number of samples 2
Method ID 00001
User
Esc OK
Sample No. 1
Sample ID
Weight [g] 1.23452
Limits [g] 0.02 - 2.0
Correction factor f 1.0
Balance 1) OK
RUN
∇
Sample No. 1
SAMPLE
Sample ID
Weight [g] 1.19831
Limits [g] 0.02 – 2.0
Correction factor f
∇
Esc
Balance 1) OK
Current sample
No. 1 of 1
Sample ID
Method ID 00001
OK
RUN
Sample No. 2
SAMPLE
Sample ID
Weight [g] 1.36712
Limits [g] 0.02 – 2.0
Correction factor f
∇
Esc
Balance 1) OK
Stir function
Wait time [s] 30
Speed [%] 80
RUN
Hold 2)
No. Status Wt/Vol. Meth. ID
1 ready 1.19831 00001
2 ready 1.36712 00001
Delete Print Modify Add
mV
E – V curve
Table Values Curve 2) Hold 2)
mL
1) appears only if a balance has been defined
2) appears only with DL55 and DL58
Result list
Method: 00001
Sample 1
R1 = 29.26 %
OK
RUN
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 1-3
Auxiliary functions: Overview
Sensor
Stirrer
Sensor
SENSOR
Stirrer
STIRRER
Measure potential
Measure temperature
Calibrate temperature sensors
OK
Titration stand Stand 1
Speed [%] 50
Modify Start
Measure potential
SENSOR
Stirrer
STIRRER
Sensor
DG111
Unit of meas. mV
Titration stand Stand 1
∇
Esc Modify Start
Titration stand Stand 1
Speed [%] 50
Stop
Measured values
SENSOR
43.7 mV
Stop
Changer
Burette
Sample changer
CHANGER
Burette
BURETTE
Change lift position
Rotate turntable
Dispense/rinse
OK
Rinse burette
Rinse tip
Dispense
OK
∇
Change lift position
CHANGER
Rinse burette
BURETTE
Lift position
bottom
Burette drive Drive 2
Esc Modify Start
Esc Modify 1) Start
Change lift position
CHANGER
Rinse burette
BURETTE
Lift position
bottom
Burette drive Drive 2
Stop
1) appears only with DL55 and DL58
Stop
1-4 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Auxiliary functions: Overview
Results
*
Report
Results 2)
RESULTS
Report 3)
REPORT
Display result list
Perform calculations
Display measured values
OK
∇
Output Printer
Results
All results
No
Yes
∇
Modify Start
Result list
Method: 00001
Sample 1
R1 = 29.26 %
Sample 2
RESULTS
OK
∇
Report
Output unit writing report
REPORT
2) The parameters appear only when results of a sample
determination are available
3) The parameters appear only when curves, results or
sample data of a sample determination are available
Misc. ...
Data
Transfer
Miscellaneous
MISC.
Data transfer
DATA TRANSFER
Define titrator settings
Adjust measuring inputs
Shorten analysis sequence
OK
∇
Memory card
Computer 4)
Remote control 4)
OK
∇
Titrator settings
MISC.
Memory card 5)
DATA TRANSFER
Date/time format
Date/time
Language
Esc Modify OK
∇
Display directory
Copy from titrator to card
Copy from card to titrator
Esc OK
∇
Formats
MISC.
Directory
DATA TRANSFER
Date format 17-OCT-1995
Time format 17:04 (24 h)
Esc Modify OK
User methods
Resources
Memory copies
Esc Print OK
4) appears only if a computer has been defined
5) appears only if memory card is formatted
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 1-5
Setup
Setup
Contents
Page
2. Setup .......................................................................................................... 2-3
2.1 Titrants ....................................................................................................... 2-4
2.1.1 Delete ....................................................................................................... 2-4
2.1.2 Modify....................................................................................................... 2-4
2.1.3 Add ........................................................................................................... 2-6
2.2 Sensors ...................................................................................................... 2-7
2.2.1 Delete ....................................................................................................... 2-7
2.2.2 Modify....................................................................................................... 2-7
2.2.3 Add ........................................................................................................... 2-10
2.3 Temperature sensors ................................................................................ 2-11
2.4 Polarized sensors ..................................................................................... 2-13
2.4.1 Modify....................................................................................................... 2-13
2.4.2 Add ........................................................................................................... 2-14
2.5 Auxiliary values ......................................................................................... 2-15
2.6 Titration stands ......................................................................................... 2-17
2.7 Peripherals................................................................................................. 2-18
2.7.1 Printer....................................................................................................... 2-18
2.7.1.1 Centronics interface: Settings for an EPSON printer ............................... 2-19
2.7.1.2 RS232 interface: Settings for an EPSON printer ..................................... 2-20
2.7.2 Balance .................................................................................................... 2-21
2.7.3 System ..................................................................................................... 2-23
2.7.4 Sample changer ....................................................................................... 2-25
2.7.5 External keyboard .................................................................................... 2-26
2.8 Solvents (with sample changer only) ...................................................... 2-27
2.8.1 Delete ....................................................................................................... 2-27
2.8.2 Modify....................................................................................................... 2-27
2.8.3 Add ........................................................................................................... 2-28
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 2-1
Setup
2-2 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Setup
Setup
2. Setup
In order to perform titrations, the titrator must be acquainted with the titrants and their
concentrations and the sensors with a feasible unit of measurement. It must know the burette
drive on which the burette is located and the input to which the sensor is connected. It must
also be familiar with the settings of the attached devices such as a balance or printer in order
to transfer data. In this menu you define all these chemical and mechanical resources and store
them: you set up your titrator.
When you press the Setup key, the list of resources appears:
Titrants
Sensors
Temperature sensors
Polarized sensors
Auxiliary values
Titration stands
Peripherals
Solvents (only with ST20A)
The titrants and METTLER TOLEDO sensors in most frequent use are already stored in the
titrator. You can not only delete these resources and modify their parameters, but also define
new ones.
Note: All information concerning automation in this titrator's Reference Handbook relates to
the earlier METTLER TOLEDO Sample changer ST20A. Its name continues to appear
in some places. Operation of other METTLER TOLEDO sample changers via your
titrator is described in the corresonding Operating Instructions.
NOTICE
All resources needed for the METTLER methods are stored! If you delete one
of these, the titrator will wait until the start of a titration of the method before
displaying the error message that the resource is not installed.
Depending on your choice of resource, its list with the corresponding parameters will be printed
out.
Modify
The list of the selected resources appears.
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 2-3
Titrants
Setup
2.1 Titrants
When you select this menu and press <F4>, the stored titrants appear with their concentration
data:
Titrants
SETUP
NaOH 0.1 mol/L
HCl 0.1 mol/L
HCl0 4 0.1 mol/L ∇
Esc Delete Add Modify OK
2.1.1 Delete
If you press <F2>, "Delete", the titrant NaOH will be deleted from the list.
2.1.2 Modify
If you press <F4>, "Modify", the parameters of the titrants appear and you can change their
names and values.
Name
NaOH
Concentration [mol/L] 0.1
Titer 1.0
Burette volume
10 mL
Burette drive Drive 2
Titer
do not check
Date/time 00-00-0000 00:00
Name
Concentration
Titer
You can change the name by pressing <F4> and selecting a different
name from the list of "Titrant names" which appears:
– Scroll the menu with the ∇ key so that the selector bar is positioned on,
e.g. "HClO 4 " and confirm with OK: the name of the titrant is now HClO 4 .
Enter the concentration of the titrant in mol/L.
Change the titer only if you know its value. If you determine the titer of the
titrant with the titrator, its value will be entered here automatically together
with the date (see Titer function, Section 3.3.14).
2-4 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Setup
Titrants
Burette volume
Select the volume of the burettes you wish to use with <F4>: 1, 5, 10 or
20 mL.
Burette drive Press <F4> to select: Drive 1.
DL50/DL53: These titrators do not have the "Modify" command.
Titer
You can choose whether to enter a date when you should check the titer
of the titrant:
– Press <F4>, select "check on" in the mask which appears and then
press <F4> again:
Check on
SETUP
Day 1
Month 1
Year 1997
Esc OK
If you select "check after" and then press <F4>, the following mask
appears:
Check after
SETUP
No. of days 10
Esc OK
When the date has lapsed, you will receive an appropriate message
during an ongoing titration method which uses this titer.
Date/time
You can neither enter nor delete the date. It refers to the titer determination
of the titrant using the method function Titer and is thus entered here
automatically together with the time (see Titer function, Section 3.3.14)
As soon as you modify a parameter of the titrant, the date/time entry will
be deleted. The expiry date of the titer is an exception: If you modify this,
date and time will not be deleted.
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 2-5
Titrants
Setup
Storage procedure
The titrant with its modified parameters is stored when you confirm both the parameter mask
and the list of titrants with OK. If you quit the list with Esc, the following message appears:
Save changes?
SETUP
No
Yes
OK
Any change to a parameter is always stored without a message if you press a Menu or Auxiliary
function key.
2.1.3 Add
With this command you add a titrant to the titrants list, e.g. NaOH of concentration 1.0 mol/L
or with a different burette volume or a titrant not yet on the list. The following parameters always
appear:
Name
NaOH
Concentration [mol/L] 0.1
Titer 1.0
Burette volume
10 mL
Burette drive Drive 2
Titer
do not check
Date/time 00-00-0000 00:00
You can modify the parameters by the procedure described in Section 2.1.2. If you define, e.g.
several NaOH solutions with the same concentration, you must also flag the names so that the
titrator can distinguish these, e.g. NaOH/1. If you do not, the following message appears:
Resource exists
SETUP
Modify name
Overwrite parameters
OK
Modify: The parameter mask for modification of the name appears.
Overwrite: The modified parameters are adopted.
2-6 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Setup
Sensors
2.2 Sensors
When you select this menu and press <F4>, the stored METTLER TOLEDO sensors with their
measurement unit appear:
Sensors
SETUP
DG111 pH
DG101 pH
DG113 mV ∇
Esc Delete Add Modify OK
You can attach the listed sensors only to a pH option (see Sections 2.4 and 10.4).
2.2.1 Delete
If you press <F2>, "Delete", the sensor DG111 will be deleted from the list.
2.2.2 Modify
If you press <F4>, "Modify", the parameters of the sensors whose names and values you can
change appear.
Name
DG111
Unit of meas.
pH
Sensor input Sensor 1
Zero point [unit] 7.0
Slope [mV/unit] –59.16
Temperature [°C] 25.0
Date/time 00–00–0000 00:00
Name
Unit of measurement
You can change the name by pressing <F4> and selecting a different
name from the list of "Sensor names" that appears.
Note: You do not need to define a reference electrode as it is part of the
sensing electrode at the corresponding sensor option (see Section
10.4).
You select possible measurement units for the sensor from the selection
menu.
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 2-7
Sensors
Setup
Unit of measurement
mV: Either there is no other suitable measurement unit with calibration
parameters for the sensor or you need only mV. If you select
the unit mV for a sensor, the calibration parameters zero point,
slope and temperature are ignored by the titrator.
pH:
pM:
pX:
%T:
If you select pH, next enter the calibration parameters for a pH
electrode or have them entered by the titrator (see page 2-9:
Notes 1 and 2).
M represents any cation. If you select pM, next enter the calibration
parameters for a pM electrode or have them entered by
the titrator (see page 2-9: Notes 1 and 2).
X represents any anion. If you select pX, next enter the calibration
parameters for a pX electrode or have them entered by the
titrator (see page 2-9; Notes 1 and 2).
If you select %T, next enter the calibration parameters for a
phototrode (measurement unit: transmission, see "Phototrode"
Operating Instructions.
A: If you select A, next enter the calibration parameters for a
phototrode (measurement unit: transmission). The extinction
[A = -log T] is calculated from this (see Section 8.4.1: Results).
µS/cm:
If you select µS/cm, next enter the calibration parameters of the
conductivity cell.
mS/cm: If you select mS/cm, next enter the calibration parameters for
the conductivity cell.
You can perform conductivity measurements and conductivity
titrations with a conductometer fitted with an analog output.
Sensor input
Press <F4> to select: Sensor 1, 2, 3 or 4. The numbers refer to the slot
of a pH option (see Section 10.4).
2-8 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Setup
Sensors
Zero point
Slope
Temperature
Date/time
The zero point is the measured value at which the sensor indicates 0 mV
voltage (measuring chain zero point).
• For the zero point of a pH electrode this is pH 0 with the unit pH.
• For the zero point of an ion selective electrode this is pM 0 with the unit
pM, or pX 0 with the unit pX.
• The zero point of the phototrode is normally 0%T (100% T =
1000 mV).
• The zero point of a conductivity cell is normally 0 µS/cm or 0 mS/cm.
The slope of the sensor is the voltage change in mV per measurement
unit.
• For the slope of a pH electrode this is the unit: mV/pH.
• For the slope of an ion selective electrode this is the unit: mV/pM or
mV/pX.
• The slope of a phototrode is normally 10 mV/%T.
• For the slope of a conductivity cell this is the unit: mV/µS * cm -1 or
mV/mS * cm -1 .
Change the temperature if necessary: The current calibration temperature
is either
• entered by you at the start of a calibration method (see Section 5.1)
or
• measured automatically if you have attached and defined a temperature
sensor (see Section 3.3.2).
You can neither enter the date nor delete it. It refers to the calibration of
the sensor using the method function Calibration and is entered here
automatically together with the time (see Calibration function, Section
3.3.13).
Notes
1. When the titrator is delivered, the defined calibration data are theoretical values of a new
sensor. You must perform a calibration of your sensor if you wish to determine accurate
values (see Calibration function, Section 3.3.13).
2. With a calibration of pH, pM or pX sensors, the calibration data (zero point, slope,
temperature) are entered here automatically together with the date.
3. As soon as you change a sensor parameter, the date/time entry is deleted.
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 2-9
Sensors
Setup
NOTICE
You should never transfer the calibration data of glass electrodes which you
obtain at sensor input 1 if you attach the electrode to, e.g. sensor input 2. To
obtain correct values, recalibrate the electrode!
Storage procedure
The sensor with the modified parameters is stored when you confirm both the parameter list
and the list of the sensors with OK. The modification of a parameter is always stored when you
press a Menu or Auxiliary function key (see corresponding section at the end of Section 2.1.2).
2.2.3 Add
With this command you add a sensor to the sensor list, e.g. a DG111 sensor with different
calibration parameters or a completely new sensor. The following parameters always appear:
Name
DG
Unit of meas.
mV
Sensor input Sensor 1
Zero point [unit] 7.0
Slope [mV/unit] –59.16
Temperature [°C] 25.0
Date/time 00–00–0000 00:00
You change the parameters by the procedure described in Section 2.2.2. If you define, e.g.
several DG111 sensors, you must also flag the name to ensure the titrator can distinguish
these, e.g. DG111/2. If you do not, a message appears (see Section 2.1.3).
2-10 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Setup
Temperature sensors
2.3 Temperature sensors
When you select this menu and press <F4>, the stored temperature sensors appear with the
sensor type:
Temperature sensors
SETUP
TEMP A Pt100
TEMP B Pt100
TEMP C Pt1000 ∇
Esc Modify OK
You can attach temperature sensors only to a pH option (see Section 10.4).
You can neither delete a temperature sensor from the list nor add a new one to it. However,
you can define, e.g. several Pt100s or Pt1000s at the same sensor input.
Select, e.g. TEMP A, press <F4> and the following parameters appear:
Name
TEMP A
Sensor type
Pt100
Sensor input Temp 1
Zero point [°C] 0.0
Date/time 00–00–0000 00:00
Name
You can not change the sensor name.
Sensor type Press <F4> to select: Pt100 or Pt1000 (see Section 9.5).
Sensor input
Zero point
Date/time
Press <F4> to select: Temp 1, 2, 3 or 4: The numbers refer to the slot of
a pH option (see Section 10.4).
The zero point of a Pt sensor is the difference between the measured
value and a reference value, e.g. 0 °C of an ice bath.
You can neither enter nor delete the date. It refers to the calibration of the
Pt sensor and is entered here automatically together with the time (see
Section 6.1.3).
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 2-11
Temperature sensors
Setup
Notes
1. The defined zero point is the theoretical value of a new Pt sensor. You must perform a
calibration of your sensor if you wish to determine the exact value. With a calibration the
zero point is entered here automatically with the date (see Section 6.1.3: Calibrate
temperature sensors).
2. As soon as you modify a parameter of the temperature sensor, the date/time entry is
deleted.
NOTICE
You should not transfer the zero point of temperature sensors which you obtain
at input Temp 1 if you attach the sensors to, e.g. input Temp 2! To obtain correct
values, recalibrate the sensor!
Storage procedure
The Pt sensor with the modified parameters is stored when you confirm both the parameter
mask and the list of the Pt sensors with OK. The modification of a parameter is always stored
when you press a Menu or Auxiliary function key (see corresponding paragraph at the end of
Section 2.1.2).
2-12 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Setup
Polarized sensors
2.4 Polarized sensors
When you select this menu and press <F4>, "DM142" appears as the only polarized METTLER
TOLEDO electrode stored in the titrator:
Polarized sensors
SETUP
DM142
Esc Add Modify OK
You can attach this sensor only to a KF option (see Section 10.4).
Note: If you have added polarized sensors to the list, the Delete command appears (<F2>).
2.4.1 Modify
If you press <F4>, "Modify", the name and sensor input of the DM142 appear:
Sensor parameters
SETUP
Name
DM142
Sensor input Sensor 2
Esc Modify OK
Name
Sensor input
You can change the name by pressing <F4> and selecting a different
name from the list of "Sensor names" that appears.
Press <F4> to select: Sensor 1, 2, 3 or 4. The numbers refer to the slot
of a KF option (see Section 10.4).
Storage procedure
The sensor with the modified parameters is stored when you confirm both the parameter list
and the list of the sensors with OK. The modification of a parameter is always stored when you
press a Menu or Auxiliary function key (see corresponding paragraph at the end of Section
2.1.2).
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 2-13
Polarized sensors
Setup
2.4.2 Add
With this command you add a polarized sensor to the sensor list. The following parameters
always appear:
Sensor parameters
SETUP
Name
DM
Sensor input Sensor 2
Esc Modify OK
You change the parameters by the procedure described in Section 2.4.1. If you define, e.g.
several DM142 sensors, you must also flag the name to ensure the titrator can distinguish
these, e.g. DM142/2. If you do not, a message appears (see Section 2.1.3).
2-14 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Setup
Auxiliary values
2.5 Auxiliary values
When you select this menu and press <F4>, the 20 memory locations for auxiliary values
appear: H1 - H20. You can use the Auxiliary value function to assign the results of a method,
such as blank values, potentials or volume or amount of substance consumption to this
memory. These values are then entered here automatically together with the date (see Section
3.3.15).
You yourself can also enter numeric values as an auxiliary value here and call them up under
the Calculation function.
Auxiliary values
SETUP
H1 1.0
H2 1.0
H3 1.0 ∇
Esc Modify OK
No auxiliary values are defined in the titrator on delivery.
You can neither delete the memory locations nor add new ones. If you select, e.g. H1 and press
<F4>, the following appears:
Auxiliary value parameters
SETUP
Auxiliary value H1
ID
Value 1.0
Date/time 00-00-0000 00:00
Esc OK
Auxiliary value You can not change H1.
ID
Value
Date/time
Enter an identification if desired.
Enter the value.
You can neither enter nor delete the date. It refers to the determination
of the auxiliary value using the method function Auxiliary value and is
entered here automatically together with the time (see Section 3.3.15).
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 2-15
Auxiliary values
Setup
Notes
1. You can delete the data of an auxiliary value only by overwriting them yourself or have the
titrator overwrite them using the Auxiliary value function.
2. The titrator does not tell you if you overwrite an auxiliary value. If you make frequent use
of the auxiliary value memory, you should thus print out a list of its values from time to time
(see page 2-3).
3. If you modify a parameter of the auxiliary value, the date/time entry is deleted.
Storage procedure
The auxiliary value with the modified parameters is stored when you confirm both the
parameter mask and the list of auxiliary values with OK. The modification of a parameter is
always stored when you press a Menu or Auxiliary function key (see appropriate paragraph
at the end of Section 2.1.2).
2-16 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Setup
Titration stands
2.6 Titration stands
When you select this menu and press <F4>, the names of the five possible titration stands
appear:
Titration stands
SETUP
Stand 1
Stand 2
ST20A
Esc Modify OK
∇
Stand 1
Stand 2
ST20A
Auto stand
External stand
is the titration stand you receive with the basic version of the titrator.
is the second titration stand of a dual titration stand.
is the sample changer titration stand.
is the name of the titration stand on which you can run a sample series
without being prompted to insert the next sample (see Section 5.1). With
this stand you can use, e.g. a robot to change the samples.
is a titration stand you have set up in addition to the above-mentioned
titration stands if you, e.g. attach a stirrer to a rod stand.
You can neither delete a titration stand from the the list nor add a new one. If you select, e.g.
Stand 1 and press <F4>, the following appears:
Titration stand parameters
Name Stand 1
Stirrer output Stirrer 1
SETUP
Esc Modify OK
Name
Stirrer output
You can not change the name of the titration stand.
Press <F4> to select the output for the stirrer: Stirrer 1, 2, 3, 4 or "Not at
DL".
1, 2, 3, 4 refer to the position of the sensor option (see Section 10.4).
Select "Not at DL" if you use a stirrer which is not attached to the titrator:
The wait time specified under the Stir function is allowed to elapse within
the method, but you yourself are responsible for arranging stirring (see
Stir function, Section 3.3.3).
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 2-17
Peripherals
Setup
2.7 Peripherals
When you select this menu and press <F4>, the names of the devices which you can attach
to the titrator via the RS232C interfaces, the Centronics interface or the DIN socket appear.
These devices are not defined by software in the factory.
Printer
Balance
System
Sample changer
External keyboard
Not defined
Not defined
Not defined
Not defined
Not defined
2.7.1 Printer
To define and set your printer, select Printer and press <F4>:
Printer
SETUP
Status
Printer type
Not defined
EPSON ESC/P
Esc Modify OK
Status
Printer type
Press <F4> to select "Defined".
Select the type from the selection menu:
EPSON ESC/P
HP PCL Level 3
IBM Proprinter
ASCII
GA42
You can change the settings for the selected printer in the parameter mask, which appears
when you press <F4>.
Note: On delivery of the titrator, the Centronics option is installed in slot 4 which has a parallel
interface for the printer connection (see Section 10.4). An RS option with a serial printer
interface has to be installed in slot 4 in order to use the METTLER TOLEDO Printer GA42.
2-18 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Setup
Peripherals
2.7.1.1 Centronics interface: Settings for an EPSON printer
Examples of default settings for an EPSON printer:
Type
LX/SX
Paper
Fanfold
Paper format 8 1 / 2 * 11"
Automatic form feed No
Frame lines
Straight
Type
Paper
Press <F4> to select : "LX/SX" or "LQ/Stylus".
Press <F4> to select : "Fanfold" or "Single sheet".
Single sheet:
- A form feed is effected at the end of a page.
- If you select Yes for "Automatic form feed", a footer and a header will
be printed out on every page.
NOTICE
You must not trigger a line or form feed at the printer if you
have selected "Automatic form feed" for "Single sheet"!
You can do this on the titrator with the following key combinations:
<Shift + F1> triggers a line feed,
<Shift + F2> triggers a form feed.
Paper format
Form feed
Frame lines
Press <F4> to select:
• DIN A4 (width = 21 cm, length = 29.7 cm)
• 8 1 / 2 * 11" (width = 8 1 / 2 inches , length = 11 inches)
• 8 1 / 2 * 12" (width = 8 1 / 2 inches, length = 12 inches)
Press <F4> to select: "Yes" or "No".
• Yes: The printer initiates a form feed after every document.
• No: The individual documents are separated by a distance of 2 lines.
Press <F4> to select the frame for the report:
• Straight: The report is framed with straight lines.
• Dotted: The report is framed with dotted lines; this speeds up the
printout by a factor of 2.
• None (no frame): Printing is fastest with this parameter.
HP printers: Selection of the lines does not influence the speed.
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 2-19
Peripherals
Setup
2.7.1.2 RS232 interface: Settings for an EPSON printer
You can also attach printers to a serial interface if you have installed an RS option in slot 4 (see
Section 10.4). You can use this interface to attach a GA42 Printer. While no software settings
are necessary for the GA42 Printer, you must set its DIP switch 2 to the ON position (see GA42
Operating Instructions).
Example of the default settings for an EPSON printer:
Type
LX/SX
Paper
Fanfold
Paper format 8 1 / 2 * 11"
Automatic form feed No
Frame lines
Straight
Baudrate 2400
Parity
Even
Number of data bits 8 Bits
Number of stop bits 1 Bit
The settings for the first 5 parameters are the same as those for the
Centronics interface (see Section 2.7.1.1).
Baud rate Press <F4> to select: 1200, 2400, 4800 or 9600.
Parity
Data bits
Stop bits
Press <F4> to select: "Even", "Odd" or "None".
Press <F4> to select: 7 or 8 bits.
Press <F4> to select: 1 or 2 bit(s).
2-20 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Setup
Peripherals
2.7.2 Balance
To define and set your balance, select Balance and press <F4>:
Balance
SETUP
Status
Transmn mode
Not defined
Unidirectional
Esc Modify OK
Status
Transmission
mode
Press <F4> to select "Defined".
Press <F4> to select: "Unidirectional" or "Bidirectional".
Unidirectional
On inquiry of the weight, the balance continuously transfers the current weight value (without
a stability check), which is displayed by the titrator. You must confirm this display with <F5>
for the value to be accepted.
Titrator
Data to titrator
Balance
Configuration: "Send Cont."
Bidirectional
On inquiry of the weight, the balance transfers the current weight value when requested by the
titrator, which then displays it (with stability check).
Many METTLER TOLEDO balances show the weight limits in the balance display (see Section
4.2). The titrator accepts the stable weight value when you press either <F5> on the titrator or
the transfer key of the balance.
Command from titrator
Titrator
Data to titrator
Balance
Configuration: "Send Stable"
("Send on Transfer")
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 2-21
Peripherals
Setup
Notes
1. On delivery of the titrator, the Centronics option is installed which has a 9-pin RS232
interface for the attachment of a balance (see Section 10.4). If you install an RS option in
slot 4, its balance interface is 8-pin. Depending on the installation, you must then have the
right connection cable (see Section 11.2: Peripherals).
RS232 METTLER TOLEDO balances with RS interface: Order No. 59759, 229029
9-pin:
METTLER TOLEDO balances with LC interface: Order No. 229065
SARTORIUS balances with RS interface: Order No. 51190363
RS232 METTLER TOLEDO balances with RS interface: Order No. 51107195, 51107196
8-pin:
METTLER TOLEDO balances with LC interface: Order No. 229185
SARTORIUS balances with RS interface: Order No. 200495
2. The data output of the METTLER TOLEDO balance you wish to attach must be set as
follows:
• Baud rate: 2400
• Parity: even
• Mode: "Send Cont." for transmission mode Unidirectional
"Send Stable" ("Send on Transfer") for transmission mode Bidirectional.
3. The following settings are important with AT balances:
• Unit: Prt on Print/Transfer command on
g
Weighing unit in g
• Int-FACE: SENd S.Stb Data transmission mode: Standby
bd 2400 Baud rate: 2400
PAr -E- Parity: even
HS OFF Handshake (XON/XOFF) off
4. With AG, PG, AB, PB and PR balances, the LC-RS8 cable must be set as follows:
For bidirectional transfer
For unidirectional transfer
• Left switch: Position 0 • Left switch: Position 7
• Middle switch: Position 3 • Middle switch: Position 3
• Right switch: Position 4 • Right switch: Position 4
5. In case of SARTORIUS balances you must select Bidirectional as the transmission mode.
The data output of the balance you wish to attach must be set as follows:
• Data output: Ext. print command/without stability
• Baud rate: 2400
• Parity: even
• Stop bit: 1
• Weight unit: g
2-22 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Setup
Peripherals
2.7.3 System
To define and set your computer or terminal, select System and press <F4>:
System
SETUP
Status
Instrument type
Not defined
Computer
Esc Modify OK
Status
Instrument type
Press <F4> to select "Defined".
Press <F4> to select from the selection menu:
Instrument type
SETUP
Computer
Color terminal
Monochrome term.
Esc Modify
You can modify the settings for the Computer in the parameter mask, which appears when
you press <F4>:
Baud rate 4800
Parity
Even
Number data bits
8 bits
Number stop bits
1 bit
Bar code to computer No
Character set
ASCII
Send mode
Spontaneous
Communication protocol Normal
Start/end character '{'/ CR
Baud rate Press <F4> to select: 1200, 2400, 4800 or 9600.
Parity
Press <F4> to select: "Even", "Odd" or "None".
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 2-23
Peripherals
Setup
Data bits
Stop bits
Bar code
Press <F4> to select: 7 or 8 bits.
Press <F4> to select: 1 or 2 bit(s).
Press <F4> to select: "Yes" or "No".
Yes: If you have attached a bar-code reader to the external keyboard, the
read-in bar code will be sent directly to the computer. Depending on
the program you have loaded, this bar code can be used to trigger
commands from the computer. (See Section 6.8.4: Bar-code string
and Section 10.8.1: Attaching a bar-code reader.)
Character set Press <F4> to select: "ASCII" or "DL" .
ASCII: The standard character set (HEX 20 to HEX 7E) for the
output of text to the computer is used.
DL:
The character set used in the titrator is used.
Notice: If you select DL, you must specify 8 for the number
of data bits!
Send mode
Communication
protocol
Press <F4> to select: "Spontaneous" or "On request".
Spontaneous: The titrator sends the computer orders and data, as well
as what has been generated (assuming that the computer
is ready).
On request: The titrator waits for the corresponding inquiry from the
computer before it sends an order or data.
Press <F4> to select: "Normal" or "Reduced".
Normal: The data received by the titrator or computer are checked
and any errors given with messages (safety mechanism
active).
Reduced: The received data are neither checked nor acknowledged
so that no error messages are given (safety mechanism
passive).
Start/end Press <F4> to select: " '{' / CR " or " STX / ETX ".
character
These start and end characters for the messages to be sent depend on
the input possibility of your computer.
The computer needs an RS232C interface (DTE). The parameters baud rate, parity, number
of data bits and number of stop bits are freely selectable. Connection cables are listed in
Section 11.2.
You will find additional information on the communication between titrator and computer in
Section 6.8.2 and in the Operating Instructions for the computer interface (see Section 11.2).
2-24 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Setup
Peripherals
You can change the settings for the selected Terminal in the parameter mask which appears
when you press <F4>:
Baud rate 9600
Parity
Even
Number data bits 8 bits
Number stop bits 1 bit
Baud rate Press <F4> to select: 1200, 2400, 4800 or 9600.
Parity
Data bits
Stop bits
Press <F4> to select: "Even", "Odd" or "None".
Press <F4> to select: 7 or 8 bits.
Press <F4> to select: 1 or 2 bit(s).
You can use a terminal of the type DEC VT340 or DEC VT241 as a color terminal and either
a DEC VT330 or DEC VT240 type as a monochrome terminal. The connection cable is listed
in Section 11.2.
You will find additional information on the configuration of the terminals and their operation in
Section 10.9.
2.7.4 Sample changer
Note: All information concerning automation in this titrator's Reference Handbook relates to
the earlier METTLER TOLEDO Sample changer ST20A. Its name continues to appear
in some places.
To define the sample changer, select Sample changer and press <F4>:
Sample changer
SETUP
Status Not defined
Esc Modify OK
Status
Press <F4> to select "Defined".
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 2-25
Peripherals
Setup
2.7.5 External keyboard
You can attach a PC keyboard which has a DIN socket using a DIN cable (see Section 10.8:
Attaching a keyboard). To define the keyboard, select External keyboard and press <F4>:
External keyboard
SETUP
Status
Layout
Not defined
US International
Esc Modify OK
Status
Layout
Press <F4> to select "Defined".
Press <F4> to select: "US International", "Swiss (German)", "French",
"German" or "Polish".
2-26 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Setup
Solvents
2.8 Solvents (with sample changer only)
The parameters of this resource are needed only if you have defined and attached a sample
changer. You can connect time-controlled pumps, electromagnetic valves or dispensers to the
sample changer to dispense solvents.
When you select this menu and press <F4>, the stored solvents appear:
Solvents (only with ST20A)
SETUP
H 2 O
CH 3 OH
CHCl 3
Esc Delete Add Modify OK
∇
2.8.1 Delete
If you press <F2>, "Delete", H 2 O will be deleted from the list.
2.8.2 Modify
If you press <F4>, "Modify", the following appears:
Solvent parameters
SETUP
Name
H 2 O
Pump rate [mL/min] 200.0
ST20A output RINSE
Esc Modify OK
Name
Pump rate
You can modify the name by pressing <F4> and choosing a different
name from the list of "Solvent names" that appears.
The titrator uses the pump rate to calculate the volume to be dispensed
for time-controlled pumps or electromagnetic valves.
You must determine the pump rate of each metering device beforehand
for every solvent:
– Add the solvent using the metering device to a measuring cylinder for
1 minute (stopwatch) and note the volume.
– Repeat this procedure for, e.g. 20, 30 and 40 seconds.
– Calculate the mean value per minute for the different volumes and
enter this value.
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 2-27
Solvents
Setup
Output
Press <F4> to select the ST20A output to which you will attach the
metering device: "DOSE", "RINSE" or "DISPENSER".
Storage procedure
The solvent with the modified parameters is stored when you confirm both the parameter mask
and the list of solvents with OK. The modification of a parameter is always stored when you
press a Menu or Auxiliary function key (see appropriate paragraph at the end of Section 2.1.2).
2.8.3 Add
You can use this command to add a solvent to the list of solvents. The following mask always
appears:
Solvent parameters
SETUP
Name
H 2 O
Pump rate [mL/min] 200.0
ST20A output RINSE
Esc Modify OK
You can modify the parameters by the procedure described in Section
2.8.2.
If you define, e.g. several solvents with the same name, you must also flag these to ensure the
titrator can distinguish them, e.g. H 2 O/2. If you do not, a message appears (see Section 2.1.3).
2-28 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Method
Method
Contents
Page
3. Methods ..................................................................................................... 3-3
3.1 Selecting methods .................................................................................... 3-4
3.1.1 Method ID.................................................................................................... 3-4
3.1.2 Standard methods ....................................................................................... 3-5
3.1.3 User methods .............................................................................................. 3-6
3.1.4 METTLER methods..................................................................................... 3-6
3.2 Modifying a method .................................................................................. 3-7
3.2.1 Storage procedure ...................................................................................... 3-7
3.3 Functions ................................................................................................... 3-10
3.3.1 Title.............................................................................................................. 3-11
3.3.2 Sample ........................................................................................................ 3-12
3.3.3 Stir ............................................................................................................... 3-16
3.3.4 Measure ...................................................................................................... 3-17
3.3.5 Dispense ..................................................................................................... 3-19
3.3.6 EQP titration (equivalence point titration).................................................... 3-20
3.3.6.1 Titrant/Sensor .............................................................................................. 3-20
3.3.6.2 Predispensing ............................................................................................. 3-21
3.3.6.3 Titrant addition............................................................................................. 3-22
3.3.6.4 Measure mode ............................................................................................ 3-25
3.3.6.5 Recognition ................................................................................................. 3-28
3.3.6.6 Termination.................................................................................................. 3-33
3.3.6.7 Evaluation ................................................................................................... 3-34
3.3.7 EP titration (end point titration).................................................................... 3-37
3.3.7.1 Titrant/Sensor .............................................................................................. 3-37
3.3.7.2 Predispensing ............................................................................................. 3-38
3.3.7.3 Titrant addition............................................................................................. 3-39
3.3.7.4 End point ..................................................................................................... 3-41
3.3.7.5 Tendency ..................................................................................................... 3-42
3.3.7.6 Termination.................................................................................................. 3-42
3.3.8 Learn titration .............................................................................................. 3-43
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 3-1
Method
Method
Page
3.3.9 EQP titration (Ipol/Upol) .............................................................................. 3-45
3.3.9.1 Titrant/Sensor .............................................................................................. 3-45
3.3.9.2 Indication ..................................................................................................... 3-46
3.3.9.3 Predispensing ............................................................................................. 3-48
3.3.9.4 Titrant addition............................................................................................. 3-48
3.3.9.5 Measure mode ............................................................................................ 3-49
3.3.9.6 Recognition ................................................................................................. 3-50
3.3.9.7 Termination.................................................................................................. 3-52
3.3.9.8 Evaluation ................................................................................................... 3-53
3.3.10 EP titration (Ipol/Upol) ................................................................................. 3-56
3.3.10.1 Titrant/Sensor .............................................................................................. 3-56
3.3.10.2 Indication ..................................................................................................... 3-57
3.3.10.3 Predispensing ............................................................................................. 3-59
3.3.10.4 Titrant addition............................................................................................. 3-60
3.3.10.5 End point ..................................................................................................... 3-60
3.3.10.6 Tendency ..................................................................................................... 3-61
3.3.10.7 Termination.................................................................................................. 3-61
3.3.11 pH/mV-stat .................................................................................................. 3-62
3.3.12 Calculation .................................................................................................. 3-66
3.3.13 Calibration ................................................................................................... 3-69
3.3.14 Titer ............................................................................................................. 3-72
3.3.15 Auxiliary value ............................................................................................. 3-73
3.3.16 Report ......................................................................................................... 3-74
3-2 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Method
Method
3. Methods
The titrator performs analyses automatically with the aid of predefined methods. When the
titrator is delivered, the Method menu already holds standard methods and methods developed
by METTLER. You can modify these methods to suit your requirements and save them
as User methods.
A complete titration method comprises sample preparation, dispensing of auxiliary solutions,
stirring and wait times, the actual titration, result calculation and a report. These sub-steps are
defined as Functions in the titrator and they are executed in succession in an analysis. The
individual functions comprise Parameters, whose values or names you can modify.
You modify a method by changing the parameters of the preset functions.
Method
Functions
Parameters
modify
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 3-3
Selecting methods
Method
3.1 Selecting methods
When you press the Method key, the following appears in the display:
Method ID _ _ _ _ _
Standard methods
User methods
METTLER methods
Under Method ID you can call up a method stored with an identification number; these are the
user and METTLER methods.
Standard methods are methods we have entered in the factory without a method identification.
User methods are the methods you have defined and saved (they are stored in the user data
memory).
METTLER methods are applications we have developed and stored for you.
Method ID:
Method groups:
The recalled method is printed out with its functions and parameters.
The list of methods of the selected group is printed out.
3.1.1 Method ID
The method identification is the 5-place, numeric identifier of a method (see Title function,
Section 3.3.1).
– If you are familiar with the identification, enter it and press <F4>: The list of the functions of
this method appears (see Section 3.2).
Methods
Method ID 00001
Standard methods
User methods
Delete Print Modify
METHOD
∇
Note: With user methods, you can enter an asterisk (*) after the first, second or third digit of
the method ID: This replaces the remaining digits, e.g. 1*: All methods whose ID starts
with 1 appear in the display. When you confirm the desired ID with OK, the function list
of the selected method appears immediately.
3-4 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Method
Selecting methods
Delete
When you press <F2>, the selection menu "Delete method?" appears in which you can confirm
or cancel the command:
• Yes: The method will be deleted.
• No: The method remains stored.
If you wish to delete a METTLER method or a user method that is stored in the sample data
memory, an error message appears. You can not delete these methods (see Section 3.1.4 and
4.1).
3.1.2 Standard methods
If you select this group, the list of 21 methods entered in the factory appears with titles (see
Section 5.2 of the Quick Guide). The standard methods are used as basic methods for the
development of your own methods and can not be called up to perform an analysis as they
have no method identification.
As soon as you allocate an identification to a standard method, it is automatically stored as a
user method with a method ID when saved and can be called up by you for the analysis.
The parameters of the other functions are defined with default values or names for all methods;
you can accept or modify these (see Sections 3.2 and 3.3).
The selected method with its functions and parameters is printed.
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 3-5
Selecting methods
Method
3.1.3 User methods
If you select this group, a list of methods with the identification (and the name) first appears
if you have stored modified standard or METTLER methods, e.g.:
User methods
METHOD
61
11152
00001 ∇
Esc Delete Print Modify
Delete
When you press <F2>, the selection menu "Delete method?" appears in which you can confirm
or cancel the command:
• Yes: The method will be deleted.
• No: The method remains stored.
Note: The "Delete" command is not available when the method is stored in the sample data
memory (see Section 4.1).
The selected method with its functions and parameters is printed.
3.1.4 METTLER methods
If you select this group, a list of applications developed by us with the identification and the
name of the method appears:
METTLER methods
METHOD
90001 Acid content
90002 Calibration pH electrode
90003 Calibration F - electrode ∇
Esc Print Modify
You can modify METTLER methods to suit your requirements, but you must then give them a
new method identification under the Title function in order to save them (see Section 3.3.1).
The selected method with its functions and parameters is printed.
3-6 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Method
Modifying a method
3.2 Modifying a method
You have selected, e.g. the "Equivalence point titration" under standard methods and then
press <F4>: The list of the functions of this method appears.
Title
Sample
Stir
EQP titration
Calculation
Calculation
Calculation
Report
You can modify the parameters of all functions:
– Select the function and press <F4>: The mask with the Parameters defined for the function
appears and you can then change their values and names (see Section 3.3).
3.2.1 Storage procedure
1. Every modified method is stored when you
• confirm the corresponding parameter masks of the individual functions and
• the list of the functions with OK.
If you switch off the titrator before doing this, your changes will be discarded: all parameters
are first written to the RAM (volatile memory) of the titrator and are not stored until the
function list has been confirmed.
After modification of the parameters of a function, you can save the method. The following
sequence is an example of the standard method "Equivalence point titration":
Method:
METHOD
Title
Sample
Stir
Esc Modify OK
∇
You modify the Title function by entering, e.g. the method identification
00001 and confirming it with OK (see also Section 3.3.1):
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 3-7
Modifying a method
Method
Method: 00001
METHOD
Title
Sample
Stir
Esc Modify OK
∇
If you now press <F5>, the mask of the method groups reappears: the
method is stored.
Methods
Method ID 00001
Standard methods
User methods
Delete Print Modify
METHOD
∇
As method 00001 is selected automatically, the list of the functions
reappears immediately when <F4> is pressed to allow modification of
other functions.
2. If you quit the list of the functions with Esc after you have modified parameters of functions,
the following appears:
Save method?
METHOD
No
Yes
OK
No: Changes are not stored.
Yes: The titrator stores all modified parameter values.
The same message appears if you press a Menu or Auxiliary function key.
3-8 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Method
Modifying a method
3. Several functions such as "EQP titration" have a list of parameters which have subparameters.
If you quit their mask with Esc after modifications, the following appears:
Save changes?
METHOD
No
Yes
OK
No: Changes are not stored.
Yes: The titrator stores the modified parameter values.
4. On storage, the titrator checks the method and, if it has found several errors, draws your
attention to the first of them.
a. When you have confirmed the error message, the list of method functions appears with
the selection bar on the first faulty function. After you have corrected its parameters and
saved the method again, the next faulty function appears etc.
b. If you have entered an existing method identification under the Title function, the following
appears:
Method ID exists
METHOD
Modify ID
Overwrite method
OK
Modify ID
The list of method functions appears; under the Title function you can
modify the method ID and then save the method.
Overwrite method
The new or modified method is stored, that with the same identification
deleted.
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 3-9
Functions
Method
3.3 Functions
You select all functions whose parameters you wish to modify by the procedure described in
Sections 3.1 and 3.2. The following explanations of all functions of the standard and METTLER
methods as well as their parameters thus pertain to the Modify command. The order of the
following sections corresponds to the below list of functions.
Title
Sample
Stir
Measure
Dispense
EQP titration
EP titration
Learn titration
EQP titration (Ipol/Upol)
EP titration (Ipol/Upol)
pH/mV-stat
Calculation
Calibration
Titer
Auxiliary value
Report
3-10 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Method
Title
3.3.1 Title
This function is used for recognition of the titration method. You must enter its method ID
parameter. You use this identification to call up the stored methods. The function has no
meaning in the sequence of the method. The following example shows the title mask of the
standard method "Equivalence point titration" (see Sections 3.2. and 3.2.1).
Title
Method ID
_ _ _ _ _
Title Equivalence point titr'n
Date/time 00-00-0000 00:00
Esc
METHOD
OK
Method ID
Title
Date/time
Enter a number with max. 5 characters (letters also admissible).
You can not enter identifications in the range 90000 to 99999: they are
reserved for METTLER methods.
Asterisks (*) are not allowed for the method ID!
Enter a title for the method.
Date and time are entered here automatically when you save a modified
method. You can neither delete not overwrite this information.
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 3-11
Sample
Method
3.3.2 Sample
Here you define the parameters needed for the entry of the sample data such as weight or
volume, the titration stand on which the sample should be titrated and whether the temperature
of the sample solution should be measured.
Sample ID
Entry type
Weight
Molar mass M 100
Equivalent number z 1
Titration stand Stand 1
Temperature sensor Manual
Note: If samples which are determined with this method are stored in the sample data memory
as "ready", you can not modify the sample ID and the entry type (see Section 4.1).
During a sample series, you can modify only the numeric parameter values for the
ST20A titration stand (see following page and Section 5.2.3).
Sample ID
Entry type
Enter the identification of the sample if required. This is adopted for all
samples of a sample series. You can also enter and modify the ID in the
Sample menu (see Section 4.1).
Select from the selection menu: "Weight", "Volume" or "Fixed volume".
Weight: Press <F4> to show the mask in which you can define the range
which the sample weight should not violate:
• Lower limit [g]
• Upper limit [g]
Volume: Press <F4> to show the mask in which you can define the range
which the added volume should not violate:
• Lower limit [mL]
• Upper limit [mL]
You do not enter the actual weight or volume until the sample preparation
or when the titrator requests this information after the start of the method
(see Section 5.1). If you violate the defined range, you receive an appropriate
message.
Fixed volume: Press <F4> to show the mask in which you can enter the
volume.
You can not modify this entry in the sample preparation!
3-12 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Method
Sample
Molar mass M
Equivalent
number z
Titration stand
Temperature
sensor
You can enter only one molar mass and one equivalent number for the
calculation. If your sample contains several substances which have to be
determined by equivalence points in the titration, you must enter their
molar mass and equivalent number in constant C under the Calculation
function (see Section 3.3.12).
Select the titration stand at which you wish to determine the sample from
the selection menu: "Stand 1", "Stand 2", "ST20A", "Auto stand" or
"External stand" (see Section 2.6).
Press <F4> to select the sensor if you have attached one; if not, select
"Manual": the temperature entered before starting the method will then be
adopted in the processing of the method (see Section 5.1).
This parameter is used for the automatic measurement or adoption of the
temperature of the sample solution before the start of the Measure,
EQP/EP/Learn titration and pH/mV-stat functions. Thus the slope of the
pH electrode is temperature-corrected and incorporated in the calculation
of the pH value (see Section 2.3: Temperature sensors).
ST20A titration stand
If you are working with a sample changer, here you define whether you wish to pump solvent,
rinse the sensor and/or condition. Press <F4> to display:
Pump
Pump
Rinse
Conditioning
No
No
No
No
Pump (1)
You have attached a pump to the sample changer that should pump
solvent into the titration vessel before every titration.
– Press <F4> to activate the parameter, in the mask that appears select
"Yes" and then press <F4> again:
Pump
METHOD
Solvent
H 2 O
Volume [mL] 10
Stir
No
Esc Modify OK
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 3-13
Sample
Method
Pump (1)
Pump (2)
Rinse
– Select the solvent from the list (press <F4>) which contains the
solvents you have defined in the Setup menu. This solvent also defines
the ST20A output to which you must attach the pump (see Section 2.8).
– Enter the volume that should be dispensed. You have defined the
pump rate of the device in the Setup menu.
– Select whether the dispensing operation should be stirred (Yes) or not
(No).
If you have attached a second pump to the sample changer, this will be
activated as soon as the first dispensing operation is at an end.
– To activate the parameters, press <F4>, select “Yes” in the mask that
appears and then press <F4> again: The same mask as described
under Pump (1) appears.
Selecting the solvent also defines the ST20A output to which you must
attach the second pump (see Section 2.8).
You have installed a rinsing unit in the titration head and attached a pump
to the sample changer to rinse the sensor and burette tip after every
titration.
– Press <F4> to activate the parameter, select "Yes" in the mask that
appears and press <F4> again:
Rinse
METHOD
Solvent
H 2 O
Volume [mL] 10
Esc Modify OK
– Select the solvent from the list (press <F4>) which contains the
solvents you have defined in the Setup menu. Selecting this solvent
also defines the ST20A output to which you must attach the pump (see
Section 2.8).
– Enter the volume which should be dispensed.
Note: If you run sample series in succession on the sample changer
(possible only with DL55/DL58) and the titrator has titrated the last
sample of the last method, the sensor and burette tip are rinsed in
the middle beaker position. The sample beaker is then raised to
ensure the sensor does not dry out, in other words it remains in the
sample solution.
To prevent this, you must insert a conditioning beaker after the last
sample beaker.
3-14 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Method
Sample
Conditioning
You must activate this parameter if you wish to condition the sensor:
– Press <F4>, select "Yes" in the mask that appears then press <F4>
again:
Conditioning
METHOD
Time [s] 10
Interval 1
Rinse
No
Esc OK
– Enter the time for conditioning.
– Enter the Interval (number of samples), e.g. 3, in other words,
conditioning must be performed after every third sample.
– If the sensor and burette tip have to be rinsed after the conditioning,
select Rinse and define the solvent and volume for this rinsing
operation (see page 3-14).
Notes
The following are defined (example): 10 s, 3 as interval and rinse 10 s
with H 2 O:
1. The ST20A conditions with the defined parameters if you place a
conditioning beaker after every 3rd sample beaker (always mark
conditioning beakers with red stopper plugs!)
S1
S2
S3
C
S4
S5
S6
C
S7
S8
S9
C
2. The ST20A conditions and rinses with the defined parameters if it
detects a conditioning beaker after a sample beaker.
S1
S2
C
S3
C
S4
S5
S6
C
S7
S8
S9
3. If the ST20A does not detect a conditioning beaker after the 3rd
sample beaker, it searches for one by rotating backward. It then
conditions for 10 s, but does not perform rinsing to avoid overflow of
the beaker contents.
C
S1
S2
S3
S4
S5
S6
S7
S8
S9
C
4. After the last sample of a series, the time and rinse parameters are not
executed anymore.
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 3-15
Stir
Method
3.3.3 Stir
You can modify the stirring speed and stirring and wait times.
Stir
METHOD
Speed [%] 50
Time [s] 10
Esc
OK
Speed
Time
•0 % → The stirrer is at a standstill;
•100% → The stirrer stirs at maximum speed.
The speed you define applies to all following functions up to the next
Stir function.
However, you can also change the speed during a titration (see Section
6.2).
• "0" means that the titrator does not wait at all,
• "10" means that it waits 10 s before it starts the next function: It then
stirs at the defined speed. The remaining stirring time appears in the
display.
3-16 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Method
Measure
3.3.4 Measure
You measure the potential of a solution under defined conditions. The titrator determines the
measured value as raw result E (see note 2 at the end of this section).
Sensor
DG111
Unit of meas.
mV
∆E [mV] 0.5
∆t [s] 1.0
t(min) mode
Fix
t(max) [s] 30
Sensor
Unit of measurement
∆E [mV]
∆t [s]
t(min) mode
Press <F4> to open the list containing the sensors you have defined in
the Setup menu (see Section 2.2). Confirm the one you want with OK.
The defined measurement unit is entered automatically.
If you select the unit from the selection menu, you should select mV or the
unit defined in the Setup menu, otherwise you will receive an error
message when the method is started.
The drift of the electrode potential must be less than ∆E/∆t (0.5 mV/s)
during the period ∆t (1 s) for the measured value to be acquired. This
happens within a defined time span t(min) and t(max). (See diagram in
Section 3.3.6.4: Equilibrium controlled measure mode).
t(min) is the earliest time for measured value acquisition. Select t(min)
from the selection menu:
t(min) mode
METHOD
Fix
E > set value
E < set value
Esc
Modify
Fix: Press <F4> to show the mask in which you can modify t(min).
E greater than set value:
E less than set value:
Instead of a fixed time, you can enter a condition:
The measured value must be greater or
less than a certain set value: Press <F4> to
show the mask in which you enter the set value
[mV, pH, ...].
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 3-17
Measure
Method
t(max) [s]
t(max) is the latest time for measured value acquisition.
If you have selected one of the conditions for t(min), the titrator starts the
next function only when the measured potential E is greater or less than
the set value and the drift condition is met, but at the latest after t(max).
Notes
1. If you have selected a temperature sensor under the Sample function, the temperature of
the sample solution is measured automatically before the titrator executes this function. If
you have not attached a sensor (parameter "Manual"), the titrator adopts the temperature
entered at the start of the method. The slope of the pH electrode is temperature-corrected
by this measured or entered value and incorporated in the calculation of the pH value.
2. The Measure, Dispense, EQP/EP/EQP (Ipol/Upol)/EP (Ipol/Upol)/EP/Learn titration and
pH/mV-stat functions generate Raw results (see Section 8.1: List of symbols). These you
can
• print out as such on the attached printer (see Sections 3.3.16 and 8.1.1 for exceptions).
• incorporate in the calculation (see Examples of formulas: Section 8.4).
• obtain as a result if you assign them to the result R: e.g. R = E (see Section 3.3.12).
At the end of a titration, you obtain only the final results in the display of the titrator, the raw
results you can only print out.
The titrator stores raw results up to the determination of the next sample within a sample
series (see Section 8.5.6).
3-18 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Method
Dispense
3.3.5 Dispense
You dispense a certain volume of a titrant with a METTLER TOLEDO burette. The titrator
determines the dispensed volume as the raw result VDISP [mL] or QDISP [mmol] (see Section
8.1: List of symbols and Section 3.3.4: Measure function).
Dispense
METHOD
Titrant
NaOH
Concentration [mol/L] 0.1
Volume [mL] 1.0
Esc Modify OK
Titrant
Concentration
Volume
Press <F4> to open the list containing the titrants you have defined in the
Setup menu (see Section 2.1). Confirm the one you want with OK.
The defined concentration of the titrant is entered automatically.
Enter the volume to be dispensed. Instead of a number, you can enter a
formula.
Note: If you do not need the Dispense function in the method, enter 0 for the volume: The
function will then be skipped during the analysis.
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 3-19
EQP titration
Method
3.3.6 EQP titration (equivalence point titration)
Here you define the control and evaluation of an equivalence point titration. The equivalence
point is that point at which exactly the same number of equivalents of titrant and analyte have
reacted. In most cases, it is virtually identical to the inflection point of the titration curve. This
inflection point is recognized and the equivalence point calculated.
Titrant/Sensor
Predispensing
Titrant addition
Measure mode
Recognition
Termination
Evaluation
The titrator determines several measured values and volumes as raw results including the half
neutralization value EHNV (see Section 8.1: List of symbols and Section 3.3.4: Measure
function).
Note: If you have selected a temperature sensor under the Sample function, the temperature
of the sample solution is measured automatically before the titrator executes this
function. If you have not attached a sensor (parameter "Manual"), the titrator adopts the
temperature entered at the start of the method. The slope of the pH electrode is
temperature-corrected by this measured or entered value and incorporated in the
calculation of the pH value.
3.3.6.1 Titrant/Sensor
To modify the titrant or sensor, press <F4>:
Titrant
NaOH
Concentration [mol/L] 0.1
Sensor
DG111
Unit of meas.
mV
Titrant
Concentration
Press <F4> to open the list containing the titrants you have defined in the
Setup menu (see Section 2.1). Confirm the one you want with OK.
The defined concentration of the titrant is entered automatically.
3-20 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Method
EQP titration: Predispensing
Sensor
Unit of measurement
Press <F4> to open the list containing the sensors you have defined in
the Setup menu (see Section 2.2.). Confirm the one you want with OK.
The defined measurement unit is entered automatically.
If you select the unit from the selection menu, you should select mV or the
unit defined in the Setup menu, otherwise you will receive an error
message when the method is started.
3.3.6.2 Predispensing
Predispensing shortens the titration time. You can select one of four predispensing modes
when you press <F4>.
to volume
to potential
to slope
to (factor x sample size)
No
Volume
Potential
Slope
Factor x
sample size
No
You dispense a specified volume [mL] which you can enter when you
press <F4>. Instead of a number, you can enter a formula.
You can also enter a wait time: After the predispensing, the titrator waits
for this time to elapse before it adds the titrant in a controlled manner.
You dispense to a specified potential [mV, pH, ....] that you can enter
when you press <F4>.
You dispense to a specified slope [mV, pH, ..../mL] of the titration curve;
this you can enter when you press <F4>.
You dispense to a specified volume that is calculated from the product of
the sample weight or volume and a factor; you can enter this factor when
you press <F4>.
You can also enter a wait time (see “Volume”).
You do not wish to predispense.
In predispensing to volumes or (sample size x factor), the titrator adds the titrant in three steps
(4/7, 2/7, 1/7 of the defined volume), which allows optimum calculation of the addition
increment in the subsequent dynamic titration. In the predispensing to potential or slope, the
titrant addition follows the selected parameters of the main titration, but the increments are
larger.
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 3-21
EQP titration: Titrant addition
Method
The titrator acquires the potential values ET1 and ET2 in the predispensing.
ET1: Potential at the start of predispensing
or the titration
ET2: Potential after predispensing
Predispensing to
a: Volume or
(factor x sample size)
b: Potential
c: Slope
3.3.6.3 Titrant addition
You select the dynamic or incremental addition mode:
Titrant addition
METHOD
Dynamic
Incremental
Esc Info Modify
You can not decide what addition mode is optimum for your method until you know the titration
curve. You can obtain general information by pressing <F3>:
• You generally select the dynamic addition mode for acid/base titrations in aqueous media
or for argentometric and redox titrations,
• the incremental for acid/base titrations in nonaqueous media or for complexometric, redox
and surfactant titrations.
Dynamic
The volume increment added by the titrator changes within the defined
smallest and largest increment: ∆V(min) and ∆V(max). This should lead
to a constant potential difference ∆E per increment.
Dynamic
METHOD
∆E(set) [mV] 8.0
∆V(min) [mL] 0.02
∆V(max) [mL] 0.15
Esc Help OK
3-22 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Method
EQP titration: Titrant addition
Dynamic
∆E ∆E ~ ∆E(soll) ∆E(set)
∆E 4
∆E 3
∆E 5
E [mV]
∆V 3
∆V 4
V [mL]
If no predispensing takes place, the titrator dispenses the first two volume
increments with ∆V(min).
Help
If you press <F3>, you can adopt the values for all three parameters from
the mask that appears. With the proposed value groups, sufficient
measured points in the vicinity of the equivalence point should be
generated to optimize its evaluation:
• For steep titration curves, a value for ∆E must be selected which is
smaller than for flat titration curves, e.g. 8 mV.
• For titration curves which show a sudden potential change, small
values for ∆E and ∆V(max) should be selected, e.g. 4 mV.
E [mV]
steep curve
flat curve
curve with sudden
potential jump
V [mL]
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 3-23
EQP titration: Titrant addition
Method
Dynamic
Note: The smallest increment that the titrator can dispense is 1/5000 of
the burette volume:
1 mL burette → 0.0002 mL 5 mL burette → 0.001 mL
10 mL burette → 0.002 mL 20 mL burette → 0.004 mL
Incremental
The volume increment added by the titrator is constant.
Incremental
METHOD
∆V [mL] 0.3
Esc Help OK
E [mV]
∆E 14
∆E 13
∆E 12
∆V 12 ∆V 14
V [mL]
Help
If you press <F3>, you can adopt a value for ∆V from the mask that
appears. The proposed values are intended to achieve the same goal as
in the dynamic titrant addition: sufficient measured points in the vicinity of
the equivalence point to optimize its evaluation:
• For steep titration curves, a smaller ∆V must be selected than for flat
curves to ensure there are sufficient measured points at the equivalence
point.
• For titration curves which exhibit a sudden potential change, a small ∆V
should be selected.
3-24 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Method
EQP titration: Measure mode
3.3.6.4 Measure mode
You select the equilibrium controlled or timed increment measured value acquisition:
Measure mode
METHOD
Equilibrium controlled
Timed increment
Esc Info Modify
These parameters are used to define the wait time up to measured value acquisition following
an incremental addition. In the equilibrium controlled measure mode, the wait time is variable,
in the timed increment it is constant. You can not decide the optimum measure mode for your
method until you know the reaction time of the components and the response time of the sensor
used. General information is available under <F3>:
• Select the equilibrium controlled measure mode for, e.g. acid/base titrations in aqueous
media, argentometric, complexometric, surfactant and redox titrations.
• the timed increment mode for acid/base titrations in nonaqueous media.
Equilibrium
controlled
∆E [mV] 0.5
∆t [s] 1.0
t(min) [s] 3.0
t(max) [s] 30.0
Before the titrator adds the next increment, an equilibrium must be
established in the solution; the measured value must stabilize.
The following are responsible for the equilibrium
• the potential change ∆E measured in the solution
within the defined time ∆t.
The wait time up to the next increment addition also depends on the
definition of the
• minimum time t(min) and
the maximum time t(max)
As soon as the potential change of the solution is less than the defined
equilibrium (∆E/∆t), the titrator acquires the measured value and adds the
next increment.
This can be at t(min) at the earliest and should be at t(max) at the latest.
At t(max) the measured value is in any case acquired, even if the
equilibrium condition is not yet met.
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 3-25
EQP titration: Measure mode
Method
Equilibrium
controlled
E [mV]
164
163
162
161
160
∆t = 2 s
a
∆E = 1 mV
159
158
157
∆t = 2 s
a
∆E = 1 mV
t(min)
156
155
b
154
153
152
t(min)
b
Increment addtion
151
Increment addition
t [s]
0 1 t(min) 5 0 1 t(min) 5
7
a: The defined equilibrium condition is not yet met.
t(min): The equilibrium condition is not met after 3 s.
b: The equilibrium condition is met for the first time after 5.4 or 6.9 s.
This measure mode results in fast increment addition in the flat part of the
titration curve and slow addition in the steep part.
In addition to the potential, the titrator acquires the time and the titrant
volume. You can print out these measured values after the titration of a
sample (see Section 3.3.16: Report).
Help
When you press <F3>, you can adopt suggested values for all four
parameters from the mask that appears:
• Fast titrations are, e.g. acid/base titrations in aqueous media,
• slow titrations, e.g. precipitation titrations in nonaqueous media.
3-26 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Method
EQP titration: Measure mode
Timed increment
After every increment addition, the titrator allows the time you have
defined to elapse before it acquires the measured value.
Timed increment
METHOD
∆t [s] 3.0
Esc Help OK
Help
When you press <F3>, you can adopt suggested values for the time
interval from the mask that appears:
• Enter small time intervals for, e.g. acid/base titrations in aqueous
media,
• large intervals for, e.g. precipitation titrations in nonaqueous media,
• 90 seconds for, e.g. TAN/TBN titrations of oils.
E [mV]
Measured
value
Measured
values
Increment addtion
Increment addition
Increment addition
∆t 1 = 3 s ∆t 2 = 3 s ∆t 3 = 3 s
t [s]
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 3-27
EQP titration: Recognition
Method
3.3.6.5 Recognition
Recognition of the equivalence point of a titration curve depends on the type of reaction of the
components and hence on the evaluation procedure (see Section 3.3.6.7):
1. With the evaluation procedures Maximum and Minimum, an equivalence point is recognized
when the greatest (smallest) potential value of the titration curve is greater (less) than
two preceding and two subsequent values.
2. With the evaluation procedures Standard and Asymmetric, an equivalence point is recognized
when the maximum of the absolute values of the 1st derivative of the titration curve
is greater than two preceding and two subsequent values.
3. With the evaluation procedure Segmented, an equivalence point is recognized when the
maximum of the absolute values of the 2nd derivative of the titration curve is greater than
two preceding and two subsequent values.
Note: For the first recognition of a possible equivalence point, the titrator needs a certain
number of measured points.
• The number depends on the evaluation procedure.
• Measured points of a predispensing are not taken into account!
Evaluation procedure First possible EQP Required number
at
of measured points
Minimum/Maximum 4th measured point 6
Standard 4th measured point 6
Asymmetric 4th measured point 10
Segmented 5th measured point 8
Example illustrating the evaluation procedure Minimum: If the lowest potential value is
found at the third measured point, it is not recognized as an equivalence point.
You must or can support the equivalence point recognition using four parameters:
Threshold 10.0
Steepest jump only No
Range
No
Tendency
None
3-28 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Method
EQP titration: Recognition
Threshold
To ensure that minor disturbances in the curve profile are not recognized
as equivalence points, you must determine a threshold value:
+/-E, ∆E/∆V or ∆ 2 E/∆V 2 . This must be exceeded.
Notes
1. The threshold value should be maximum half as large as the expected
maximum value of the first or second derivative at the equivalence
point.
2. The threshold value depends on so many factors (solvent, concentration,
sensor, type of reaction, etc.) that you can not define the "correct"
threshold value until you have performed the first titration.
Info
You can obtain general information on threshold values in the standard
evaluation if you press <F3>:
• For steep, normal and flat titration curves, ranges of threshold
values are given for the first derivative in "pH/mL" and "mV/mL".
With the other evaluation procedures, you can enter one of the following
values for the first titration:
Evaluation procedure mV pH/pM/pX %T
Minimum/Maximum 0 0 0
Asymmetric 10 0.2 1
Segmented 10 0.2 1
Using the corresponding titration curve or the table of measured
values (there is no table for the 2nd derivative), you can read off the
potential value, the values for ∆E/∆V or the values for ∆ 2 E/∆V 2 and
then enter the threshold value.
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 3-29
EQP titration: Recognition
Method
Threshold
Example of threshold value of a titration curve for the evaluation procedure Minimum
E [mV]
+100
The minimum of this titration curve lies
at -150 mV. You can enter, for example,
-30 as the threshold value. The
sign for the mV value must be entered.
0
-30
Threshold value
-100
Minimum
V [mL]
Example of threshold value of a titration curve for the evaluation procedures Standard
and Asymmetric
Titration curve
E [mV]
V [mL]
1st derivative
∆E/∆V [mV/mL]
Maximum (absolute)
Measured value
Threshold
value
V [mL]
3-30 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Method
EQ titration: Recognition
Threshold
Example of threshold value of a titration curve for the evaluation procedure Segmented
Titration curve
E [mV]
V [mL]
1st derivative
∆E/∆V [mV/mL]
V [mL]
2nd derivative ∆ 2 E/∆V 2 [mV/mL 2 ]
Maximum (absolute)
Measured value
Threshold value
V [mL]
Steepest jump
only
You can also select this parameter for recognition of the equivalence
point (press <F4>): The titrator then recognizes only the steepest jump of
the titration curve.
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 3-31
EQP titration: Recognition
Method
Range
You can also define a potential range for recognition of the equivalence
point. Equivalence points which lie outside this range are not recognized.
– Press <F4>, select "Yes" in the mask that appears and then press <F4>
again:
Range
METHOD
Limit A [mV,pH,..] 100
Limit B [mV,pH,..] -100
Esc
OK
Instead of a number, you can also enter a potential stored as an auxiliary
value or a formula.
The equivalence point range always refers to the titration curve, in other
words it is independent of the evaluation procedure.
Tendency
As a further aid in recognition of the equivalence point, you can define the
tendency, namely in what part of the titration curve – descending or
ascending – the equivalence point should be located (see diagram).
– Use <F4> to select the tendency: "None", "Positive" or "Negative".
Example illustrating selection of a range and positive tendency. Equivalence points
which lie outside this range and which do not have the right tendency are ignored.
Notice: A positive mV change means a negative pH change!
E [mV]
Limit A = +200
Limit B = +100
EQP
Tendenz: Tendency: positive
0
V [mL]
3-32 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Method
EQP titration: Termination
3.3.6.6 Termination
You can define how a titration should be terminated by selection of five different parameters.
The titrator can terminate the titration either
• when the first of the selected conditions is met, or
• when all the selected conditions are met.
• An exception is the maximum volume: when this is reached, the titration is immediately
terminated!
at maximum volume [mL] 10.0
at potential
No
at slope
No
after number EQPs
No
comb. termination conditions No
Maximum volume You must enter the maximum volume. It is intended as a safety precaution:
if the titration is faulty, an excessive amount of titrant is not dispensed
unnecessarily as the titration is always terminated.
Potential
The titrator terminates the titration at the defined potential.
– Press <F4>, select "Yes" in the mask that appears then press <F4>
again:
At potential
METHOD
Potential [mV,pH,..] 0.0
Esc
OK
Instead of a number, you can also enter a potential stored as an auxiliary
value or a formula.
Slope
The titrator terminates the titration when the slope of the titration curve is
less than a certain value. The measured slope must exceed this absolute
value once and drop below it twice to terminate the titration.
– Press <F4>, select "Yes" and press <F4> again: Enter the slope [mV,
pH,.../mL] in the mask that appears.
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 3-33
EQP titration: Evaluation
Method
Number EQPs
Comb. terminatetion
criteria
The titrator terminates the titration after a certain number n of equivalence
points has been found.
– Press <F4>, select "Yes" and press <F4> again: Enter the number in
the mask that appears.
The titrator does not terminate the titration until all defined termination
conditions are met (exception: maximum volume, see above).
– Select "Yes" with <F4>.
3.3.6.7 Evaluation
You determine which evaluation procedure is used to calculate the equivalence points found
and select potential values you wish to have evaluated.
Procedure
Potential 1
Potential 2
Stop for reevaluation
Standard
No
No
No
Procedure
Use <F4> to select the procedure among several calculation modes
which matches the titration curve (see Section 8.3).
Standard Evaluation procedure for all S-shaped titration curves
Minimum Determination of the minimum of a titration curve
Maximum Determination of the maximum of a titration curve
Segmented Evaluation procedure for titration curves with segments
(segmented curve)
Asymmetric Evaluation procedure for S-shaped, highly asymmetric
titration curves.
Note: If an evaluation is not possible with the standard procedure, the
equivalence point is calculated by interpolation (see Section 8.3.1).
If an evaluation is not possible with the asymmetric procedure, the
equivalence point is calculated by the standard procedure or by
interpolation (see Section 8.3.4).
In both cases, the evaluation procedure will be documented for
each sample in the “raw results” report if it differs from the defined
procedure. If you have to work according to GLP guidelines, you
should therefore always record the raw results (see Section
3.3.16).
3-34 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Method
EQP titration: Evaluation
Potential 1
Potential 2
In the selection of potentials, the titrator evaluates not only the equivalence
points, but also the titrant consumption in mmol or mL required for
attainment of these potential values (see Section 8.1: List of symbols).
– Press <F4>, select "Yes" and press <F4> again: Enter the first potential
[mV, pH,...] in the mask that appears.
Instead of a number, you can also enter a potential stored as an
auxiliary value or a formula (see Section 2.5: Auxiliary values).
If you wish to have a second potential evaluated, proceed as described
for potential 1.
Notes for Potential 1/2
1. If you have selected potential 1 and/or 2, the titrator does not test all
termination conditions– except maximum volume – until the potentials
are reached (see Section 3.3.6.6).
2. When you select the two potentials, you also define the tendency
(from P1 to P2) and the order: If the titrator finds P2 first, it will no longer
search for P1.
Stop for
reevaluation
The EQP titration function is interrupted immediately before its completion
if the termination conditions and the condition defined here are met.
This allows you to modify parameters of the equivalence point recognition
(all) and the evaluation (potential 1 and 2). All evaluations are performed
again using the modified parameters (see Section 5.2.2).
– Press <F4>, select "Yes" in the mask that appears then press <F4>
again:
Reevaluation
METHOD
Condition
neq=0
Esc Symbol Modify OK
The condition "neq = 0" means that the EQP titration function is interrupted
immediately before its completion if no equivalence point has been
found.
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 3-35
Method
Stop for
reevaluation
Modify
Use <F4> to select a different condition in the list that appears:
neq > 1: If more than one,
neq < 2: less than two, or
3 > neq > 0: one or two equivalence points are found,
the titration will always be interrupted.
Symbol
Press <F3> to show the list of symbols (see Section 8.1). You can select
symbols from this list to define your own condition, e.g. VEQ > 0.6:
- Select "neq > 1" (under "Modify" command)
- delete "neq"
- select "VEQ" from the symbol list
- then position cursor on "1" and replace by "0.6".
3-36 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Method
EP titration
3.3.7 EP titration (end point titration)
Here you define the control and evaluation of an end point titration and thus titrate to a particular
value of the selected measurement unit.
NOTICE
Before an end point titration to a particular pH value, you should calibrate the pH
electrode used!
Titrant/Sensor
Predispensing
Titrant addition
End point
Tendency
Termination
The titrator determines several measured values and volumes as raw results (see Section 8.1:
List of symbols and Section 3.3.4: Measure function).
Note: If you have selected a temperature sensor under the Sample function, the temperature
of the sample solution is measured automatically before the titrator executes this
function. If you have not attached a sensor (parameter "Manual"), the titrator adopts the
temperature entered at the start of the method. The slope of the pH electrode is
temperature-corrected by this measured or entered value and incorporated in the
calculation of the pH value.
3.3.7.1 Titrant/Sensor
To modify the titrant or sensor, press <F4>:
Titrant
NaOH
Concentration [mol/L] 0.1
Sensor
DG111
Unit of meas.
mV
Titrant
Concentration
Sensor
Press <F4> to open the list containing the titrants you have defined in the
Setup menu (see Section 2.1). Confirm the one you want with OK.
The defined concentration of the titrant is entered automatically.
Press <F4> to open the list containing the sensors you have defined in
the Setup menu (see Section 2.2). Confirm the one you want with OK.
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 3-37
EP titration: Predispensing
Method
Unit of measurement
The defined unit of measurement is entered automatically.
When you select the unit from the selection menu, you should select mV
or the unit defined in the Setup menu, otherwise you will receive an error
message when the method is started.
3.3.7.2 Predispensing
Predispensing shortens the titration time. You can select one or two predispensing modes
when you press <F4>.
Predispensing
METHOD
to volume
to (factor x sample size)
No
Esc Modify
Volume
Factor x
sample size
No
You dispense to a specified volume [mL] that you can enter when you
press <F4>. Instead of a number, you can enter a formula.
You can also enter a wait time: After the predispensing, the titrator waits
for this time to elapse before it adds the titrant in a controlled manner.
You dispense to a specified volume that is calculated from the product of
the sample weight or volume and a factor; you can enter this factor if you
press <F4>.
You can also enter a wait time (see “Volume”).
You do not wish to predispense.
In the subsequent continuous titrant addition, the titrator dispenses the volume in one step. It
acquires the potential values ET1 and ET2 (see Section 3.3.6.2).
In the subsequent dynamic titrant addition, it dispenses the volume in three steps (4/ 7, 2/ 7,
1/ 7 of the defined volume), which allows optimum calculation of the addition increment. It
acquires the potential values ET1 and ET2.
3-38 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Method
EP titration: Titrant addition
3.3.7.3 Titrant addition
You can select a dynamic or continuous addition mode:
Titrant addition
METHOD
Dynamic
Continuous
Esc Info Modify
Press <F3> for general information on what addition mode you should select:
• In general, select the dynamic addition mode for acid/base titrations in aqueous media,
argentometric and redox titrations,
• the continuous mode for acid/base titrations in nonaqueous media.
Dynamic
For dynamic titrant addition in the EP titration, the same conditions apply
as in dynamic addition in the EQP titration (see Section 3.3.6.3). The
measured values are acquired with equilibrium control as in the EQP
titration (see Section 3.3.6.4).
∆E(set) [mV] 8.0
∆V(min) [mL] 0.02
∆V(max) [mL] 0.15
∆E [mV] 1.0
∆t [s] 1.0
t(min) [s] 2.0
t(max) [s] 10.0
dynamic
titrant addition
equilibrium
controlled
measured value
acquisition
Help (titrant addition)
If you have selected the first three parameters and press <F3>, you can
adopt suggested values for the dynamic addition from the mask that
appears.
• For steep titration curves, a smaller value of ∆E must be selected than
for flat titration curves.
• For titration curves that exhibit a sudden jump, small values for ∆E and
∆V(max) should be selected.
In the absence of predispensing, the titrator dispenses the first two
volume increments with ∆V(min).
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 3-39
EP titration: Titrant addition
Method
Dynamic
Help (measured value acquisition)
If you have selected one of the last four parameters and press <F3>, you
can adopt suggested values for the equilibrium condition from the mask
that appears:
• Fast titrations are, e.g. acid/base titrations in aqueous media,
• slow titrations are, e.g. precipitation titrations in nonaqueous media.
The equilibrium conditions apply only in the end point range defined by
the following formula:
EP range = EP ± {1.5 * ∆E(set)}.
Example: If the end point is -30 mV, the condition holds for above values
of -15 to -45 mV.
Outside this range: ∆E (outside) = 4 * ∆E holds.
If the end point is reached: ∆E = ∆E/2 holds.
Continuous
Continuous
Control band [mV, pH,...] 100.0
∆V(min) [mL] 0.01
METHOD
Esc Help OK
The titrator dispenses the titrant slowly at the start then at the maximum
rate until the defined control band. Within the control range, the dispensing
rate decreases exponentially. In the vicinity of the end point, the
titrator adds the increment ∆V(min) (the smallest increment that the
titrator can dispense is 1/5 000 of the burette volume).
E [mV, pH]
Start of the control range
+200
+100
4
5
6
Control band = 250 mV
0 7
(4.3 pH)
8
-100 End point
V [mL]
3-40 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Method
EP titration: End point
Continuous
Help
When you press <F3> you can adopt suggested values for the continuous
addition from the mask that appears (confirm with OK).
• For titration curves with a steep control range, the control band must be
larger than that for the flat control range.
For ∆V(min) you can enter a formula instead of a number.
The titrator measures the initial potential ET1 after 1 - 3 seconds and for
the table of measured values it acquires a measured value every 5
seconds.
3.3.7.4 End point
You have a choice between two end point titrations.
End point
METHOD
EP absolute
EP relative
Esc Modify
EP absolute
EP relative
The absolute end point is the value on the electrode signal scale referred
to zero.
– Press <F4> and enter the potential [mV, pH, ...] in the mask that
appears.
Instead of a number, you can enter a formula or call up a potential stored
as an auxiliary value.
Example: You have stored the value of "EPOT" of H 3 PO 4 for the 1st
equivalence point as H4. If you wish to titrate to this potential,
enter H4.
The relative end point is the difference between the electrode signal at the
start of the titration and that at the end.
– Press <F4> and enter the potential [mV, pH, ...] in the mask that
appears.
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 3-41
EP titration: Tendency/Termination
Method
3.3.7.5 Tendency
A titration exhibits a positive or negative mV or pH change (see diagram). You have to define
this.
– Press <F4> and select "Positive" or "Negative" in the mask that appears (press <F4>).
NOTICE
A positive mV change means a negative pH change!
E [mV – pH]
+ 300
+ 200
+ 100
3
4
5
6
Start potential
Tendency positive (mV)
negative (pH)
Tendency negative (mV)
positive (pH)
EP relative = 310 mV (pH 5.2)
0
7
End potential
8
EP absolute = - 30 mV (pH 7.5)
V [mL]
3.3.7.6 Termination
You define the termination of a titration by selecting two parameters.
Termination
METHOD
Maximum volume [mL] 10.0
Delay [s] 10
Esc
OK
Maximum volume You must define the volume. It is intended as a safety precaution: if the
titration is faulty, an excess amount of titrant is not added unnecessarily.
Delay
is the time from attainment of the end point up to definitive termination of
the titration. If the measured value of the end point decreases within the
specified time, the titrator adds additional increments.
3-42 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Method
Learn titration
3.3.8 Learn titration
If you are not sure what parameters you should enter for the titration function, select the
standard method Learn titration and define the parameters for the titrant and sensor under the
Learn titration function.
Note: The titrator always titrates the entire volume of the burette used. You thus have to select
weight/volume of your sample in accordance with the burette volume.
Learn titration
METHOD
Titrant/Sensor
Esc Modify OK
To modify the titrant or sensor, press <F4>:
Titrant
NaOH
Concentration [mol/L] 0.1
Sensor
DG111
Unit of meas.
mV
Titrant
Concentration
Sensor
Unit of measurement
Press <F4> to open the list containing the titrants you have defined in the
Setup menu (see Section 2.1). Confirm the one you want with OK.
The defined concentration of the titrant is entered automatically.
Press <F4> to open the list containing the sensors you have defined in
the Setup menu (see Section 2.2). Confirm the one you want with OK.
The defined measurement unit is entered automatically.
If you select the unit from the selection menu, you should select mV or the
unit defined in the Setup menu. Otherwise you will receive an error
message when the method is started.
When you start the method, the titrator executes one function after another. As soon as it has
completed the Learn titration function, it calculates the parameters from the response
behavior of the sensor, the shape of the titration curve and with consideration of the burette
volume.
It stores these as an EQP titration function and sends the data to the printer. While the data
are being printed, the titrator executes the remaining functions: The "Learn titration" has given
rise to an equivalence point titration!
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 3-43
Learn titration
Method
Printout of a learned EQP titration (example: acid determination with NaOH)
LEARN TITRATION
Method 12 Learn titration
Version 28-Oct-1996 14:36
EQP titration
Titrant/Sensor
Titrant ............................... NaOH
Concentration [mol/L] ................. 0.1
Sensor ................................ DG111
Unit of meas. ......................... pH
Predispensing ............................. to volume
Volume [mL] ........................... 1.0
Wait time [s] ......................... 0
Titrant addition .......................... Dynamic
∆E(set) [mV] .......................... 12.0
∆V(min) [mL] .......................... 0.08
∆V(max [mL] ........................... 0.4
Measure mode .............................. Equilibrium controlled
∆E [mV] ............................... 1.0
∆t [s] ............................... 1.0
t(min) [s] ............................ 2.0
t(max) [s] ............................ 20.0
Recognition
Threshold ............................. 2.0
Steepest jump only .................... No
Range ............................... No
Tendency .............................. None
Termination
at maximum volume [mL] ................ 10.0
at potential .......................... No
at slope .............................. No
after number EQPs ..................... Yes
n = ................................ 1
comb. termination conditions .......... No
Evaluation
Procedure ............................. Standard
Potential 1 ........................... No
Potential 2 ........................... No
Stop for reevaluation ................. No
• If the titrator finds more than one equivalence point, it calculates the parameters for the one
with the steepest jump.
• If the titrator does not find any equivalence points, it interrupts the method and an error
message appears (see Section 5.2.2).
• The titrator first recognizes a possible equivalence point when this lies at the sixth measured
point (see Section 3.3.6.5).
• As an EQP titration emerges immediately from the Learn titration function, you can titrate a
series of samples with the same content. If you have selected "Statistics" under the Calculation
function, the result of the first sample will then be discarded!
3-44 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Method
EQP titration (Ipol/Upol)
3.3.9 EQP titration (Ipol/Upol)
Here you define the control and evaluation of a voltametric or amperometric equivalence point
titration. The equivalence point is that point at which exactly the same number of equivalents
of titrant and analyte have reacted. In most cases, it is virtually identical to the inflection point
of the titration curve. This inflection point is recognized and the equivalence point calculated.
Titrant/Sensor
Indication
Predispensing
Titrant addition
Measure mode
Recognition
Termination
Evaluation
The titrator determines several measured values and volumes as raw results (see Section 8.1:
List of symbols and Section 3.3.4: Measure function).
3.3.9.1 Titrant/Sensor
To modify the titrant or sensor, press <F4>:
Titrant/Sensor
METHOD
Titrant 1 / 2 I 2
Concentration [mol/L] 0.1
Sensor
DM142
Esc Modify OK
Titrant
Concentration
Sensor
Press <F4> to open the list containing the titrants you have defined in the
Setup menu (see Section 2.1). Confirm the one you want with OK.
The defined concentration of the titrant is entered automatically.
Press <F4> to open the list containing the sensors you have defined in
the Setup menu (see Section 2.4.). Confirm the one you want with OK.
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 3-45
EQP titration (Ipol/Upol): Indication
Method
3.3.9.2 Indication
To select the voltametric or amperometric indication, press <F4>:
Indication
METHOD
Voltametric
Amperometric
Esc Modify
Voltametric
The potential difference between two metal electrodes is measured by
polarizing them with a constant current (in the case of the DM142 electrode,
the two pins of the platinum electrode are polarized).
– Press <F4> to enter the current Ipol. Its value depends among other
things on:
• the electrode itself (e.g. distance between the platinum pins in the
case of the DM142 electrode)
• the dissolved substance to be determined and its concentration
• the solvent
• the reaction with the titrant.
You can obtain an initial starting point for the current intensity to be
selected by measuring the corresponding solutions with different current
intensities (see Section 6.1.4) and recording the voltage value. This
should lie within the measurement range ±1500 mV for both the initial
and the equivalence point potential (see Section 12.1: Technical Data).
E [mV]
Example of a vitamin C determination in a multivitamin drink
with DPI [0.01 mol/L]
V [mL]
3-46 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Method
EQP titration (Ipol/Upol): Indication
Amperometric
The potential difference between two metal electrodes is measured by
polarizing them with a constant voltage (in the case of the DM142 electrode,
the two pins of the platinum electrode are polarized).
– Press <F4> to enter the voltage Upol. Its value depends on the same
factors mentioned under the parameter "Voltametric".
You can obtain an initial starting point for the voltage to be selected by
measuring the corresponding solutions with different voltage values (see
Section 6.1.4) and recording the current intensity. This should lie within
the measurement range ±150 µA for both the initial and the equivalence
point potential (see Section 12.1: Technical Data).
E [µA]
Example of a vitamin C determination in a multivitamin drink
with DPI [0.01 mol/L]
V [mL]
Note: Higher current or voltage values result in higher jumps (a steeper curve) in the equivalence
point region, however the platinum pins of the electrode become contaminated
quicker.
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 3-47
EQP titration (Ipol/Upol): Predispensing/Titrant addition
Method
3.3.9.3 Predispensing
Predispensing shortens the titration time. You can select one of two predispensing modes
when you press <F4>.
Predispensing
METHOD
to ∆E volume ~ ∆E(set)
to (factor x sample size)
No
Esc
Change
Volume
Factor x
sample size
No
You dispense a specified volume [mL] which you can enter when you
press <F4>. Instead of a number, you can enter a formula.
You can also enter a wait time: After the predispensing, the titrator waits
for this time to elapse before it adds the titrant in a controlled manner.
You dispense to a specified volume that is calculated from the product of
the sample weight or volume and a factor; you can enter this factor when
you press <F4>.
You can also enter a wait time (see “Volume”).
You do not wish to predispense.
In predispensing to volumes or (sample size x factor), the titrator adds the titrant in three steps
(4/7, 2/7, 1/7 of the defined volume).
3.3.9.4 Titrant addition
You can enter only a constant increment for the volume (see Section 3.3.6.3: Incremental
titrant addition):
Titrant addition
METHOD
∆V [mL] 0.05
Esc
OK
3-48 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Method
EQP titration (Ipol/Upol): Measure mode
3.3.9.5 Measure mode
You select the equilibrium controlled or timed increment measured value acquisition:
Measure mode
METHOD
Equilibrium controlled
Timed increment
Esc Modify
These parameters are used to define the wait time up to measured value acquisition following
an incremental addition. In the equilibrium controlled measure mode, the wait time is variable,
in the timed increment it is constant. You can not decide the optimum measure mode for your
method until you know the reaction time of the components and the response time of the sensor
used.
Equilibrium
controlled
∆E [mV, µA] 0.5
∆t [s] 1.0
t(min) [s] 3.0
t(max) [s] 30.0
Before the titrator adds the next increment, an equilibrium must be
established in the solution; the measured value must stabilize.
The following are responsible for the equilibrium
• the potential change ∆E measured in the solution
within the defined time ∆t.
The wait time up to the next increment addition also depends on the
definition of the
• minimum time t(min) and
the maximum time t(max)
As soon as the potential change of the solution is less than the defined
equilibrium (∆E/∆t), the titrator acquires the measured value and adds the
next increment.
This can be at t(min) at the earliest and should be at t(max) at the latest.
At t(max) the measured value is in any case acquired, even if the equilibrium
condition is not yet met (see diagram in Section 3.3.6.4).
This measure mode results in fast increment addition in the flat part of the
titration curve and slow addition in the steep part.
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 3-49
EQP titration (Ipol/Upol): Recognition
Method
Timed increment
After every increment addition, the titrator allows the time you have
defined to elapse before it acquires the measured value (see diagram in
Section 3.3.6.4).
Timed increment
METHOD
∆t [s] 3.0
Esc
OK
For both measure modes, the titrator acquires the potential, the time and the titrant volume.
You can print out these measured values after the titration of a sample (see Section 3.3.16:
Report).
3.3.9.6 Recognition
Recognition of the equivalence point of a titration curve depends on the type of reaction of the
components and hence on the evaluation procedure (see Section 3.3.9.8):
1. With the evaluation procedures Maximum and Minimum, an equivalence point is recognized
when the greatest (smallest) potential value of the titration curve is greater (less) than
two preceding and two subsequent values.
2. With the evaluation procedures Standard and Asymmetric an equivalence point is recognized
when the maximum of the absolute values of the 1st derivative of the titration curve
is greater than two preceding and two subsequent values.
3. With the evaluation procedure Segmented, an equivalence point is recognized when the
maximum of the absolute values of the 2nd derivative of the titration curve is greater than
two preceding and two subsequent values.
Note: For the first recognition of a possible equivalence point, the titrator needs a certain number
of measured points (see Section 3.3.6.5).
You must or can support the equivalence point recognition using four parameters:
Threshold 10.0
Steepest jump only No
Range
No
Tendency
None
3-50 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Method
EQP titration (Ipol/Upol): Recognition
Threshold
To ensure that minor disturbances in the curve profile are not recognized
as equivalence points, you must determine a threshold value:
+/-E, ∆E/∆V oder ∆ 2 E/∆V 2 . This must be exceeded.
Notes
1. The threshold value should be maximum half as large as the expected
maximum value of the first or second derivative at the equivalence
point.
2. The threshold value depends on so many factors (solvent, concentration,
sensor, type of reaction, etc.) that you can not define the "correct"
threshold value until you have performed the first titration.
3. You will find examples of the threshold value for the different evaluation
procedures in Section 3.3.6.5.
Steepest jump
only
Range
You can also select this parameter for recognition of the equivalence
point (press <F4>): The titrator then recognizes only the steepest jump of
the titration curve.
You can also define a potential range for recognition of the equivalence
point. Equivalence points which lie outside this range are not recognized.
– Press <F4>, select "Yes" in the mask that appears and then press <F4>
again:
Range
METHOD
Limit A [mV, µA] 100
Limit B [mV, µA] 200
Esc
OK
Instead of a number, you can also enter a potential stored as an auxiliary
value or a formula.
The equivalence point range always refers to the titration curve, in other
words it is independent of the evaluation procedure.
Tendency
As a further aid in recognition of the equivalence point, you can define the
tendency, namely in what part of the titration curve – descending or
ascending – the equivalence point should be located (see diagram on
page 3-32).
– Use <F4> to select the tendency: "None", "Positive" or "Negative".
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 3-51
EQP titration (Ipol/Upol): Termination
Method
3.3.9.7 Termination
You can define how a titration should be terminated by selection of five different parameters.
The titrator can terminate the titration either
• when the first of the selected conditions is met, or
• when all the selected conditions are met.
• An exception is the maximum volume: when this is reached, the titration is immediately
terminated!
at maximum volume [mL] 10.0
at potential
No
at slope
No
after number EQPs
No
comb. termination conditions No
Maximum volume You must enter the maximum volume. It is intended as a safety precaution:
if the titration is faulty, an excessive amount of titrant is not dispensed
unnecessarily as the titration is always terminated.
Potential
The titrator terminates the titration at the defined potential.
– Press <F4>, select "Yes" in the mask that appears then press <F4>
again:
At potential
METHOD
Potential [mV, µA] 0.0
Esc
OK
Instead of a number, you can also enter a potential stored as an auxiliary
value or a formula.
Slope
The titrator terminates the titration when the slope of the titration curve is
less than a certain value. The measured slope must exceed this absolute
value once and drop below it twice to terminate the titration.
– Press <F4>, select "Yes" and press <F4> again: Enter the slope
[mV/mL] or [µA/mL] in the mask that appears.
3-52 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Method
EQP titration (Ipol/Upol): Evaluation
Number EQPs
Comb. termination
criteria
The titrator terminates the titration after a certain number n of equivalence
points has been found.
– Press <F4>, select "Yes" and press <F4> again: Enter the number in
the mask that appears.
The titrator does not terminate the titration until all defined termination
conditions are met (exception: maximum volume, see above).
– Select "Yes" with <F4>.
3.3.9.8 Evaluation
You determine which evaluation procedure is used to calculate the equivalence points found
and select potential values you wish to have evaluated.
Procedure
Potential 1
Potential 2
Stop for reevaluation
Standard
No
No
No
Procedure
Use <F4> to select the procedure among several calculation modes
which matches the titration curve (see Section 8.3).
Standard Evaluation procedure for all S-shaped titration curves;
can be used for curves with steep jumps in titrations with
voltametric and amperometric indication.
Minimum Determination of the minimum of a titration curve
Maximum Determination of the maximum of a titration curve
Segmented Evaluation procedure for titration curves with segments
(segmented curve); can be used for titrations with amperometric
indication.
Asymmetric Evaluation procedure for S-shaped, highly asymmetric
titration curves.
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 3-53
EQP titration (Ipol/Upol): Evaluation
Method
Procedure
Notes
1. In an evaluation with the standard procedure is not possible, the
equivalence point is calculated by interpolation (see Section 8.3.1).
If an evaluation is not possible with the asymmetric procedure, the
equivalence point is calculated by the standard procedure or by
interpolation (see Section 8.3.4).
In both cases, the evaluation procedure will be documented for each
sample in the "raw results" report if it differs from the defined procedure.
If you have to work according to GLP guidelines, you should
therefore always record the raw results (see Section 3.3.16).
2. If you determine the same sample with a voltametric and amperometric
EQP titration, in many cases you must change not only the
parameters of the indication, measured value acquisition and recognition,
but also the evaluation procedure. An example is the vitamin C
determination in a multivitamin drink with DPI [0.01mol/L]:
Voltametric titration: Standard
Amperometric titration: Segmented
E [mV]
E [µA]
E-V curve
E-V curve
∆E/∆V-V curve
∆E/∆V-V curve
∆ 2 E/∆V 2 -V curve
∆ 2 E/∆V 2 -V curve
V [mL]
V [mL]
3-54 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Method
EQP titration (Ipol/Upol): Evaluation
Procedure
3.If you obtain a so-called Z curve in an equivalence point determination
(see diagram), we recommend performing an end point titration.
E [mV]
Example of a titer determination of iodine solution [0.1 mol/L]
with L-ascorbic acid (vitamin C)
V [mL]
Potential 1
Potential 2
In the selection of potentials, the titrator evaluates not only the equivalence
points, but also the titrant consumption in mmol or mL required for
attainment of these potential values (see Section 8.1: List of symbols).
– Press <F4>, select "Yes" and press <F4> again: Enter the first potential
[mV, µA] in the mask that appears.
Instead of a number, you can also enter a potential stored as an
auxiliary value or a formula (see Section 2.5: Auxiliary values).
If you wish to have a second potential evaluated, proceed as described
for potential 1.
Notes for Potential 1/2
1. If you have selected potential 1 and/or 2, the titrator does not test all
termination conditions– except maximum volume – until the potentials
are reached (see Section 3.3.9.7).
2. When you select the two potentials, you also define the tendency
(from P1 to P2) and the order: If the titrator finds P2 first, it will no
longer search for P1.
Stop for
reevaluation
The EQP titration (Ipol/Upol) function is interrupted immediately before
its completion if the termination conditions and the condition defined here
are met (see Section 3.3.6.7).
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 3-55
EP titration (Ipol/Upol)
Method
3.3.10 EP titration (Ipol/Upol)
Here you define the control and evaluation of an end point titration with voltametric or amperometric
indication.
Titrant/Sensor
Indication
Predispensing
Titrant addition
End point
Tendency
Termination
The titrator determines several measured values and volumes as raw results (see Section 8.1:
List of symbols and Section 3.3.4: Measure function).
3.3.10.1 Titrant/Sensor
To modify the titrant or sensor, press <F4>:
Titrant/Sensor
METHOD
Titrant 1 / 2 I 2
Concentration [mol/L] 0.1
Sensor
DM142
Esc Modify OK
Titrant
Concentration
Sensor
Press <F4> to open the list containing the titrants you have defined in the
Setup menu (see Section 2.1). Confirm the one you want with OK.
The defined concentration of the titrant is entered automatically.
Press <F4> to open the list containing the sensors you have defined in
the Setup menu (see Section 2.4). Confirm the one you want with OK.
3-56 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Method
EP titration (Ipol/Upol): Indication
3.3.10.2 Indication
Press <F4> to select the voltametric or amperometric titration :
Indication
METHOD
Voltametric
Amperometric
Esc Modify
Voltametric
The potential difference between two metal electrodes is measured by
polarizing them with a constant current (in the case of the DM142
electrode, the two pins of the platinum electrode are polarized).
– Press <F4> to enter the current Ipol. Its value depends among other
things on:
• the electrode itself (e.g. distance between the platinum pins in the
case of the DM142 electrode)
• the dissolved substance to be determined and its concentration
• the solvent
• the reaction with the titrant.
You can obtain an initial starting point for the current intensity to be
selected by measuring the corresponding solutions with different current
intensities (see Section 6.1.4) and recording the voltage value. This
should lie within the measurement range ±1500 mV for both the initial
and the end point potential (see Section 12.1: Technical Data).
E [mV]
Example of a Karl Fischer titration with Hydranal
(2-component system)
V [mL]
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 3-57
EP titration (Ipol/Upol): Indication
Method
Amperometric
The potential difference between two metal electrodes is measured by
polarizing them with a constant voltage (in the case of the DM142 electrode,
the two pins of the platinum electrode are polarized).
– Press <F4> to enter the voltage Upol. Its value depends on the same
factors mentioned under the parameter "Voltametric".
You can obtain an initial starting point for the voltage to be selected by
measuring the corresponding solutions with different voltage values (see
Section 6.1.4) and recording the current intensity. This should lie within
the measurement range ±150 µA for both the initial and the end point
potential (see Setion 12.1: Technical Data).
E [µA]
Example of a Karl Fischer titration with Hydranal
(2-component system)
V [mL]
3-58 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Method
EP titration (Ipol/Upol): Predispensing
3.3.10.3 Predispensing
Predispensing shortens the titration time. You can select one or two predispensing modes
when you press <F4>.
Predispensing
METHOD
to volume
to (factor x sample size)
No
Esc Modify
Volume
Factor x
sample size
No
You dispense to a specified volume [mL] that you can enter when you
press <F4>. Instead of a number, you can enter a formula.
You can also enter a wait time: After the predispensing, the titrator waits
for this time to elapse before it adds the titrant in a controlled manner.
You dispense to a specified volume that is calculated from the product of
the sample weight or volume and a factor; you can enter this factor if you
press <F4>.
You can also enter a wait time (see "Volume").
You do not wish to predispense.
With the predispensing, the titrator dispenses the volume in one step.
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 3-59
EP titration (Ipol/Upol): Titrant addition/End point
Method
3.3.10.4 Titrant addition
The addition can only be continuous (see Section 3.3.7.3: Continuous titrant addition):
Titrant addition
METHOD
Control band [mV, µA] 500.0
∆V(min) [mL] 0.01
Esc
OK
The titrator dispenses the titrant slowly at the start then at the maximum
rate until the defined control band. Within the control range, the dispensing
rate decreases exponentially. In the vicinity of the end point, the
titrator adds the increment ∆V(min) (see diagram in Section 3.3.7.3); the
smallest increment that the titrator can dispense is 1/5 000 of the burette
volume.
For ∆V(min) you can enter a formula instead of a number.
The titrator measures the initial potential ET1 after 1 - 3 seconds and for
the table of measured values it acquires a measured value every 5
seconds. The titrator can store maximum 300 measured values. If this
limit is exceeded, it reduces the number of measured values by half by
leaving only every second measured value in the memory. The new
measured values continue to be acquired every 5 seconds.
Note: The control band determines the control and the speed of the end point titration. To enter
a "reasonable" value, you should perform an equivalence point titration with timedincrement
measured value acquisition. Using the table of measured values of this
titration, you can determine both the end point potential and the control band exactly.
3.3.10.5 End point
You enter the end point for the titration.
End point
METHOD
Potential [mV, µA] 100.0
Esc
OK
Instead of a number, you can enter a formula or call up a potential stored
as an auxiliary value.
3-60 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Method
EP titration (Ipol/Upol): Tendency/Termination
3.3.10.6 Tendency
A titration exhibits a positive or negative mV or µA change. You have to define this.
Tendency
METHOD
Tendency
Negative
Esc Modify OK
Note: The tendency of a voltametric titration with a positive current intensity is negative, with
a negative current intensity it is positive. The tendency of an amperometric titration with
a positive voltage is positive, with a negative voltage it is negative.
3.3.10.7 Termination
You define the termination of a titration by selecting two resp. three parameters.
Termination
Maximum volume [mL] 10.0
Delay [s]
Yes
Maximum time [s] No
Esc
METHOD
OK
Maximum volume You must define the volume. It is intended as a safety precaution: if the
titration is faulty, an excess amount of titrant is not added unnecessarily.
Delay
Maximum time
is the time from attainment of the end point up to definitive termination of
the titration. If the measured value of the end point decreases within the
specified time, the titrator adds additional increments.
Instead of or in addition to the delay time, you can also enter a time when
the titration should be stopped.
Enter a maximum time when, e.g. you perform a stating (see METTLER
method 90016: Drift determination for KF titrations).
If you define a value for both termination parameters, the titration will be
terminated when one of the two values is reached.
Note: If the titration is terminated because the maximum time has been
reached, no end point will be calculated. VEND (VEX) is recorded
as the raw result.
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 3-61
pH/mV-stat
Method
3.3.11 pH/mV-stat
With the aid of this function you can perform a pH-stating. You determine the parameters which
assure the constancy of a potential value, which lead to termination of the pH-stating and which
are used for the evaluation.
Titrant/Sensor
Pretitration
End point
Tendency
Termination
Storage interval
Evaluation
The titrator determines the following raw results (see also 8.1: List of symbols):
• the titrant consumption in mL or mmol up to termination of the pH-stating.
• the titrant consumption in mL or mmol up to attainment of defined time limits.
• the mean titrant consumption in mL/min or mmol/min within the defined time limits.
• the correlation coefficient CSTAT, which results from calculation of the mean titrant
consumption through linear regression.
Notes
1. If you have selected a temperature sensor under the Sample function, the temperature of
the sample solution is measured automatically before the titrator executes this function. If
you have not attached a sensor (parameter "Manual"), the titrator adopts the temperature
entered at the start of the method. The slope of the pH electrode is temperature-corrected
by this measured or entered value and incorporated in the calculation of the pH value.
2. In the recording of a titration curve, you obtain only the actual pH-stating. You can follow
the graphical plot of the pretitration only on the display.
Titrant/Sensor
To modify the titrant or the sensor, press <F4>:
Titrant
NaOH
Concentration [mol/L] 0.1
Sensor
DG111
Unit of meas.
mV
3-62 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Method
pH/mV-stat
Titrant
Concentration
Sensor
Unit of measurement
Press <F4> to open the list containing the titrants you have defined in the
Setup menu (see Section 2.1). Confirm the one you want with OK.
The defined concentration of the titrant is entered automatically.
Press <F4> to open the list containing the sensors you have defined in
the Setup menu (see Section 2.2). Confirm the one you want with OK.
The defined measurement unit is entered automatically.
If you select the unit from the selection menu, you should select mV or the
unit defined in the Setup menu, otherwise you will receive an error message
when the method is started.
Pretitration
Select the pretitration if the potential value of your solution differs from the value you need for
the pH-stating:
– Press <F4> and select "Yes" in the mask that appears.
– Press <F4> and enter the control band [mV, pH, ...] in the mask that appears (see Section
3.3.7.3: EP titration, Titrant addition).
End point
End point
Potential [mV, pH,...] 8
Control band [mV, pH,...] 0.7
METHOD
Esc
OK
Potential
Control band
The value you enter here is kept constant during the pH-stating. Instead
of a number, you can enter a formula or call up a potential stored as an
auxiliary value (see example in Section 3.3.7.4).
The defined value controls the end point range: The lower the value, the
faster the titrator reacts to a deviation from the potential of the defined end
point.
Help
You can accept suggested values for the value of the control band: Press
<F3> and confirm a value in the mask that appears with OK.
• For pH-statings whose end point lies in a steep control range, the
control band must be larger than for statings whose end point lies in a
flat control range.
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 3-63
pH/mV-stat
Method
Tendency
The tendency you have to define refers to the titrant which is added. The kinetics of the reaction
determined by pH-stating are always in the opposite direction!
– Press <F4> and select "Positive" or "Negative" in the mask that appears.
Termination
You define the termination of the pH-stating by the following parameters:
Maximum volume [mL] 10
t(min) [s] 60
t(max) [s] 600
Minimum consumption [mL] 0.1
Time span [s] 10
Maximum volume The entry is intended as a safety precaution: if the titration is faulty, an
excess amount of titrant is not added unnecessarily.
t(min)
t(max)
is the earliest time a termination can take place; it depends on the
minimum consumption within the defined time range.
is the time when the pH-stating should be terminated.
Min. consumption The stating is terminated when not more than 0.1 mL titrant are consumed
Time span within 10 seconds (example of default parameters).
This condition can be effective at the earliest after 60 seconds [t(min)].
After 600 seconds [t(max)], the stating is terminated even if the condition
is not met.
Storage interval
You define a time interval for storage of the measurement data.
– Press <F4> and enter the interval in the mask that appears.
With the default value, the current measured value and the associated volume are stored after
10 s.
The titrator can store and print out maximum 300 measured values. If the defined interval is
too small, after 300 measured values the titrator first eliminates the values of the pretitration.
If the limit of 300 measured values is then again exceeded, it reduces their number by half by
leaving only every second measured value in the memory. The measured values continue to
be stored at the defined interval.
3-64 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Method
pH/mV-stat
Evaluation
The following parameters define the evaluation of the pH-stating:
Evaluation
METHOD
Time limit t1 [s] 100.0
Time limit t2 [s] 200.0
Esc
OK
Time limit t1
Time limit t2
The titrant consumption is calculated up to the defined times t1 and t2
(see Section 8.1: List of symbols).
The time limits are not restricted to the values t1 and t2 entered here. In
one of the additional Calculation functions, you can calculate, e.g.
• the mean titrant consumption within different time limits with
R = QSTAT (100, 200).
• the titrant consumption up to a different time limit with
R = VT (200).
• the correlation coefficient within a different time limit with
R = CSTAT (300, 600).
In this manner you can also define additional results with new time limits
following a pH-stating (see Section 6.5.3: Perform calculations).
Example of the progress of a pH-stating with time
∆V/∆t [mL/s]
3
1: Pretitration
2: When the end point is reached, the titrator
waits 5 s before sending an audio signal and
the message "Pretitration complete: Please
add sample!". The titrator continues to exert
control and stir.
3: You enter your sample and confirm the
message with OK: The titration starts. If you
have to remove the titration vessel for addition,
you can stop the stirrer during this time
(see Section 6.2)
4: The termination condition of ∆V/∆t is met,
the pH-stating is terminated.
2
1
t(min)
t(max)
4
t [s]
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 3-65
Calculation
Method
3.3.12 Calculation
You can use this function to calculate one result R for every sample.
Formula
R=Q*C/m
Constant
C=M/(10*z)
Decimal places 3
Result unit %
Result name
Statistics
No
Formula
Constant
Press <F4> to open the "List of formulas" and select a different formula
or enter the one you need for your calculation (see also Section 8.4:
Examples of formulas).
Press <F4> to open the "List of constants" and select a different constant
or enter the one you need for your calculation.
Symbol
You can select the symbols you need for your calculation from the list that
appears by pressing <F3>. It contains all symbols possible for the
individual functions (see also Section 8.1: List of symbols).
You can link together all available parameters and numeric values.
Example illustrating the entry of the formula "R=QSTAT(120,180)":
- Press <F4> and accept "R=VSTAT(100,200)",
- position the cursor on "V", delete it and select "Q" from the list,
- replace "100" by "120",
- skip the comma with the cursor and
- replace "200" by "180".
The following operations are available for calculations:
on the keypad
in "List of formulas"
• Addition: + • Logarithmic function to base 10: lg(x)
• Subtraction: – • Logarithmic function to base e: ln(x)
• Multiplication: * • Exponential function to base 10: pw(x)
• Division: / • Exponential function to base e: ex(x)
• Squaring function:
sq(x)
• Square root function:
sr(x)
3-66 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Method
Calculation
Decimal places
Result unit
Result name
Statistics
Enter the number of decimal places you require in the result.
Press <F4> to select the unit from the selection menu.
You can enter the name with an attached keyboard.
Use <F4> to select whether a statistical calculation should be performed.
"Yes": in the analysis of a sample series the mean value x, the standard
deviation s and the relative standard deviation srel are calculated and
recorded.
Notes
1. Some standard methods include three Calculation functions. If you need only one result,
delete the formula "Rn = ..." with CE. This skips the function when the method is processed.
2. For the calculation of a titration
curve with 3 equivalence points
(acidic mixture of 3 substances),
you must define the parameters
of the three Calculation functions.
E
R3
R2
R1
Q1
Q2
Q3
V
Formula Constant Unit
1st Calculation function: R1 = Q1∗ C1 / m C1 = M / (10 ∗ z) [%]
2nd Calculation function: R2 = Q2 ∗ C2 / m C2 = (60.01 ∗ 1000) / 1 [ppm]
3rd Calculation function: R3 = Q3 ∗ C3 / m C3 = 53.5 / 1 [mg/g]
Another example is the METTLER method 90001: In addition to the NaOH consumption in
mL as a result, the acid content of HCl in mol/L and in g/L are calculated.
3. R and C are indexed according to the preset sequence. If you delete, e.g. the formula
"R2 =...", 3 remains as the index for the third calculation if you do not change it yourself (see
Section 8.2: Use of indices).
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 3-67
Calculation
Method
4. You must enter the molar mass M and the equivalent number z of the 2nd and 3rd
Calculation function as numeric values or insert as Hj if you have stored them as auxiliary
values (see Section 2.5): M and z are defined in the Sample function for the calculation
of the first equivalence point (see Section 3.3.2).
5. If you do not know the number of equivalence points, you can have Q recognized by a
condition instead of defining indices for Q (see "List of formulas").
Example: R = Q (200 < EPOT < 300) means that the Q used for the calculation is that whose
equivalence point potential lies between 200 and 300 mV. If the titrator detects more than
one equivalence point in this range, it calculates the first (see also Section 8.4.3).
6. The titrator checks your entries when you quit the parameter mask with OK. If you have
entered wrong formulas or constants, you immediately receive an appropriate error
message, e.g.
Error No. 3
Wrong formula
Modify
Terminate
METHOD
OK
Modify: You can change the entry.
Terminate: The entry last stored reappears.
If you have used raw results (e.g. R = QDISP) which the titrator can not determine as the
method does not include the Dispense function, you do not receive an error message until
the method is stored.
7. The titrator stores results until you start a new sample series or switch off the titrator (see
also Section 8.5.6).
8. If you abort the method with Reset, an evaluation of the data obtained to date follows. If
you were to abort the titration sequence (see point 2), e.g. shortly before the second
equivalence point, you would receive result R1 but not R2 and R3 (see Section 5.2.4).
9. Under the auxiliary function Results, you can
• perform additional calculations after the titration of a sample (see Section 6.5.3)
• eliminate results (outliers) on completion of a sample series and thus modify the statistical
evaluation (see Section 6.5.5).
3-68 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Method
Calibration
3.3.13 Calibration
You can use this function to calibrate an electrode: its zero point and its slope are calculated.
You determine the buffers which you use for calibration of the sensor. The potential of the
buffer solutions is acquired using the Measure function (see example in the Quick Guide).
Notes
1. As the slope of an electrode depends on the temperature, it is important to incorporate the
temperature in the calibration.
• Before the Measure function, the temperature of the buffer solution is measured
automatically if you have attached a temperature sensor; otherwise, the temperature you
entered before the start of the calibration is used (see Sample function, Sections 3.3.2
and 5.1).
If you later measure the pH value of a solution at a different temperature, the slope of the
sensor is temperature-corrected by the titrator.
2. The calibration data (zero point, slope and calibration temperature) are entered automatically
with the date in the parameter mask of the sensor (see Section 2.2.2).
3. Depending on the number of buffer solutions measured, the titrator performs the following
calibration:
• With one buffer solution, it calculates the zero point of the sensor. The slope remains
unchanged.
• With several buffer samples, it calculates the zero point and the slope of the sensor by
means of a linear regression through the measured points.
You can modify the following parameters:
Sensor
DG111
Buffer type
pH (DIN/NIST)
Result R 1
Minimum slope [mV/unit] -55.0
Maximum slope [mV/unit] -65.0
Sensor
Buffer type
Press <F4> to open the list containing the sensors you have defined in
the Setup menu (see Section 2.2). Confirm the one you want with OK.
Press <F4> and select the type in the mask that appears:
• pH (DIN/NIST),
• pH (MERCK),
• pH (METTLER TOLEDO) or
• pH, pM, pX.
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 3-69
Calibration
Method
Buffer type
The titrator has 8 values each for DIN/NIST, MERCK Titrisol and METTLER
TOLEDO buffer solutions stored which you can select to calibrate pH
electrodes. Under "pH, pM, pX (free selection)" enter values of userselected
buffer solutions to calibrate pH or ion-selective electrodes (see
Note 2).
Example illustrating selection of pH values of a MERCK buffer. Press
<F4> to display the selection menu:
pH (MERCK Titrisol)
METHOD
First buffer pH 4
Second buffer pH 4
Third buffer pH 4 ∇
Esc Modify OK
– If your first buffer solution does not have a pH value of 4, press <F4>
(otherwise, select 2nd buffer):
MERCK buffers
pH 3
pH 4
pH 5
pH 6
Esc
METHOD
OK
∇
– Select a different buffer value from this list, e.g. pH 3 and confirm with
OK.
– In the "pH (MERCK Titrisol)" mask select the 2nd buffer and again
press <F4> to select its pH value, e.g. pH 7 etc.
Notes
1. The buffer values apply to a temperature of
• 25 °C for DIN/NIST buffers
• 20 °C for MERCK Titrisol buffers
• 25 °C for METTLER TOLEDO buffers.
If you calibrate at a different temperature, these buffer values are
automatically temperature-corrected and recorded on the printout.
If you calibrate at a temperature that lies outside the defined range for
the corresponding buffer, "Temperature outside limits" appears in the
report; however, the calibration data will be recorded.
3-70 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Method
Calibration
Buffer type
2. For the pH, pM, pX buffer values you can also enter formulas.
3. Positive ions result in a negative slope, negative ions in a positive
slope.
Result R
Minimum slope
Maximum slope
"1" (result index): The measured values E of the buffer solutions used are
stored under the first Calculation function: R1 = E (see Section 8.2: Use
of indices). The titrator calculates the zero point and the slope of the
electrode by linear regression by assigning the measured values to the
standard concentrations of the buffer solutions. The index must thus be
the same as under the Calculation function.
Enter the minimum and maximum slope; the slope of an electrode is a
measure of its quality.
If you obtain a slope, e.g. with the pH buffer types in a pH calibration which
lies outside the limits you have defined, the calibration data are not
entered for the corresponding sensor. The error message "Data not
transferred" appears on the report.
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 3-71
Titer
Method
3.3.14 Titer
Titer is an assignment function: The result R or its calculated mean value x of the titration of
a titrant is assigned to the titer by t = R or t = x and automatically entered in the parameter
mask of the titrant with the date (see Section 2.1.2 and the example in the Quick Guide).
Titer
METHOD
Titrant
NaOH
Concentration [mol/L] 0.1
Formula t = x
Esc Modify OK
Titrant
Concentration
Formula t =
Press <F4> to open the list containing the titrants you have defined in the
Setup menu (see Section 2.1). Confirm the one you want with OK.
The defined concentration of the titrant is entered automatically.
Press <F4> to select from the selection menu:
"x", if you determine the titer with more than one sample so that the
calculated mean value is assigned to the titer. The statistical calculation
must be selected in the Calculation function for this.
"R" only if you determine the titer with a single sample.
Note: If you delete, e.g. one result of these titer determinations under the auxiliary function
Results, the new calculated mean value will not be entered in the parameter mask of
the titrant!
3-72 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Method
Auxiliary value
3.3.15 Auxiliary value
Auxiliary value is an assignment function. The result (R or Ri) or its calculated mean value
(x or x [k]) or a raw result of the titration function is assigned to the auxiliary value Hj and
entered automatically with the date in the parameter mask of the auxiliary value (see Section
2.5 and Section 8.2: Use of indices).
20 auxiliary value memories are available: H1 - H20.
You can store, e.g. the following as auxiliary values:
• the blank value of a titration (example of the standard method: "Blank by EQP titration")
• a raw result that you obtain under the EQP or EP titration function, e.g. "H6 = VEQ".
You can call up these auxiliary values in the Calculation function.
Auxiliary value
METHOD
ID
Formula
H20=R1
Esc
OK
ID
Formula
Enter an identification if desired.
Enter "H = R" or "Hj = Ri" if you determine the auxiliary value with a single
sample.
Enter "H = x " or "Hj = x [k]" only if you determine the auxiliary value with
more than one sample so that the calculated mean value is assigned to
the auxiliary value. The Statistics calculation must be selected in the
Calculation function for this.
You can also enter, e.g. "H (Hj) = VEND * 1.5" as a formula.
Symbol
If "Formula" is selected, you can open the list of symbols with <F3> (see
also Section 8.1). Select the symbol from this list which should be
assigned to the auxiliary value, e.g. for the formula "H6 = VEQ":
- Position the cursor on "2",
- replace "20" by "6",
- position the cursor on "R" and delete "R1",
- select "VEQ" from the symbol list.
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 3-73
Report
Method
3.3.16 Report
You use this function to determine what data should be recorded.
Note: If your method contains two Titration functions, e.g. standard method "2 Step titrations
(EQP)", a Report function also follows the first Titration function so that you can record,
e.g. the table of measured values and titration curve of the first function. The reason for
this is that the titrator stores only the measured values of the last Titration function.
The titrator stores the raw results of all functions up to the determination of the next
sample within a series (see Section 8.5.6).
Report
METHOD
Output Printer
Results
Yes
All results No ∇
Esc Modify OK
Output
Use <F4> to select the unit(s) or card to which the data should be sent:
• Printer,
• Memory card (not with DL50),
• Computer,
• Printer + memory card (not with DL50) or
• Printer + computer.
Printer is stored as the default parameter (for the printer you have
defined, see Section 2.7.1).
If you have not defined a printer and/or computer, the function can not be
executed. This case neither initiates an error message nor does it
influence the titration.
If you have defined and switched on a printer, but this is set to "offline",
the titrator waits until you switch the printer to "online" and then starts to
transfer its data.
If you have defined and attached a computer, but have not switched it on,
an appropriate error message appears.
If you have selected memory card but have not inserted one, an appropriate
error message appears.
3-74 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Method
Report
Results
All results
Raw results
Table of measured
values
Sample data
E – V curve
∆E/∆V – V curve
log ∆E/∆V – V
curve
∆ 2 E/∆V 2 – V
curve
E – t curve
V – t curve
∆V/∆t – t curve
All results defined in the Calculation function are recorded for the current
sample after every titration. Use <F4> to select "No" if you do not want a
report.
After the last titration of a sample series, the results of all samples are
recorded if you select "Yes" with <F4>. These include the statistics,
calibration, titer and auxiliary value data if the parameter or the functions
have been defined in the method.
After every titration the raw results such as VEQ or VDISP of the current
sample are recorded if you select "Yes" with <F4>.
After every titration the table of measured values of the current sample
is recorded if you select "Yes" with <F4>.
After the last titration of a sample series the sample data of all samples
are recorded if you select "Yes" with <F4> (ID, molar mass, equivalent
number, weight/volume, correction factor).
After every titration the titration curve Potential vs Volume of the current
sample is printed out if you select "Yes" with <F4>.
After every titration the 1st derivative of the titration curve Potential vs
Volume of the current sample is recorded if you select "Yes" with <F4>.
The ordinate representation is linear.
After every titration the 1st derivative of the titration curve Potential vs
Volume of the current sample is recorded if you select "Yes" with <F4>.
The ordinate representation is logarithmic.
After every titration the 2nd derivative of the titration curve Potential vs
Volume of the current sample is recorded if you select "Yes" with <F4>.
The ordinate representation is linear.
After every titration the titration curve Potential vs Time of the current
sample is recorded if you select "Yes" with <F4>.
After every titration the titration curve Volume vs Time of the current
sample is recorded if you select "Yes" with <F4>.
After every titration the 1st derivative of the titration curve Volume vs
Time of the current sample is recorded if you select "Yes" with <F4>.
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 3-75
Sample
Contents
Page
4. Sample data memory ................................................................................ 4-3
4.1 Entering sample data ................................................................................ 4-4
4.1.1 Deleting sample data ............................................................................... 4-6
4.1.2 Printing sample data ................................................................................ 4-7
4.1.3 Modifying sample data ............................................................................. 4-7
4.1.4 Adding sample data ................................................................................. 4-8
4.1.4.1 DL50/DL53 ............................................................................................... 4-8
4.1.4.2 DL55/DL58 ............................................................................................... 4-8
4.2 Weight transfer from a balance ............................................................... 4-12
4.3 Requesting sample data from a computer ............................................. 4-13
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 4-1
Sample
4. Sample data memory
This menu is used for the entry and storage of sample data. You can enter weight or volume,
identification, correction factor and temperature for maximum 60 samples. The data remain
stored even if you switch off the titrator; data of analyzed samples, on the other hand, are
deleted.
The menu is always accessible during the analysis to allow the user
• to see which samples have already been titrated
• to modify or delete data of samples not yet analyzed
• to enter data for additional samples.
The performance of the four titrators differs in this menu as follows:
DL50/DL53:
DL55/DL58:
You can enter data for only one sample series.
You can enter data for three sample series; these series can be run with
the same method or with three different methods.
You can enter the data of a so-called urgent sample.
With the Run key, method(s) and sample data are called up automatically and analyzed in
succession by the titrator (see Section 5).
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 4-3
Sample
4.1 Entering sample data
When you press the Sample key, the (blank) sample data list appears:
No. Status Wt/vol.
Meth. ID
Add
Add
SAMPLE
New sample series
New sample series (data: computer) 1)
Esc OK
1) appears if you have defined and attached a
computer (see Section 4.3).
Sample entry
SAMPLE
Number of samples 1
Method ID 00001
User
Esc OK
Number samples Enter, e.g. 3.
Method ID
User
Enter, e.g. 00001 (example for the standard method "Equivalence point
titration", see Section 5.1): This calls up the method to be used for
analysis of the samples.
If you do not know the identification of the method, press <F4> "Modify":
The method groups appear.
– Confirm either user or METTLER methods: The list of these methods
appears.
– Confirm the method you need for the analysis.
Enter your name if a keyboard is attached to the titrator.
4-4 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Sample
Note: The titrator also accepts methods stored on an inserted memory card or in an attached
computer: When you enter an identification whose method is not stored in the titrator,
the method is first searched for on the memory card and entered (not possible with the
DL50). If the method is not stored on the card, it is requested from the computer and
entered (see Operating Instructions: Computer Interface Description).
Press <F5> to show the sample data mask (weight entry as example):
Sample No. 1
SAMPLE
Sample ID
Weight [g] 0.0
Limits [g] 0.02 - 2.0
Correction factor f 1.0 ∇
Esc Balance* OK
* "Balance" appears if you have defined a balance in the Setup menu.
Sample ID
Enter or modify or delete if you have already defined the identification in
the Sample function (Section 3.3.2).
Weight Enter or transfer from an attached balance (see Section 4.2).
(Volume) • If you can not enter the weight or volume until during the titration, e.g.
as you do not wish to lose time through back weighing, do not enter
anything: As the titrator does not need the sample data until it starts to
calculate, the sample data mask appears again in the display during the
titration (see Section 5.1).
You can also enter the weight during the titration in the sample data
memory (see Section 5.2: Sequence of a pH-stat titration).
• If you have defined a fixed volume in the Sample function, no change
is possible.
• If you have violated the lower or upper limits of the weight/volume, but
still confirm this entry, the following appears:
Entry outside limits
SAMPLE
Modify entry
Save entry
Modify: The sample data mask reappears.
Save: The weight is transferred. The sample data mask for the next
sample or the sample data list appears.
OK
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 4-5
Sample
Limits
Correction factor
Temperature
You can not change the limits. These data do not appear if you have
defined a fixed volume in the Sample function.
You can enter a value for every sample that is reconciled in the
calculations if you enter f in the formula of the Calculation function.
Example: Each substance has a different, known moisture content
whose value you can insert for f:
1st sample: Moisture content = 4% → f = 0.96 → R = Q * C/(m * f)
2nd sample: Moisture content = 3% → f = 0.97 → R = Q * C/(m * f)
Enter the temperature of the solution to be titrated. You can not change
the value if you have defined a temperature sensor in the Sample function.
When you confirm the entries, the sample data mask for the second and the third sample
appears, followed by the sample data list:
No. Status Wt/vol. Meth. ID
2 ready 1.23454 00001 45321 ∆
3 ready 1.67115 00001 45322
Delete Print Modify Add
4.1.1 Deleting sample data
When you press <F2>, the following appears:
Delete sample(s)
SAMPLE
Selected sample
Sample series
Esc OK
Selected sample
The sample highlighted by the selection bar when you confirmed "Delete"
is deleted.
Sample series DL50/DL53: All entered data are deleted.
DL55/DL58: All data of the sample series which contains the sample
highlighted by the selection bar when you confirmed
"Delete" are deleted.
4-6 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Sample
Notes
1. You can not delete sample data of a series which are done until all defined samples have
been analyzed. When you then press <F2>, you can delete only the series.
2. If the titrator is controlled by a computer, you can not delete the sample data (see Section
6.8.3).
4.1.2 Printing sample data
A list of all sample data contained in the sample data memory is printed out.
4.1.3 Modifying sample data
The sample data mask of the selected sample appears.
• As long as the sample is ready, you can modify all parameters.
• If the sample is being titrated, in other words is active, you can modify the parameters as
long as the weight/volume of the sample has not yet been used for the calculation.
• When the sample is titrated, in other words ready, you can not modify any parameters.
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 4-7
Sample
4.1.4 Adding sample data
4.1.4.1 DL50/DL53
You add data of a new sample to the existing data of a sample series:
Add
SAMPLE
Sample to series 1)
Esc OK
1) If you have transferred the existing data from a computer, "Sample to series (data:
computer)" appears, see Section 4..
If the series has been processed, "New sample series" also appears.
Sample to series
The sample data mask for sample No. 4 of method 00001 appears
(example in Section 4.1.1).
As long as you do not delete the data of a series classed as done, you
can add data for a new sample to the series.
The data of a processed series are deleted when you
• switch off the titrator
• select "New sample series" in place of "Sample to series"
• modify the method used to run the series.
4.1.4.2 DL55/DL58
You either add the data of an additional sample or that of a new series to the existing data of
a series. The series are processed by the titrator in the defined sequence.
Add
SAMPLE
New sample series
New sample series (data: computer) 1)
Sample to series ∇ 1)
Esc OK
1) appears when you have defined and attached a computer (see Section 4.3). If you
have transferred existing data from the computer. "Sample to series (data: computer)"
also appears → scroll display
4-8 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Sample
New sample
series
You can add maximum two additional series to an existing series.
Before the first sample series (possible only if its method has not yet
been started):
No. Status Wt/vol. Meth. ID
1 ready 1.36182 00001 45320
2 ready 1.23454 00001 45321 ∇
Add
– Select the blank line above the first sample and press <F5>: The "Add"
mask appears (see above and the sequence in Section 4.1).
Between two sample series
No. Status Wt/vol. Meth. ID
2 ready 1.23454 00001 45321 ∆
3 ready 1.67115 00001 45322
1 ready 5 2
2 ready 5 2 ∇
Add
– Select a sample data line of the first series and press <F5> : The "Add"
mask appears (see above and sequence in Section 4.1).
After the last sample series
– Select a sample data line of the last series and press <F5>: The "Add"
mask appears.
Sample to series
The sample data mask for sample No. 4 of method 00001 appears (example
in Section 4.1.1).
You can add samples to a completed series before the titrator starts to
process the next series, in other words before you confirm the result list
of the last sample with OK.
As soon as you have confirmed the result list of the last sample of a series,
all its data are deleted when the next series is defined. You can then enter
the data for a new sample series.
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 4-9
Sample
Urgent sample
If you have to analyze one sample immediately while a sample series is being processed by
the DL55/DL58, you can insert this after the active series sample. The sample series is
interrupted and then continued after the determination.
Example: The titrator is processing sample 2 of method 00001.
– Press the Sample key and select the active sample:
No. Status Wt/vol. Meth. ID
1 done 1.36181 00001 45320
2 active 1.23454 00001 45321
3 ready 1.67115 00001 45322
1 ready 5 2 ∇
Add
Add
SAMPLE
New sample series 1)
Sample to series
Urgent sample
Esc OK
1) appears if three series have not yet been defined.
– Confirm "Urgent sample" with OK.
Sample entry
SAMPLE
Number of samples 1
Method ID 00001
User
Esc OK
Number samples
Method ID
User
You can not modify these: You can determine only one urgent sample at
a time!
Enter the identification of the method which should be used to determine
the sample.
Enter your name if a keyboard is attached to the titrator.
4-10 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Sample
After entry and confirmation of the sample data, the following appears:
No. Status Wt/vol. Meth. ID
1 done 1.36181 00001 45320
2 active 1.23452 00001 45321
EX ready 0.41894 8
3 ready 1.67115 00001 45322 ∇
Delete Print Modify Add
• When sample 2 is done, method 8 for the urgent sample is called up
(see Section 5.4: Running sample series).
• When the urgent sample is done, method 00001 for sample 3 is called
up. The data of the urgent sample are deleted.
Note: When a defined sample series has been analyzed and you confirm the Add command
to add a sample to the series, "Urgent sample" is also available in the mask that appears
(see previous page).
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 4-11
Sample
4.2 Weight transfer from a balance
If you have attached a balance to the titrator, you must have defined this in the Setup menu.
If not, "Balance" does not appear in the sample data mask.
Sample No. 1
SAMPLE
Sample ID
Weight [g] 0.0
Limits [g] 0.02 - 2.0
Correction factor f 1.0 ∇
Esc Balance OK
When you press <F4>, the value displayed on the balance appears, e.g.
Weight:
SAMPLE
> 1.36181
Esc OK
The weight is transferred with OK.
If the weight violates the lower limit, three minus signs (– – –) always
appear after "Weight", if it exceeds the upper limit three plus signs
(+ + +) appear. You can change or confirm the weight (see Section 4.1:
Weight parameter).
With METTLER TOLEDO balances which have a key to transfer the weight,
you can enter the weight values for a sample series from the balance
without pressing a key on the titrator. The setting "Bidirectional transmission
mode" is required for this.
Some balances show the sample number, others also show weight limits
(DeltaTrac).
– With weight "0.0" in the first sample data mask, press <F4>.
– Press the TARE key of the balance to display the current weight.
– Activate the weight transfer with the appropriate key of the balance: The
weight is entered in the sample data mask; at the same time, the titrator
automatically starts the weight entry for the next sample.
4-12 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Sample
4.3 Requesting sample data from a computer
If you have loaded the DLWin or the LabX titration software or an appropriate program on the
attached computer, you can request the sample data stored on it. With the "Add" command,
the following appears automatically:
Add
New sample series
New sample series (data: computer)
SAMPLE
Esc OK
Series entry
SAMPLE
ID
User
Esc OK
ID
User
You enter the identification of the sample series stored in the computer.
This is transferred to the sample data memory with OK. These sample
data can only be printed out or displayed ("Modify" command).
Enter your name if a keyboard is attached to the titrator.
Note: For reports which are sent to the printer, the computer and the
memory card,
• the name entered in the titrator is always adopted,
• the name entered in the computer is adopted if you do not enter
a name in the titrator.
If you wish to add the data of a new sample to a series transferred from the computer, confirm
"Sample to series (data: computer)" in the "Add" mask with OK. The sample data are added
to the sample data list.
You will find further information in the Operating Instructions for the computer interface or the
DLWin or LabX titration software.
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 4-13
Run
Run
Contents
Page
5. Analysis ..................................................................................................... 5-3
5.1 Menu sequence up to insertion of the sample ....................................... 5-4
5.2 Analysis sequence of a sample ............................................................... 5-7
5.2.1 Fading out analysis ..................................................................................... 5-11
5.2.2 Analysis is interrupted ................................................................................. 5-11
5.2.3 Modifying method of the current sample series .......................................... 5-12
5.2.4 Aborting analysis ......................................................................................... 5-12
5.3 DL55/DL58: Modifying ongoing method ................................................ 5-13
5.4 DL55/DL58: Running sample series ........................................................ 5-15
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 5-1
Run
Run
5. Analysis
You perform titrations with a selected method. The data required for this are stored in the
Method and Setup menus. You prepare your sample, titrate it and receive the result.
You enter the sample data such as weight or volume directly before the start of each sample
determination. These data are also entered in the sample data memory. To modify them during
the titration or to add samples to a defined sample series, you can open the Sample menu.
While a method is running, you can call up the auxiliary functions "Stirrer" and "Results"; all
other auxiliary functions and the Setup and Method menus are blocked.
In this menu, the functions of the Run and command key <F5> are identical, in other words you
can press one or other key.
The performance of the four titrators differs in this menu as follows:
DL50/DL53:
DL55/DL58:
One titration curve is displayed (EQP/EP titration: Potential vs. Volume,
pH/mV titration: Potential vs. Time).
Seven titration curves for the display can be selected.
You can interrupt the analysis of a sample to modify parameters of the
ongoing method.
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 5-3
Menu sequence
Run
5.1 Menu sequence up to insertion of the sample
When you press the Run key, there are two possibilities for the display (as an example of the
representation, the standard method "Equivalence point titration" is used. This has been
changed for the determination of CaCl 2 and stored with the identification 00001):
No data in sample data memory:
All sample data in sample data memory (see
Section 4.1):
Samples to be analyzed
RUN
Samples to be analyzed
RUN
Number of samples 3
Method ID 00001
User
Start
3 samples, Method 00001
Delete Start
Defined are
RUN
Defined are
RUN
Stirrer 1: Stand 1
Sensor 1: DM141
Drive 2: 0.1 mol/L AgNO 3
Esc Print OK
Stirrer 1: Stand 1
Sensor 1: DM141
Drive 2: 0.1 mol/L AgNO 3
Esc Print OK
Sample No. 1
RUN
Sample ID 45320
Weight [g] 1.23452
Limits [g] 0.02 - 2.0
Correction factor f 1.0
∇
Balance* OK
* "Balance" appears, if you have defined a
balance in the Setup menu
Current sample
No. 1 of 3
Sample ID 45320
Method ID 00001
OK
RUN
Current sample
No. 1 of 3
Sample ID 45320
Method ID 00001
OK
RUN
The Stir function is being processed
Notes
1. If you have defined an expiry date for the titrant and this has lapsed, an appropriate
message follows the "Defined are" mask. In order to determine a titer before the sample
analysis, you can confirm the message with Esc.
2. When the sample data mask or that of the "Current sample" appears, the sample is active
in the sample data memory. The titrator executes the Sample function of the method and
starts to acquire the time for every determination as the raw result TIME (see Section 5.2:
Example of an analysis sequence and Section 8.1: List of symbols).
5-4 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Run
Menu sequence
3. The titrator also accepts methods which are stored on an inserted memory card or in an
attached computer: When you enter an identification whose method is not stored in the
titrator, the method is first sought on the memory card and entered (not possible with the
DL50). If the method is not stored there, it is requested from the computer and entered
(see Operating Instructions: Computer Interface Description).
Mask: Samples to be analyzed
Number samples
Method ID
User
Enter, e.g. 3; this enters a sample series.
Enter, e.g. 00001: This calls up the method which should be used for
analysis of the samples.
If you do not know the identification of the method, press <F4> (Modify):
The method groups appear.
– Confirm either user or METTLER methods: The list of these methods
appears.
– Confirm the method you need for the analysis.
Enter your name if a keyboard is attached.
Delete
If, e.g. you have called up a method which you do not wish to use for the
analysis, you can delete it. You have this possibility available with direct
start if you confirm the "Defined are" mask with Esc (see page 5-4).
Mask: Defined are
This mask appears so that you can check whether you have attached the resources required
by the method.
• You can exclude the mask from the sequence (see Section 6.7.3).
• If you have not defined a resource or there is no RS or sensor option available for the defined
resource, an appropriate error message appears.
Mask: Sample No. 1 (sample data mask)
This mask does not appear if you
• have defined "Fixed volume" as entry type in the Sample function
• have not defined a predispensing to "factor x sample size" in the EQP/EP titration function
• have defined results such as "R = Q" in the Calculation function(s) which do not require the
sample size m.
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 5-5
Menu sequence
Run
If you still have to enter sample data such as the temperature for buffer solutions to be
measured in a sensor calibration, you can do this for every sample in the Sample menu (see
Section 4.1).
Sample ID
Enter or modify or delete if you have already defined the identification in
the Sample function (see Section 3.3.2).
Weight Enter or transfer from an attached balance (see Section 4.2).
(Volume) If you can not enter the weight or volume until during the titration as, e.g.
you do not wish to lose time through back weighing, do not enter anything.
As the titrator does not need the sample size until it performs calculations,
the sample data mask appears again in the display during the titration:
• either when the titrator reaches the EQP/EP titration function under
which a predispensing to (factor x sample size) has been defined
• or when it reaches the Calculation function under which the sample size
m is needed for the result.
You can also enter weight/volume in the Sample menu (see Section 5.2:
Example of a pH-stating).
If you violate the upper or lower limits of the weight/volume but still confirm
the entry, an appropriate message appears (see Section 4.1).
Limits
Correction factor
Temperature
You can not change the limits. This information does not appear if you
have defined a fixed volume in the Sample function.
You can enter a value for each sample which is reconciled in the
calculations if you insert f in the formula of the Calculation function (see
Section 4.1).
Enter that of the solution to be titrated. You can not change the value if
you have defined a temperature sensor in the Sample function.
Mask: Current sample
• If you work with a sample changer, in other words you have defined "ST20A" as a titration
stand, the mask always appears, but is confirmed automatically (no OK command).
• If you have defined "Auto stand" as the titration stand, the mask does not appear as a system,
e.g. a robot undertakes insertion or changing of the sample.
5-6 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Run
Analysis sequence
5.2 Analysis sequence of a sample
1st example: Method 00001 (see Section 5.1)
Sample No. 1
Sample ID 45320
Weight [g] 1.23452
Limits [g] 0.02 - 2.0
Correction factor f 1.0
Current sample
No. 1 of 3
Sample ID 45320
Method ID 00001
RUN
∇
Balance OK
RUN
OK
The titrator starts to process the method with the
Sample function.
(– Enter ID.)
– Weigh in sample, enter the weight for the first
sample or have it transferred from a balance.
(– Enter correction factor and temperature.)
– Press <F5> or the Run key.
– Before you confirm this mask, fix the sample beaker
to the titration stand.
Stir function
Wait time [s] 30
Speed [%] 80
RUN
Hold 1)
The titrator stirs with 80% of the maximum speed for
30 seconds (CaCl 2 dissolves during this time). The
elapse of time is displayed.
1) appears only with the DL55/DL58 (see Section 5.3)
mV
E – V curve
The titrator executes the EQP titration function.
mL
Table Values Curve 1) Hold 2)
1) appears only with DL55/DL58 (see next page)
2) appears only with DL55/DL58 (see Section 5.3)
Result list
Method: 00001
Sample 1
R1 = 29.26 %
OK
RUN
After the last method function, in other words before
the result list appears, the titrator ends the time
acquisition for this sample determination (see Note 2,
Section 5.1): The sample is present in the sample
data memory as done.
Sample No. 2
Sample ID 45321
Weight [g] 1.08723
Limits [g] 0.02 - 2.0
Correction factor f 1.0
Balance OK
RUN
∇
The sample data mask for the second sample appears:
– Procede as described under sample 1.
Current sample
No. 2 of 3
Sample ID 45321
Method ID 00001
OK
RUN
You are "prompted" to insert the second sample etc.
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 5-7
Analysis sequence
Run
At the end of the series, the results of all samples appear in the display.
• If you confirm this result list, the mask "Samples to be analyzed" again appears. The method
used to perform the last titration is entered as method.
• In the sample data memory, all samples of this series are present as done. You can now add
samples to this series here (see Section 4.1.4).
• As soon as you start a new series or an individual sample, all data of the executed series
will be deleted.
Notes
1. Changing stirrer speed
During a titration you can always activate the Stirrer key to change the stirrer speed (see
Section 6.2).
2. Method with two different titrants/sensors
If, e.g. you have to use two titrants for a back titration and have only one burette drive, the
following message (example) appears:
Change resources
RUN
Drive 2: 0.1 mol/L HCl
OK
If you have to use two sensors to, e.g. set the pH before the actual determination with a
phototrode and have only one sensor option, "Sensor 1: DP660" appears.
The message always appears before the function in which the second resource is needed.
3. DL55/DL58: Curve display
During the EQP/EP titration or EQP/EP titration (Ipol/Upol) or pH/mV-stat function, the
command <F4> (Curve) remains active if the curve representation has been selected. You
can select the following curves for the display:
E – V curve
Potential vs. Volume
∆E/∆V – V curve
1st derivative (Potential vs. Volume)
log ∆E/∆V – V curve
1st derivative (Potential vs. Volume, logarithmic)
∆ 2 E/∆V 2 – V curve
2nd derivative (Potential vs. Volume)
E – t curve
Potential vs. Time
V – t curve
Volume vs. Time
∆V/∆t – t curve
1st derivative (Volume vs. Time)
5-8 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Run
Analysis sequence
4. Result list
a. During a titration, you can always activate the Results key to view the results generated
to date (see Section 6.5.1).
b. All results of a learn titration are marked in the display by an exclamation mark(!).
c. With a sample series, the titrator skips the list for each sample you have excluded from
the analysis sequence (see Section 6.7.3).
d. With a sample series, all menus and auxiliary functions between the individual
determinations are again accessible if you do not confirm the result list of the sample.
You can then, e.g. also modify the method used to analyze the samples of this series
(see Section 5.2.3).
This is no longer possible if you exclude the "Result list for each sample" from the
sequence (see point c.).
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 5-9
Analysis sequence
Run
2nd example: pH-stating
This sequence of a pH-stating in which the titrator should start the titration immediately
after sample addition shows the entry of the sample data in the Sample menu during the
titration:
Sample No. 1
RUN The titrator executes the Sample function.
Table Values Curve 1) Hold 2)
Sample ID 592
Weight [g] 0.0
Limits [g] 0.02 - 2.0
Correction factor f 1.0
∇
Balance OK
Current sample
RUN
No. 1 of 5
Sample ID 592
Method ID 33
OK
mV
E – t curve
s
– Confirm the sample data mask without entering the
weight.
– Add the sample and immediately confirm with OK.
The titrator starts the pH-stating (the Stir function
does not appear as the stirring time is defined with 0
(zero).
1) appears only with DL55/DL58 (see page 5-8)
2) appears only with DL55/DL58 (see Section 5.3)
Press
Sample
No. Status Wt/Vol. Meth. ID
The sample data list appears.
1 active 0.0 33 592
2 ready 0.0 33 593
3 ready 0.0 33 594
Print Modify Add
Sample No. 1
SAMPLE
Sample ID 592
Weight [g] 0.1456
Limits [g] 0.02 - 2.0
Correction factor f 1.0
∇
Balance OK
The sample data mask appears.
– Enter the weight.
Press
Run
mV
E – t curve
s
Table Values Curve Hold
If the pH-stating is not yet complete, the E - t curve of
the pH/mV-stat function is again displayed. Otherwise,
the function executed by the titrator appears.
5-10 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Run
Analysis sequence
5.2.1 Fading out analysis
During the titration you can
• enter sample data: press the Sample key (see Section 4.1 and example on previous page)
• change the stirrer speed: press the Stirrer key (see Section 6.2)
• view all results generated to date: press the Results key (see Section 6.5.1).
The titration continues to run in the background.
– To display it again, press the Run key: The display shows the method function which is
currently being executed by the titrator.
5.2.2 Analysis is interrupted
The titrator immediately interrupts an analysis in the following situations:
1. After the Learn titration function if it has not found any equivalence points (see Section
3.3.8).
– Confirm the error message: The representation of the Titration function shown before the
interruption appears in the display (see example below).
If you wish the titrator to execute the remaining functions of the method,
– press <F5>, "Proceed": The titrator executes the remaining functions.
2. At the immediate end of an EQP titration or EQP titration (Ipol/Upol) function when the
condition for the defined parameter Stop for reevaluation is met (see Sections 3.3.6.7 and
3.3.9.8). The following display appears:
Reevaluation required
RUN
EQP titr. [1] neq=0
Confirm with OK and perform
reevaluation (RESULTS)
OK
(Example)
Measured values
6.789 mL
265.9 mV
RUN
Table Curve Proceed
You can now confirm "Proceed" (the analysis is continued) or press the Results key to
perform the reevaluation, see Section 6.5.2.
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 5-11
Analysis sequence
Run
Note: If a method is interrupted which is not stored in the titrator but has been requested from
the computer or a memory card, a reevaluation is not possible (see Section 5.1: Note
3 and Section 6.5.2).
5.2.3 Modifying method of the current sample series
After every sample determination, you can modify the method used to perform the series. A
requirement is that you do not exlude the "Result list for each sample" from the analysis
sequence (see Section 6.7.3).
If you wish to modify the functions of the method,
• all parameters of the Sample function are blocked with the exception of the numeric
parameters you have defined for the ST20A titration stand.
• with all other functions you can change only the numeric parameter values.
When you save the method, the printer automatically records the complete method with the
modified values: these remain stored for the method.
– To continue the sample series, press the Run key twice: The prompt to insert the next sample
appears.
Note: You can not change a method which is not stored in the titrator but is requested from
the computer or the memory card.
5.2.4 Aborting analysis
If you wish to abort the sample which is currently being titrated, press the Reset key. The mask
"METTLER TOLEDO/DL5x/Version x.x" appears.
• If you abort the sample at a time at which a result has already been calculated this is
displayed in the result list (auxiliary function "Results"). If no result has been calculated, only
the sample number will be listed.
Individual sample
The sample is present in the sample data memory as done and is deleted as soon as you start
the analysis of a new sample.
Sample series
A series is not aborted with Reset: The sample is present in the sample data memory as done.
With Run the display for the remaining samples of the series appears (example):
Samples to be analyzed
RUN
2 samples, Method 00001
Delete Start
5-12 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Run
DL55/DL58: Modifying ongoing method
5.3 DL55/DL58: Modifying ongoing method
If you confirm "Hold" during a sample determination, the titrator stops the function it is currently
executing. You can then modify the method used to analyze the sample:
- Press the Method key (example):
Methods
METHOD
Method ID 00001
Print Modify
Notice
METHOD
Parameters that you can
not modify are blocked.
OK
With OK the functions of the interrupted method appear.
• All parameters of the Sample function are blocked with the exception of the numeric
parameter values you have defined for the ST20A titration stand.
• With all other functions you can modify only the numeric parameter values.
If you save the method, the printer automatically records the complete method with the modified
values; these remain stored for the method.
To continue the sample series,
– press the Run key: The representation shown before the interruption appears in the display.
– Press the Run key or <F5>: The analysis sequence is continued.
In the case of an interrupted sample determination, the modifications apply only to the
functions which the titrator has not yet started to execute. Exceptions are the Stir, EQP/EP
titration, EQP/EP titration (Ipol/Upol) and pH/mV-stat functions.
1st example: Stir function
If you interrupt the method at this function, you can modify not only the stirring and wait times,
but also the rotational speed. On continuation of the determination, the titrator
• adds the difference between the defined and modified times to the stirring time remaining:
the entire stirring time corresponds to the modified time and
• stirs at the modified speed.
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 5-13
DL55/DL58: Modifying ongoing method
Run
2nd example: Dispense function
If you interrupt the method at this function, change the volume and then continue the
determination, the new volume is not added until the next sample determination.
3rd example: EQP titration function
If you interrupt the method at this function, you can modify, e.g. the value for the maximum
volume or the potential value for termination of the titration if this termination parameter is
defined within the method.
On continuation of the method, the titrator executes the modified parameters of the EQP
titration.
4th example: pH/mV-stat function
If you interrupt the method at this function, you can modify, e.g. the value for the control band,
the volume for the minimum consumption or the time for the defined termination criteria.
On continuation of the method, the titrator executes the modified parameters of the pH-stat
function.
Note: You can not modify a method which is not stored in the titrator but is requested from the
computer or the memory card (see Section 5.1: Note 3).
5-14 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Run
DL55/DL58: Running sample series
5.4 DL55/DL58: Running sample series
You can store three sample series in the sample data memory of the DL55/DL58. These are
processed in this menu by the titrator in sequence with your help. An attached sample changer
automates this procedure.
You can check the ongoing method in the Sample menu (see Section 4).
As soon as the next sample series is started, all data of the executed series are deleted. At
the end of the processed sample series, only the data of the last series remain stored. They
are deleted as soon as you start the determination of a new sample or series.
With all titration stands, the sample data mask appears only if you have not entered the weight
(volume) for the samples before the start of the method.
If you work with titration stand 1, 2 or an external stand, you must confirm the mask "Current
sample" for every sample.
If you have attached a sample changer, the titrator titrates the samples of all measurement
methods without the need for your intervention. A requirement is that
• you have entered all sample data
• you have excluded the mask "Defined are" from the analysis sequence (see Section 6.7.3)
• the methods use the same resources or their setup data are defined at the other drive
(titrants), at a different sensor option (sensors) or at the other ST20A output (solvents) (see
comparison of the titration sequences on the following page).
If you work with the auto stand, the mask "Current sample" does not appear, in other words
each sample determination is started without confirmation. As a result, you can attach a robot
to handle sample changing. The same prerequisites hold for the automatic running of sample
series as for the use of the sample changer.
Note: All information concerning automation in this titrator's Reference Handbook relates to
the earlier METTLER TOLEDO Sample changer ST20A. Its name continues to appear
in some places. Operation of other METTLER TOLEDO sample changers via your
titrator is described in the corresonding Operating Instructions.
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 5-15
DL55/DL58: Running sample series
Run
Comparison of the analysis sequences for three sample series at the ST20A titration stand or at stand 1, 2 or the
external stand. (The masks "Defined are" and "Result list for each sample" are excluded from the sequence for
stand 1/2/external.)
Samples to be analyzed
RUN
ST20A
3 samples, Method 00001
6 samples, Method 2
10 samples, Method 8
Delete Start
Stand 1/2/external
Current sample
No. 1 of 3
Sample ID 45320
Method ID 00001
RUN
Current sample
No. 1 of 3
Sample ID 45320
Method ID 00001
OK
RUN
The mask "Current sample" appears for every sample,
but is confirmed automatically.
The titrator titrates all samples of method 00001.
It then starts to titrate the samples of method 2 (the
setup data are the same as for method 00001).
It then starts to titrate the samples of method 8 (the
setup data are the same as for method 00001). At the
end of all sample determinations of method 8, the
following appears:
Result list
Method: 8
Sample 1
R1 = 97.386 %
Sample 2
OK
RUN
∇
The titrator titrates all samples of method 00001. At the
end of the 3 determinations, the following appears:
Result list
Method: 00001
Sample 1
R1 = 29.26 %
Sample 2
Current sample
No. 1 of 6
Sample ID 11
Method ID 2
Result list
Method: 2
Sample 1
R1 = 4.234 mmol
R2 = 2.31 g/L
OK
OK
OK
RUN
∇
RUN
The titrator titrates all samples of method 2. At the end
of the 6 determinations, the following appears:
RUN
∇
Current sample
No. 1 of 10
Sample ID 432148
Method ID 8
OK
RUN
The titrator titrates all samples of method 8. At the end
of the 10 determinations, the result list with the results
of all samples of this method appears (see mask in
ST20A column).
5-16 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Run
DL55/DL58: Running sample series
If the titrator has to process methods which have the same burette drives or sensor inputs in succession on the
sample changer, you have to intervene:
Samples to be analyzed
RUN
4 samples, Method 00001
7 samples, Method 111
Delete Start
Current sample
No. 1 of 4
Sample ID 45320
Method ID 00001
RUN
The titrator titrates all samples of method 00001.
It then starts to execute the functions of the first
sample of method 111: a different titrant and a different
sensor are required. The following message appears
before the Titration function in which the titrant
and the sensor are defined:
Change resources
Drive 2: 0.1 mol/L EDTA
Sensor 1: DP660
OK
RUN
– Place the burette with the EDTA solution on burette
drive 2 and attach the DP660/DP5 Phototrode
to sensor input 1 (pH option in slot 1) and
plug into titration stand.
If you confirm the message, the titrator executes the
remaining functions of the method, it then titrates the
other samples. At the end of the 7 determinations, the
result list with the results of all samples of this method
appears.
The mask "Current sample" appears for every sample,
but is confirmed automatically.
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 5-17
Auxiliary functions
Contents
Page
6. Auxiliary functions .................................................................................... 6-3
6.1 Sensor ........................................................................................................ 6-4
6.1.1 Measure potential........................................................................................ 6-4
6.1.2 Measure temperature .................................................................................. 6-5
6.1.3 Calibrate temperature sensors .................................................................... 6-6
6.1.4 Measure voltage/current (Ipol/Upol) ............................................................ 6-7
6.2 Stirrer ......................................................................................................... 6-8
6.3 Sample changer ........................................................................................ 6-9
6.3.1 Change lift position...................................................................................... 6-9
6.3.2 Rotate turntable........................................................................................... 6-9
6.3.3 Dispense/rinse ............................................................................................ 6-10
6.4 Burette........................................................................................................ 6-11
6.4.1 Rinse burette ............................................................................................... 6-11
6.4.2 Rinse tip ...................................................................................................... 6-12
6.4.3 Dispense ..................................................................................................... 6-12
6.4.4 Dispense continuously ................................................................................ 6-13
6.4.5 Titrate manually ........................................................................................... 6-14
6.4.4 Dispense continuously (Ipol/Upol)............................................................... 6-15
6.4.6 Titrate manually (Ipol/Upol) ......................................................................... 6-16
6.5 Results ....................................................................................................... 6-18
6.5.1 Display result list ......................................................................................... 6-18
6.5.2 Perform reevaluation ................................................................................... 6-18
6.5.3 Perform calculations.................................................................................... 6-20
6.5.4 Display measured values ............................................................................ 6-21
6.5.5 Modify statistics ........................................................................................... 6-22
6.6 Report......................................................................................................... 6-24
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 6-1
Auxiliary functions
Page
6.7 Miscellaneous............................................................................................ 6-26
6.7.1 Define titrator settings ................................................................................. 6-26
6.7.1.1 Date/time format.......................................................................................... 6-26
6.7.1.2 Date/time ..................................................................................................... 6-27
6.7.1.3 Language .................................................................................................... 6-27
6.7.1.4 Report header ............................................................................................. 6-27
6.7.1.5 Titrator ID .................................................................................................... 6-27
6.7.1.6 Audio signal................................................................................................. 6-28
6.7.2 Adjust measuring inputs .............................................................................. 6-28
6.7.3 Shorten analysis sequence ......................................................................... 6-30
6.7.4 Define user level ......................................................................................... 6-31
6.7.5 Activate control input ................................................................................... 6-32
6.7.6 Activate control output................................................................................. 6-33
6.8 Data transfer .............................................................................................. 6-36
6.8.1 Memory card ............................................................................................... 6-36
6.8.2 Computer .................................................................................................... 6-40
6.8.3 Remote control ............................................................................................ 6-42
6.8.4 Bar-code string ............................................................................................ 6-43
6-2 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Auxiliary functions
6. Auxiliary functions
You can use these symbol keys to perform the activities mentioned below. These auxiliary
functions are independent of a titration method, but they can support or supplement a method.
While a titration is running, you can activate only the functions of the Stirrer and Results keys,
all other keys are blocked.
Sensor
• Measure potential or temperature of a solution
• Calibrate temperature sensors
• Measure voltage/current (Ipol/Upol)
Stirrer
• Switch stirrer on or off, change stirrer speed
Changer
Burette
Results
Report
*
• Operate sample changer manually
• Rinse burette and burette tip
• Dispense a specified volume
• Dispense continuously
• Titrate manually
• Dispense continuously (Ipol/Upol)
• Titrate manually (Ipol/Upol)
• Display result list of all samples of a series
• Modify evaluation of an EQP titration
• Perform additional calculations of the sample last titrated
• Display measured values of the sample last titrated
• Modify statistical evaluation of the series last titrated
• Print out additional reports of the series last performed or sample last
titrated
Misc. ...
Data
Transfer
• Define date and language
• Enter calibration data of the measuring inputs
• Modify analysis sequence
• Change user level
• Activate control inputs and outputs
• Copy data from titrator to the memory card and vice versa (not possible
with DL50)
• Transfer data from the titrator to the computer
• Select whether titrator should be controlled by the computer
• Enter bar-code string (computer transfer)
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 6-3
Sensor
6.1 Sensor
When you press the Sensor key, the following auxiliary functions appear:
Measure potential
Measure temperature
Calibrate temperature sensors
Measure voltage/current (Ipol/Upol)
6.1.1 Measure potential
You can measure the potential of solutions (mV, pH, %T, etc.).
Sensor
DG111
Unit of meas.
mV
Titration stand Stand 1
Speed [%] 50
Temperature [°C] 25.0
Temperature sensor Manual
Report
No
∆t [s] 5
Sensor
Unit of measurement
Titration stand
Speed
Temperature
Press <F4> to open the list containing the sensors you have defined in
the Setup menu (see Section 2.2). Confirm the one you want with OK.
The defined measurement unit is entered automatically.
You can select a different measurement unit from the selection menu
(press <F4>).
Press <F4> to select: Stand 1, Stand 2, ST20A, Auto stand or External
stand; this tells the titrator what stirrer it has to control.
• 0% → The stirrer is at a standstill.
• 100% → The stirrer stirs at maximum speed.
Enter the temperature of the solution to be measured if you have not
attached a temperature sensor.
Temp. sensor Press <F4> to select: Manual, TEMP A, TEMP B, TEMP C or TEMP D.
Report
∆t [s]
Use <F4> to select "Yes" if desired: A measured value will be recorded
every 5 seconds if you accept the time interval of "5".
6-4 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Sensor
You can start and stop the measurement with <F5>.
Start: The titrator starts to measure the potential of the solution. The
measured values are shown in the display and are recorded.
During the measurement, all other keys are blocked.
6.1.2 Measure temperature
You can measure the temperature of solutions (°C, °F or K).
Temperature sensor TEMP A
Unit of meas. °C
Titration stand Stand 1
Speed [%] 50
Report
No
∆t [s] 5
Temp. sensor Press <F4> to select: TEMP A, TEMP B, TEMP C or TEMP D.
Unit of meas. Press <F4> to select: °C, °F or K.
Titration stand
Speed
Report
∆t [s]
Press <F4> to select: Stand 1, Stand 2, ST20A, Auto stand or External
stand; this tells the titrator what stirrer it has to control.
• 0% → The stirrer is at a standstill.
• 100% → The stirrer stirs at maximum speed.
Press <F4> to select "Yes" if desired: A measured value is recorded every
5 seconds if you accept a time interval of "5".
Start and stop the measurement with <F5>.
Start: The titrator starts to measure the temperature of the solution. The
measured values are shown in the display and are recorded.
During the measurement, all other keys are blocked.
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 6-5
Sensor
6.1.3 Calibrate temperature sensors
You can calibrate your Pt100 or Pt1000 sensors. The zero point determined by this procedure
is entered together with the date in the parameter mask of the corresponding temperature
sensor (see Section 2.3).
Calibrate temperature sensors
SENSOR
Temperature sensor TEMP A
Esc Modify Start
Temp. sensor Press <F4> to select: TEMP A, TEMP B, TEMP C or TEMP D.
– Press <F5>: the mask with the prompt to immerse the sensor in ice
water appears. (We advise you to use a Dewar vessel.)
– When you are certain that the sensor has cooled to 0 °C, confirm this
with OK:
• The calibration was successful if Modify and Start reappear in the
display. You can calibrate the next Pt sensor.
• If the measured temperature value does not lie between -2 °C and
+2 °C, an appropriate error message appears.
– Confirm this message and first "check" the temperature value given
by the sensor using the auxiliary function Measure temperature
(see Section 6.1.2).
6-6 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Sensor
6.1.4 Measure voltage/current (Ipol/Upol)
You can measure the voltage or the current of solutions with polarized sensors (mV, µA).
Sensor
DM142
Indication
Voltametric
Titration stand Stand 1
Speed [%] 50
Report
No
∆t [s] 5
Sensor
Indication
Titration stand
Speed
Report
∆t [s]
Press <F4> to open the list containing the polarized sensors you have
defined in the Setup menu (see Section 2.4). Confirm the one you want
with OK.
Press <F4> to select: "Voltametric" or "Amperometric":
Voltametric: Enter the current value which should be applied to the
sensor to measure the voltage.
Amperometric: Enter the voltage value which should be applied to the
sensor to measure the current.
Press <F4> to select: Stand 1, Stand 2, ST20A, Auto stand or External
stand; this tells the titrator what stirrer it has to control.
• 0% → The stirrer is at a standstill.
• 100% → The stirrer stirs at maximum speed.
Use <F4> to select "Yes" if desired: A measured value will be recorded
every 5 seconds if you accept the time interval of "5".
You can start and stop the measurement with <F5>.
Start: The titrator starts to measure the voltage or current of the solution.
The measured values are shown in the display and are recorded.
During the measurement, all other keys are blocked.
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 6-7
Stirrer
6.2 Stirrer
You can start the stirrer and change its speed, e.g. to dissolve a substance.
Stirrer
STIRRER
Titration stand Stand 1
Speed [%] 50
Modify
Start
Titration stand
Speed
Press <F4> to select: Stand 1, Stand 2, ST20A, Auto stand or External
stand; this tells the titrator what stirrer it has to control.
• 0% → The stirrer is at a standstill.
• 100% → The stirrer stirs at maximum speed.
Start and stop the stirrer with <F5>.
Start: The stirrer stirs at the specified speed.
Changing the speed during a titration
To change the speed during a sample determination:
– press the Stirrer key (this does not interrupt the method):
Stirrer
STIRRER
Speed [%] 50
OK
– Change the speed and confirm with OK: the stirrer stirs at the new speed or is at a standstill
if you have entered zero.
– To follow the analysis sequence in the display again, press the Run key.
6-8 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Changer
6.3 Sample changer
Note: All information concerning automation in this titrator's Reference Handbook relates to
the earlier METTLER TOLEDO Sample changer ST20A. Its name continues to appear
in some places. Operation of other METTLER TOLEDO sample changers via your
titrator is described in the corresonding Operating Instructions.
You can control the sample changer (titration stand: ST20A) and the attached pumps/
dispensers manually. For this you must define the sample changer in the Setup menu and have
attached it (see Section 2.7.4).
Sample changer
Change lift position
Rotate turntable
Dispense/rinse
CHANGER
OK
6.3.1 Change lift position
Change lift position
CHANGER
Lift position
bottom
Esc Modify Start
Position
Use <F4> to select: bottom, middle or top.
Press <F5> to move the lift to the desired position.
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 6-9
Changer
6.3.2 Rotate turntable
Rotate turntable
CHANGER
Direction
forward
Number of positions 1
Esc Modify Start
Direction
Number of
positions
Press <F4> to select: forward or backward.
Enter the number of positions the turntable should move forward or
backward.
Press <F5> to move the turntable the specified number of positions
forward or backward.
6.3.3 Dispense/rinse
If you have attached a pump or a dispenser to an ST20A output, you can dispense or rinse the
sensor in the titration vessel.
Dispense/rinse
CHANGER
ST20A output
RINSE
Esc Modify Start
ST20A output
Use <F4> to select: DOSE, RINSE or DISPENSER.
Start and stop the dispensing or rinsing process with <F5>.
Note: The relay at the "DISPENSER" output closes for 0.5 s and triggers
the dispensing.
6-10 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Burette
6.4 Burette
When you press the Burette key, the following auxiliary functions appear:
Rinse burette
Rinse tip
Dispense
Dispense continuously
Titrate manually
Dispense continuously (Ipol/Upol)
Titrate manually (Ipol/Upol)
6.4.1 Rinse burette
Rinse burette
BURETTE
Burette drive Drive 2
Esc Modify Start
Burette drive Press <F4> to select : Drive 1
DL50/DL53: The Modify command is missing.
Start and stop the rinsing process with <F5>.
Start: The titrator discharges 1 burette volume. The piston of the burette
moves below its zero position to ensure that air bubbles "hidden"
in the stopcock are drawn in then expelled.
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 6-11
Burette
6.4.2 Rinse tip
Rinse tip
BURETTE
Burette drive Drive 2
Esc Modify Start
Burette drive Press <F4> to select: Drive 1
DL50/DL53: The Modify command is missing.
Start: 5% of the burette volume are discharged to eliminate any diffusion
with solvent in the dispensing tube tip.
6.4.3 Dispense
You can dispense a defined volume.
Dispense
BURETTE
Burette drive Drive 2
Burette volume 10 mL
Volume [mL] 1.0
Esc Modify Start
Burette drive Press <F4> to select: Drive 1
DL50/DL53: The Modify command is missing.
Burette volume
Volume
Press <F4> to select: 20, 1, 5 or 10 mL.
Enter the volume to be dispensed.
Start: The titrator dispenses the specified volume.
6-12 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Burette
6.4.4 Dispense continuously
You can dispense stepwise and follow the potential change as a function of the dispensing in
the display. You must define the following parameters:
Burette drive Drive 2
Burette volume
10 mL
Rate [mL/min] 1.0
Sensor
DG111
Titration stand Stand 1
Speed [%] 50
Temperature [°C] 25.0
Temperature sensor Manual
Burette drive Press <F4> to select: Drive 1
DL50/DL53: The Modify command is missing.
Burette volume
Rate
Sensor
Titration stand
Speed
Temperature
Press <F4> to select: 20, 1, 5 or 10 mL.
is the rate of dispensing.
Press <F4> to open the list containing the sensors you have defined in
the Setup menu. Confirm the one you want with OK. The sensor acquires
the measured values in the measurement unit you have defined for this
sensor (see Section 2.2.2).
Use <F4> to select: Stand 1, Stand 2, ST20A, Auto stand or External
stand; this tells the titrator what stirrer it has to control.
• 0% → The stirrer is at a standstill.
• 100% → The stirrer stirs at maximum speed.
Enter the temperature of the solution to be measured if you have not
attached a temperature sensor.
Temp. sensor Press <F4> to select: Manual, TEMP A, TEMP B, TEMP C or TEMP D.
Start and stop the dispensing with <F5>.
Start: The titrator dispenses continuously and acquires the potential
values of the changing solution. The measured values "mL" and
"mV, pH,..." are displayed.
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 6-13
Burette
6.4.5 Titrate manually
You can titrate manually and follow the potential change as a function of the dispensed volume
in the display. You must define the following parameters:
Titrant
NaOH
Concentration [mol/L] 0.1
Sensor
DG111
Unit of meas.
mV
Titration stand Stand 1
Speed [%] 50
Temperature [°C] 25.0
Temperature sensor Manual
Titrant
Concentration
Sensor
Unit of measurement
Titration stand
Speed
Temperature
Press <F4> to open the list containing the titrants you have defined in the
Setup menu (see Section 2.1). Confirm the one you want with OK.
The defined concentration of the titrant is entered automatically.
Press <F4> to open the list containing the sensors you have defined in
the Setup menu (see Section 2.2). Confirm the one you want with OK.
The defined measurement unit is entered automatically.
You can select a different measurement unit from the selection menu
(press <F4>).
Press <F4> to select: Stand 1, Stand 2, ST20A, Auto stand or External
stand; this tells the titrator what stirrer it has to control.
• 0% → The stirrer is at a standstill.
• 100% → The stirrer stirs at maximum speed.
Enter the temperature of the solution to be measured if you have not
attached a temperature sensor.
Temp. sensor Press <F4> to select: Manual, TEMP A, TEMP B, TEMP C or TEMP D.
6-14 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Burette
Use <F5> to start, interrupt and continue the titration, use <F1> to
terminate it.
Start: The titrator starts to dispense and increases the dispensing
rate continuously. The measured values "mL" and "mV, pH,..."
are displayed.
Hold: The dispensing is interrupted, but measurement of the potential
value continues.
Proceed: The titrator starts again to dispense at the smallest dispensing
rate, in other words you control the addition in the vicinity of the
desired end point with the <F5> key.
End: The titration is terminated.
The results are displayed: the volume in "mL" and "mmol" and the potential
in the selected measurement unit. You can print out these results.
6.4.6 Dispense continuously (Ipol/Upol)
You can dispense stepwise and follow the voltage and current change as a function of the
dispensing in the display. You must define the following parameters:
Burette drive Drive 2
Burette volume
10 mL
Rate [mL/min] 1.0
Sensor
DM142
Indication
Voltametric
Titration stand Stand 1
Speed [%] 50
Burette drive Press <F4> to select: Drive 1
DL50/DL53: The Modify command is missing.
Burette volume
Rate
Sensor
Press <F4> to select: 20, 1, 5 or 10 mL.
is the rate of dispensing.
Press <F4> to open the list containing the polarized sensors you have
defined in the Setup menu (see Section 2.4). Confirm the one you want
with OK.
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 6-15
Burette
Indication
Titration stand
Speed
Use <F4> to select:"Voltametric" or "Amperometric":
Voltametric: Enter the current value which should be applied to the
sensor to measure the voltage.
Amperometric: Enter the voltage value which should be applied to the
sensor to measure the current.
Use <F4> to select: Stand 1, Stand 2, ST20A, Auto stand or External
stand; this tells the titrator what stirrer it has to control.
• 0% → The stirrer is at a standstill.
• 100% → The stirrer stirs at maximum speed.
Start and stop the dispensing with <F5>.
Start: The titrator dispenses continuously and acquires the voltage or
current values of the changing solution. The measured values "mL"
and "mV" or "µA" are displayed.
6.4.5 Titrate manually (Ipol/Upol)
You can titrate manually and follow the voltage or current change as a function of the dispensed
volume in the display. You must define the following parameters:
Titrant 1 / 2 I 2
Concentration [mol/L] 0.1
Sensor
DM142
Indication
Voltametric
Titration stand Stand 1
Speed [%] 50
Titrant
Concentration
Sensor
Press <F4> to open the list containing the titrants you have defined in the
Setup menu (see Section 2.1). Confirm the one you want with OK.
The defined concentration of the titrant is entered automatically.
Press <F4> to open the list containing the polarized sensors you have
defined in the Setup menu (see Section 2.4). Confirm the one you want
with OK.
6-16 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Burette
Indication
Titration stand
Speed
Use <F4> to select:"Voltametric" or "Amperometric":
Voltametric: Enter the current value which should be applied to the
sensor to measure the voltage.
Amperometric: Enter the voltage value which should be applied to the
sensor to measure the current.
Press <F4> to select: Stand 1, Stand 2, ST20A, Auto stand or External
stand; this tells the titrator what stirrer it has to control.
• 0% → The stirrer is at a standstill.
• 100% → The stirrer stirs at maximum speed.
Use <F5> to start, interrupt and continue the titration, use <F1> to
terminate it.
Start: The titrator starts to dispense and increases the dispensing
rate continuously. The measured values "mL" and "mV" or
"µA" are displayed.
Hold: The dispensing is interrupted, but measurement of the voltage
or current value continues.
Proceed: The titrator starts again to dispense at the smallest dispensing
rate, in other words you control the addition in the vicinity of the
desired end point with the <F5> key.
End: The titration is terminated.
The results are displayed: the volume in "mL" and "mmol" and the voltage
value in "mV" or the current value in "µA". You can print out these results.
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 6-17
Results
*
6.5 Results
You can also press the Results key during a titration. The following auxiliary functions can be
executed:
Display result list
Perform reevaluation 1)
Perform calculations
Display measured values
Modify statistics
1) appears only when the titrator interrupts the analysis (see Section 6.5.2)
6.5.1 Display result list
You can view the results generated in the titration of a sample or a sample series. The list is
displayed at the earliest (press <F5>) during the first titration (only "Method ID" and "Sample
1" appear if no result is available).
6.5.2 Perform reevaluation
The parameters for the reevaluation of an EQP titration are displayed when the titrator
interrupts the current EQP titration or EQP titration (Ipol/Upol) function immediately before
completion as the defined condition for the Stop for reevaluation parameter is met (see
Sections 3.3.6.7 and 3.3.9.8). In such a case, the message "Reevaluation required" appears
(see Section 5.2.2) and you can modify the following parameters under this auxiliary function
(example):
Threshold 1000
Steepest jump only No
Range
Yes
Tendency
Positive
Potential 1
No
Potential 2
No
Maximum volume 5
– Modify the appropriate parameters and press <F5>.
A mask appears in which you can choose whether you wish to save the modifications of these
method parameters for this sample or sample series permanently or only temporarily.
6-18 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Results
The titrator then reevaluates the titration curve.
At the same time, the printer records this EQP titration/ EQP titration (Ipol/Upol) function with
the modified parameters. (If your changes still do not meet the defined condition, the message
"Reevaluation required" reappears.)
*
– To continue the interrupted sample determination, press the Run key: The representation
of the EQP titration/EQP titration (Ipol/Upol) function shown before the interruption appears
in the display.
– Press <F5>: The titrator executes either the EQP titration/EQP titration (Ipol/Upol) function
(if, e.g. dispensing must be continued owing to the modified parameters) or the next function.
Notes
1. To evaluate the effects of your changes, you can
• view and print out the table of measured values or curve (see Sections 6.5.4 and 6.6).
• check the new raw results of the EQP titration/EQP titration (Ipol/Upol) function:
– select the auxiliary function Perform calculations (see Section 6.5.3)
– assign the raw results to Rx or Cx , e.g. Rx = VEQ or Cx = VP1.
2. "Perform reevaluation" does not appear if the interrupted method is not stored in the titrator
but on the memory card or in the computer (see Section 5.1).
3. With the DL55/DL58, "Perform reevaluation" also appears if you interrupt the current EQP
titration/EQP titration (Ipol/Upol) function of a user method after a possible predispensing.
The auxiliary function does not appear if the interrupted method is a METTLER method.
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 6-19
Results
*
6.5.3 Perform calculations
At the end of a titration, you can perform additional calculations for these samples. The function
appears when you press the Results key on completion of the titration. In the case of a sample
series, this is possible only if the "Result list for each sample" is not excluded from the sequence
of a series (see Section 6.7.3).
DL55/DL58:
The auxiliary function also appears if you interrupt the titration and then press
the Results key.
When you select the auxiliary function, the following parameters appear:
Formula Rx =
Q*Cx/m
Constant Cx =
M/(10*z)
Decimal places 3
Result unit %
Result name
Formula
Constant
Press <F4> to open the "List of formulas" and accept a different formula
from the list or enter the one you need for your calculation (see also
Section 8.4: Examples of formulas).
Press <F4> to open the "List of constants" and accept a different constant
from the list or enter the one you need for your calculation.
Symbol
You can take the symbols you need for your calculation from the list you
obtain when you press <F3>. This includes all symbols possible for the
individual functions (see Section 3.3.12: Calculation function and Section
8.1: List of symbols).
Decimal places
Result unit
Result name
Enter the number of decimal places you wish to have in your result.
Press <F4> to select the unit from the selection menu.
You can enter the name using an attached keyboard.
Press <F5> to calculate and display the result(s), press <F3> to print it
(them) out.
6-20 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
6.5.4 Display measured values
Results
*
On completion of every titration, you can display the table of measured values or view the
titration curve of this sample again. The function appears when you press the Results key at
the end of the titration. With a sample series, this is possible only if the "Result list for each
sample" is not excluded from the sequence of the series (see Section 6.7.3).
When you select the auxiliary function, the following appears (example):
Measured values
5.621 mL
-267.3 mV
RESULTS
Table Curve 1) OK
You can view either the table of measured values, the measured value
pair Potential/Volume or the titration curve.
1) DL50: With the "(display) Curve" command, the E-V curve is shown in the case of an EQP/
DL53: EP titration and the E-t curve with a pH-stat titration.
1) DL55: The "(display) Curve" command remains active when you have selected it. Press
DL58: <F4> to show the selection menu from which you can select one of the possible
curve representations:
E – V curve
∆E/∆V – V curve
log ∆E/∆V – V curve
∆ 2 E/∆V 2 – V curve
E – t curve
V – t curve
∆V/∆t – t curve
Potential vs Volume
1st derivative (Potential vs Volume)
1st derivative (Potential vs Volume, logarithmic)
2nd derivative (Potential vs Volume)
Potential vs Time
Volume vs Time
1st derivative (Volume vs Time)
When you confirm one of the curves with OK, it is displayed.
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 6-21
Results
*
6.5.5 Modify statistics
If you have selected Statistics: Yes in the method under the Calculation function(s), the mean
value is calculated with a sample series of 2 samples, with more than 2 samples you also obtain
the standard deviation and the relative standard deviation. This auxiliary function is then
accessible under the Results menu to modify the statistical evaluation, i.e. to allow elimination
of outliers.
The function appears when you press the Results key on completion of every titration. This is
possible only if the "Result list for each sample" is not excluded from the sequence of the series
(see Section 6.7.3).
Modify statistics
Display evaluation
Delete result
Undo deletion
Esc
RESULTS
OK
Display
evaluation
Example of a statistical evaluation with 6 samples
Statistical evaluation
RESULTS
R1 n = 6
x = 1.234 %
s = 0.01268 % ∇
OK
Delete result
Using the result list you decide what sample result is an outlier and then
delete it.
Delete result
RESULTS
Sample No. 1
Result R 1
Esc
OK
– Enter the number of the sample whose result should be deleted, e.g.
2.
– Enter the number (index) of the result defined by the Calculation function.
6-22 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Delete result
Results
If you confirm OK, the result is deleted and the statistical data are
calculated with the new number of samples. You can delete several
sample results and always view the final result in the mask "Statistical
evaluation" (example):
*
Statistical evaluation
s = 0.00346 %
srel = 0.112 %
Outlier
Sample No. 2 deleted
RESULTS
OK
You can use the parameter "All results" of the auxiliary function Report
to print out the new evaluation (see Section 6.6).
Undo deletion
As long as the result list of this sample series remains stored, you can
undo the deletion for all sample results:
Undo deletion
RESULTS
Sample No. 1
Result R 1
Esc
OK
– Enter the number of the sample and that of the corresponding result R:
The result of this sample will be reincorporated in the statistical
calculation.
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 6-23
Report
6.6 Report
On completion of each sample series, you can print out additional reports or transfer them to
a computer (see also Section 3.3.16: Report function). In the case of a sample series, this is
possible only if you have not excluded the "Result list for each sample" from the sequence of
the series (see Section 6.7.3).
DL55/DL58: You can activate the auxiliary function if you interrupt the titration and then press
the Report key.
Report
REPORT
Output Printer
Results
No
All results No ∇
Modify Start
Output
Results
All results
Raw results
Table of
measured values
Sample data
E – V curve
Press <F4> to select the output unit to which the data should be sent:
• Printer,
• Memory card (not with the DL50)
• Computer,
• Printer + memory card (not with the DL50) or
• Printer + computer.
All results of the sample last determined defined in the Calculation
functions are recorded when you select "Yes" with <F4> and press <F5>.
Results of all samples generated to date are recorded when you select
"Yes" with <F4> and press <F5>. They also include the statistical data if
these have been calculated (see Section 6.5.5).
The raw results of the sample last determined are recorded when you
select "Yes" with <F4> and press <F5>.
The table of measured values of the sample last determined is recorded
when you select "Yes" with <F4> and press <F5>.
The sample data of all samples of the current series are recorded when
you select "Yes" with <F4> press <F5>.
The titration curve Potential vs Volume of the sample last determined is
recorded when you select "Yes" with <F4> and press <F5>.
6-24 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Report
∆E/∆V – V curve
log ∆E/∆V – V
curve
∆ 2 E/∆V 2 – V
curve
E – t curve
V – t curve
∆V/∆t – t curve
The 1st derivative of the titration curve Potential vs Volume of the sample
last determined is recorded when you select "Yes" with <F4> and press
<F5>. The ordinate representation is linear.
The 1st derivative of the titration curve Potential vs Volume of the sample
last determined is recorded when you select "Yes" with <F4> and press
<F5>. The ordinate representation is logarithmic.
The 2nd derivative of the titration curve Potential vs Volume of the sample
last determined is recorded when you select "Yes" with <F4> and press
<F5>. The ordinate representation is linear.
The titration curve Potential vs Time of the sample last determined is
recorded when you select "Yes" with <F4> and press <F5>.
The titration curve Volume vs Time of the sample last determined is
recorded when you select "Yes" with <F4> and press <F5>.
The 1st derivative of the titration curve Volume vs Time of the sample last
determined is recorded when you select "Yes" with <F4> and press <F5>.
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 6-25
Misc. ...
6.7 Miscellaneous
When you press the Misc. key, the following appears
Define titrator settings
Adjust measuring inputs
Shorten analysis sequence
Define user level
Activate control input
Activate control output
6.7.1 Define titrator settings
You can or must select or define the following settings for the titrator:
Date/time format
Date/time
Language
Report header
Titrator ID
Audio signal
6.7.1.1 Date/time format
You have several possibilities for the date and time notation. If you wish to change the existing
formats, select this menu line and press <F4>:
Formats
MISC.
Date format 17-OCT-1995
Time format 17:04 (24 h)
Esc Modify OK
Date format
Time format
Press <F4> to select one of the four possible formats.
Press <F4> to select the other format.
If you change the formats, e.g. after a month they are transferred to all
data stored at the time.
6-26 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Misc. ...
6.7.1.2 Date/time
To change the date or the time, select this menu line and press <F4>; the mask appears in
which you can overwrite the date.
6.7.1.3 Language
In the titrator all texts are stored in English, German, French, Italian and Spanish. To change
the active language, select this menu line and press <F4>:
Language
MISC.
Active language English
Esc Modify OK
Active language
Press <F4> to select one of the languages.
6.7.1.4 Report header
If you wish to enter an identifier or a text (with attached keyboard) which should appear in the
header of every report of a titration method, select this menu line and press <F4>:
Report header
MISC.
Text:
Text:
Esc OK
6.7.1.5 Titrator ID
If you wish to enter an identification for your titrator – this is always recorded – select this menu
line and press <F4>; the mask appears in which you can enter an identification.
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 6-27
Misc. ...
6.7.1.6 Audio signal
An audio signal confirms either every keystroke or draws your attention to instructions, notes
and error messages. When the titrator is delivered, all these parameters are assigned a signal.
To change this, press <F4>:
Audio signal
MISC.
Keystroke Yes
Results Yes
Messages Yes
Esc Modify OK
Keystroke
Results
Messages
Yes: A brief signal sounds after every keystroke.
If you do not wish to hear this, press <F4>: → "No“.
Yes: A brief signal sounds after every result that appears in the display.
If you do not wish to hear this, press <F4>: → "No“.
Yes: A signal draws your attention to error messages, notes and instructions.
If you do not wish to hear this, press <F4>: → "No“.
6.7.2 Adjust measuring inputs
The measuring inputs for sensors and temperature sensors must be adjusted to ensure the
titrator measures potential and temperature values correctly.
The adjustment of the measuring inputs is performed by METTLER TOLEDO and the calibration
data of the characteristic line of each input recorded on a data sheet, which is enclosed with
every sensor option (see Section 10.5.1).
The measuring inputs of the pH option installed in slot 1 of the titrator in the factory are adjusted;
the calibration data are stored and can be printed out (see mask).
Measuring inputs
MISC.
Option in slot 1
Option in slot 2 1)
Esc Print Modify OK
1) appears only if a sensor option is installed in the slot
6-28 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Misc. ...
Notes
1. If you receive a data sheet with a pH option which contains instructions for the entry of the
calibration data, you must follow these: If, for example, you install this or an additional option
in slot 2, you must enter the data of the data sheet for slot 2 (overwrite default values!).
Calibration data
MISC.
Sensor 1: Offset 0
Slope 30950
Temp 1 (Pt100): Offset -15463 ∇
Esc OK
2. If you receive a data sheet with a pH option which contains only calibration data, these are
stored on the option itself. They are automatically adopted for every slot and can not be
deleted. In this case you can only view the data:
Calibration data
MISC.
Sensor 1: Offset 0
Slope 30950
Temp 1 (Pt100): Offset -15463 ∇
Esc OK
3. You always receive a data sheet with a KF option which contains only calibration data, in
other words the data are stored on the option and are automatically adopted for each slot;
they can not be deleted:
Calibration data
MISC.
Sensor 2 (Ipol): Offset 0
Slope 26025
Sensor 2 (Upol): Offset 0 ∇
Esc OK
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 6-29
Misc. ...
6.7.3 Shorten analysis sequence
During the sequence of an analysis, information is shown in the display which you must confirm
with OK to allow the titrator to continue working:
• the mask "Defined are" appears at the start (see Section 5.1), and
• on completion, the "Result list" for every titrated sample within a series (see Section 5.2).
You can exclude these two masks from the analysis sequence to accelerate it.
Shorten analysis sequence
MISC.
Defined resources
Result list for each sample
Yes
Yes
Esc Modify OK
Defined
resources
Result list
for each sample
Yes: The mask "Defined are" appears before the determination of a
sample or sample series. If this is not required, select "No" with
<F4>.
Yes: Within a sample series the result list appears after the determination
of each sample if you are working with titration stand 1, 2 or external.
If this is not required, select "No" with <F4>.
With series which you perform at a sample changer (titration stand:
ST20A) or at an auto stand, the mask is skipped automatically (see
Sections 5.1 and 5.4).
6-30 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Misc. ...
6.7.4 Define user level
You can decide if all people who work with the titrator should also have access to the Setup
or Method menu, i.e. have the possibility to delete resources or modify methods. If you have,
e.g. temporary staff who can perform only routine analyses, it is good practice to block their
access to these menus.
Define user level
MISC.
User level
Expert
Esc Modify OK
Expert
Routine
Expert
In the factory setting of the titrator, you have access to all menus. To
change this, select "Routine" with <F4>.
• In the Setup menu you can only print out the defined resources.
• In the Method menu you can only print out the methods.
• In the Data transfer menu, you can not
- transfer data to the computer
- remote-control the titrator
- transfer data to and from the memory card
- delete data on the memory card
- format a memory card.
• In the Analysis menu you can not modify an ongoing method (DL55/
DL58: see Section 5.3).
To switch to the expert level at a later date,
– open this mask,
Define user level
MISC.
User level
Routine
Esc OK
– and while holding down the Shift key, press key 1 three times: "Expert"
and the "Modify" command reappear.
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 6-31
Misc. ...
6.7.5 Activate control input
The RS option has a socket whose two TTL inputs you can activate (see Section 10.7). To do
this, you must have
• installed the option in the third slot,
• attached the devices whose signal should be scanned.
Activate control input
MISC.
Input In 3.1
Esc Modify Start
Input Press <F4> to activate the other input: "In 3.2".
Start: The input signal of the attached device is scanned until you press
<F5> (Stop). The input signal Off corresponds to a voltage of 5 V,
On to a voltage of 0 V.
Example: A level sensor is attached via a TTL input to check the amount
of liquid in a container. The setup is designed for a maximum level of
500 mL:
TTL socket
Level sensor
500 mL
• If the container holds more than 500 mL, the sensor sends the input
signal Off.
• If the container holds less than 500 mL, the sensor sends the input
signal On to indicate that the container should be filled.
6-32 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Misc. ...
6.7.6 Activate control output
The RS option has a socket whose four TTL outputs you can control (see Section 10.7). To
do this, you must have
• installed the option in the third slot
• attached the devices you intend to control.
Note: TTL are signals which do not transfer power. To operate devices such as a pump,
you must connect an amplifier in the circuit (see Section 10.7)
Activate control output
MISC.
Output Out 3.1
Control mode On/Off
Esc Modify Start
Output Press <F4> to select: "Out 3.2", "Out 3.3", "Out 3.4" or Out 3.1".
Control mode
The following modes are possible (press <F4>):
• On/Off
• Input controlled
• Sequential
On/Off
Start: The attached device is controlled until you press <F5> (Stop). The
output signal On (Start) corresponds to a voltage of 0 V, Off (Stop)
to a voltage of 5 V.
Example: A pump is attached via a TTL output. The pump is switched on
(it dispenses) with "Start" and switched off with "Stop".
Input controlled
Input controlled
MISC.
Input In 3.1
Output signal Normal
Esc Modify OK
Press <F4> to select the other input: "In 3.2".
Press <F4> to select the other signal possibility: "Inverted" (see following
page).
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 6-33
Misc. ...
Control mode
Example: A pump is attached via a TTL output, a level sensor via a TTL
input:
TTL socket
Pump with
supply vessel
Container with
level sensor
The titrator checks at the start whether the signal of the selected input is
in Section 6.7.5).
With this voltage of 0 V the titrator controls the pump (normal output
signal) so that it starts to pump liquid from the supply vessel into the
container with the level sensor.
As soon as this container holds more than 500 mL, the pump is switched
off automatically.
Inverted: The pump can be controlled only with the 5 V signal instead of
the normal output signal of 0 V. When the pump is switched off,
the voltage is 0 V.
Sequential
You can use this special mode to control, e.g. a dispenser.
Sequential
Number of pulses 10
Pulse duration [s] 1.0
Interval [s] 1.0
Esc
MISC.
OK
Each pulse causes the dispenser to dispense a quantity of liquid. This
depends on the pulse duration, which you can find on the data sheet of
the dispenser.
6-34 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Misc. ...
Control mode
1st pulse
Interval
Off
3rd pulse
Pulse duration
On
Start: The dispensing starts and is stopped automatically when the
sequence has been processed. You can also terminate the dispensing
with <F5> (Stop).
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 6-35
Data
Transfer
6.8 Data transfer
With the DL53, DL55 and DL58 you can back up your data on a memory card. You can also
transfer the data to a computer or control the titrator from the computer; for this you can enter
a bar-code string.
Memory card
Computer 1)
Remote control 1)
Bar-code string 1)
1) These parameters do not appear until you have defined the computer in the Setup
menu (see Section 2.7.3)
6.8.1 Memory card
You can use a memory card with the DL53, DL55 and DL58 to
1. • store user methods and resources when the user data memory is full
• back up all data of the user data memory, i.e. resources and user methods.
You can reload these data when needed on the same or on another titrator of the same
type. The cards you can use are listed in Section 12.8, insertion information in Section
10.6).
2. • store reports of analyses.
You can print these out from the card at any time.
DL50: You can use only a memory card supplied by METTLER TOLEDO to load a new software
version.
Formatting a new card
– Remove the write protection of the card and insert card.
– Confirm "Memory card" with OK: The message "Card not readable" appears. If you confirm
this with OK, the following appears:
Format card
DATA TRANSFER
All data will be deleted!
Esc
Start
6-36 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Data
Transfer
Start: The card is formatted and the end of the operation displayed.
Note: Formatting of a 1 MB SRAM card takes approx. 30 seconds, that of a 2 MB Flash card
approx. 4 minutes.
The formatting generates a directory and commands.
– Confirm "Memory card" in the Data Transfer mask again:
Display directory
Copy from titrator to card
Copy from card to titrator
Format card
Directory
Directory
DATA TRANSFER
User methods
Resources
Memory copies
Reports
Esc Print OK
Print: A list of the stored methods is printed out.
OK: A list of the methods, resources or memory copies appears if they
are on the card. You can then print out or delete an individual
method, the list of selected resources or the list of the memory
copies.
Reports: If you have selected "Memory card" as output in the function or
auxiliary function Report, the selected results or curves are stored under
the method ID and the date, e.g.
Reports (series)
DATA TRANSFER
90001 18-Oct-1996 11:48
235 22-Oct-1996 14:34
90002 22-Oct-1996 10:03 ∇
Esc Delete OK
<F5> shows the list of the reports for each sample of the series (see
following page):
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 6-37
Data
Transfer
Directory
Reports (samples)
DATA TRANSFER
1 [1] 18-Oct-1996 11:48
2 [1] 18-Oct-1996 12:01
3 [1] 18-Oct-1996 12:23
Esc Delete Print
1, 2 and 3 refer to the sample No., [1] to the number of the Report function
(index, relevant only with special methods). [*] appears if the report has
been sent by the auxiliary function Report (Report key).
<F4> prints out the results and/or curves defined in the function.
From titrator...
Copy to card
DATA TRANSFER
User methods
Resources
Memory copy
Esc OK
User methods: With <F5> the following appears:
Copy method
DATA TRANSFER
Method ID
Esc Modify Start
Enter the identification of the method; if you do not know this, press <F4>:
The list of user methods appears from which you can select the one you
wish to copy.
• If a method with the same ID is stored on the card, an appropriate
message appears.
Resources: <F5> is used to show the list of resources from which you
can select those you wish to copy. Resources stored on the card will be
overwritten.
6-38 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Data
Transfer
From titrator...
Memory copy: With <F5> the following appears:
Copy
DATA TRANSFER
All data of the user data memory
will be copied on the card.
Esc Start
You use this procedure to generate a backup copy of the user data
memory.
Start: All stored resources and user methods are copied and stored as
"memory copies" on the card with date and time (see following
page). During the copying operation, the message "Data are transferred"
appears.
From card...
Copy to titrator
DATA TRANSFER
User methods
Resources
Memory copies
Esc OK
User methods: Use <F5> to show the mask of the method ID. If you do
not know the identification, press <F4>: The list of user methods appears
from which you can select the one you wish to copy.
• If a method with the same ID is stored in the user data memory, an
appropriate message appears.
• If the sample data list contains data, you can not copy the method until
you have deleted these data (see Section 4).
Resources: Use <F5> to show the list of resources from which you can
select those you wish to copy.
• Titrants stored in the titrator are overwritten if their names and concentrations
are the same as the titrants being copied. All other titrants
remain stored.
• Sensors and solvents stored in the titrator are overwritten if their names
are the same as those being copied.
• All other resources are overwritten (temperature sensors, auxiliary
values, titration stands, peripherals).
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 6-39
Data
Transfer
From card…
Memory copies: With <F5> the following appears (example):
Load memory copy
10-Oct-1996 16:45
11-Nov-1996 10:12
12-Dec-1996 14:34
Esc
DATA TRANSFER
∇
Start
Load
DATA TRANSFER
Notice: All data of the user data
memory will be overwritten!
Esc Start
Start: All data of the user data memory are overwritten by those of the
memory copy of Oct. 10, 1996.
• If the sample data list contains data, you can not load the copy
until you have deleted these data (see Section 4).
Format card
To delete all data on the card in one operation, you can reformat the card
(see start of this section).
6.8.2 Computer
You can transfer user methods and/or defined resources from the titrator to a computer. A
requirement for this is that you have
• defined the computer in the Setup menu and attached it
• loaded the DLWin or the LabX titration software or an appropriate program.
– Switch on the computer and start required program.
– Confirm "Computer" with OK:
Computer
DATA TRANSFER
Transfer from titrator to computer
Esc OK
6-40 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Data
Transfer
From titrator... Transfer to computer DATA TRANSFER
User methods
Resources
Memory copy
Esc OK
User methods: Use <F5> to show the mask with the method ID. If you
do not know the identification, press <F4>: the list of user methods
appears from which you can select the one you wish to transfer.
Resources: Use <F5> to show the list of resources from which you can
select those you wish to transfer.
Memory copy: With <F5> the following appears:
Transfer
DATA TRANSFER
All data of the user data memory
will be transferred to the computer
Esc Start
Start: All data except those in the sample data memory are transferred.
During the operation, the message "Data are transferred" appears.
Note: You can also transfer data stored in the computer back to the titrator. You will find further
information in the Operating Instructions of the computer interface or in the Operating
Instructions of the DLWin or the LabX titration software.
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 6-41
Data
Transfer
6.8.3 Remote control
You determine whether the titrator is controlled exclusively by a computer. The requirement
for this is that you have loaded the DLWin or the LabX titration software or an appropriate
program.
Remote control
DATA TRANSFER
Status
Off
Esc Modify OK
Status
Press <F4> to select "On": The titrator can be controlled only by the
computer, i.e. you can neither enter data on the titrator, nor start an
analysis nor execute an auxiliary function.
When you have confirmed "On", all that appears is:
Data transfer
DATA TRANSFER
Remote control
Bar-code string
Esc OK
If the sample data list contains data or an auxiliary function is running, an
appropriate error message appears beforehand.
To follow the entries and commands of the computer in the display of the titrator, you can use
the following keys:
• Arrow keys → Scroll parameters of a mask
• Run key → Display of the active function of an ongoing method
• Sample key → Sample data list
• Data transfer key → Switch off remote control
Enter a bar-code string (see next section).
You will find further information on remote control in the Operating Instructions of the computer
interface or in the Operating Instructions of the DLWin or the LabX titration software.
6-42 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Data
Transfer
6.8.4 Bar-code string
This parameter is intended for automatic systems. The string which is read in by a defined and
attached bar-code reader is immediately transferred to the computer (see Section 2.7.3).
If you do not have a bar-code reader, you can enter the string here with the titrator keyboard.
Enter bar-code string
DATA TRANSFER
>
Esc OK
When you confirm the entry with OK, the string is transferred to the
computer.
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 6-43
DL58
DL58
Contents
Page
7. The DL58 .................................................................................................... 7-3
7.1 Additional resources ................................................................................ 7-4
7.1.1 Control inputs (only with TTLIO) .............................................................. 7-4
7.1.1.1 Modify....................................................................................................... 7-4
7.1.1.2 Add ........................................................................................................... 7-5
7.1.2 Control outputs (only with TTLIO) ............................................................ 7-5
7.1.2.1 Delete ....................................................................................................... 7-6
7.1.2.2 Modify....................................................................................................... 7-6
7.1.2.3 Add ........................................................................................................... 7-6
7.2 Special methods........................................................................................ 7-7
7.2.1 Additional commands ............................................................................... 7-7
7.3 Additional functions ................................................................................. 7-10
7.3.1 Temperature ............................................................................................. 7-10
7.3.2 Instruction................................................................................................. 7-11
7.3.3 Pump ........................................................................................................ 7-11
7.3.4 Two-phase titration ................................................................................... 7-12
7.3.4.1 Titrant/Sensor ........................................................................................... 7-12
7.3.4.2 Predispensing .......................................................................................... 7-13
7.3.4.3 Titrant addition.......................................................................................... 7-14
7.3.4.4 Measure mode ......................................................................................... 7-14
7.3.5 Control input ............................................................................................. 7-15
7.3.6 Control output........................................................................................... 7-16
7.3.7 Sync ......................................................................................................... 7-18
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 7-1
DL58
DL58
7. The DL58
This section explains the resources, commands and functions for the development of methods
which the other three titrators do not possess. With the preloaded METTLER and standard
methods, you can change only the parameters of the functions. The DL58 memory holds
additional, Special methods which allow the insertion of identical or new functions or the
deletion of existing functions thus enabling you to develop completely new methods.
Setup menu
Method menu
Special methods
Resources Commands Functions
Control inputs Add Temperature
Control outputs Delete Instruction
Pump
Two-phase titration
Control input
Control output
Sync
Note: You can also run the special methods with the DL53 and DL55 if you purchase a dongle
and the special methods available on memory cards (see Section 11.2).
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 7-3
Control inputs
DL58
7.1 Additional resources
"Control inputs" and "Control outputs" are available in the Setup menu as additional resources
which you must define.
7.1.1 Control inputs (only with TTLIO)
You can use the parameters of this resource only if you have installed an RS option in the third
slot. The two inputs In 3.1 and In 3.2 of the TTLIO socket are used for the attachment of devices
whose signal should be scanned (see Sections 6.7.5 and 10.7).
If you select this menu and press <F4>, the following appears:
Control inputs (only with TTLIO)
SETUP
Fill level sensor
Esc Add Modify OK
Notes
1. When the titrator is delivered, the English name for the fill level sensor is defined. Should
you wish to store the Spanish name, use the Modify command (see Section 7.1.1.1).
2. If you have added device names to the list, the Delete command appears (<F2>).
7.1.1.1 Modify
Press <F4>, "Modify", to show the parameters of the control input whose name and input you
can change.
Control input parameters
SETUP
Name Fill level sensor
Input In 3.1
Esc Modify OK
Name
You can change the name by pressing <F4> and selecting a new name
from the list of input names which appears.
Input Select with <F4>: "In 3.1" or "In 3.2".
7-4 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
DL58
Control outputs
Storage procedure
The definition of the control input is stored when you confirm both the parameter mask and the
list of names of the control inputs with OK. The modification of a parameter is always stored
when you press a Menu or Auxiliary function key (see corresponding paragraph at the end of
Section 2.1.2).
7.1.1.2 Add
With this command you can add a new device to the list of the defined devices. The following
mask always appears:
Control input parameters
SETUP
Name _ _ _ _ _ _ _ _ _ _ _ _ _ _
Input In 3.1
Esc Modify OK
You can modify the parameters by the procedure described in Section 7.1.1.1.
If you define, e.g. several devices with the same name, you must provide these with an
additional identification so that the titrator can distinguish them, e.g. Fill level sensor/2. If you
do not, a message appears (see Section 2.1.3).
7.1.2 Control outputs (only with TTLIO)
You can use the parameters of this resource only if you have installed an RS option in the third
slot. The four outputs Out 3.1...Out 3.4 of the TTLIO socket are used for the attachment of, e.g.
dosing equipment (see Sections 6.7.6 and 10.7).
If you select this menu and press <F4>, the following appears:
Control outputs (only with TTLIO)
SETUP
Pump
Dispenser
Valve
Esc Delete Add Modify OK
Note: When the titrator is delivered, the English name for the stored devices is defined. Should
you wish to store the Spanish name, use the Modify command (see Section 7.1.2.1).
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 7-5
Control outputs
DL58
7.1.2.1 Delete
If you press <F2>, "Delete", the device for the defined output will be deleted from the list.
7.1.2.2 Modify
Press <F4>, "Modify", to show the parameters of the control output whose name and output
you can change.
Control output parameters
SETUP
Name Pump
Output Out 3.1
Esc Modify OK
Name
You can change the name by pressing <F4> and selecting a new name
from the list of output names which appears.
Output Select with <F4>: "Out 3.1", "Out 3.2", "Out 3.3" or Out "3.4".
Storage procedure
The definition of the control output is stored when you confirm both the parameter mask and
the list of names with OK. The modification of a parameter is always stored when you press
a Menu or Auxiliary function key (see corresponding paragraph at the end of Section 2.1.2).
7.1.2.3 Add
With this command you can add a new device to the list of defined devices. The following mask
always appears:
Control output parameters
SETUP
Name _ _ _ _ _ _ _ _ _ _ _ _ _ _
Output Out 3.1
Esc Modify OK
You can change the parameters by the procedure described in Section 7.1.2.2.
If you define, e.g. several devices with the same name, you must provide these with an
additional identification so that titrator can distinguish them, e.g. Pump/2. If you do not, a
message appears.
7-6 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
DL58
Special methods
7.2 Special methods
If you press the Method key and scroll the menu downward (Shift and ∇-key), the following
appears:
Methods
METTLER methods
Special methods
METHOD
∇
Print OK
Special methods
METHOD
91001 Surfactant determination
91002 Vitamin C determination
Esc Delete Print Modify
You can change these two special methods to meet your requirements and store them under
a different method ID. Each new method is stored in the "Special methods" group.
7.2.1 Additional commands
You can modify the special method by changing or deleting its functions or by adding new
functions. Virtually all functions can occur more than once in a method.
Special method
Functions
delete
add
Parameters
modify
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 7-7
∇
Special methods
DL58
– Press <F4> when the surfactant determination is selected: The list of functions appears.
Method: 91001
METHOD
Title
Sample
Stir
Esc Add Modify OK
∇
Method: 91001
METHOD
Title
Sample
Stir
Two-phase titration
∇
Esc Add Modify OK
Method: 91001
METHOD
Sample
Stir
Two-phase titration
Calculation
∇
Esc Delete Add Modify OK
Add
Delete
You can insert a new function after every selected function (see following
page). If you select a wrong order when inserting functions (e.g. you can
not place the EQP titration function before the Sample function), an error
message appears, but not until the method is stored.
You can not delete the Title function as it is used to identify the method
and must always be placed at the head of the functions.
You can not delete the Sample function as it provides the titrator with
information regarding the titration stand at which titration is performed.
7-8 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
DL58
Special methods
If, e.g. Stir is selected and you press <F3>, the list of functions you can add appears:
Stir .............................. (6) see Section 3.3.3
Measure .............................. (10) see Section 3.3.4
Temperature .............................. (10) see Section 7.3.1
Instruction .............................. (10) see Section 7.3.2
Dispense .............................. (6) see Section 3.3.5
Pump .............................. (6) see Section 7.3.3
EQP titration .............................. (*) see Section 3.3.6
EP titration .............................. (*) see Section 3.3.7
Learn titration .............................. (*) see Section 3.3.8
EQP titration (Ipol/Upol) .............................. (*) see Section 3.3.9
EP titration (Ipol/Upol) .............................. (*) see Section 3.3.10
Two-phase titration .............................. (*) see Section 7.3.4
pH/mV-stat .............................. (2) see Section 3.3.11
Control input .............................. (6) see Section 7.3.5
Control output .............................. (6) see Section 7.3.6
Calculation .............................. (12) see Section 3.3.12
Auxiliary value .............................. (20) see Section 3.3.15
Titer .............................. (1) see Section 3.3.14
Calibration .............................. (1) see Section 3.3.13
Report .............................. (8) see Section 3.3.16
Sync .............................. (20) see Section 7.3.7
– Select the function you wish to add and press <F5>. It will be inserted under the Stir function.
Note: The number in brackets is the maximum possible number of this function a special
method can contain.
(*): A special method can contain maximum 6 titration functions.
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 7-9
Temperature
DL58
7.3 Additional functions
7.3.1 Temperature
You can use this function to measure the temperature of a solution under defined conditions.
The titrator acquires the measured value as raw result T (see Note a. in Section 3.3.4).
Sensor
Temp A
Unit of meas. °C
∆T [°C, °F, K] 0.02
∆t [s] 2.0
t(min) mode
Fix
t(max) [s] 30.0
Sensor
Press <F4> to select the temperature sensor: TEMP A, TEMP B, TEMP C
or TEMP D.
Unit of meas. Press <F4> to select the unit: °C, °F or K.
∆T
∆t
t(min) mode
The temperature drift must be less than ∆T/∆t (0.01 °C/s) during the
period ∆t (2 s) if the temperature is to be acquired as a measured value.
This occurs within a defined time interval t(min) and t(max).
t(min) is earliest time for the acquisition of the measured value. Select
t(min) from the selection menu:
t(min) mode
METHOD
Fix
T > set value
T < set value
Esc
Modify
Fix: With <F4> the mask appears in which you can change t(min).
T greater than set value: Instead of a fixed time you can select a condition:
The measured value must be greater or
T less than set value: less than a specified set value: <F4> is used
to show the mask in which you can enter the
set value [°C, °F, K].
t(max)
is the latest time for the acquisition of the measured value.
If you have selected one of the conditions for t(min), the titrator starts the
next function when the measured temperature T is greater than (less
than) the set value and the drift condition is met, but at the latest after
t(max).
7-10 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
DL58
Instruction/Pump
7.3.2 Instruction
This function allows you to intervene manually in the sequence of the titration method, in other
words the method is interrupted. The entered text appears at the desired position during the
current method in the display. The method does not continue until you confirm the instruction
with Run.
Instruction
METHOD
Text
Text
Text
Esc
_ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _
OK
Text
Enter the desired instruction.
7.3.3 Pump
With the aid of this function you can add solvent with a pump attached to the sample changer.
Pump
METHOD
Solvent
H 2 O
Volume [mL] 10
Stir
No
Esc Modify OK
Solvent
Volume
Stir
With <F4> select the solvent from the list which contains those you have
defined in the Setup menu (see Section 2.8).
Enter the volume that should be added. You have defined the pump rate
of the device in the Setup menu.
Select whether (Yes) or not (No) stirring should be effected during the
dosing operation.
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 7-11
Two-phase titration
DL58
7.3.4 Two-phase titration
This section explains the first 4 parameter groups of this titration function. The explanations
for those of the group Recognition, Termination and Evaluation can be found in Sections
3.3.6.5 through 3.3.6.7.
Titrant/Sensor
Predispensing
Titrant addition
Measure mode
Recognition
Termination
Evaluation
The titrator records several measured values and volumes as raw results (see Section 8.1).
7.3.4.1 Titrant/Sensor
To modify the titrant or sensor, press <F4>:
Titrant
Hyamine
Concentration [mol/L] 0.004
Sensor
DP550
Unit of meas.
mV
Titrant
Concentration
Sensor
Unit of measurement
Press <F4> to open the list containing the titrants you have defined in the
Setup menu (see Section 2.1). Confirm the one you want with OK.
The defined concentration of the titrant is entered automatically.
Press <F4> to open the list containing the sensors you have defined in
the Setup menu (see Section 2.2.). Confirm the one you want with OK.
The defined measurement unit is entered automatically.
If you select the unit from the selection menu, you should select mV or the
unit defined in the Setup menu, otherwise you will receive an error
message when the method is started.
7-12 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
DL58
Two-phase titration
7.3.4.2 Predispensing
Predispensing shortens the titration time. You can select one of two predispensing modes
when you press <F4>.
Predispensing
METHOD
to volume
to (factor x sample size)
No
Esc
Modify
Volume
Factor x
sample size
No
You dispense a specified volume [mL] which you can enter when you
press <F4>. Instead of a number, you can enter a formula.
You can also enter a wait time: After the predispensing, the titrator waits
for this time to elapse before it adds the titrant in a controlled manner.
Notice: The wait time includes the time of the dosing operation!
You dispense to a specified volume that is calculated from the product of
the sample weight or volume and a factor; you can enter this factor when
you press <F4>.
You can also enter a wait time (see "Volume").
You do not wish to predispense.
The titrator adds the volume in one step. It adopts the potential values
ET1 and ET2.
E [mV]
ET1: Potential at the start of predispensing
or the titration
ET2: Potential after predispensing
ET2
ET1
V [mL]
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 7-13
Two-phase titration
DL58
7.3.4.3 Titrant addition
For the two-phase titration, only the incremental addition mode is available.
Titrant addition
METHOD
∆V [mL] 0.05
Esc
OK
∆V
The volume increment that the titrator adds is constant (see Section
3.3.6.3: Incremental titrant addition).
7.3.4.4 Measure mode
You define the duration of the mixing and separation time as well as the stirrer speed during
the separation time.
Measure mode
Mixing time [s] 15.0
Separation time [s] 50.0
Stirrer speed (sepn time) [%] 10
Esc
METHOD
OK
Mixing time
Separation
time
After the addition of every increment, the two phases are stirred for this
time at the speed selected in the Stirrer function.
During this time, the two phases should separate so that a stable
measured value of the corresponding phase can then be acquired.
Stirrer speed You can set your "own" speed for the separation time or set it to 0.
Note: If you use the auxiliary function Stirrer to
• change the speed during the mixing time, this applies only to the
mixing time during the two-phase titration and for any functions
after the two-phase titration.
• change the speed during the separation time, this applies only
to the separation time during the two-phase titration.
7-14 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
DL58
Control input
7.3.5 Control input
With this function you can influence the progress of a method externally via the inputs of the
TTLIO socket. This is possible only with an RS option installed in the third slot (see Section
10.7). The titrator waits for the defined signal to be detected or the maximum time to elapse
before it processes the next function.
Control input
METHOD
Name
_ _ _ _ _ _ _ _ _ _
Input signal Rising
Max. time [s] 10.0
Esc Modify OK
Name
Input signal
Max. time
With <F4> open the list containing the names of the inputs you have
defined in the Setup menu (see Section 7.1.1). Confirm the one you wish
to use with OK.
With Rising or Falling you specify the direction of change of the input
signal when the signal you have defined is detected.
This time defines the wait time for detection of the signal. On elapse of this
time, processing of the method continues even if no signal has been
detected.
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 7-15
Control output
DL58
7.3.6 Control output
With this function you can generate TTL signals. This is possible only with an RS option
installed in the third slot (see Section 10.7). The nature of the signal depends on the control
mode you have selected. The standard status of the control outputs is "high".
Control output
METHOD
Name
_ _ _ _ _ _ _ _ _ _
Control mode Fixed time
Instruction
Esc Modify OK
Name
Control mode
With <F4> open the list containing the names of the outputs you have
defined in the Setup menu (see Section 7.1.2). Confirm the one you wish
to use with OK.
The mode determines the number and nature of the generated TTL
signals. The following modes are possible (press <F4> ): "Fixed time",
"On/Off", "Input controlled" or "Sequential".
Fixed time
Select this mode when you wish to operate a device that is attached to the
control output over a fixed time period.
– Enter the time during which the control output should be switched on.
On/Off
With this mode you can, e.g. switch on a device attached to the control
output at the start of the method and switch it off at the end of the method.
– Select whether the control output should be switched on or off. The
status of the control output will be automatically set to off on completion
of the method.
Input controlled
This mode allows a device attached to the control output to be switched
on or off by a device attached to the control input (e.g. pushbutton). The
function is ended when the signal at the defined control input changes or
after elapse of a maximum time.
7-16 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
DL58
Control output
Control mode
Input controlled
METHOD
Name
_ _ _ _ _ _ _ _ _ _
Output signal Normal
Max. time [s] 10.0
Esc Modify OK
Name: With <F4> open the list containing the names of the inputs you
have defined in the Setup menu (see Section 7.1.1). Use OK to confirm
the one which should control the control output function.
Output signal "Normal": The signal is routed directly from the control
input to the control output.
Output signal "Inverted": The signal of the control input is first inverted
and then routed to the control output.
Max. time: The maximum time defines the wait time for the signal change.
On elapse of this time, the method is processed further even if no signal
has been sent.
Sequential
Use this mode to select devices which you can use to either trigger
various actions through different sequences or repeat particular actions
through single pulses of the sequence (see Section 6.7.6).
Sequential
Number of pulses 10
Pulse duration [s] 1.0
Interval [s] 1.0
Esc
METHOD
OK
Number of pulses: Enter the number of pulses of the desired sequence.
Pulse duration: Enter the time during which a pulse should be switched
on.
Interval: Enter the wait time between the pulses.
Instruction
You can enter a text which is shown while the function is active.
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 7-17
Sync
DL58
7.3.7 Sync
Under this function you specify the parameters for the synchronization of the titrator with an
external control unit. The synchronization takes place via the system interface (RS option in
the slot 3!).
Sync
METHOD
Sync mode Send
Code 1
Comment
Esc Modify OK
Sync mode
Code
Comment
Press <F4> to select: "Send" or "Send/Wait".
Send: The titrator transmits the numeric code you have defined to the
external control unit and then begins to process the next method function
immediately.
Send/Wait: The titrator transmits the numeric code you have defined to
the external control unit and then waits until this sends back the same
code. Only then does the titrator start to process the next method function.
Enter a number between 1 and 32, which is sent to the external control
unit.
Enter a comment, which appears on the display when the function is
executed.
Notes
1. You will find additional information regarding the communication between the titrator and
the external control unit in the Operating Instructions "DL50/DL53/DL55/DL58: RS232C
Interface Description".
2. Two titrators can also be synchronized by means of the system interface using the Sync
function. You must then connect the two titrators using a null modem.
7-18 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Symbols and explanations
Contents
Page
8. Symbols and explanations ........................................................................ 8-3
8.1 List of symbols ........................................................................................... 8-3
8.1.1 Compilation of the raw results ...................................................................... 8-8
8.2 Use of indices ............................................................................................. 8-9
8.2.1 Compilation of the symbols by indexing forms ............................................. 8-12
8.3 Evaluation procedures .............................................................................. 8-14
8.3.1 Standard....................................................................................................... 8-14
8.3.2 Minimum/maximum ...................................................................................... 8-14
8.3.3 Segmented ................................................................................................... 8-15
8.3.4 Asymmetric .................................................................................................. 8-16
8.4 Examples of formulas ................................................................................ 8-17
8.4.1 Results ......................................................................................................... 8-17
8.4.2 Constants ..................................................................................................... 8-18
8.4.3 Formulas for restriction of the equivalence point ......................................... 8-19
8.5 Restrictions in the analysis....................................................................... 8-20
8.5.1 Maximum number of samples ........................................................................8-20
8.5.2 Maximum number of sample series in the sample data memory ................. 8-20
8.5.3 Maximum number of equivalence points per sample determination ............ 8-20
8.5.4 Maximum number of measured values per Titration function ...................... 8-20
8.5.5 Maximum number of results per sample series ........................................... 8-20
8.5.6 How long does the titrator store data? ......................................................... 8-20
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 8-1
8-2 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
List of symbols
8. Symbols and explanations
In this section you will find additional information which supplements Sections 3 to 7.
8.1 List of symbols
This list shows you the abbreviations of all symbols of the parameters and the raw results
determined by the titrator together with their definition. If you enter the symbols using an
external keyboard, you must comply with the uppercase or lowercase notation.
Function/Resource Symbol Definition
Titrant c Nominal concentration of the titrant in mol/L
t
Titer of the titrant
N
c * t: actual equivalent concentration of the titrant
in mol/L (ACTUAL value); it is calculated automatically
by the titrator.
Sensor ZERO Zero point of the pH electrode
SLOPE Slope of the pH electrode
Auxiliary value H Auxiliary value : H1 to H20
Sample
(Sample data)
m Sample size of the weight in g or volume in mL (for
Calculation function)
M
Molar mass in g/mol
z
Equivalent number: Number of reaction entities
compared to the titrant
Note: For technical reasons, the notation z* of DIN
standard 32 625 has not been used for the titrator.
f
Correction factor
Current sample TIME Time of a sample determination in seconds
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 8-3
List of symbols
Function Symbol Definition
EQP titration
Learn titration
EP titration(Ipol/Upol)
ET1 Initial potential of the measured solution before the
first addition of titrant in the defined unit of the
sensor used
EQP titration (Ipol/Upol)
ET2 Potential after predispensing in the defined unit of
the sensor used
Two-phase titration*
VEQ Titrant consumption in mL up to the equivalence
point or end point; the equivalence points are
numbered consecutively
Q Titrant consumption in mmol up to the equivalence
point or end point; the equivalence points are
numbered consecutively. Q = VEQ * c * t
VEX Calculated excess of titrant in mL
QEX Calculated excess of titrant in mmol
VEND Total titrant consumption in mL (VEQ + VEX)
QEND Total titrant consumption in mmol (Q + QEX)
EPOT Calculated equivalence point potential of the equivalence
or end point VEQ in the defined unit of the
sensor used
EQP titration EHNV Calculated half neutralization value (potential at
EQP titration (Ipol/Upol)
VEQ/2) referred to the equivalence point VEQ in
the defined unit of the sensor used
Two-phase titration*
neq Number of equivalence points found (applies to
one Titration function)
P1 Potential 1 in the defined unit of the sensor used
(evaluation parameter)
P2 Potential 2 in the defined unit of the sensor used
(evaluation parameter)
VP1 Titrant consumption in mL up to potential P1
VP2 or potential P2
QP1 Titrant consumption in mmol up to potential P1
QP2 or potential P2
* Two-phase titration: Special methods
8-4 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
List of symbols
Predispensing to
a: Volume or
(factor x sample size)
b: Potential
c: Slope
Example of an EQP titration
with 3 equivalence points
E [mV]
EPOT3
EHNV3
EPOT2
EHNV2
EPOT1
EHNV1
E [mV]
neq = 3
2
1
VEQ1 VEQ2
VEQ3
Q1 Q2 Q3
VEND
QEND
3
VEX
QEX
V [mL]
+200
P1
+100
0
-100
-200
P2
VP1/QP1
VP2/QP2
V [mL]
Example of an EQP titration
with evaluation of the defined
potential values
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 8-5
List of symbols
Function Symbol Definition
pH/mV-stat VTOT Total titrant consumption in mL
QTOT Total titrant consumption in mmol
VT1 Titrant consumption in mL up to time limit t1
VT2 or time limit t2
VT
Titrant consumption in mL up to a different time
limit
QT1 Titrant consumption in mmol up to time limit t1
QT2 or time limit t2
QT Titrant consumption in mmol up to a different time
limit
CSTAT Correlation coefficient of the regression lines between
t1 and t2 of the V – t curve
VSTAT Mean titrant consumption in mL/min within the
time limits t1 and t2
QSTAT Mean titrant consumption in mmol/min within the
time limits t1 and t2
V [mL]
∆V/∆t V/t [mL/s]
VTOT
QTOT
VT2
QT2
VT1
QT1
VSTAT
QSTAT
t1
t2
t [s]
t1
t2
t [s]
Note: VSTAT/QSTAT is determined by means of linear regression of the measured values
between t1 and t2. CSTAT is the correlation coefficient of this linear regression. For a
correct result, CSTAT should be greater than 0.95.
8-6 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
List of symbols
Function Symbol Definition
Measure E Measured value in mV or in the defined unit of the
sensor used (does not apply to polarized sensors)
Dispense VDISP Total dispensed volume in mL
QDISP Total dispensed amount of substance in mmol
Calculation R Calculated result
C
Calculation constant
Cx
Calculation constant (applies only to the auxiliary
function "Perform calculations")
x
Mean value of the statistics evaluation (calculated
only when the Statistics parameter is defined)
s
Standard deviation
srel Relative standard deviation in %
Temperature* T Measured temperature in °C, °F or K of the Pt
sensor used
* Temperature: Special methods
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 8-7
List of symbols
8.1.1 Compilation of the raw results
Analysis/Function Raw results Printed out when "Raw
results" is selected in the
report function
Current sample TIME no
Measure E yes
Dispense VDISP, QDISP yes
EQP/EP titration ET1, ET2 no
Learn titration
VEQ, Q
yes
EQP/EP titration (Ipol/Upol)
VEX, QEX
yes
Two-phase titration*
VEND, QEND
no
EPOT
yes
EQP titration EHNV no
EQP titration (Ipol/Upol) neq
no
Two-phase titration*
P1, P2 yes
VP1, QP1
yes
VP2, QP2
yes
pH/mV-stat VTOT, QTOT yes
VT1, QT1
no
VT2, QT2
no
VT, QT
no
VSTAT, QSTAT
yes
CSTAT
no
Temperature* T yes
Notes:
1. * Two-phase titration, Temperature: Special methods
2. To obtain the raw results which are not printed out as standard, assign these to the result
R, e.g. R = QP1, R = VT1.
8-8 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Use of indices
8.2 Use of indices
As some functions can occur a number of times within a method and parameters can also
appear a number of times within a function, indices are needed in addition to the parameter
symbol (see Section 8.2.1: Compilation of the symbols by indexing forms).
The titrator recognizes 4 different indexing forms:
1. Parameters without index
These parameters include, e.g. c, t, M, ZERO, SLOPE, TIME. The parameters always refer
to the current titrant, the current sensor or the current sample determination.
The parameters of the functions which occur only once within the standard methods also
do not need an index, e.g. E of the Measure function or VSTAT of the pH/mV-stat function.
2. Parameters of the form Xi
Parameters which use an index without brackets are all those whose link with the method
function is meaningless. This indexing form includes R, C and H.
Example: R3: third result (see fourth indexing form)
Note: R = R1 holds, in other words if the index = 1 it may be omitted. This also applies to
the indexing forms under points 3 and 4.
3. Parameters of the form X [j]
Index j is the function counter. This indexing form includes, e.g. VDISP, QEX and x.
Example: QDISP[2]: Measured value of the second Dispense function (extract from
standard method "2 Step titration (EQP)")
Dispense
Stir
After confirmation of the "Current sample", the titrator dispenses
and determines the raw result QDISP of the first Dispense function.
The titrator executes the Stir function, then
EQP titration
the EQP titration followed by the
Calculation
Calculation
first calculation function. To obtain the measured value QDISP as
a result, set R1 = QDISP[1].
The titrator executes the second calculation function, then
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 8-9
Use of indices
Report
the Report function before
Dispense
Stir
it dispenses again and determines the raw result QDISP of the
second Dispense function.
The titrator runs the second Stir function, then
EQP titration
the second EQP titration, followed by the
Calculation
third Calculation function. To obtain the measured value QDISP
as a result, set R3 = QDISP[2].
4. Parameters of the form Xi [j]
Index i is the number of times a parameter should be run within a function. Index j is the
counter for the number of functions within a method. This indexing form includes Q, VEQ,
EHNV and EPOT.
Example: Q2[2]: mmol consumption up to the second equivalence point of the second
EQP titration function. Extract from the standard method "2 Step titration
(EQP)", which has been adapted for the determination of HCl, CH 3 COOH
and NH 4 Cl with NaOH. (The 2 titration functions were selected as other
values have been defined for the equilibrium controlled measured value
acquisition in the determination of CH 3 COOH and NH 4 Cl.)
EQP titration
E [mV]
The titrator executes the first EQP
titration function and titrates until the
equivalence point of HCl. It determines,
among other things, the mmol
consumption Q1[1] and the mmol
excess QEX[1].
EQP1
Q1[1] ≡ Q
QEX[1] ≡ QEX.
V [mL]
Q1[1]
QEX[1]
8-10 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Use of indices
Calculation
Formula .................................. R1 = Q1[1]
Constant.................................
Decimal places ...................... 4
Result unit .............................. mmol
Result name ........................... HCl
Statistics................................. Yes
In the first Calculation function, you
define the mmol consumption for
HCl.
As you can omit the Index 1, the
formula can also be R = Q.
EQP titration
E [mV]
EQP1
The titrator executes the second
EQP titration function. It determines,
among other things, the mmol consumption
Q1[2] up to the first equivalence
point (CH 3 COOH), then the
mmol consumption Q2[2] up to the
second equivalence point (NH 4 Cl).
EQP2
V [mV]
Q1[2]
Q2[2]
Calculation
Formula .................................. R2 = Q1[2]+QEX[1]
Constant.................................
Decimal places ...................... 4
Result unit .............................. mmol
Result name ........................... CH 3 COOH
Statistics................................. Yes
In the second (third) Calculation function,
you define the mmol consumption
for acetic acid.
Here, you take into account the titrated
excess of the first EQP titration
function.
Calculation
Formula .................................. R3 = Q2[2]
Constant.................................
Decimal places ...................... 4
Result unit .............................. mmol
Result name ........................... NH 4 Cl
Statistics................................. Yes
In the third (fourth) Calculation function
you define the mmol consumption
for NH 4 Cl.
Note: The two Dispense functions of the standard method are skipped (0 mL has been defined
as volume in the method); the second Calculation function is skipped (no formula has
been defined in the method for the result).
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 8-11
Use of indices
8.2.1 Compilation of the symbols by indexing forms
Analysis/Resource / Symbol Unit Index Examples
Function
Current sample TIME [s]
Titrant c [mol/L]
t
Sensor ZERO [mV, pH ...]
SLOPE [mV/ mV, pH ...]
– –
Sample m [g] or [mL]
M
[g/mol]
z
f
Auxiliary value H H1…H20
Calculation R Xi R3
C
C2
Measure** E [mV, pH ...]
Temperature* T [°C, °F, K]
Dispense VDISP [mL]
QDISP [mmol] X[j] QDISP[2]
EQP/EP titration ET1 [mV, pH, µA...]
EQP/EP titration
ET2 [mV, pH, µA...]
(Ipol/Upol)** VEX [mL]
Two-phase titration*
QEX [mmol] QEX[2]
VEND [mL]
QEND [mmol]
8-12 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Use of indices
Function Symbol Unit Index Examples
EQP titration neq neq[3]
EQP titration(Ipol/Upol)** P1
[mV, pH, µA...]
Two-phase titration* VP1 [mL]
QP1 [mmol]
P2
[mV, pH, µA...]
VP2 [mL]
QP2 [mmol]
EP titration VEQ [mL]
EP titration (Ipol/Upol)** Q [mmol] Q[2]
EPOT [mV, pH, µA...]
pH/mV-stat** VTOT [mL]
X[j]
QTOT [mmol]
VT1 [mL] VT1[2]
QT1 [mmol]
VT2 [mL]
QT2 [mmol]
VSTAT [mL/min]
QSTAT [mmol/min]
CSTAT
Calculation x x[3]
(Statistics)
s
srel
EQP titration VEQ [mL]
EQP titration(Ipol/Upol)** Q [mmol] Q2[1]
Two-phase titration* EPOT [mV, pH, µA ...] Xi[j] EPOT1[2]
EHNV [mV, pH, µA ...]
Note:
The parameters of the functions which occur only once in a method do not need an
index in "[ ]".
* Temperature, Two-phase titration: Special methods;
** Measure, pH/mV-stat, EQP titration (Ipol/Upol), EP titration (Ipol/Upol): two or more
of these functions are possible only with special methods.
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 8-13
Evaluation procedures
8.3 Evaluation procedures
The titrator has various evaluation procedures for precise determination of the equivalence
points of a titration curve:
• Standard
• Minimum/maximum
• Segmented
• Asymmetric
8.3.1 Standard
You use this procedure for all S-shaped titration curves. The evaluation uses an iterative
procedure (nonlinear regression) [1]. The titration of a strong acid with a strong base is used
as a mathematical model. With this model, the determined equivalence point always lies in the
vicinity of the inflection point.
E
Inflection point
At least five measured points around
the the inflection point are used for the
evaluation. The slope from measured
point to measured point must increase
or decrease. If this condition is not met,
the iterative procedure can not be employed.
In such a case, the inflection
point of the titration curve is determined
by interpolation. A reference to
this effect appears in the report of the
"raw results".
V
8.3.2 Minimum/maximum
The result of this evaluation is the calculated minimum (maximum) from the measured points
of the titration. The classical example of a titration curve with a minimum is the determination
of surfactants with photometric indication.
The minimum (maximum) is calculated by a polynomial approximation of the titration curve in
the region of the minimum (maximum). The equivalence point is recognized directly using the
data of the titration curve.
8-14 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Evaluation procedures
8.3.3 Segmented
Different indication methods (e.g. photometry, conductometry and amperometry) generate
titration curves with linear or approximately linear segments (segmented curves). The titrator
also evaluates these curves. The evaluation procedure used is based on the following
consideration:
E
∆E/∆V
∆ 2 E/∆V 2
Titration Titrierkurve curve
V
1st derivative
1. Ableitung
V
2nd derivative 2. Ableitung
V
The first derivative of a segmented
curve has the typical shape of an S-
shaped curve, its inflection point represents
a good approximation of the
equivalence point.
Segmented curves are evaluated with
the Standard procedure, which uses
the calculated data of the first derivative
instead of the data points of the
titration curve.
The equivalence point recognition is
thus effected not with the calculated
first derivative but with the calculated
second derivative. The threshold for
the equivalence point recognition also
refers to data of the second derivative.
The individual segments do not need
to be exactly linear. Decisive for an
accurate determination of the equivalence
point is the presence of a clear
kink between the individual segments
of the titration curves.
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 8-15
Evaluation procedures
8.3.4 Asymmetric
With pronounced asymmetric curves, the standard procedure can lead to a systematic error.
The difference between the true equivalence point and the inflection point can then be greater
than the precision normally attained. For these cases, the titrator is equipped with the evaluation
based on the Tubbs method [2].
This empirical approximation method is a tried and tested procedure for the evaluation of
asymmetric, analog recorded titration curves. It can also be applied to titration curves
measured digitally. The result of the Tubbs evaluation provides a better approximation of the
true equivalence point than the inflection point.
The evaluation procedure is based on the following considerations:
E
M
1
M
2
Intersection
point EQP
V
A circle of curvature with minimum radius
can be drawn on both branches of the
titration curve. The ratio of the two radii
is determined by the asymmetry of the
curve. The mean intersection point of
the line connecting the centers of the
circles M1 and M2 with the titration curve
represents the sought-after equivalence
point. Theoretical calculations show that
the true equivalence point with asymmetric
titration curves always lies between
the inflection point and that branch
of the titration curve which has the greater
curvature (the smaller radius of curvature).
In the titrator the calculation is implemented according to Ebel [3]. This involves approximating
those parts of the titration curve which lie in the region of the greatest curvature by a hyperbola.
The vertex is determined for each approximated hyperbola. This point on the hyperbola lies
at the site of greatest curvature. The mid-points of the assigned smallest radii of curvature are
the focal points of the two hyperbolae. As in the graphical variant, the intersection point of the
line connecting the two focal points with the titration curve results in the required equivalence
point.
For the evaluation at least six measured points both before and after the inflection point of the
titration curve are required in the region of the greatest curvature. If the titration curve has a
profile which does not allow calculation of the radii of curvature, the titrator calculates the
equivalence point by the standard procedure. This will be indicated in the report of the "raw
results".
[1] K. Waldmeier and W. Rellstab, Fres.Z.Anal.Chem., 264, 337, (1973)
[2] C.F. Tubbs, Anal. Chem., 26, 1670 (1954)
[3] S. Ebel, E. Glaser, R. Kantelberg and B. Reyer, Fres. Z. Anal. Chem., 312, 604 (1982)
8-16 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Examples of formulas
8.4 Examples of formulas
8.4.1 Results
R = Q ∗ C / m (standard formula) Content of a sample with weighing
C = f (M, z, unit) % – ppm, – mg / g, – TAN [mg KOH / g], –
mol / kg, – mmol / g
Content of a sample solution
% (g/mL), – g / L, – ppm, – mg / L, – g / 100mL,
mol / L, – mmol / L
R = m / (VEQ ∗ c ∗ C)
R = Q ∗ C
R = Q
R = VEQ
R = VEQ / m
R = (QDISP – Q) ∗ C / m
R = (Q – Hj) ∗ C / m
R = (Q / m – Hj) ∗ C
R = ET1 [2]
R = pw(–E) ∗ 1000
R = lg(–E)
Titer determined with primary standard
Titer determined with volumetric standard
Content per sample
mmol consumption as result
mL consumption as result
mL consumption per unit weight or per unit volume
as result
Back titration:
QDISP: dispensed amount of substance in mmol of
the Dispense function
Q: mmol consumption up to the equivalence point
or end point of an EQP or EP titration function.
Solvent blank value included in the calculation
(blank value is stored as auxiliary value Hj )
Matrix blank value included in the calculation
[mmol / g] (matrix blank value is stored as auxiliary
value Hj)
Initial potential of the second EQP or EP titration
function as result
Ion concentration in mmol/L, measured with an ionselective
electrode
Extinction of a solution, measured as % transmission
with a phototrode
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 8-17
Examples of formulas
8.4.2 Constants
Weight or volume of the sample known
C = M / z
Unit
mg / g or g/L
C = M / (10 ∗ z) %
C = M / (z ∗ 10 ∗ ρ) ... (ρ = density of the solution)
C = M ∗ 1000 / z
% (g/100 mL)
ppm (mg/kg)
C = M ∗ 1000 / (z ∗ ρ)...(ρ = density of the solution) ppm (mg / L)
C = 1 / z mol / kg (mmol / g)
C = 56.1 TAN and TBN (mg KOH / g)
C = M / (1000 ∗ z)
C = 1000
C = 1
Titer
mmol / L
mol / L
Volume and weight of the sample unknown
C = M / z
C = 1 / z
mg
mmol
8-18 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Examples of formulas
8.4.3 Formulas for restriction of the equivalence point
To calculate the titrant consumption in mmol of a particular equivalence point when a titration
curve has several equivalence points, you can have Q identified by a condition, e.g.
R1 = Q(EPOT > -50)
R1 = Q(EPOT > -50)[2]
R1 = Q(EPOT ~ 100)[2]
R1 = Q(EPOT ~ P1)
R1 = Q(EPOT ~ H11)
R1 = Q(P1 < EPOT < 300)
Titrant consumption up to the equivalence point of the 1st
EQP titration function whose potential is above -50 mV. Here,
the equivalence point which first meets this condition is determined.
Equivalence points whose potential lies between, e.g.
-51 and -300 mV are not considered.
Titrant consumption up to the equivalence point of the 2nd
EQP titration function whose potential lies above -50 mV.
Titrant consumption up to the equivalence point of the 2nd
EQP titration function whose potential is nearest to 100 mV.
Titrant consumption up to the equivalence point of the 1st
EQP titration whose potential is nearest to potential P1.
Titrant consumption up to the equivalence point of the 1st
EQP titration function whose value is nearest to the value
stored under auxiliary value H11.
Titrant consumption up to the equivalence point of the 1st
EQP titration lying within the range of potential P1 and 300 mV.
The equivalence point which first meets this condition is determined.
Note: You can also define these conditions with VEQ or EPOT, e.g. R 1= EPOT(VEQ ~ 5).
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 8-19
Restrictions
8.5 Restrictions in the analysis
8.5.1 Maximum number of samples
The sample data of maximum 60 samples can be entered.
8.5.2 Maximum number of sample series in the sample data memory
DL50/DL53:
DL55/DL58:
Maximum 1 sample series can be entered.
Maximum 3 sample series can be entered.
8.5.3 Maximum number of equivalence points per sample determination
Per sample determination, the titrator can determine maximum 16 equivalence points (EQP
titration), divided between one or two EQP titration functions. If there are more than 16
equivalence points, these will not be determined, but the determination will be continued.
8.5.4 Maximum number of measured values per Titration function
The titrator can store 300 measured values per titration function (EQP, EP, Learn, EQP (Ipol/
Upol), Two-phase); it then aborts the function. With the titration functions EP (Ipol/Upol) and
pH/mV-stat, the number of measured points will be reduced automatically (see Sections
3.3.10 and 3.3.11).
8.5.5 Maximum number of results per sample series
The titrator can store 180 results per sample series; it then aborts the series.
Example: If you titrate 40 samples with a method, the titrator can store 4 results for each
individual sample.
8.5.6 How long does the titrator store data?
1. Measured values of the titration functions
The titrator stores the measured values up to the next titration function or up to the start of
the next sample determination.
2. Raw results
The titrator stores all raw results up to the titration of the next sample within a series.
3. Results
The titrator stores all results of a sample series up to the start of a new series.
8-20 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Error messages and malfunctions
Contents
Page
9. Error messages and malfunctions .......................................................... 9-3
9.1 Error messages from the titrator ............................................................. 9-3
9.2 Other errors and malfunctions ................................................................ 9-6
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 9-1
Error messages and malfunctions
9-2 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Error messages and malfunctions
9. Error messages and malfunctions
9.1 Error messages from the titrator
You yourself can rectify all errors which the titrator displays as messages. Some of these
messages refer to this section (errors 1...6).
However, if the EPROM or RAM test following startup of the titrator failed ("test failed"), the only
measure available is: "Press any key to continue". The program memory or the program data
memory is faulty.
After any key has been pressed, the titrator attempts to continue the startup process: This may
be successful, but does not necessarily guarantee 100% perfect operation.
Measure: Please contact METTLER TOLEDO service!
1. Internal ERROR: ... ← (error which should be noted down, e.g. PB, M1)
Parts of the hardware may be faulty.
Measure: – Before you contact METTLER TOLEDO service,
• note the fault and the operation which initiated it,
• note the instrument number of the titrator (rear panel)
• print out the system information, which provides details of the titrator
equipment:
– Press the shift and <F4> keys to print out the information.
• Finally, inform the service of these points and the attached peripherals with
the defined configuration.
2. Memory faulty
Parts of the user data memory are faulty.
Measure: – Call METTLER TOLEDO service to have the memory changed. Meanwhile,
you can still continue working with the titrator.
3. Faulty data deleted
a. The titrator has stored only parts of a method, e.g. as the power failed during storage
of this method. It deletes the entire method (it is also possible that it has deleted several
methods). As a result, the sample data of this method which are stored in the sample
data memory will also be deleted.
Measure: – Confirm the message.
– Check which of your methods have been deleted and reenter.
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 9-3
Error messages and malfunctions
3 b. The titrator has stored only parts of parameters of a resource, e.g. as the power failed
during storage of these parameters. It deletes the entire list of this resource (e.g. all
titrants or all sensors).
Measure: – Confirm the message.
The titrator now loads the default list of the deleted resource, e.g. all
titrants stored in the titrator in the factory.
– Check which list has been changed:
If the user data memory has insufficient storage space, only the titrants
which can be accommodated in the available space will be stored.
If the memory is full, the titrator will not store any titrants. In such a case,
you must delete data of other resources or one of your methods to create
memory space.
– Switch the titrator off then on again.
– Check that all resources are again available.
If the error messages appears repeatedly, inform METTLER TOLEDO service!
4. Storage not possible
a. The titrator can not assign the titer, the auxiliary value or the calibration data to the
corresponding resources as the list is missing (titrant, sensors or auxiliary values).
Measure: – Confirm the message.
– Check whether the list of the resource has been deleted.
– Switch the titrator off then on again: The titrator will then reload the default
list of the corresponding resource, e.g. all the titrants which were stored
in the titrator in the factory.
If the user data memory has insufficient storage space, only the titrants
which can fit in the available space will be stored.
If the memory is full, the titrator will not store any titrants. In such a case,
you must delete other resources or one of your methods to create
storage space.
– Switch the titrator off then on again.
– Check that all resources are again available.
b. The titrator can no longer store, e.g. data as the user data memory is full. This is
possible
• with resources which you wish to add in the Setup menu
• in the case of a method you generate in the Method menu
• with sample data you wish to enter in the Sample menu
• on storage of a changed, active method
• on storage of the evaluation parameters of the titration functions EQP, EQP (Ipol,
Upol) and Two-phase of an active method (parameter: "Stop for reevaluation").
9-4 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Error messages and malfunctions
Measure: – Confirm the message.
– Delete methods or resource data or copy to a memory card.
4 c. The titrator performs a learn titration. If the method for this is stored on the memory
card or in the computer, the parameters of the function are not stored.
Measure: – Confirm the message.
– Always perform a learn titration with a method stored in the titrator and
do not remote-control the titrator!
If the error message appears repeatedly, contact METTLER TOLEDO service!
5. Error: Memory card
a. An error appears during copying.
Measure: – Confirm the message.
– Check that the card is inserted properly and restart the copying process.
– If the process is aborted again, reformat the memory card and attempt
to repeat the copying process (the data on the card will be deleted).
– If the process is again aborted, use a new memory card.
b. An error appears when a memory copy is loaded.
Measure: – Confirm the message.
– Check that the card is inserted properly and restart the copying process.
– If the process is again aborted, switch the titrator off then on again.
– Use a new memory card.
6. Measuring inputs not adjusted (refers only to a pH option on which the calibration
data are not stored, see Section 6.7.2).
The calibration data for the characteristic line of the measuring inputs are missing.
Measure: – Enter the data from the data sheet enclosed with the pH option (see Section
6.7.2).
When the measuring inputs are adjusted, this message appears if system data have been
deleted in the user data memory.
Measure: – Switch titrator off then on again: the system data will be reloaded.
– Reenter zero point and slope for all measuring inputs.
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 9-5
Error messages and malfunctions
9.2 Other errors and malfunctions
The following compilation of errors and malfunctions are not reported by the titrator and may
frequently help you rectify the faults yourself without the need to contact METTLER TOLEDO
service.
If you have to contact the service, please print out the system information to obtain details of
the titrator equipment and pass these data on to the service:
– Press the Shift and <F4> keys to print out the information.
Fault Possible cause Rectification
No display on titrator Titrator not connected to power Connect to power supply,
supply
if fault persists contact
METTLER TOLEDO Service
Several points of the display
missing
Contact METTLER TOLEDO
Service
Display does not match the
pressed key
Contact METTLER TOLEDO
Service
Stirrer does not rotate Stirrer not properly assembled Check stirrer and seating of
or sensors block it at the
the sensors
titration stand
Transfer error to attached Peripheral faulty or switched Check attached device is
peripheral off functioning properly
Device (printer, balance, Device not switched on Switch on device
terminal) at Centronics or
RS interface do not react Wrong settings Settings and
Configuration (switch settings) configuration must match
wrong (see Section 2.7)
Burette does not move to Burette drive faulty Contact METTLER TOLEDO
zero position when switched on
service
Wrong potential or Electrode faulty Check electrode (see
pH values
electrode data sheet)
Calibration data wrong
Check defined data
Use new electrode
No dispensing, the titrant Burette tip clogged Clean burette tip
is discharged from stopcock
or piston Follower cam on burette Insert follower cam correctly
mounted wrongly (see Section 10.2.3)
9-6 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Installation and maintenance
Installation and maintenance
Contents
Page
10. Installation and maintenance ......................................................................10-3
10.1 Installing the burette drive ..........................................................................10-3
10.2 Installing and maintaining the interchangeable burette...........................10-4
10.2.1 The DV1001, DV1005, DV1010, DV1020 burettes ........................................10-4
10.2.2 Equipping the burette .....................................................................................10-5
10.2.3 Inserting the burette .......................................................................................10-6
10.2.4 Maintaining the burette parts..........................................................................10-7
10.2.5 General information........................................................................................10-8
10.3 Equipping the titration stand ......................................................................10-9
10.4 Rear view of the titrator .............................................................................10-10
10.5 Installing the sensor, Centronics or RS option ....................................... 10-11
10.5.1 Calibration data for the characteristic curve of the measuring inputs ..........10-12
10.6 Inserting the memory card ........................................................................10-12
10.7 Using TTL inputs and outputs ..................................................................10-13
10.8 Attaching a keyboard.................................................................................10-15
10.8.1 Attaching a bar-code reader.........................................................................10-16
10.9 Configuration of the terminal (DEC VT340) .............................................10-16
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 10-1
Installation and maintenance
Burette drive
10. Installation and maintenance
– Switch the titrator off and disconnect the power cable before you open the
housing! An electric shock could be lethal.
– You can clean the housing of the titrator with a cloth moistened with ethanol.
10.1 Installing the burette drive
Position 2
Screw
– Unscrew the burette guide at position 2.
– Pull out the ribbon cable that is located in the titrator
below this second opening.
Note
Please place the burette drive on the second opening.
DL50/DL53
The burette drive is defined as the middle position by the software
(drive 2), in other words you can not control a drive installed at
position 1.
DL55/DL58
For the first titration, which you perform in accordance with the
instructions in the Quick Guide, the 2nd burette drive is defined.
– Plug the coupling of the ribbon cable in the connector
of the burette drive.
Coupling
Connector
Ribbon cable
Ribbon cable
– Position the burette drive on the second opening of
the titrator – from the first opening press the ribbon
cable downward at an angle (ensure it is not pinched!)
– and fasten with the screw.
– Screw the burette guide to the first opening.
DL55/DL58
If you wish to use a second burette drive, follow the
above procedure:
– Unscrew the burette guide at position 1 etc.
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 10-3
Interchangeable burette
Installation and maintenance
10.2 Installing and maintaining the interchangeable burette
10.2.1 The DV1001, DV1005, DV1010, DV1020 burettes
The 5, 10 and 20 mL burettes differ only in the size of their cylinder, centering ring and piston,
whereas the 1 mL burette has a different construction: its piston is longer, piston guide and
cylinder holder ''replace'' the centering ring of the other burettes. Instead of the locking screw,
its glass cylinder is fastened with an O-ring and a knurled nut.
1 mL burette 5 / 10 / 20 mL burette
Connection tubing
Knurled nut
O-ring
Stopcock
Stopcock
Cylinder holder
Locking screw
Glass cylinder
Piston
Piston rod
Piston guide
Protection tube
Piston
Piston rod
Burette housing
Lip seals
Centering
ring
10-4 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Installation and maintenance
Interchangeable burette
10.2.2 Equipping the burette
– Lay gasket on bottle and screw on burette holder.
– Insert stopper (or a drying tube with holder).
Caution: If you use a stopper as protection for the titrant, always use the stopper with flat
side! Otherwise a partial vacuum forms in the bottle (order no. 23646).
– Push suction tubing into the bottle, slide the red PVC tubing over the fitting as kink protection,
and fasten other end to left connection of burette head.
– Screw dispensing tubing into connection at right and place burette tip in tip holder.
Suction tube
Dispensing tube
Tip
holder
Burette
assembled
Burette holder
Stopper
Gasket
1 L bottle
Fitting
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 10-5
Interchangeable burette
Installation and maintenance
10.2.3 Inserting the burette
2
3
1
– Slide the burette onto burette drive 2 with
the stop of arrestment knob (1) located on
left side.
– Check that piston rod (2) is correctly positioned
in push rod (3).
– Fix burette by turning the arrestment knob
to the right.
Note
Before sliding the burette onto the titrator,
check the exact position of the stopcock and
the piston rod.
If the piston has been pushed too far into the
cylinder, carefully take it out a short way. Then
press the burette onto the burette holder thus
positioning the piston exactly. Piston must
project 7 mm!
Position of the
valve
Position of the
piston
10-6 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Installation and maintenance
Interchangeable burette
10.2.4 Maintaining the burette parts
Depending on the titrant, you should clean the burette cylinder, piston, stopcock and tubing
relatively often.
– Slide the burette off the titrator, invert it so that the stopcock points towards you and carefully
take out the piston; this causes the burette contents to flow out through the suction tubing
(waste or titrant bottle!).
– In the same position, turn the stopcock through 90° clockwise and any liquid in the stopcock
will flow out through the dispensing tubing (waste bottle!).
– Unscrew the suction, dispensing and connection tubing.
1 mL burette
– Unscrew the knurled screw of the burette
and remove the O-ring from the glass cylinder
using tweezers.
– Unscrew the holder of the glass cylinder
and take out the cylinder.
Caution: Do not misplace O-ring!
5, 10, or 20 mL burette
– Unscrew the locking screw of the burette
and take out the glass cylinder.
Caution: Do not misplace the centering
ring of the burette housing!
– Press the cam of the stopcock inward and
lift the stopcock up and out.
Cam
• Depending on the contamination caused by the titrant, rinse cylinder and tubing with acids
or deionized H 2 O then with ethanol and finally dry the parts with oil-free compressed air or
vacuum.
• Rinse the stopcock with solvents or deionized H 2 O only! Then dry it with oil-free compressed
air.
• Never place O-rings in organic solvents!
• Never attempt to remove any crystals in the cylinder by scratching with a hard object! Pipe
cleaners or Q tips are more suitable.
• Never put the parts in a drying oven whose temperature is above 40 °C!
– Replace the piston if it leaks or is badly scored at the edge. Pay special attention to crystal
formation between the lip seals of the piston if you work with NaOH/KOH and KF solutions!
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 10-7
Interchangeable burette
Installation and maintenance
10.2.5 General information
You can remove air bubbles at the piston surface by taking the burette out of the guide and
lightly tapping its base. This causes air bubbles to rise to the top. Then rinse the burette.
You can remove air bubbles in the tubes by tapping the tubes with your fingers while titrant is
being siphoned off or dispensed. If this does not help, undo the suction tube so that titrant flows
back into the bottle, screw it on again and rinse the burette (air purging). In obstinate cases
undo both tubes, rinse with deionized H 2 O and ethanol and dry using oil-free compressed air
or vacuum.
Titrants such as KMnO 4 or KOH in MeOH can easily crystallize in the tip of the burette and block
it. If you have no immediate use for a titrant, it is best to empty and clean the dispensing tube:
Unscrew connection to let the titrant flow out. Check the threaded connection and wipe off any
drops.
KF titrants are subject to outgassing, above all at elevated room temperatures (SO 2 ), which
leads to the formation of air bubbles in the tubes and in the burette (stopcock). You should thus
rinse these burettes before titrations!
10-8 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Installation and maintenance
Titration stand
10.3 Equipping the titration stand
3
9a
– Screw on titration arm (1) with nurled
screw (2).
– Insert stirring rod (3) and attach propeller
stirrer (4) from below to rod.
2
13
1
8
10
5
6
4
7
9
12
11
– Press spacing ring (5) onto titration
head, place clamping ring (6) in
threaded ring (7) and screw into
place.
– To install a titration vessel, turn the
threaded ring half a turn to the left,
attach the vessel and tighten threaded
ring.
– Insert burette tip (8) and electrode
(9) diagonally opposed – this ensures
better control in the titration –
and close the remaining openings
with stoppers (11 & 12).
(9a) is the plug-in electrode cable.
(10): This opening is intended for the
rinsing bottle connection.
(13) is the electrode holder.
– Always test the titration vessel for firm seating in the titration head! If it falls
off, you could injure yourself if working with toxic titrants and solvents or
strong acids or bases.
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 10-9
Rear view
Installation and maintenance
10.4 Rear view of the titrator
Slot:
No. 1 No. 2 No. 3 No. 4
Input:
Reference electrode
Sensor inputs:
Sensor 1
Sensor 2
Temperature sensor
input: Temp 1
Stirrer outputs:
Stirrer 1
Stirrer 2
Analog outputs
TTLIO socket
TTLIO socket (used by titrator)
RS232 interface for sample changer
RS232 interface for balance
Centronics interface for printer
RS232 interface for system
pH KF RS Centronics option
Connector for
power cable
On/off
switch
The diagram shows a possible occupancy of all slots. The basic version of the titrator has the
following installed
• a pH option at slot No. 1
• an Centronics option at slot No. 4.
pH and KF options
You can install a sensor option in any slot. Each option has a number which is also affixed to
the data sheet enclosed with each sensor option (see Section 10.5.1).
DL53/55/58: You can use the inputs and outputs of maximum four sensor options, e.g. two pH
options at slot No. 1 and 2, two KF options at slot No. 3 and 4.
DL50: You can use the inputs and outputs of maximum one sensor option.
Centronics option
You can use the Centronics option only if it is installed in slot No. 4. The TTL inputs and outputs
of the socket can be used only for the commands preallocated by the titration software (see
Section 10.7).
10-10 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Installation and maintenance
Installing options
RS options
You can install an RS option either in slot No. 3 or slot No. 4. Although the options are identical,
the functions of the connections are not!
Slot No. 3: You can freely control the TTL inputs and outputs of the socket.
To the upper RS interface you can attach only a sample changer,
to the lower only a computer or a terminal.
Slot No. 4: You can use the TTL inputs and outputs of the socket only for the commands
stipulated by the titration software (see Section 10.7).
To the upper RS interface you can attach only a balance,
to the lower only a printer.
10.5 Installing the sensor, Centronics or RS option
– Switch the titrator off and disconnect the power cable before you unscrew
the cover plate of the slot! An electric shock could be fatal.
– Unscrew the cover plate of the slot.
– Insert the option in the two guide rails, push in option and
fasten with the two screws.
Notes
1. When you have installed a pH option that has not stored
the calibration data you should
• enter the calibration data for the characteristic line of the
measuring inputs (see Section 10.5.1).
• calibrate the electrodes/temperature sensors at the new
inputs (see Sections 2.2 and 6.1.3).
Soldering scheme
2. To attach temperature sensors of other companies, you
can use a Lemo cable plug (4-pin) and solder on the
appropriate cable (see Section 11.2: Accessories).
2
3
1
4
Shield
Pt100/Pt1000 sensor
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 10-11
Memory card
Installation and maintenance
10.5.1 Calibration data for the characteristic curve of the measuring inputs
When the titrator is delivered, the measuring inputs are calibrated for a sensor and temperature
sensor (Pt100 and Pt1000) of the pH option. The data sheet filed in Section 14 documents the
data for the characteristic line of the measuring inputs of this pH option.
If the data are not stored on the actual option, you must reenter them (see Section 6.7.2)
• if system data are lost (see Section 9.1)
• if you install this option in a different slot.
If you order an additional pH option, you receive a data sheet with the number of the option
and the calibration data for its measuring inputs. You must enter these data for correct potential
and temperature measurements if they are not stored on the actual option (see Section 6.7.2).
The data for the characteristic line of the measuring inputs are always stored on KF options!
Note: As the characteristic line of the measuring inputs can exhibit long-term drift, we advise
you to have the inputs recalibrated every 2 years by METTLER TOLEDO Service.
10.6 Inserting the memory card
Note: Ensure that you are not electrostatically charged when you perform this operation! We
recommend you touch the metallic part of the titrator housing before you insert the
memory card to ensure the charge is not transferred to the titrator interior.
1 2
– Swing open the cover (1) of the slot for the
memory card.
– Push in the memory card until knob (2) springs
out.
– To remove the card, press the knob back in.
10-12 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Installation and maintenance
TTLIO
10.7 Using TTL inputs and outputs
The TTLIO socket of the Centronics option has two inputs and four outputs (In 1/In 2 and
Out 1...Out 4). These are assigned by the titrator to commands which are triggered during an
analysis. You can wire these inputs and outputs mechanically or electronically to trigger the
defined commands by auxiliary units or systems (see following examples).
The RS option has the same TTLIO socket as the Centronics option. When it is installed in slot
3, the inputs and outputs are always available and freely configurable (see Sections 6.7.5 and
6.7.6). At slot 4, they are assigned by the titrator to commands which you trigger during an
analysis:
Socket
2
1
3
8 4
5
7 6
Assignment (see Section 12.10.2: Technical data)
GND (Pin 1)...... signal ground
In 1 (Pin 2)........
In 2 (Pin 3)........
triggers the start of the analysis or confirms the mask
"Current sample" (titration stand 1, 2 or external)
triggers the abort of the analysis (Reset)
Out 1 (Pin 4) .....
Out 2 (Pin 5) .....
Out 3 (Pin 6) .....
VCC (Pin 7) ......
Out 4 (Pin 8) .....
gives a pulse after the sample function
the output signal assumes a "low" status during the
titration; this does not change until the sample is analyzed
(on display of the result list)
the output signal assumes a "low" status in the case of an
error message; this does not change until the message
has been confirmed with OK
supply voltage
on request for the next sample of a series, this triggers a
pulse (only with titration stand 1, 2 or external)
Wiring mechanically
If you wire, e.g. pin 1 (ground) with pin 2 (input 1) by
attaching a manual event sensor, you can initiate the
start of the analysis with this switch. The mask
"Defined are" must be blocked from the analysis
sequence for this as it can not be confirmed (see
Section 6.7.3).
In
GND
Event sensor
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 10-13
TTLIO
Installation and maintenance
Wiring electronically
Input In 1 or In 2 Output Out 1, Out 2, Out 3 or Out 4
Titrator
external
Titrator
external
Out
In
GND
GND
GND
Output Out 1, Out 2, Out 3 or Out 4
Relay supply internal Relay supply external
(relay operate voltage: max. 3.5 V);
Icc: max. 60 mA
Titrator
external
Titrator
external
VCC
Relay
Relay
Out
Out
+
–
GND
Note: TTL are signals which do not transfer power. To operate, e.g. a pump it is necessary
to connect an amplifier in the circuit.
10-14 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Installation and maintenance
External keyboard
10.8 Attaching a keyboard
A DIN socket is located on the left side of the titrator and can be used for the attachment of an
external keyboard for alphanumeric entries.
– Attach the keyboard using the appropriate DIN cable (see Section 12.10: Technical data).
The following table shows the keys and key combinations which correspond to the command,
menu, auxiliary function and entry keys of the titrator:
Titrator
External keyboard
key key <Alt> + key <Ctrl>+ key
F1
F1
F2
F2
F3
F3
F4
F4
F5
F5
Setup
F12
Method
F10
Sample
F11
Run
F9
Sensor
e
Stirrer
s
Changer
c
Burette
b
Results
r
Report
p
Misc...
m
Data Transfer
d
Reset
Break (Pause)
Arrow key ∆ / Shift + ∆
↑ / Page up
Arrow key ∇ / Shift + ∇
↓ / Page down
Shift
Shift
1...9, . 1...9, . (b) 1...9 (a) 1...9
Del
Del
CE (Shift + Del)
Shift + Del
The characters /, -, +, =, etc., which are written with the Shift key, depend on the assignment
of the keyboard. Additional characters not on the titrator keypad are:
x → <Alt Gr> + x µ → <Alt Gr> + u ∆ → <Alt Gr> + d
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 10-15
Bar-code reader/terminal
Installation and maintenance
10.8.1 Attaching a bar-code reader
You can attach a bar-code reader to the external keyboard. To ensure this is recognized by
the titrator, you must program a "header" and "terminator" with the external keyboard (see
Operating Instructions of the keyboard in question). The titrator requires the key sequence
"Shift"/"Enter" for both parameters.
You can then read in, e.g. the bar code for the sample identification without having to define
the bar-code reader in the Setup menu.
You must define the bar-code reader if you wish to use it for direct data transfer to a computer
(see Section 2.7.3).
10.9 Configuration of the terminal (DEC VT340)
You have attached the terminal to the titrator and defined it (see Section 2.7.3). To configure
it, proceed as follows:
– Switch off the titrator.
– Switch on the terminal and wait until the message VT340 OK appears.
– Press the Set-Up key: The SET-UP DIRECTORY appears.
a. If you have already configured the terminal for other connections, first select Recall
Factory Default Settings to reactivate the default settings.
b. If you have not yet used the terminal, you can change some of the default parameters
directly.
– Select General Set-UP and change the following parameters:
Terminal mode
VT300-8bit
– Select Display Set-Up and change the following parameters:
Scrolling
jump
Status Display
host writable
– Select Communications Set-Up and change the following parameters:
Transmit Speed 9600 *
Receive XOFF Point 512
Character Format 8 bits, even parity *
* These parameters must match the defined terminal settings!
(See Section 2.7.3)
– Select Keyboard Set-Up and change the following parameters:
Keypad mode
numeric, so that the numeric keys of the numeric key field (on
extreme right of keyboard) are active.
10-16 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
∆
Installation and maintenance
Terminal
– Select Save Current Settings to save the modified parameters.
– Press the Set-Up key to quit the SET-UP DIRECTORY.
– Switch on the titrator: The terminal screen is now built up by the titrator.
If you have switched off both instruments,
– first switch on the terminal, then the titrator.
You can use the keyboard of the terminal to operate the titrator. The keys have the following
assignments:
Titrator
Terminal
key key numeric key field
F1
F2
F3
F4
F5
Setup
Method
Sample
Run
F6
F7
F8
F9
F10
F14
F12
F13
F11
Sensor 8
Stirrer 5
Changer 4
Burette 7
Results 3
Report 2
Misc... 1
Data Transfer 6
Reset
Arrow key ∆ / Shift + ∆
Arrow key ∇ / Shift + ∇ ∇ ∇
F20
↑ / Prev. screen
↓ / Next screen
Shift
Shift
1...9, . 1...9, .
Del
Remove
Shift +
←
Shift +
→
∆
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 10-17
Accessories
Contents
Page
11. Accessories .................................................................................................. 11-3
11.1 Standard equipment .................................................................................... 11-4
Small accessories .......................................................................................... 11-6
11.2 Optional accessories ................................................................................... 11-7
Burette drive ................................................................................................... 11-7
Interchangeable burettes ............................................................................... 11-7
Additional parts for the burette ....................................................................... 11-9
Titration stands ............................................................................................... 11-9
Additional parts for the titration stand........................................................... 11-11
Sensors ........................................................................................................ 11-13
Temperature sensors ................................................................................... 11-15
Accessories for Karl-Fischer titrations.......................................................... 11-15
Peripherals ................................................................................................... 11-16
Titration software .......................................................................................... 11-16
Miscellaneous .............................................................................................. 11-17
Options ......................................................................................................... 11-19
Documentation ............................................................................................. 11-19
11.3 METTLER TOLEDO literature and application brochures ...................... 11-20
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 11-1
Accessories
11. Accessories
All instrument components of and working aids for the titrator that form part of
• the standard equipment and
• the optional accessories
are listed below.
Each part that is listed with an order number can be ordered from METTLER TOLEDO.
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 11-3
Accessories
11.1 Standard equipment
Interchangeable burette
(DL50 Graphix/
DL50 Rondolino only)
Propeller stirrer
Rondolino
(DL50 Rondolino only)
Burette drive
○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○
Titration stand
(DL50 Graphix only)
Electrode holder
(DL53/DL55/DL58 only)
11-4 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Accessories
DL53
DL50 DL50 DL55
Graphix Rondolino DL58
Titration stand x
Rondolino
x
Electrode holder
x
Burette drive DV90 x x x
Interchangeable burette
(10 mL) DV1010* x* x*
Propeller stirrer
incl. 2 stirring rods x x x
pH option x x x
Centronics option x x x
RS option x x x
Titration Software LabX light x x x
Power cable x x x
Electrode cable x x x
Printer cable x x x
RS connection cable x x x
Reference Handbook x x x
Quick Guide x x x
Memo card x x x
Application broschure No. 12 x x x
Small accessories
(see next page)
* see optional accessories
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 11-5
Accessories
Small accessories
DL53
DL50 DL50 DL55
(Order No.) Graphix Rondolino DL58
2 knurled screws (25650) x x x
Threaded ring (25652)
Clamping ring (25653)
Spacing ring (23842)
Titration vessel (Set) x x
x
x
x
2 electrode holder (25654) x x x
3 stoppers ST 14.5 (23451) x x x
2 stoppers ST 7.5 (23452) x x x
Stopper holder (25662) x x x
Phillips screwdriver No. 2 (73072) x x x
Operating Instructions LabX light x x x
11-6 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Accessories
11.2 Optional accessories
The numbers enclosed in parentheses refer to the purchase order, e.g. of an interchangeable
burette. In case of additional orders some parts are available only in multipack form or as a
minimum quantity.
Burette drive
Order No.
Burette drive
DV90
Interchangeable burettes
Interchangeable burette, complete 1 mL DV1001
5 mL DV1005
10 mL DV1010
20 mL DV1020
comprising:
1 L brown glass bottle (1) 71296
and
Burette holder (1) 23645
and
Piston (1) DV1001 51107535
DV1005 51107115
DV1010 51107116
DV1020 51107117
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 11-7
HCl
HCl
c= mol/L c= mol/L
Date:
Date:
1/2 H2SO4
c= mol/L
AgNO3
c= mol/L
Date:
c= mol/L
Date:
c= mol/L
Date:
c= mol/L
Date:
c= mol/L
Date:
c= mol/L
Date:
c= mol/L
Date:
c= mol/L
Date:
c= mol/L
Date:
c= mol/L
Date:
c= mol/L
Date:
Date:
c= mol/L
Date:
c= mol/L
Date:
c= mol/L
Date:
c= mol/L
Date:
c= mol/L
Date:
c= mol/L
Date:
c= mol/L
Date:
c= mol/L
Date:
c= mol/L
Date:
c= mol/L
Date:
NaOH
c= mol/L
Date:
1/2 H2SO4
c= mol/L
Date:
c= mol/L
Date:
c= mol/L
Date:
c= mol/L
Date:
c= mol/L
Date:
c= mol/L
Date:
c= mol/L
Date:
c= mol/L
Date:
c= mol/L
Date:
c= mol/L
Date:
c= mol/L
Date:
NaOH
AgNO3
c= mol/L c= mol/L
Date:
Date:
KF
KF
c= mg H2O/mL c= mg H2O/mL
Date:
Date:
c= mol/L c= mol/L
Date:
Date:
c= mol/L
Date:
c= mol/L
Date:
c= mol/L
Date:
c= mol/L
Date:
c= mol/L
Date:
c= mol/L
Date:
c= mol/L
Date:
c= mol/L
Date:
c= mol/L
Date:
c= mol/L
Date:
c= mol/L
Date:
c= mol/L
Date:
c= mol/L
Date:
c= mol/L
Date:
c= mol/L
Date:
c= mol/L
Date:
c= mol/L
Date:
c= mol/L
Date:
Accessories
Order No.
and
Stopcock (1) with valve disk 1) made of:
PTFE (light-gray) (standard equipment of interchangeable burette) 51107537
ceramic (dark-brown) (optional accessories) 51107525
1) Both valve disks are chemically resistant. For 24-hour use with titrants that
have a tendency to cristallize out, we recommend the ceramic disk.
and
Light protection tube (1) 23644
and
Dispensing tube (1) with siphon tip 0.70 m 25687
and
Suction tube (1) 0.83 m 25688
and
Gasket (1) min. order quantity: 5 23981
for 1 L brown glass bottle
and
Tip holder (1) for burette tip 23960
and
Stopper ST 14.5 (1) min. order quantity: 5 23646
and
Set of labels (1) 51107506
11-8 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Accessories
Order No.
Additional parts for the burette
Dispensing tube with siphon tip 1.00 m 25961
Drying tube with cover 23961
Drying tube holder 23915
Molecular sieves 250 g 71478
Siphon tip set set of 5 23240
Adapter for bottles of: Merck, DE 23774
Fisher, US 23787
Bottle rack for two 1-liter 51107065
bottles with burettes
Titration stands
Titration stand, complete 51108760
Dual titration stand, complete
DV92
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 11-9
Accessories
Order No.
Titration stands 51108760 DV92
comprising:
Angle bracket (1) x 25655
Propeller stirrer x 51109150
incl. 2 stirring rods
Titration arm (1) x x 25651
Spacing ring (1) x x 23842
Clamping ring (1) x x 25653
Threaded ring (1) x x 25652
Knurled screw (2) x 25650
Electrode holder (2) x 25654
min. order quantity: 5
Stopper ST 14.5 (3) x 23451
min. order quantity: 5
Stopper ST 7.5 (2) x 23452
min. order quantity: 5
Titration vessel 100 mL x x 101974
polypropylene (2)
set of 1400
11-10 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Accessories
Order No.
Additional parts for the titration stand
Titration vessel 100 mL set of 1400 25777
polypropylene, red colored
Titration vessel 80 mL, glass set of 20 101446
Titration vessel 250 mL, glass set of 10 23515
Titration vessel 5-20 mL, glass 23516
Titration vessel for two-phase titration 51107655
incl. stop cock, threaded ring and gasket
Thermostatable titration vessel 80 mL, glass 23517
Plastic cover for titration vessels set of 16 101448
Heat exchanger for thermostating 23834
incl. adapter with taper joint
Gas inlet 23721
Rinsing unit, complete 23821
with titration head insert and stoppers
for unused openings in titration head
1 set of stoppers for rinsing unit 101230
Electrode holder 51108730
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 11-11
Accessories
Order No.
Propeller stirrer 51109150
incl. 2 stirring rods
Propeller stirring rod 101229
Micropropeller stirring rod 655073
(for titration vessel 23516)
Adapter cable 0.50 m 51107216
(mini DIN male / RCA female)
Peristaltic pump SP250 51108016
with Novoprene tubes, adapters, hose clamps
Novoprene tubes (1x 1 m + 10x 120 mm) 51190969
Fluorosilicone elastomer tubes (5x 120 mm) 51108149
for SP250
Diaphragm pump 51108012
incl. suction tube
Heating system DH100
110 V 51108779
230 V 51108780
Dispensing unit DU200
EU version 51370200
US version 51370210
11-12 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Accessories
Order No.
Sensors
Combined pH electrode for
titrations in aqueous solutions
DG111-SC
Combined pH electrode for small volumes
in small titration vessels in aqueous solutions
DG101-SC
Combined glass electrode with movable sleeve
frit for titrations in nonaqueous solutions
DG113-SC
Combined glass electrode with movable sleeve
frit for titrations in aqueous solutions
DG114-SC
Combined glass electrode with sleeve
frit for titrations in aqueous solutions
DG115-SC
Combined platinum ring electrode for
redox titrations
DM140-SC
Combined silver ring electrode for
argentometric titrations
DM141-SC
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 11-13
Accessories
Order No.
Ion selective measuring electrodes
• Fluoride ISE DX219 51089931
• Chloride ISE DX235 51089933
• Nitrate ISE DX262 51089934
• Sodium ISE DX223 51089930
• Potassium ISE DX239 51089932
• Lithium ISE DX207 51107673
• Ammonia GSE DX217 51107677
• Ammonium ISE DX218 51107679
• Magnesium ISE DX224 51107684
• Cyanide ISE DX226 51107681
• Sulfide ISE DX232 51107675
• Calcium ISE DX240 51107683
• Bromide ISE DX280 51107671
• Fluoroborate ISE DX287 51107676
• Silver ISE DX308 51107682
• Cadmium ISE DX312 51107672
• Iodide ISE DX327 51107680
• Barium ISE DX337 51107674
• Surfactant sensitive electrode DS500 51107670
Reference electrode for
• ion selective electrodes DX200 51089935
• surfactant sensitive electrodes Inlab 301 52000128
Triaxial cable (Electrode cable with LEMO connector)
Cable SC-LEMO-60 0.60 m 89601
Cable SC-LEMO-100 1.00 m 89602
Cable SC-LEMO-160 1.60 m 51108034
Phototrode (incl. power supply unit)
for color-indicated titrations
Transmission measurement
at 555, 660, 520, 620, 590 nm
DP5
Adapter cable 1) (DIN-LEMO) for 89600
attachment of electrodes or phototrode
with DIN connector to the titrator
1) If sensors with DIN connectors are attached via the adapter cable to the titrator,
the advantages offered by the triaxial cable – high level of protection against
electrostatic interference – are in part lost. In the case of sensors with a very high
resistance, e.g. DG113 in nonaqueous solutions, we advise against use of the
adapter cable.
11-14 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Accessories
Order No.
Temperature sensors
Pt1000 sensor
DT1000
Lemo cable connector (4 pin) 88321
for temperature sensors (non-METTLER)
Accessories for Karl-Fischer titrations
KF option 51107269
Double-pin platinum electrode
DM143-SC
Electrode cable (SC/Banana) 1.00 m 51108061
External titration stand
DV705
Thermostatable beaker 51107497
(for KF titration vessel)
Stand rod 51107495
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 11-15
GA42 Printer
17.375 g
19.319 g
8. 03 g
7. 73 g
6. 54 g
10.506 g
8.097 g
5.876 g
3.205 g
1.098 g
OUT 1 OUT 2
24 VDC 1.2 A
1 0-240 V
50/60 Hz
unfused
max. 9 A
OUT 3
OUT 4
IN 1
IN 2
ON
TBox DR42
ME TLER TOLEDO
Accessories
Order No.
Peripherals
METTLER TOLEDO balances with data output
AG, AM, PM, AT
AX, AB, PB, PR, XP
Printer DIN A4 / US Letter
on request
Printer
RS-P42
OUT 1
OUT 2
OUT 3
OUT 4
TBox (control of external devices via TTLIO)
DR42
Sample changer
Rondolino
Sample changer Rondo 60
Connection cable for Rondolino or TBox 51107424
Connection cable for Rondo 60 51108304
Titration software
- LabX pro titration (for Windows 2000/XP)
- Instrument licenses for Titrators
11-16 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Accessories
Order No.
Miscellaneous
Keypad cover set of 3 51107667
Adapter cable 0.23 m 25914
(banana sockets-DIN connector)
Memory Card
for DL53 and DL55
USER DATA
ME-51107230
Memory card (Flash type, 2 MB) 51107230
Dongle for software updates, DL53 ⇒ DL55 51107210
upgrading the titrator DL53 ⇒ DL55+ 51107211
and special methods DL53 ⇒ DL53+ 51107212
DL55 ⇒ DL55+ 51107213
DL53 ⇒ DL58 51107275
DL55 ⇒ DL58 51107276
Connection cable: Centronics option
Connection cable for AM, PM, AT balances 229029
(D sub 9-pin female, 15-pin male)
Connection cable for AX balances 11101051
Connection cable for balances with Option 011 59759
(D sub 9-pin female, 25-pin male)
Connection cable for AG, AB, PB, PR balances 229065
(LC-RS9 cable)
Connection cable for SARTORIUS balances 51190363
(D sub 9-pin female, 25-pin male)
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 11-17
Accessories
Order No.
Connection cable: RS option
Connection cable for AM, PM, AT balances 51107196
Connection cable for balances with Option 51107195
03 and 011 (RS8 male, D sub 25-pin male)
Connection cable for AG, AB, PB, PR balances
(LC-RS8 cable) 229185
Connection cable for SARTORIUS balances 200495
(RS8 male, D sub 25-pin male)
Connection cable (RS232C) for printer 200495
(RS8 male, D sub 25-pin female)
Connection cable (RS232C) for GA42 printer 201508
(DTE, 9-pin female)
Connection cable (RS232C) for computer
(DTE, 9-pin female) 201508
Connection cable (RS232C) for computer/terminal
(DTE, 25-pin female) 201507
11-18 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Accessories
Order No.
Options
RS option: 2 RS232C interfaces (DCE) 51107172
plus 1 TTLIO socket
Centronics option: 1 Centronics and 51107289
1 RS232C interface plus 1 TTLIO socket
KF option: 2 inputs for polarized electrode, 1 stirrer 51107269
output and 1 analog output
pH option: One input each for reference electrode, 51107152
sensor, temperature sensor plus stirrer and
analog output
Documentation
Reference Handbook • German 51709613
• English 51709614
• French 51709615
Quick Guide • German 51709616
• English 51709617
• French 51709618
• Spanish 51709619
• Italian 51709620
Memo card • German 51709621
• English 51709622
• French 51709623
• Spanish 51709624
• Italian 51709625
Application Brochure (No. 12) • German 51724764
• English 51724765
Computer Interface • German 51709495
Description • English 51709496
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 11-19
Accessories
11.3 METTLER TOLEDO literature and application brochures
Basics of Titration 51725008
Basiswissen in der Titration 51725007
Fundamentals of Titration 704153
Grundlagen der Titration 704152
Bases du Titrage 704154
Guide to pH Measurement 51300047
Anleitung zur pH-Bestimmung 51300058
Guide to Ion Selective Measurement 51300075
Anleitung zur ionenselektiven Messung 51300201
Guide to Conductivity and Dissolved Oxygen 51724716
Anleitung zur Leitfähigkeits- und Sauerstoffmessung 51724715
Guide de mésure da la conductivité et de l'oxygène dissous 51724717
DL5x Application Brochure 12 (30 Selected Applications for DL5x Titrators) 51724765
DL5x Applikationsbroschüre 12 (30 ausgewählte Applikationen für DL5x Titratoren) 51724764
DL5x Application Brochure 22 (Surfactant Titration) 51725015
DL5x Applikationsbroschüre 22 (Titration von Tensiden) 51725014
DL5x / DL7x Application Brochure 29 (Rondo 60 Sample Changer) 51710082
Memory Card Application package for METTLER TOLEDO Titrators DL53/DL55/DL58 51107233
Memory Card and Application Brochure 18 (51724917): Standardization of Titrants
Memory Card Applikationssammlung für METTLER TOLEDO Titratoren DL53/DL55/DL58: 51107234
Speicherkarte und Applikationsbroschüre 18 (51724916): Titerbestimmungen
Memory Card Application package for METTLER TOLEDO Titrators DL53/DL55/DL58 51107235
Memory Card and Application Brochure 19 (51725012): Determinations in Beverages
Memory Card Applikationssammlung für METTLER TOLEDO Titratoren DL53/DL55/DL58: 51107236
Speicherkarte und Applikationsbroschüre 19 (51725013): Getränkebestimmungen
Memory Card Application package for METTLER TOLEDO Titrators DL53/DL55/DL58 51107237
Memory Card and Application Brochure 17 (51724915): Pulp and Paper Industry
Memory Card Application package for METTLER TOLEDO Titrators DL53/DL55/DL58 51107238
Memory Card and Application Brochure 20 (51725020): Petroleum Industry
11-20 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Accessories
We have developed applications for the METTLER TOLEDO DL770, DL67, DL70ES and DL77 Titrators, which
have a method concept similar to that of the DL50, DL53, DL55 and DL 58 series. You will find these useful when
defining the parameters of your method functions.
DL70 Application Brochure 1 (18 Customer Methods) 724492
DL70 Applikationsbroschüre 1 (18 Kundenapplikationen) 724491
DL70 Application Brochure 2 (Various Methods) 724557
DL70 Applikationsbroschüre 2 (Verschiedene Beispiele) 724556
DL70 Application Brochure 3 (TAN and TBN) 724559
DL70 Applikationsbroschüre 3 (TAN und TBN) 724558
DL70 Application Brochure 4 (Electroplating) 724561
DL70 Applikationsbroschüre 4 (Galvanik) 724560
DL7x Application Brochure 5 (Determinations in Water) 51724634
DL7x Applikationsbroschüre 5 (Bestimmungen in Wasser) 51724633
DL7x Application Brochure 6 (Direct Measurement with Ion Selective Electrodes) 51724646
DL7x Applikationsbroschüre 6 (Direktmessung mit ionenselektiven Elektroden) 51724645
DL7x Application Brochure 7 (Incremental Techniques with Ion Selective Electrodes) 51724648
DL7x Applikationsbroschüre 7 (Additionsverfahren mit ionenselektiven Elektroden) 51724647
DL7x Application Brochure 8 (Standardization of Titrants I) 51724650
DL7x Applikationsbroschüre 8 (Titerbestimmungen I) 51724649
DL7x Application Brochure 9 (Standardization of Titrants II) 51724652
DL7x Applikationsbroschüre 9 (Titerbestimmungen II) 51724651
DL7x Application Brochure 13 (Nitrogen Determination by Kjeldahl Digestion) 51724769
DL7x Applikationsbroschüre 13 (Stickstoffbestimmung nach Kjeldahl Aufschluss) 51724768
DL7x Application Brochure 14 (Good Labatory Practice in the Titration Lab) 51724908
DL7x Applikationsbroschüre 14 (Gute Laborpraxis im Titrationslabor) 51724907
DL7x Application Brochure 15 (Guidelines for Result Check, 51724910
Method Validation and Instrument Certification)
DL7x Applikationsbroschüre 15 (Leitfaden zur Resultatkontrolle, 51724909
Methodenvalidierung und Gerätezertifizierung)
DL7x Application Brochure 16 (Validation of Titration Methods) 51724912
DL7x Applikationsbroschüre 16 (Validierung von Titrationsmethoden) 51724911
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 11-21
Technical Data
Contents
Page
12. Technical data ........................................................................................... 12-3
12.1 Measurement system................................................................................ 12-3
12.2 Outputs ...................................................................................................... 12-4
12.3 Burette drive module ................................................................................ 12-5
12.4 Interchangeable burettes ......................................................................... 12-5
12.5 Propeller stirrer ......................................................................................... 12-5
12.6 Display ....................................................................................................... 12-5
12.7 Memories ................................................................................................... 12-5
12.8 Interface for memory card ........................................................................ 12-6
12.9 Attachment possibilities for peripheral units ......................................... 12-6
12.10 Sockets and connectors........................................................................... 12-7
12.10.1 DIN socket................................................................................................... 12-7
12.10.2 TTLIO socket............................................................................................... 12-7
12.10.3 RS232C connector (Centronics option) ...................................................... 12-7
12.10.4 Centronics socket (Centronics option) ........................................................ 12-8
12.10.5 RS232C socket ........................................................................................... 12-9
12.11 Additional data .......................................................................................... 12-9
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 12-1
Technical Data
12. Technical data
12.1 Measurement system
End point indication
• potentiometric
• photometric
• change of the refractive index and/or the turbidity
• voltametric/amperometric
• conductometric
Third-party instrument with analog output
Temperature compensation
Temperature input via keypad or automaticalfor
pH/pM/pX measurement ly with the aid of a temperature sensor
Sensor input: pH option
(differential amplifier)
Triaxial socket (LEMO)
• Input resistance
>10 12 Ω (protected up to 1000 V capacitatively)
• Offset current
<1 pA (20 °C)
• Measurement range
- pH value ±27.6 pH
- voltage ±2050.0 mV
• Resolution
0.1 mV, 0.002 pH (pM, pX)
• Maximum permissible error 0.1%
• Zero point drift <30 µV/ °C
• Additional reference input
Banana socket, diameter: 4 mm
Temperature sensor input
4-pin socket (LEMO)
• Measurement range -30 °C to 130 °C
• Resolution 0.1 °C
• Maximum permissible error 0.2 °C
• Measurement principle
4-line measurement technique: Pt100 or
Pt1000
• Zero point drift <0.01 °C/ °C
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 12-3
Technical Data
Sensor input: KF option
Sensor input
2 Banana sockets, diameter: 4 mm
Voltage source
• Range
-1270 to +1270 mV
• Resolution
10 mV in the range -1270 to +1270 mV
1 mV in the range -127 to +127 mV
• Measurement range -150 to +150 µA
• Resolution
0.01µA (16-bit AD converter)
• Maximum permissible error 0.1%
Current source
• Range -127 to +127 µA
• Resolution 1 µA in the range -127 to +127 µA
0.1 µA in the range -12.7 to +12.7 µA
• Measurement range
-1500 to +1500 mV
• Resolution
0.1mV (16-bit AD converter))
• Maximum permissible error 0.1%
12.2 Outputs
Stirrer output
5-pin socket (mini DIN)
• Voltage range
0 – 18 V in no-load operation
• Speed stabilized Typically: <5%
the output is protected electronically against
overload, Imax: 300 mA
Analog output
2 banana sockets, diameter: 2 mm
• Gain 1
• Offset: max. permissible error
35 mV
• Range
±2000 mV
• Range: max. permissible error 1%
• Output resistance
220 Ω
12-4 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Technical Data
12.3 Burette drive module
With direct current motor
Equipment DL50/DL53: max. 1, DL55/DL58: max. 2
Resolution
1/5000 of the burette volume
Maximum permissible error
0.3% relative to the respective burette volume
of 5, 10 and 20 mL
Filling time
20 s
Discharge time
Minimum 20 s
12.4 Interchangeable burettes
Volume
Materials which come into contact with
the titrant
1, 5, 10 and 20 mL
Fluoroplastic, borosilicate glass, ceramics
12.5 Propeller stirrer
Maximum speed
3800 rpm
Power consumption P max. : 6 W
P typical : 1.2 W at 12 V
12.6 Display
LCD with graphics capability
Languages
6 lines, 39 characters per line
240 x 64 pixels, backlighting through cold
cathode fluorescent lamp
English, German, French, Spanish and Italian
12.7 Memories
Database
User data memory (EEPROM)
METTLER methods, standard methods
User methods, resource and sample data, space
for approx. 50 standard methods
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 12-5
Technical Data
12.8 Interface for memory card
Slot for 68-pin memory card of type 1
Standards supported:
Memory cards which can be used:
1. Flash memory card series 2
Intel with the designation:
Mitsubishi with the designation:
2. SRAM Melcard
Mitsubishi with the designation:
• PCMCIA, version 2.0
• JEIDA, version 4.1
• iMC002FLSA - 20 (2 MB storage capacity)
• iMC004FLSA - 20 (4 MB storage capacity)
• MF82M1-G7DAT 01 (2 MB storage capacity)
• MF84M1-G7DAT 01 (4 MB storage capacity)
• MF31M1-LCDAT 01 (1 MB storage capacity)
12.9 Attachment possibilities for peripheral units
Balance
Printer
Sample Changer
Terminal or computer
(LIMS: in-house laboratory
information management system)
Auxiliary units (third-party units)
External keyboard (third-party unit,
e.g. commercial PC keyboard)
Attachment via RS232C interface of the Centronics
or an RS option at slot No. 4 for all
METTLER TOLEDO and SARTORIUS balances
Attachment via Centronics interface of the Centronics
option or via RS232C interface of an RS
option at slot No. 4 for various commercial
graphics printers (on use of the RS option with
XON/XOFF protocol)
Attachment via RS232C interface of the RS
option at slot No. 3 (available as an option)
Attachment via RS232C interface of the RS
option at slot No. 3 (available as an option)
Attachment via TTLIO socket of the RS option
at slot No. 3 (available as an option)
Attachment via the DIN socket, 5-pin
12-6 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Technical Data
12.10 Sockets and connectors
12.10.1 DIN socket
5-pin, DIN 180°, following DIN 41524
Pin assignment
5
2
4
CLOCK
DATA
1
2
3
1
–
GND
VCC
Titrator
+
4
5
VCC: 5 V
12.10.2 TTLIO socket
8-pin connector, series 712 waterproof (TTL: Transistor-Transistor-Logic)
The socket has 2 TTL inputs and 4 TTL outputs.
Pin assignment
1
8
7
2
6
3
4
5
GND
In 1
In 2
Out 1
Out 2
Out 3
VCC
Out 4
Titrator
+5 V
1
2
3
4
5
6
7
8
VCC: +5 V, max. 60 mA
In: TTL; short circuit proof
Pulse ≥150 ms (the input signal
must be at least 150 ms for
the titrator)
Out: Open collector
• Vce max. = 24 V
• Ic max. = 20 mA
12.10.3 RS232C connector (Centronics option)
9-pin, D-Sub
5
9
4
8
3
7
2
6
1
Shielding
RxD (In)
TxD (Out)
DSR
RSGND
+12 V
Titrator
BR
330 Ω
Pin assignment
1
2
3
4
5
Data to titrator
Data from titrator
Handshake from titrator,
fixed to operational readiness
Electrically isolated signal
ground
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 12-7
Technical Data
12.10.4 Centronics socket (Centronics option)
19-pin connector, series 723 waterproof
Pin assignment
/STROBE
A: Data ready
D0
B:
D1
C:
D2
D:
L
A
K
B
U
M
I
T
N
C
H
S
R
P
O
D
G
E
F
D3
D4
D5
D6
D7
/ACK
BUSY
PE
SLCT
/AUTO FDXT
E:
F:
G:
H:
I:
Data
K: Data adopted
L: Printer occupied
M: End of paper
N: Printer active
O: Automatic line feed
/ERROR
P: Printer error
/INIT
R: Initializing printer
/SLCTIN
S: Printer selected
GND
GND
T:
U:
Signal ground
Titrator
12-8 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Technical Data
12.10.5 RS232C socket
8-pin connector DIN 45326, series 723 waterproof
6
8
7
BR
Shielding
1
TxD (Out)
4
2 RxD (In)
5 RSGND
3 330 Ω
DSR +12 V
Titrator
Pin assignment
1
2
3
6
8
Data to titrator
Data from titrator
Electrically isolated signal
ground
Handshake from titrator
fixed to operational readiness
12.11 Additional data
Housing
Titration stand
Keypad
Dimensions
Weight
Power supply
Fuse
Frequency
Maximum power consumption
Polyester
Polypropylene
Polyester, splashwater-proof
Width: 260 mm, depth: 395 mm, height: 270 mm
Approx. 8 kg with one burette drive
100 – 240 VAC ±10% (automatic range adjustment)
2 x T1,6L250V (not exchangeable)
50/60 Hz
400 mA
Ambient conditions • Ambient temperature: +5 °C to +40 °C
• Maximum relative atmospheric humidity of 80% for
temperatures of up to 31 °C, decreasing linearly to
50% relative atmospheric humidity at 40 °C.
• Use indoors
• Height up to 2000 m
• Overvoltage category II
• Pollution degree 2
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 12-9
Index
Index
Numbers incorporating dashes refer to sections in the Reference Handbook (example: 3-38),
Numbers without dashes refer to the Quick Guide (example: 9).
<F1>...<F5> 9, 10
∆E
EP titration function 3-39
EQP titration function 3-25
EQP titration function (Ipol/Upol) 3-49
measure function 3-17
∆E(set)
EP titration function 3-39
EQP titration function 3-22
∆t
EP titration function 3-39
EQP titration function 3-25
EQP titration function (Ipol/Upol) 3-49
measure function 3-17
Accessories
optional accessories 11-7
standard equipment 11-4
Activate
control input 6-32
control output 6-33
Adjust measuring inputs 6-28
Amperometric indication
EP titration function (Ipol/Upol) 3-58
EQP titration function (Ipol/Upol) 3-47
Analysis
aborting (Reset) 5-12
fading out 5-11
is interrupted 5-11
menu sequence 5-4
"Run" menu 5-3
start 5-4
Analysis sequences (examples)
CaCl 2 determination 5-7
calibration of a pH electrode 27
comparison: titration stand 1/ST20A 5-16
determination of the HCl content 12
pH-stating 5-10
titer determination (NaOH) 32
with the sample changer 5-17
Arrow keys 7
Asterisk (*) 3-4, 3-11
Asymmetric (evaluation procedure)
equivalence point recognition 3-28, 3-50
evaluation parameter 3-34, 3-53
explanation 8-16
Audio signal 6-28
Auto stand
definition 2-17
sample series 5-6, 5-15
Auxiliary functions
explanation 9, 6-3
keys 6-3
overview 1-4, 1-5
Auxiliary value(s)
entering 2-15
entry through determination 2-15
function 3-73
memories 3-73
resource 2-15
storage procedure 2-16
Balance(s)
configuration (METTLER TOLEDO) 2-22
configuration (SARTORIUS) 2-22
connection 10-10
defining 2-21
Bar-code reader
connection 10-16
defining 2-24
Bar-code string 6-43
Baud rate
balances 2-22
computer 2-23
printer 2-20
terminal 2-25
Bidirectional transmission mode 2-21, 4-12
Buffer solutions
DIN/NIST buffer 3-70
MERCK buffer 3-69
METTLER TOLEDO buffer 3-69
Burette
air bubbles 10-8
equipping 10-5
filling 13
inserting 10-6
installing 10-4
maintaining 10-4, 10-7
order No. 11-4
rinsing 6-11
rinsing tip 6-12
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 13-1
Index
Burette drive
inserting 10-3
selecting 2-5
Burette volume
selecting 2-5
smallest increment 3-24
Burette tip 6-12
C (constant) 3-66, 8-7
c ( nominal concentration) 8-3
c * t (actual concentration) 8-3
Calculation(s)
formulas 8-17
function 3-66
indices 8-9
performing (auxiliary function) 6-20
Calibration
function 3-69
pH electrodes 27
sensors 3-69
temperature sensors 6-6
Calibration data
measuring inputs 6-29
sensors 2-9, 2-10, 3-69
temperature sensors 2-12
Centronics interface
GA42 printer 2-18
printer configuration 2-19
Centronics option
connection possibilities 10-9
installing 10-10
Changer 6-9
Character set (computer) 2-24
Clock (titrator) 6
Code (sync function) 7-18
Combined termination conditions 3-34, 3-53
Command keys 9
Communication protocol (computer) 2-24
Computer
bar-code reader 2-24
bar-code string 6-43
configuration 2-23
connection 10-10
data transfer 6-40
remote control 6-42
Concentration (titrant) 2-4, 8-3
Conditioning 3-15
Conductivity measurement 2-8
Conductometer 2-8
Constants
calculation function 3-66
calculations (auxiliary function) 6-20
examples 8-18
Continuous (titrant addition) 3-40, 3-60
Control band
EP titration function 3-40
EP titration function (Ipol/Upol) 3-60
pH/mV-stat function 3-63
Control input
activating (auxiliary function) 6-32
function 7-15
Control inputs
defining 7-4
resource 7-4
Control mode (activate control output)
input controlled 6-33
on/off 6-33
sequential 6-34
Control mode (control output function)
fixed time 7-16
input controlled 7-16
on/off 7-16
sequential 7-17
Control output
activating (auxiliary function) 6-33
function 7-16
Control outputs
defining 7-5
resource 7-5
Control range
EP titration function 3-40
EP titration function (Ipol/Upol) 3-60
pH/mV-stat function 3-63
Correction factor f 4-6, 5-6
Correlation coefficient 3-62, 3-65, 8-6
CSTAT (correlation coefficient) 3-62, 8-6
Current measurement (Ipol/Upol) 6-7
Current sample (mask)
notes 5-6
Cursor 10
Cx (calculation constant) 8-7
Database 12-5
Data bits
computer 2-24
printer 2-20
terminal 2-25
Data sheet
KF option 10-12
pH option 10-12
Data Transfer 6-36
13-2 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Index
Date
entering 6-27
selecting format 6-26
Date specification
auxiliary values 2-15
methods 3-11
sensors 2-9
temperature sensors 2-11
titrants 2-5
Defined are (mask)
exclusion 6-30
meaning 5-5
Delay 3-42, 3-61
Directory (memory card) 6-37
Dispense
auxiliary function (burette) 6-12
auxiliary function (ST20A) 6-10
continuously 6-13
continuously (Ipol/Upol) 6-15
function 3-19
DL58
commands 7-3
functions 7-3
resources 7-3
DLWin software 4-13, 6-41
Documentation
information 4
Drift (potential measurement) 3-17
Dynamic (titrant addition)
EP titration function 3-39
EQP titration function 3-22
E (potential) 3-17, 8-7
EHNV (half neutralization value) 3-20, 8-4
Electrodes (METTLER TOLEDO) 2-7, 2-13
End point
EP titration function 3-41
EP titration function (Ipol/Upol) 3-60
pH/mV-stat function 3-63
range 3-40, 3-60
titration 3-37, 3-56
Entry keys 10
EP
absolute 3-41
relative 3-41
EP titration (function) 3-37
EP titration (Ipol/Upol) (function) 3-56
EPOT 8-4
EQP titration (function) 3-20
EQP titration (Ipol/Upol) (function) 3-45
Equilibrium controlled measured value acquisition
EP titration function 3-39
EQP titration function 3-25
EQP titration function (Ipol/Upol) 3-49
measure function 3-17
Equivalence point(s)
explanation 3-20, 3-45, 8-14
formulas for restriction 8-19
maximum number/sample determination 8-20
range 3-32, 3-51
recognition 3-28, 3-50
Equivalence point titration 3-20, 3-45
Equivalent number
calculation function 3-68
entering 3-13
Error messages
"EPROM test failed" 9-3
learn titration 5-11
"RAM test failed" 9-3
referring to Section 9.1 9-3
Esc command (<F1>) 9
ET1 3-22, 3-38, 8-4
ET2 3-22, 3-38, 8-4
Evaluation
EQP titration function 3-34
EQP titration function (Ipol/Upol) 3-53
pH/mV-stat function 3-65
Evaluation procedure
EQP titration function 3-28, 3-34
EQP titration function (Ipol/Upol) 3-50, 3-53
explanation 8-14
Expert (user level) 6-31
Expiry date (titrant) 2-5, 5-4
External keyboard
connection 10-15
defining 2-26
key assignment 10-15
External stand 2-17
Extinction 2-8
f (correction factor) 4-6, 8-3
Fixed time (control output function ) 7-16
Fixed volume
note 5-5
selecting 3-12
Form feed (printer) 2-19
Formula(s)
auxiliary value function 3-73
calculation function 3-66
calculations (auxiliary function) 6-20
constants 8-18
for restriction of the equivalence point 8-19
results 8-17
titer function 3-72
Frame lines (report) 2-19
Function(s)
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 13-3
Index
adding 7-8
deleting 7-8
DL58 7-3
explanation 20, 3-3
maximum number 7-9
modifying 3-10
overview 24
selecting 3-7
H (auxiliary value) 2-15, 8-3
Half neutralization value 3-20
In 3.1/3.2 (TTL inputs) 6-32
Incremental (titrant addition)
EQP titration function 3-24
EQP (Ipol/Upol) titration function 3-48
two-phase titration function 7-14
Indexing forms
compilation 8-12
examples 8-9, 8-10
Indication
EP titration function (Ipol/Upol) 3-57
EQP titration function (Ipol/Upol) 3-46
Indices 8-9
Inflection point 8-14
Initial potential ET1 3-22
Input controlled
activate control output (auxiliary function) 6-33
control output function 7-16
Input signal (control input function) 7-15
Installation and maintenance 10-3
Instruction
control output function 7-17
function 7-11
Interchangeable burette: see burette 10-4
Interpolation (evaluation) 8-14
Interval (conditioning) 3-15
Karl-Fischer titrations
accessories 11-15
Keyboard
external 2-26
titrator 7
Key combinations 10
KF option
inputs and outputs 10-10
installing 10-11
measuring inputs (calibration data) 6-29
polarized sensors 2-13
LabX titration software 4-13, 6-41
Language (changing) 11, 6-27
Learn titration
function 3-43
results 5-9
Lift position (sample changer) 6-9
Line feed (printer) 2-19
Lists (explanation) 8
M (molar mass) 3-68, 8-3
m (sample size) 8-3
Maintenance
burette 10-7
titrator 10-3
Malfunctions 9-6
Manual
temperature entry 3-13
titration 6-14, 6-16
Masks (explanation) 8
Max. time (control input function) 7-15
Maximum (evaluation procedure)
equivalence point recognition 3-28, 3-50
evaluation parameter 3-34, 3-53
explanation 8-14
Maximum time (EP titration function (Ipol/Upol))
3-61
Maximum volume
EP titration function 3-42
EP titration function (Ipol/Upol) 3-61
EQP titration function 3-33
EQP titration function (Ipol/Upol) 3-52
pH/mV-stat function 3-64
Mean value x
auxiliary value function 3-73
calculation function 3-67
titer function 3-72
Measure (function) 3-17
Measure mode
EQP titration function 3-25
EQP titration function (Ipol/Upol) 3-49
two-phase titration function 7-14
Measured points for equivalence point recog-nition
3-28, 3-44, 3-50
Measured values
displaying 5-7, 6-21
maximum number/titration function 8-20
storage specification 3-64, 3-74, 8-20
Measuring inputs
adjusting 6-28
calibration data 6-29
13-4 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Index
Memory card
directory 6-37
formatting 6-36, 6-40
inserting 10-11
interface 12-6
types 12-6
use 6-36
Memory copy(ies)
copying on memory card 6-39
loading to titrator 6-40
transferring to computer 6-41
Menus
explanation 8
keys 8
overview 1-2, 1-3
Method
concept 19
deleting 3-5
explanation 3-3
generating 24
modifying 3-7
modifying (current sample series) 5-12
modifying (ongoing method: DL55/DL58) 5-13
on computer 4-5
on memory card 4-5
printing 3-4
storage procedure 3-7
Methods
menu 3-3
printing 3-4
selecting 3-4
Method ID
defining 3-11
entering 4-4, 5-5
explanation 3-4
METTLER methods 3-6
modifying 3-9
standard methods 3-5
METTLER methods
90001: performance 14
90002: performance 27
explanation 19
modifying 25, 3-6
printing 3-6
Minimum (evaluation procedure)
equivalence point recognition 3-28, 3-50
evaluation parameter 3-34, 3-53
explanation 8-14
Misc. ... 6-26
Mixing time (two-phase titration function ) 7-14
Modify: command (<F4>) 10
Molar mass M
calculation function 3-68
entering 3-13
neq (number of equivalence points) 8-4
OK command (<F5>) 9
On/Off
activate control output (auxiliary function) 6-33
control output function 7-16
Operating concept 7
Options 10-9
Out 3.1...3.4 (TTL outputs) 6-33
Output (reports) 3-74, 6-24
P1/P2 8-6
Paper (printer)
fanfold 2-19
format 2-19
single sheet 2-19
Parameters (explanation) 20
Parity
balances 2-22
computer 2-23
terminal 2-25
printer 2-20 t
Peripherals (setup menu) 2-18
pH electrode
calibrating 26
slope 26
zero point 26
pH measurement 35
pH option
inputs and outputs 10-10
installing 10-11
measuring inputs (calibration data) 6-29
sensors 2-7
pH/mV-stat (function) 3-62
pH-stating (example) 3-65
Piston
assembling 10-4
inserting 10-4
Polarized sensors
adding 2-14
KF option 2-13
modifying 2-13
resource 2-13
storage procedure 2-13
Potential 1/2 (evaluation parameters) 3-35, 3-54
Potential measurement 6-4, 6-7
Predispensing
EP titration function 3-38
EP titration function (Ipol/Upol) 3-59
EQP titration function 3-21
EQP titration function (Ipol/Upol) 3-48
two-phase titration function 7-13
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 13-5
Index
Pretitration (pH/mV-stat function) 3-63
Primary standard 30, 31
Printer
ASCII 2-18
configuration 2-19, 2-20
connection 10-10
defining 2-18
Pt100/Pt1000 sensor 2-11
Pump
function 7-11
titration stand parameters 3-13
rate (solvents) 2-27
Q 8-4
QDISP 3-19, 8-7
QEND 8-4
QEX 8-4
QP1/QP2 8-4
QSTAT 8-6
QT1/QT2 8-6
QTOT 8-6
R (result) 3-66, 8-7
Range (equivalence point) 3-32, 3-51
Raw results
compilation 8-8
explanation 3-18
recording 3-75, 6-24
storage specification 8-20
Rear view of titrator 10-10
Recalculations 6-20
Recognition (equivalence point) 3-28, 3-50
Reevaluation
condition met 5-11
defining condition 3-35, 3-36, 3-55
performing 6-18
Reference electrode 2-7
Relative standard deviation srel 3-67
Remote control 6-42
Report
auxiliary function 6-24
function 3-74
header 6-27
Reset key 7, 5-12
Resources
changing 5-8
copying on memory card 6-39
copying to titrator 6-39
deleting 2-3
DL58 7-3
explanation 2-3
list 2-3
modifying 2-3
printing 2-3
transferring to computer 6-41
Result(s)
auxiliary function 6-18
deleting 6-22
examples 3-67, 8-17
learn titration 5-9
maximum number/sample series 8-20
R 3-66, 3-71
recording 3-75, 6-24
Rx 6-20
storage specification 8-20
units 3-67, 8-17
Result list
displaying 6-18
exclusion 6-30
notes 5-9
Rinse
auxiliary function (ST20A) 6-10
tip (burette) 6-12
titration stand parameters 3-14
Routine (user level) 6-31
RS option
connection possibilities 10-11
installing 10-11
RS232 interface
GA42 printer 2-20
printer configuration 2-20
Run
analysis menu 5-3
key function 5-3
s (standard deviation) 8-7
Safety measures
for operational safety 5
for your protection 5, 10-3
Safety notes 1-1
Sample
function 3-12
ID (identification) 4-5, 5-6
13-6 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Index
Samples
maximum number 8-20
status 4-7
Sample changer
auxiliary function 6-9
connection 10-10
defining 2-25
sample series 5-6, 5-15
titration stand 2-17
titration stand parameters 3-13
Sample data
adding 4-8
deleting 4-6
entering 4-4, 5-5, 5-10
list 4-4, 4-6
memory (sample menu) 4-3
modifying 4-7
on computer 4-13
printing 4-7
recording 3-75, 6-24
storage specification 36, 4-3
Sample data mask
analysys menu 5-5
sample menu 4-5
Sample series
deleting 4-6
entering 4-4
with auto stand 5-6, 5-15
with ST20A 5-6, 5-15
Sample series (DL55/DL58)
entering 4-9
executing 5-15
Sample size 3-21, 3-38, 3-48, 3-59, 8-3
Save (storage procedure)
auxiliary values 2-16
devices for control intput 7-5
devices for control output 7-6
methods 3-7
polarized sensors 2-13
sensors 2-10
solvents 2-28
temperature sensors 2-12
titrants 2-6
Segmented (evaluation procedure)
equivalence point recognition 3-28, 3-50
evaluation parameter 3-34, 3-53
explanation 8-15
Send mode (computer) 2-24
Sensor(s)
adding 2-10
auxiliary function 6-4
deleting 2-7
modifying 2-7
order No. 11-10
pH option 2-7
resource 2-7
storage procedure 2-10
Sensor inputs
polarized sensors 2-13
sensors 2-8
temperature sensors 2-11
Separation time (two-phase titration function) 7-14
Sequential
activating control output (auxiliary function) 6-34
control output function 7-17
Setup (menu) 2-3
Shorten analysis sequence 6-30
SLOPE 8-3
Slope
calculation 3-69
entry through determination 2-9
explanation 2-9
minimum/maximum value 3-71
pH electrode 26
theoretical value 2-9
Slots 10-10
Software LabX 4-13, 6-41
Software version 1-1
Solvents
adding 2-28
deleting 2-27
modifying 2-27
storage procedure 2-28
Special methods
DL58 7-7
explanation 7-3
Speed
defining 3-16
modifying 5-8, 6-8
srel (relative standard deviation) 8-7
ST20A
connection 10-10
outputs 2-28
Stand 1/2 (titration stands) 2-17
Standard (evaluation procedure)
equivalence point recognition 3-28, 3-50
evaluation parameter 3-34, 3-53
explanation 8-14
Standard deviation s 3-67
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 13-7
Index
Standard method(s)
application examples 22
explanation 19, 3-5
functions 20
modifying 25, 30
number 20
printing 3-5
Start/end character (computer) 2-24
Statistics
modifying 6-22
selecting for calculations 3-67
Status
remote control 6-42
peripherals 2-18
samples 4-7
Steepest jump only 3-31
Stir (function) 3-16
Stirrer
auxiliary function (stirrer) 6-8
output 2-17
speed 5-8, 6-8
Stirring time 3-16
Stop bits
computer 2-24
printer 2-20
terminal 2-25
Stop for reevaluation 3-35, 3-55, 5-11
Storage interval (pH/mV-stat function) 3-64
Storage procedure
measured values 3-64, 3-74, 8-20
raw results 8-20
results 8-20
sample data 4-3
Symbols
explanation 8-3
list 8-3
Sync (function) 7-18
Synchronization mode (Sync function) 7-18
System
computer/terminal 2-23
titrator (information) 9-3, 9-6
T (temperature) 7-10
t (titer) 3-72, 8-3
t(max)
EP titration function 3-39
EQP titration function 3-25
EQP titration function (Ipol/Upol) 3-49
measure function 3-18
temperature function 7-10
t(min)
EP titration function 3-39
EQP titration function 3-25
EQP titration function (Ipol/Upol) 3-49
measure function 3-17
temperature function 7-10
Table of measured values
displaying 5-7, 6-21
recording 3-75, 6-24
Technical data
burette 12-5
burette drive 12-5
Centronics socket 12-8
DIN socket 12-7
display 12-5
measurement system 12-3
memories 12-5
outputs 12-4
peripherals 12-6
RS232C connector (Centronics option) 12-7
RS232C socket (RS option) 12-9
stirrer 12-5
TTLIO sockets 12-7
Temperature
correction 3-13, 3-69
entering 2-9, 4-6, 5-6
function 7-10
measurement 2-9, 3-13, 3-18, 3-20, 3-37,
3-62, 3-69, 7-10
measuring (auxiliary function) 6-5
Temperature sensor(s)
calibrating 6-6
Lemo cable plug 10-10
Pt100/Pt1000 2-11
resource 2-11
selecting 3-13
storage procedure 2-12
Tendency
EP titration function 3-42
EP titration function (Ipol/Upol) 3-61
EQP titration function 3-32
EQP titration function (Ipol/Upol) 3-51
pH/mV-stat function 3-64
13-8 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Index
Terminal
configuration (keyboard) 10-16
configuration (titrator) 2-25
connection 10-11
key assignment 10-17
Termination (parameters)
EP titration function 3-42
EP titration function (Ipol/Upol) 3-61
EQP titration function 3-33
EQP titration function (Ipol/Upol) 3-52
pH/mV-stat function 3-64
Threshold 3-29, 3-51
Threshold value
evaluation procedure "Asymmetric" 3-30
evaluation procedure "Minimum/Maximum" 3-30
evaluation procedure "Segmented" 3-31
evaluation procedure "Standard" 3-30
TIME 5-4, 8-3
Time
entering 6-27
sample function (conditioning) 3-15
selecting format 6-26
stir function 3-16, 5-7
Time specification
auxiliary values 2-15
methods 3-11
sensors 2-9
temperature sensors 2-11
titrants 2-5
Timed increment measured value acquisition
EP titration function 3-41
EP titration function (Ipol/Upol) 3-60
EQP titration function 3-27
EQP titration function (Ipol/Upol) 3-49
Titer
checking 2-5
determination (NaOH solution) 32
entering 2-4
entry through determination 2-4
explanation 30
function 3-72
Title
function 3-11
METTLER methods 19
standard methods 19
Titrant
adding 2-6
concentration 2-4, 8-3
deleting 2-4
modifying 2-4
storage procedure 2-6
Titrant addition
EP titration function 3-39
EP titration function (Ipol/Upol) 3-60
EQP titration function 3-22
EQP titration function (Ipol/Upol) 3-48
pH/mV-stat function 3-63
two-phase titration function 7-14
Titrate
manually 6-14
manually (Ipol/Upol) 6-16
Titration curves
1st derivative 3-30, 3-31, 3-54, 8-15
2nd derivative 3-31, 3-54, 8-15
displaying 5-8, 5-8, 6-21
recording 3-75, 6-24
Titration stand
equipping 10-9
meaning 7-8
order No. 11-7
resource 2-17
selecting 3-13
Titrator
battery 6
cleaning housing 10-3
connections 6, 10-10
documentation 4
front view 6
ID (identification) 6-27
internal clock 6
keypad 7
opening housing 10-3
rear view 6, 10-10
remote control 6-42
self test 7
settings 6-26
Transmission mode (balance)
bidirectional 2-21
unidirectional 2-21
TTL inputs
activating 6-32
wiring electronically 10-14
wiring mechanically 10-13
TTL outputs
activating 6-33
wiring electronically 10-14
TTLIO socket(s)
inputs and outputs 10-13
technical data 12-7
used by titrator 10-13
Turntable (sample changer) 6-10
Two-phase titration
function 7-12
special method 7-12
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 13-9
Index
Unidirectional transmission mode 2-21
Units 3-67, 8-17, 8-18
Units of measurement (sensors) 2-8
Urgent sample (DL55/DL58) 4-10
User data memory 12-5
User level 6-31
User method(s)
copying on memory card 6-38
copying to titrator 6-39
deleting 3-6
explanation 19, 3-6
printing 3-6
"titer by EQP titration" 34
transferring to computer 6-41
VDISP 3-19, 8-7
VEND 8-4
VEQ 8-4
VEX 8-4
Voltage measurement (Ipol/Upol) 6-7
Voltametric indication
EP titration function (Ipol/Upol) 3-57
EQP titration function (Ipol/Upol) 3-46
Volume
entering 4-5, 5-6
selecting 3-12
Volume limits
at volume entry 4-5, 5-6
entering 3-12
VP1/VP2 8-4
VSTAT 8-6
VT 3-65, 8-6
VT1/VT2 8-6
VTOT 8-6
Wait time
predispensing 3-21, 3-38, 3-48, 3-59, 7-13
stir function 3-16, 5-7
Weight
entering 4-5, 5-6
selecting 3-12
transferred by balance 4-12
Weight limits
at weight entry 4-5, 5-6
entering 3-12
x (mean value) 8-7
z (equivalent number) 3-13, 3-68, 8-3
ZERO 8-3
Zero point (sensor)
calculation 3-69
entry through determination 2-9
explanation 2-9
pH electrode 26
theoretical value 2-9
Zero point (temperature sensor)
entry through determination 2-12
explanation 2-11
theoretical value 2-12
13-10 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Indice
Indice
Los números separados por un guión (ejemplo 3-38) se refieren al capítulo correspondiente
en el Manual de Instrucciones, las cifras aisladas a la Guía de Manejo (ejemplo 9).
<F1>...<F5> 9, 10
∆E
func. medir 3-17
func. valoración EP 3-39
func. valoración EQP 3-25
func. valoración EQP (Ipol/Upol) 3-49
∆E (nom.)
func. valoración EP 3-39
func. valoración EQP 3-22
∆t
func. medir 3-17
func. valoración EP 3-39
func. valoración EQP 3-25
func. valoración EQP (Ipol/Upol) 3-49
Accesorios
accesorios facultativos 11-7
equipo básico 11-4
Acondicionamiento (ST20A) 3-15
Activar
entrada mando a distancia 6-32
salida mando a distancia 6-33
Acumulador 6
Adición de reactivo
func. pH/mV-Stato 3-63
func. valoración EP 3-39
func. valoración EP (Ipol/Upol) 3-60
func. valoración EQP 3-22
func. valoración EQP (Ipol/Upol) 3-48
func. valoración en dos fases 7-14
Adquisición datos medidos: incremental tiempo
func. valoración EP 3-41
func. valoración EP (Ipol/Upol) 3-60
func. valoración EQP 3-27
func. valoración EQP (Ipol/Upol) 3-49
Agitador
función auxiliar (Stirrer) 6-8
seleccionar salida 2-17
Agitar (función) 3-16
Ajustar entradas de medida 6-28
Análisis
empezar 5-4
menú "Run" 5-3
se interrumpe 5-11
secuencia menú 5-4
terminar (Reset) 5-12
visualizar 5-11
Asimétrica (procedimiento de evaluación)
explicación 8-16
parámetro de evaluación, 3-34, 3-53
reconocimiento punto de equivalencia 3-28,
3-50
Asterisco (*) 3-4, 3-11
Avance
de línea (impresora) 2-19
de página (impresora) 2-19
Aviso de errores
con indicaciones en cáp. 9.1 9-3
"EPROM test failed" 9-3
"RAM test failed" 9-3
valoración de aprendizaje 5-1
Avisos de seguridad 1-1
Balanza (s)
conectar 10-10
configuración (METTLER TOLEDO) 2-22
configuración (SARTORIUS) 2-22
definir 2-21
Banda de control
func. pH/mV-Stato 3-63
func. valoración EP 3-40
func. valoración EP (Ipol/Upol) 3-60
Base de datos 12-5
Baudios
balanza 2-22
impresora 2-20
ordenador 2-23
terminal 2-25
Bits de datos
impresora 2-20
ordenador 2-24
terminal 2-25
Bits de parada
impresora 2-20
ordenador 2-24
terminal 2-25
Bomba
función 7-11
parámetros cabeza de valoración 3-13
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 13-11
Indice
Bureta
burbujas de aire 10-8
función auxiliar 6-11
insertar 10-6
instalar 10-4
llenar 13
llenar (lavar) 6-11
mantenimiento 10-4, 10-7
montaje 10-5
C (constante) 3-66, 8-7
c (concentración nominal) 8-3
c * t (concentración real) 8-3
Cabeza de val. auto
definición 2-17
serie(s) de muestras 5-6, 5-15
Cabeza de val. externa 2-17
Cabeza de valoración
medios auxiliares 2-17
montaje 10-8
número de pedido 11-7
seleccionar 3-13
significado 7-8
Cálculo posterior 6-20
Cálculo(s)
fórmulas 8-17
función 3-66
indicativos 8-9
realización (función auxiliar) 6-20
Calibración
electrodo de pH 27
función 3-69
sensores 3-69
sondas de temperatura 6-6
Cambiador de muestras
cabeza de valoración 2-17
conectar 10-10
definir 2-25
función auxiliar (Changer) 6-9
parámetros cabeza de valoración 3-14
serie de muestras(s) 5-6, 5-15
Cambiar: mando (<F4>) 10
Cambio de buretas: ver buretas
Cambio de lengua 11, 6-27
Centronics: ver interface/tarjeta
Código (func. sincronización) 7-18
Código de barras 6-43
Coeficiente de correlación 3-62, 3-65, 8-6
Combinación de teclas 10
Concentración (reactivo) 2-4, 8-3
Concepto
de manejo 7
de métodos 19
Conductímetro 2-8
Constantes
ejemplo 8-18
función auxiliar cálculo 6-20
función cálculo 3-66
Continua (adición de reactivo) 3-40, 3-60
Control de entrada
activar salida mando a distancia (función auxiliar)
6-33
func. salida mando a distancia 7-16
Control de equilibrio (adquisición de los valores
medidos)
func. medir 3-17
func. valoración EP 3-39
func. valoración EQP 3-25
func. valoración EQP (Ipol/Upol) 3-49
Copia de la memoria
carga al titulador 6-39
copia a la tarjeta memoria 6-39
transferir al ordenador 6-41
Criterios de terminación combinados 3-34, 3-53
CSTAT (coeficiente de correlación) 3-62, 8-6
Curso del análisis (ejemplo )
calibración electrodos de pH 27
comparación: Cabeza de valoración 1/ST20A
5-16
con cambiador de muestras 5-17
determinación CaCl 2 5-7
determinación del contenido HCl 12
determinación del título (NaOH) 32
regulación del pH 5-10
Cursor 10
Curvas de valoración
1ª derivada 3-30, 3-31, 3-54, 8-15
2ª derivada 3-31, 3-54, 8-15
impresión 3-75, 6-24
visualización 5-7, 5-8, 6-21
Cx (constantes de cálculo) 8-7
Data Transfer (transmisión de datos) 6-36
Datos de calibración
entradas de medida 6-29
sensores 2-9, 2-10, 3-69
sondas de temperatura 2-12
Datos de muestra
añadir 4-8
borrar 4-6
cambiar 4-7
condiciones de memorización 36, 4-3
desde ordenador 4-13
entrar 4-4, 5-5, 5-10
imprimir 4-7
informe 3-75, 6-24
lista 4-4, 4-6
memoria (menú Sample) 4-4, 5-5, 5-10
13-12 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Indice
Datos técnicos
agitador 12-5
buretas 12-5
conector Centronics 12-8
conector DIN 12-7
conector RS232C (tarjeta RS) 12-9
conector TTLIO 12-7
enchufe RS232C (tarjeta Centronics) 12-7
memorias 12-5
motor de bureta 12-5
pantalla 12-5
periféricos 12-6
salidas 12-4
sistema de medida 12-3
Deriva (medida del potencial) 3-17
Desviación estándar relativa (srel) 3-67
Desviación estándar (s) 3-67
Dinámica (adición de reactivo)
func. valoración EP 3-39
func. valoración EQP 3-22
Directorio (tarjeta memoria) 6-37
Diversos (Misc. ...) 6-26
DLWin Software 4-13, 6-41
DL58
funciones 7-3
mandos 7-3
medios auxiliares 7-3
Documentación
información 4
Dosificación
continua 6-13
continua (Ipol/Upol) 6-15
Dosificación previa
func. valoración en dos fases 7-13
func. valoración EP 3-38
func. valoración EP (Ipol/Upol) 3-59
func. valoración EQP 3-21
func. valoración EQP (Ipol/Upol) 3-48
Dosificar
función 3-19
función auxiliar (buretas) 6-12
función auxiliar (ST20A) 6-10
E (potencial) 3-17, 8-7
EHNV (valor de semineutralización) 3-20, 8-4
Electrodo de referencia 2-7
Electrodos (METTLER TOLEDO) 2-7, 2-13
Electrodos de pH
calibrar 26
pendiente 26
punto cero 26
Electrodos polarizados
almacenar 2-13
añadir 2-14
cambiar 2-13
tarjeta KF 2-13
medios auxiliares 2-13
Embolo
colocación 10-4
montaje 10-4
Encabezamiento de informe 6-27
Entrada mando a distancia
activar (función auxiliar) 6-32
función 7-15
Entradas de sensores
func. salida mando a distancia 7-17
sensores 2-8
sensores polarizados 2-13
Entradas mando a distancia
definir 7-4
medios auxiliares 7-4
Entrar tiempo
métodos 3-11
reactivo 2-5
sensor de temperatura 2-11
sensores 2-9
valores auxiliares 2-15
EP
absoluto 3-41
relativo 3-41
EPOT 8-4
Esc (salir) mando (<F1>) 9
Estadísticas
cambiar 6-22
incluir en el cálculo 3-67
Estado
mando a distancia 6-42
muestras 4-7
periféricos 2-18
Están definidos (máscara)
bloquear 6-30
significado 5-5
Estándar (procedimiento de evaluación)
explicación 8-14
parámetro de evaluación 3-34, 3-53
reconocimiento punto de equivalencia 3-28,
3-50
ET1 3-22, 3-38, 8-4
ET2 3-22, 3-38, 8-4
Evaluación
func. pH/mV-Stato 3-65
func. valoración EQP 3-34
func. valoración EQP (Ipol/Upol) 3-53
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 13-13
Indice
Evaluación posterior
condición satisfecha 5-11
definir una condición 3-35, 3-36, 3-55
ejecución 6-18
Experto (nivel de usuario) 6-31
Extinción 2-8
f (factor de corrección) 4-6, 5-6, 8-3
Fallos 9-6
Fecha
entrar 6-27
seleccionar formato 6-26
Fecha caducidad (reactivo) 5-4
Forma de control (activar salida mando a distancia)
control por entrada 6-33
marcha/parada 6-33
secuencial 6-34
Forma de control (func. salida mando a distancia)
control por entrada 7-16
marcha/parada 7-16
secuencial 7-17
tiempo fijo 7-16
Fórmula(s)
constantes 8-18
función auxiliar cálculos 6-20
función cálculo 3-66
función título 3-72
función valores auxiliares 3-73
para limitar los puntos de equivalencia 8-19
resultados 8-17
Función(es)
añadir 7-8
borrar 7-8
cambiar 3-10
DL58 7-3
explicación 20, 3-3
número máximo 7-9
seleccionar 3-7
visión de conjunto 24
Funciones auxiliares
explicación 9, 6-3
teclas 6-3
visión de conjunto 1-4, 1-5
Funciones de cálculo 3-66
H (valor auxiliar) 2-15, 8-3
Hoja de datos
tarjeta KF 10-12
tarjeta pH 10-12
ID métodos
cambiar 3-9
definir 3-11
entrar 4-4, 5-5
explicación 3-4
métodos estándar 3-5
métodos METTLER 3-6
ID muestras 4-5, 5-6
Impresora
ASCII 2-18
conectar 10-10
configuración 2-19, 2-20
definir 2-18
In 3.1/3.2 (entradas TTL) 6-32
Incremental (adición de reactivo) 3-24, 3-48, 7-14
Indicación
func. valoración EP (Ipol/Upol) 3-57
func. valoración EQP (Ipol/Upol) 3-46
Indicación amperométrica
func. valoración EP (Ipol/Upol) 3-58
func. valoración EQP (Ipol/Upol) 3-47
Indicación de fecha
métodos 3-11
reactivo 2-5
sensores 2-9
sondas de temperatura 2-11
valores auxiliares 2-15
Indicación voltamétrica
func. valoración EP (Ipol/Upol) 3-57
func. valoración EQP (Ipol/Upol) 3-46
Indicativos (forma)
ejemplos 8-9, 8-10
resumen 8-12
Informe
función 3-74
función auxiliar (Report) 6-24
Informe de comunicación (ordenador) 2-24
Instalación (menú Setup) 2-3
Instalación y mantenimiento 10-3
Instrucción
función 7-11
función salida mando a distancia 7-17
Interface Centronics
configuración impresora 2-19
impresora GA42 2-18
Interface RS232
configuración impresora 2-20
impresora GA42 2-20
Interpolación (evaluación) 8-14
Intervalo (acondicionamiento) 3-15
Intervalo de memorización (func. pH/mV-Stato)
3-64
13-14 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Indice
Juego de caracteres (ordenador) 2-24
LabX Software 4-13, 6-41
Lector de código de barras
conectar 10-16
definir 2-24
Lengua (cambiar) 11, 6-27
Límites de peso
entrar 3-12
en entrada de muestra 4-5, 5-6
Límites volumen
en entrada de muestras 4-5, 5-6
entrar 3-12
Lista (aclaración) 8
Lista datos de muestras 4-4, 4-6
Lista de resultado
avisos 5-9
bloquear 6-30
visualizar 6-18
Llenar (lavar)
función auxiliar (ST20A) 6-10
parámetros cabeza de valoración 3-14
Llenar punta (bureta) 6-12
M (masa molar) 3-68, 8-3
m (tamaño muestra) 8-3
Mando a distancia 6-42
Mantenimiento
bureta 10-7
titulador 10-3
Manual (entrada de temperatura) 3-13
Marcha/Parada
activar salida mando a distancia (función auxiliar)
6-33
func. salida mando a distancia 7-16
Márgenes (informe) 2-19
Masa molar M
entrar 3-13
func. cálculo 3-68
Máscara (aclaración) 8
Máscara datos de muestras
menú análisis 5-5
menú muestras 4-5
Máximo (procedimiento de evaluación)
explicación 8-14
parámetro evaluación 3-34, 3-53
reconocimiento punto de equivalencia 3-28,
3-50
Medición del voltaje (Ipol/Upol) 6-7
Medida
de la conductividad 2-8
de la corriente (Ipol/Upol) 6-7
del pH 35
del potencial 6-4, 6-7
Medidas de seguridad
en el puesto de trabajo 5
para su protección 5, 10-3
Medios auxiliares
borrar 2-3
cambiar 2-3, 5-8
copiar en titulador 6-39
DL58 7-3
explicación 2-3
imprimir 2-3
lista 2-3
transferir a ordenador 6-41
Medir (función) 3-17
Memoria de datos
de muestras (menú Sample) 4-3
de usuario 12-5
Memorización
datos de muestras 4-3
electrodos polarizados 2-13
instrumento para entrada mando a distancia 7-5
instrumento para salida mando a distancia 7-6
métodos 3-7
reactivos 2-6
resultados 8-20
resultados sin elaborar 8-20
sensores 2-10
solventes 2-28
sondas de temperatura 2-12
valores auxiliares 2-16
valores medidos 3-64, 3-74, 8-20
Memorizar valor auxiliar 3-73
Menús
explicación 8
teclas 8
visión de conjunto 1-2, 1-3
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 13-15
Indice
Método
almacenar 3-7
borrar 3-5
cambiar 3-7
cambiar ( método en curso: DL55/DL58) 5-13
cambiar (actual serie de muestras) 5-12
desde ordenador 4-5
desde tarjeta memoria 4-5
elaboración 24
explicación 3-3
imprimir 3-4
Métodos
imprimir lista 3-4
menú 3-3
seleccionar 3-4
Método(s) del usuario
borrar 3-6
copiar en la tarjeta memoria 6-38
copiar en titulador 6-39
explicación 19, 3-6
imprimir 3-6
título con valoración EQP 34
transferir a ordenador 6-41
Método(s) estándar
cambiar 25, 30
ejemplo de uso 22
explicación 19, 3-5
funciones 20
imprimir 3-5
número 20
Métodos especiales
DL58 7-7
explicación 7-3
Métodos METTLER
90001: Ejecución 14
90002: Ejecución 27
cambiar 25, 3-6
explicación 19
imprimir 3-6
Mínimo (procedimiento de evaluación)
explicación 8-14
parámetro de evaluación 3-34, 3-53
reconocimiento punto de equivalencia 3-28,
3-50
Misc. ... (Diversos) 6-26
Motor de bureta
instalar 10-3
seleccionar 2-5
Muestra
función 3-12
menú memoria de datos de muestras 4-3
en curso (máscara): aviso 5-6
urgente (DL55) 4-10
Muestras
estado 4-7
número máximo 8-20
neq (número punto(s) de equivalencia) 8-4
Nivel de usuario 6-31
Nombre
función 3-11
métodos estándar 19
métodos METTLER- 19
Número de equivalencia
entrada 3-13
func. cálculo 3-68
OK mando (<F5>) 9
Ordenador
código de barras (string) 6-43
conectar 10-10
configuración 2-23
lector de código de barras 2-24
mando a distancia 6-42
transmisión de datos 6-40
Out 3.1...3.4 (salidas TTL) 6-33
P1/P2 8-6
Papel (impresora)
formato 2-19
hojas únicas 2-19
papel sin fin 2-19
Parada para evaluación posterior 3-35, 3-55, 5-11
Parada pospuesta 3-42, 3-61
Parámetros (aclaración) 20
Paridad
balanza 2-22
impresora 2-20
ordenador 2-23
terminal 2-25
Parte trasera del titulador 10-9
Pendiente
cálculo 3-69
electrodo de pH 26
explicación 2-9
registrar 2-9
valor máximo/mínimo 3-71
valor teórico 2-9
13-16 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Indice
Periféricos (menú Setup) 2-18
Peso
entrar 4-5, 5-6
seleccionar 3-12
transmisión desde balanza 4-12
pH-stato (ejemplo) 3-65
pH/mV-Stato (función) 3-62
Plato giratorio (cambiador de muestras) 6-10
Posición elevador (cambiador de muestras) 6-9
Potencial
1/2 (parámetros de evaluación) 3-35, 3-54
inicial ET1 3-22
Proceso de evaluación
explicación 8-14
func. valoración EQP 3-28, 3-34
func. valoración EQP (Ipol/Upol) 3-50, 3-53
Punto cero (sensor)
cálculo 3-69
electrodo de pH 26
explicación 2-9
registrar 2-9
valor teórico 2-9
Punto cero (sensor de temperatura)
explicación 2-11
registrar 2-12
valor teórico 2-12
Punto de equivalencia
explicación 3-20, 3-45, 8-14
fórmulas para restricción 8-19
número máx./determinación muestras 8-20
rango 3-32, 3-51
reconocimiento 3-28, 3-50
Punto de inflexión 8-14
Punto final
func. pH/mV-Stato 3-63
func. valoración EP 3-41
func. valoración EP (Ipol/Upol) 3-60
Puntos de medida para reconocimiento del punto
de equivalencia 3-28, 3-44, 3-50
Q 8-4
QDISP 3-19, 8-7
QEND 8-4
QEX 8-4
QP1/QP2 8-4
QSTAT 8-6
QT1/QT2 8-6
QTOT 8-6
R (Resultado) 3-66, 8-7
Rango
punto de equivalencia 3-32, 3-51
punto final 3-40, 3-60
Rango de regulación
func. pH/mV-Stato 3-63
func. valoración EP 3-40
func. valoración EP (Ipol/Upol) 3-60
Ranuras 10-9
Reactivo
almacenar 2-6
añadir 2-6
borrar 2-4
cambiar 2-4
concentración 2-4, 8-3
Reconocimiento (punto de equivalencia) 3-28,
3-50
Reducción curso del análisis 6-30
Registro (tarjeta memoria) 6-37
Reloj (titulador) 6
Report (informe) 6-24
Reset (tecla) 7, 5-12
Resultado(s)
borrar 6-22
condiciones de memorización 8-20
ejemplos 3-67, 8-17
función auxiliar (Results) 6-18
hacer un informe 3-75, 6-24
número máximol/serie de muestras 8-20
R 3-66, 3-71
Rx 6-20
unidades 3-67, 8-17
valoración de aprendizaje 5-9
Resultados sin elaborar
condiciones de memorización 8-20
explicación 3-18
hacer un informe 3-75, 6-24
resumen 8-8
Results (Resultados) 6-18
Run
menú análisis 5-3
teclas función 5-3
Rutina (nivel de usuario) 6-31
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 13-17
Indice
s (desviación estándar) 8-7
Salida mando a distancia
activar (función auxiliar) 6-33
función 7-16
Salidas mando a distancia
definir 7-5
medios auxiliares 7-5
Secuencial
activar salida mando a distancia (función auxiliar)
6-34
función salida mando a distancia 7-16
Segmentada (procedimiento de evaluación)
explicación 8-15
parámetro de evaluación 3-34, 3-53
reconocimiento punto de equivalencia 3-28,
3-50
Señal
acústica 6-28
de entrada (func. entrada mando a distancia)
7-15
Sensor(es)
almacenar 2-10
añadir 2-10
borrar 2-7
cambiar 2-7
función auxiliar 6-4
medios auxiliares 2-7
pH tarjeta 2-7
Serie de muestras
borrar 4-6
con cabeza val. auto 5-6, 5-15
con ST20A 5-6, 5-15
entrar 4-4
Serie de muestras (DL55/DL58)
ejecución 5-15
entrar 4-9
Setup (menú instalación) 2-3
Símbolo inicial/final (ordenador) 2-24
Símbolos
explicación 8-3
lista 8-3
Sincronización (función) 7-18
Sistema
información 9-3, 9-6
ordenador/terminal 2-23
SLOPE (pendiente de electrodos) 8-3
Software LabX 4-13, 6-41
Sólo salto más abrupto 3-31
Soluciones tampón
tampones DIN/NIST 3-70
tampones MERCK 3-69
tampones METTLER TOLEDO 3-69
Solventes
almacenar 2-28
añadir 2-28
borrar 2-27
cambiar 2-27
Sonda de temperatura
almacenar 2-12
calibrar 6-6
enchufe Lemo 10-10
medios auxiliares 2-11
Pt100/Pt1000 2-11
seleccionar 3-13
srel (desviación estándar relativa) 8-7
ST20A
conectar 10-10
salidas 2-28
Stand 1/2 (cabezas de valoración) 2-17
Stirrer (agitador) 6-8
Sustancia patrón 30, 31
T (temperatura) 7-10
t (título) 3-72, 8-3
t(máx)
func. medir 3-18
func. temperatura 7-10
func. valoración EP 3-39
func. valoración EQP 3-25
func. valoración EQP (Ipol/Upol) 3-49
t(mín)
func. medir 3-17
func. temperatura 7-10
func. valoración EP 3-39
func. valoración EQP 3-25
func. valoración EQP (Ipol/Upol) 3-49
Tabla de valores medidos
hacer un informe 3-75, 6-24
visualizar 5-7, 6-21
Tamaño de muestras 3-21, 3-38, 3-48, 3-59, 8-3
Tarjeta Centronics
installación 10-10
posibilidades de conexión 10-9
13-18 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Indice
Tarjeta KF
electrodos polarizados 2-13
entradas de medida: datos de calibración 6-29
entradas y salidas 10-10
instalar 10-11
Tarjeta memoria
directorio 6-37
formatización 6-36, 6-40
inserción 10-11
interface 12-6
tipos 12-6
uso 6-36
Tarjeta pH
entradas de medida: datos de calibración 6-29
entradas y salidas 10-10
instalar 10-11
sensores 2-7
Tarjeta RS
instalación 10-11
posibilidades de conexión 10-11
Tecla(s)
de entrada 10
flecha 7
Reset 7, 5-12
de mandos 9
Teclado (titulador) 7
Teclado externo
asignación de las teclas 10-15
conectar 10-15
definir 2-26
Temperatura
corrección 3-13, 3-69
entrar 2-9, 4-6, 5-6
función 7-10
medir 2-9, 3-13, 3-18, 3-20, 3-37, 3-62, 3-69,
7-10
medir (función auxiliar) 6-5
Tendencia
func. pH/mV-Stato 3-64
func. valoración EP 3-42
func. valoración EP (Ipol/Upol) 3-61
func. valoración EQP 3-32
func. valoración EQP (Ipol/Upol) 3-51
Terminación (parámetros)
func. pH/mV-Stato 3-64
func. valoración EP 3-42
func. valoración EP (Ipol/Upol) 3-61
func. valoración EQP 3-33
func. valoración EQP (Ipol/Upol) 3-52
Terminal
asignación de teclas 10-17
conectar 10-11
configuración 10-16
sincronización 2-25
Tiempo
entrar 6-27
func. agitar 3-16, 5-7
func. muestra (acondicionar) 3-15
seleccionar formato 6-26
Tiempo
de agitación 3-16
de espera 3-16
de mezcla (func. valoración en dos fases) 7-14
de separación (func. valoración en dos fases)
7-14
fijo (func. salida mando a distancia) 7-16
máx. (func. entrada mando a distancia) 7-15
máximo (func. valoración EP (Ipol/Upol)) 3-61
TIME 5-4, 8-3
Tipo de medición
func. valoración en dos fases 7-14
func. valoración EQP 3-25
func. valoración EQP (Ipol/Upol) 3-49
Tipo de sincronización (func. sincronización) 7-18
Tipo de transmisión (balanza)
bidireccional 2-21
unidireccional 2-21
Tipo de transmisión (ordenador) 2-24
Titulador
acumulador 6
apertura de la carcasa 10-3
autotest 7
conexiones 6, 10-10
configuración 6-26
documentación 4
ID (identificación ) 6-27
limpieza de la carcasa 10-3
mando a distancia 6-42
reloj interno 6
teclado 7
vista frontal 6
vista trasera 6, 10-10
Título
controlar 2-5
determinación (disolución NaOH) 32
entrar 2-4
explicación 30
función 3-72
registrar 2-4
Transmisión de datos (Data Transfer) 6-36
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 13-19
Indice
TTL (entradas)
activar 6-32
cableado electrónico 10-14
cableado mecánico 10-13
TTL (salidas)
activar 6-33
cableado electrónico 10-14
TTLIO (conectores)
datos técnicos 12-7
entradas y salidas 10-13
usado con titulador 10-13
Umbral 3-29, 3-51
Unidad de medida (sensores) 2-8
Unidades 3-67, 8-17, 8-18
Valoración
a punto de equivalencia 3-20, 3-45
a punto final 3-37, 3-56
de aprendizaje función 3-43
en dos fases función 7-12
EP (función) 3-37
EP (Ipol/Upol) (función) 3-56
EQP (función) 3-20
EQP (Ipol/Upol) (función) 3-45
manual 6-14
manual (Ipol/Upol) 6-16
método especial 7-12
previa (func. pH/mV-Stato) 3-63
resultados 5-9
Valoraciones Karl-Fischer
accesorios 11-12
Valor del umbral
procedimiento de evaluación estándar 3-30
procedimiento de evaluación: asimétrica 3-30
procedimiento de evaluación: mínimo/máximo
3-30
procedimiento de evaluación: segmentada 3-31
Valor de semineutralización 3-20
Valor medio x
func. cálculo 3-67
func. título 3-72
func. valor auxiliar 3-73
Valor(es) auxiliar(es)
almacenar 2-16
entrar 2-15
función 3-73
medios auxiliares 2-15
registrar 2-15
Valores medidos
instrucciones para memorización 3-64, 3-74,
8-20
número máximo/func. valoración 8-20
visualizar 5-7, 6-21
VDISP 3-19, 8-7
Velocidad bombeo (solventes) 2-27
Velocidad de agitación
cambiar 5-8, 6-8
definir 3-16
VEND 8-4
VEQ 8-4
Version software 1-1
VEX 8-4
Volumen
entrar 4-5, 5-6
seleccionar 3-12
Volumen bureta
incremento mínimo 3-24
seleccionar 2-5
Volumen fijo
aviso 5-5
seleccionar 3-12
Volumen máximo
func. pH/mV-Stato 3-64
func. valoración EP 3-42
func. valoración EP (Ipol/Upol) 3-61
func. valoración EQP 3-33
func. valoración EQP (Ipol/Upol) 3-52
VP1/VP2 8-4
VSTAT 8-6
VT 3-65, 8-6
VT1/VT2 8-6
VTOT 8-6
x (valor medio) 8-7
z (valencia, nº de equivalencia) 3-13, 3-68, 8-3
ZERO (punto cero, electrodo) 8-3
13-20 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Indice analitico
Indice analitico
I numeri separati da un trattino (esempio: 3-38) si riferiscono alle sezioni e pagine del manuale
d'uso, mentre i numeri senza trattino si riferiscono alle istruzioni di base (esempio: 9).
<F1>...<F5> 9, 10
∆E
funzione Misurare 3-17
funzione Titolazione EP 3-39
funzione Titolazione EQP 3-25
funzione Titolazione EQP (Ipol/Upol) 3-49
∆E(nom.)
funzione Titolazione EP 3-39
funzione Titolazione EQP 3-22
∆t
funzione Misurare 3-17
funzione Titolazione EP 3-39
funzione Titolazione EQP 3-25
funzione Titolazione EQP (Ipol/Upol) 3-49
Abbreviare l’analisi 6-30
Accessori
accessori facoltativi 11-7
equipaggiamento di base 11-4
Accumulatore 6
Aggiunta di titolante
funzione pH/mV-Stat 3-63
funzione Titolazione a due fasi 7-14
funzione Titolazione EP 3-39
funzione Titolazione EP (Ipol/Upol) 3-60
funzione Titolazione EQP 3-22
funzione Titolazione EQP (Ipol/Upol) 3-48
Agitare (funzione) 3-16
Agitatore
funzione ausiliaria (Stirrer) 6-8
selezionare l’uscita 2-17
Allineamento degli ingressi di misura 6-28
Analisi
arresto 5-11
eclissare 5-11
inizio 5-4
interrompere (Reset) 5-12
menu „Run“ 5-3
svolgimento del menu 5-4
Arresto (parametri di)
funzione pH/mV-Stat 3-64
funzione Titolazione EP 3-42
funzione Titolazione EP (Ipol/Upol) 3-61
funzione Titolazione EQP 3-33
funzione Titolazione EQP (Ipol/Upol) 3-52
Arresto per rivalutazione 3-35, 3-55, 5-11
Asimmetrico (procedura di valutazione)
parametro di valutazione 3-34, 3-53
riconoscimento del punto d’equivalenza 3-28,
3-50
spiegazione 8-16
Asterisco (*) 3-4, 3-11
Attivare
entrata comando a distanza 6-32
uscita comando a distanza 6-32
Autocampionatore
collegare 10-10
definire 2-25
funzione ausiliaria (Changer) 6-9
parametri dello stativo di titolazione 3-13
serie di campioni 5-6, 5-15
stativo di titolazione 2-17
Avanzamento
della pagina (stampante) 2-19
delle linee (stampante) 2-19
Banca dati 12-5
Banda di controllo
funzione pH/mV-Stat 3-63
funzione Titolazione EP 3-40
funzione Titolazione EP (Ipol/Upol) 3-60
Bidirezionale (modo di trasmissione) 2-21, 4-12
Bilancia(e)
collegare 10-10
configurazioni (METTLER TOLEDO) 2-22
configurazioni (SARTORIUS) 2-22
definire 2-21
Buretta
bolle d’aria 10-8
inserire 10-6
installare 10-4
lavare 6-11
manutenzione 10-4, 10-7
montare 10-5
riempire 13
Burette (buretta, funz. ausil.) 6-11
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 13-21
Indice analitico
C (costante) 3-66, 8-7
c (conc. nominale) 8-3
c * t (conc. reale) 8-3
Calcoli supplementari 6-20
Calcolo(i)
eseguire (funzione ausiliaria) 6-20
formule 8-17
funzione 3-66
indici 8-9
Calibrare
elettrodo pH 27
funzione 3-69
sonde 3-69
sonde della temperatura 6-6
Cambiare la lingua 11, 6-27
Campione (funzione) 3-12
Campione attuale (maschera)
indicazioni 5-6
Campione urgente (DL55/DL58) 4-10
Campioni
numero massimo 8-20
stato 4-7
Campo (punto di equivalenza) 3-32, 3-51
Campo di regolazione
funzione pH/mV-Stat 3-63
funzione Titolazione EP 3-40
funzione Titolazione EP (Ipol/Upol) 3-60
Carattere iniziale/finale (computer) 2-24
Carta (stampante)
formato 2-19
infinito 2-19
pagina singola 2-19
Changer (autocampionatore, funz. ausil.) 6-9
Code (funzione Sync) 7-18
Codice a barre 6-43
Coefficiente di correlazione 3-62, 3-65, 8-6
Comando a distanza 6-42
Combinazioni di tasti 10
Computer
codice a barre 6-43
collegare 10-10
comando a distanza 6-42
configurazione (titolatore) 2-23
lettore di codici a barre 2-24
trasmissione dei dati 6-40
Concentrazione (titolante) 2-4, 8-3
Concetto
d’impiego 7
di metodo 19
Condizionare (ST20A) 3-15
Condizioni di arresto combinate 3-34, 3-53
Conduttimetro 2-8
Connettore(i) TTLIO
dati tecnici 12-7
ingressi ed uscite 10-13
utilizzati dal titolatore 10-13
Continua (aggiunta di titolante) 3-40, 3-60
Controllo dall’ingresso
attivare l’uscita comando a distanza (funzione
ausiliaria) 6-33
funzione Uscita comando a distanza 7-16
Copia(e) della memoria
caricare sul titolatore 6-39
copiare su scheda di memoria 6-39
trasferire al computer 6-41
Costanti
esempi 8-18
funzione ausiliaria Calcoli 6-20
funzione Calcolo 3-66
CSTAT (coefficiente di correlazione, pH-Stat)
3-62, 8-6
Cursore 10
Curve di titolazione
1 a derivata 3-30, 3-31, 3-54, 8-15
2 a derivata 3-31, 3-54, 8-15
documentare 3-75, 6-24
visualizzare 5-7, 5-8, 6-21
Cx (costante di calcolo) 8-7
Data
introdurre 6-27
selezionare il formato 6-26
Data (indicazione)
metodi 3-11
sonde 2-9
sonde della temperatura 2-11
titolanti 2-5
valori ausiliari 2-15
Data di scadenza (titolante) 5-4
Data Transfer (trasferimento dei dati, funz. ausil.)
6-36
Dati di calibrazione
ingressi di misura 6-29
sonde 2-9, 2-10, 3-69
sonde della temperatura 2-12
Dati dei campioni
aggiungere 4-8
cancellare 4-6
condizioni di memorizzazione 36, 4-3
definire dal computer 4-13
documentare 3-75, 6-24
introdurre 4-4, 5-5, 5-10
modificare 4-7
stampare 4-7
13-22 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Indice analitico
Dati tecnici
agitatore 12-5
buretta 12-5
connettore Centronics 12-8
connettore RS232C (Opzione RS) 12-9
connettori TTLIO 12-7
display 12-5
connettore DIN 12-7
memoria 12-5
motore della buretta 12-5
periferiche 12-6
sistema di misura 12-3
spina RS232C (Opzione Centronics) 12-7
uscite 12-4
Deriva (misura del potenziale) 3-17
Deviazione standard relativa srel 3-67
Deviazione standard s 3-67
Dinamica (aggiunta di titolante)
funzione Titolazione EP 3-39
funzione Titolazione EQP 3-22
Disturbi 9-6
DL58
comandi 7-3
funzioni 7-3
mezzi ausiliari 7-3
DLWin Software 4-13, 6-41
Documentazione
informazioni 4
Dosaggio
continuo 6-13
continuo (Ipol/Upol) 6-15
Dosare
funzione 3-19
funzione ausiliaria (buretta) 6-12
funzione ausiliaria (ST20A) 6-10
E (potenziale) 3-17, 8-7
ESC (<F1>) 9
EHNV (valore di semineutralizzazione) 3-20, 8-4
Elettrodi (METTLER TOLEDO) 2-7, 2-13
Elettrodo di riferimento 2-7
Elettrodo per pH
calibrare 26
pendenza 26
punto zero 26
Emissione (protocolli) 3-74, 6-24
EP
assoluto 3-41
relativo 3-41
EPOT 8-4
Esperto (livello utente) 6-31
Estinzione 2-8
ET1 3-22, 3-38, 8-4
ET2 3-22, 3-38, 8-4
f (fattore di correzione) 4-6, 5-6, 8-3
Formula(e)
costanti 8-18
funzione ausiliaria Calcoli 6-20
funzione Calcolo 3-66
funzione Titolo 3-72
funzione Valore ausiliario 3-73
per circoscrivere il punto di equivalenza 8-19
risultati 8-17
Funzione(i)
aggiungere 7-8
cancellare 7-8
DL58 7-3
modificare 3-10
numero massimo 7-9
riassunto 24
selezionare 3-7
spiegazione 20, 3-3
Funzioni ausiliarie
riassunto 1-4, 1-5
spiegazione 9, 6-3
tasti 6-3
Funzioni di calcolo 3-66
H (valore ausiliario) 2-15, 8-3
ID (identificazione) dei metodi
definire 3-11
introdurre 4-4, 5-5
metodi METTLER 3-6
metodi standard 3-5
modificare 3-9
spiegazione 3-4
ID del campione 4-5, 5-6
In 3.1/3.2 (ingressi TTL) 6-32
In funzione/stop (modo di controllo)
attivare uscita comando a distanza (funzione
ausiliaria) 6-33
funzione Uscita comando a distanza 7-16
Incrementale (aggiunta di titolante) 3-24, 3-48,
7-14
Indicazione
funzione Titolazione EP (Ipol/Upol) 3-57
funzione Titolazione EQP (Ipol/Upol) 3-46
Indicazione amperometrica
funzione Titolazione EP (Ipol/Upol) 3-58
funzione Titolazione EQP (Ipol/Upol) 3-47
Indicazione del tempo
metodi 3-11
sonde 2-9
sonde della temperatura 2-11
titolanti 2-5
valori ausiliari 2-15
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 13-23
Indice analitico
Indicazione voltametrica
funzione Titolazione EP (Ipol/Upol) 3-57
funzione Titolazione EQP (Ipol/Upol) 3-46
Indicazioni di sicurezza 1-1
Indice (scheda di memoria) 6-37
Indici
esempi 8-9, 8-10
riassunto 8-12
Ingressi del comando a distanza
definire 7-4
mezzi ausiliari 7-4
Ingressi TTL
attivare 6-32
avvolg. elettronico („wiring“) 10-14
avvolg. meccanico 10-13
Ingresso del comando a distanza
attivare (funzione ausiliaria) 6-32
funzione 7-15
Ingresso delle sonde
sonde 2-8
sonde della temperatura 2-11
sonde polarizzate 2-13
Installazione (menu "Setup") 2-3
Installazione e manutenzione 10-3
Interfaccia Centronics
configurazioni delle stampanti 2-19
stampante GA42 2-18
Interfaccia RS232
configurazioni delle stampanti 2-20
stampante GA42 2-20
Interpolazione (valutazione) 8-14
Intervallo (condizionare) 3-15
Intervallo di memorizzazione (funz. pH/mV-Stat)
3-64
Istruzione
ESC (<F1>) 9
funz. Uscita comando a distanza 7-17
funzione 7-11
OK (<F5>) 9
modificare (<F4>) 10
LabX Software 4-13, 6-41
Lavare
funzione ausiliaria (ST20A) 6-10
parametri dello stativo di titolazione 3-14
la punta 6-12
Lettore di codici a barre
collegare 10-16
definire 2-24
Limiti di peso
introdurre 3-12
introduzione dei dati dei campioni 4-5, 5-6
Limiti di volume
introdurre 3-12
introduzione dei dati dei campioni 4-5, 5-6
Linee di cornice (protocollo) 2-19
Lingua (cambiare) 11, 6-27
Lista dei dati dei campioni 4-4, 4-6
Lista dei risultati
bloccare 6-30
indicazioni 5-9
visualizzare 6-18
Liste (spiegazione) 8
Livello utente 6-31
M (massa molare) 3-68, 8-3
m (quantità del campione) 8-3
Manuale (introduzione della temperatura) 3-13
Manutenzione
buretta 10-7
titolatore 10-3
Maschera dei dati del campione
menu Analisi 5-5
menu "Sample" 4-5
Maschere (spiegazione) 8
Massa molare M
introdurre 3-13
funzione Calcolo 3-68
Massimo (procedura di valutazione)
parametri di valutazione 3-34, 3-53
riconoscimento del punto d’equivalenza 3-28,
3-50
spiegazione 8-14
Memoria dei dati dei campioni (menu "Sample")
4-3
Memoria dei dati utente 12-5
Memoria dei valori ausiliari 3-73
Memorizzare
dati dei campioni 4-3
metodi 3-7
risultati 8-20
risultati grezzi 8-20
sensori polarizzati 2-13
solventi 2-28
sonde 2-10
sonde della temperatura 2-12
strumenti per ingresso comando a distanza 7-5
strumenti per uscita comando a distanza 7-6
titolanti 2-6
valori ausiliari 2-16
valori misurati 3-64, 3-74, 8-20
Menus
riassunto 1-2, 1-3
spiegazione 8
tasti 8
13-24 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Indice analitico
Messaggi di errore
con riferimento al cap. 9.1 9-3
"EPROM test failed" 9-3
"RAM test failed" 9-3
Titolazione in auto-studio 5-11
Metodi
menu 3-3
selezionare 3-4
stampare la lista 3-4
Metodi METTLER
90001: Esecuzione 14
90002: Esecuzione 27
modificare 25, 3-6
spiegazione 19
stampare 3-6
Metodi speciali
DL58 7-7
spiegazione 7-3
Metodi standard
esempi di applicazione 22
funzioni 20
modificare 25, 30
numero 20
spiegazione 19, 3-5
stampare 3-5
Metodi utente
cancellare 3-6
copiare su scheda di memoria 6-38
copiare sul titolatore 6-39
spiegazione 19, 3-6
stampare 3-6
titolo con titolazione EQP 34
trasferire al computer 6-41
Metodo
a partire dal computer 4-5
cancellare 3-5
memorizzare 3-7
modificare 3-7
modificare (metodo in corso: DL55/DL58) 5-13
modificare (serie di campioni in corso) 5-12
realizzare 24
spiegazione 3-3
stampare 3-4
sulla scheda di memoria 4-5
Mezzi ausiliari
cambiare 5-8
cancellare 2-3
copiare su scheda di memoria 6-39
copiare sul titolatore 6-39
DL58 7-3
lista 2-3
modificare 2-3
spiegazione 2-3
stampare 2-3
trasferire al computer 6-41
Minimo (procedura di valutazione)
parametro di valutazione 3-34, 3-53
riconoscimento del punto di equivalenza 3-28,
3-50
spiegazione 8-14
Misc. ... (Varia) 6-26
Misura controllata all’equilibrio
funzione Misurare 3-17
funzione Titolazione EP 3-39
funzione Titolazione EQP 3-25
funzione Titolazione EQP (Ipol/Upol) 3-49
Misura in modo incrementale (tempo)
funzione Titolazione EP 3-41
funzione Titolazione EP (Ipol/Upol) 3-60
funzione Titolazione EQP 3-27
funzione Titolazione EQP (Ipol/Upol) 3-49
Misurare
funzione 3-17
il pH 35
il potenziale 6-4, 6-7
la corrente (Ipol/Upol) 6-7
la tensione (Ipol/Upol) 6-7
Misure di conducibilità 2-8
Misure di sicurezza
per la sicurezza sul lavoro 5
per la Sua incolumità 5, 10-3
Modificare: istruzione (<F4>) 10
Modo di controllo (attivare uscita TTL)
controllato dall’entrata 6-33
in funzione/stop 6-33
sequenziale 6-34
Modo di controllo (uscite TTL)
controllato dall’entrata 7-16
in funzione/stop 7-16
sequenziale 7-17
tempo fisso 7-16
Modo di misura
funzione Titolazione a due fasi 7-14
funzione Titolazione EQP 3-25
funzione Titolazione EQP (Ipol/Upol) 3-49
Modo di sincronizzazione (funzione Sync) 7-18
Modo di trasmissione (computer) 2-24
Modo di trasmissione (bilancia)
bidirezionale 2-21
unidirezionale 2-21
Motore della buretta
inserire 10-3
selezionare 2-5
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 13-25
Indice analitico
neq (numero di punti di equivalenza) 8-4
No. bits dei dati
computer 2-24
stampante 2-20
terminale 2-25
No. bits di stop
computer 2-24
stampante 2-20
terminale 2-25
Numero di equivalenza
funzione Calcolo 3-68
introdurre 3-13
OK (<F5>) 9
Opzione Centronics
installare 10-10
possibilità di collegamento 10-9
Opzione KF
ingressi di misura: dati di calibrazione 6-29
ingressi ed uscite 10-10
installare 10-11
sonde polarizzate 2-13
Opzione pH
ingressi di misura: dati di calibrazione 6-29
ingressi ed uscite 10-10
installare 10-11
sonde 2-7
Opzione RS
installare 10-11
possibilità di collegamento 10-11
Orologio (titolatore) 6
Out 3.1...3.4 (uscite TTL) 6-33
P1/P2 8-6
Parametri (spiegazione) 20
Parità
bilance 2-22
computer 2-23
stampante 2-20
terminale 2-25
Pendenza
calcolo 3-69
elettrodo per pH 26
spiegazione 2-9
trascrizione automatica 2-9
valore minimo/massimo 3-71
valore teorico 2-9
Periferiche (menu "Setup") 2-18
Peso
introdurre 4-5, 5-6
selezionare 3-12
trasferire dalla bilancia 4-12
pH-Stat (esempio) 3-65
pH/mV-Stat (funzione) 3-62
Piatto rotante (autocampionatore) 6-10
Pistone
inserire nella buretta 10-4
montare 10-4
Pompe
funzione 7-11
parametri dello stativo di titolazione 3-13
Posizione del lift (autocampionatore) 6-9
Posizioni delle opzioni 10-9
Potenziale
1/2 (parametri di valutazione) 3-35, 3-54
iniziale ET1 3-22
Predosaggio
funzione Titolazione a due fasi 7-13
funzione Titolazione EP 3-38
funzione Titolazione EP (Ipol/Upol) 3-59
funzione Titolazione EQP 3-21
funzione Titolazione EQP (Ipol/Upol) 3-48
Pretitolazione (funzione pH/mV-Stat) 3-63
Procedura di valutazione
funzione Titolazione EQP 3-28, 3-34
funzione Titolazione EQP (Ipol/Upol) 3-50, 3-53
spiegazione 8-14
Protocollo
funzione 3-74
funzione ausiliaria (Report) 6-24
di comunicazione (computer) 2-24
Punti di misura
condizioni di memorizzazione 3-64, 3-74, 8-20
numero massimo/funz. Titolazione 8-20
visualizzare 5-7, 6-21
Punti di misura per il riconoscimento del punto di
equivalenza 3-28, 3-44, 3-50
Punto di inflessione 8-14
Punto finale
funzione pH/mV-Stat 3-63
funzione Titolazione EP 3-41
funzione Titolazione EP (Ipol/Upol) 3-60
Punto zero (sonda)
calcolo 3-69
elettrodo per pH 26
spiegazione 2-9
trascrizione automatica 2-9
valore teorico 2-9
Punto zero (sonda della temperatura)
spiegazione 2-11
trascrizione automatica 2-12
valore teorico 2-12
13-26 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Indice analitico
Punto(i) di equivalenza
campo 3-32, 3-51
formule per circoscrivere 8-19
numero massimo/determinazione di campioni
8-20
riconoscimento 3-28, 3-50
spiegazione 3-20, 3-45, 8-14
Q 8-4
QDISP 3-19, 8-7
QEND 8-4
QEX 8-4
QP1/QP2 8-4
QSTAT 8-6
QT1/QT2 8-6
QTOT 8-6
Quantità del campione 3-21, 3-38, 3-48, 3-59, 8-3
R (risultati) 3-66, 8-7
Report (protocollo) 6-24
Reset (tasto) 7, 5-12
Results (risultati) 6-18
Riconoscimento (punto di equivalenza) 3-28, 3-50
Risultati grezzi
condizioni di memorizzazione 8-20
documentare 3-75, 6-24
riassunto 8-8
spiegazione 3-18
Risultato(i)
cancellare 6-22
condizioni di memorizzazione 8-20
documentare 3-75, 6-24
esempi 3-67, 8-17
funzione ausiliaria (Results) 6-18
numero massimo/serie di campioni 8-20
R 3-66, 3-71
Rx 6-20
titolazione in auto-studio 5-9
unità 3-67, 8-17
Ritardo nell’arresto 3-42, 3-61
Rivalutazione
condizione soddisfatta 5-11
definire la condizione 3-35, 3-36, 3-55
eseguire 6-18
Routine (livello utente) 6-31
Run
funzione dei tasti 5-3
menu Analisi 5-3
s (deviazione standard) 8-7
Sample (menu Memoria dei dati dei campioni) 4-3
Scheda di memoria
formattare 6-36, 6-40
indice 6-37
inserire 10-11
interfaccia 12-6
tipi 12-6
uso 6-36
Scheda tecnica
opzione KF 10-12
opzione pH 10-12
Segmentato (procedura di valutazione)
parametro di valutazione 3-34, 3-53
riconoscimento del punto di equivalenza 3-28,
3-50
spiegazione 8-15
Segnale
acustico 6-28
d’ingresso (funz. ingresso comando a dist.) 7-15
Sequenziale
attivare l’uscita comando a distanza (funzione
ausiliaria) 6-34
funzione Uscita comando a distanza 7-17
Serie di campioni
cancellare 4-6
con ST20A 5-6, 5-15
con stativo auto 5-6, 5-15
introdurre 4-4
Serie di campioni (DL55/DL58)
eseguire 5-15
introdurre 4-9
Setup (menu Installazione) 2-3
simboli
lista 8-3
spiegazione 8-3
Sistema
computer/terminale 2-23
informazioni 9-3, 9-6
SLOPE (pendenza degli elettrodi) 8-3
Software LabX 4-13, 6-41
Soglia 3-29, 3-51
Solo il salto più ripido 3-31
Soluzioni tampone
tamponi DIN/NIST 3-70
tamponi MERCK 3-69
tamponi METTLER TOLEDO 3-69
Solventi
aggiungere 2-28
cancellare 2-27
memorizzare 2-28
modificare 2-27
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 13-27
Indice analitico
Sonda Pt100/Pt1000 2-11
Sonda(e)
aggiungere 2-10
cancellare 2-7
funzione ausiliaria 6-4
mezzi ausiliari 2-7
modificare 2-7
opzione pH 2-10
Sonda(e) della temperatura
calibrare 6-6
memorizzare 2-12
mezzi ausiliari 2-11
Pt100/Pt1000 2-11
selezionare 3-13
spina per connettore Lemo 10-10
Sonda(e) polarizzate
aggiungere 2-14
memorizzare 2-13
mezzi ausiliari 2-13
modificare 2-13
opzione KF 2-13
Sono definiti (maschera)
bloccare 6-30
significato 5-5
Sostanza standard primaria 30, 31
srel (deviazione standard relativa) 8-7
ST20A
collegare 10-10
uscite 2-28
Stampante
ASCII 2-18
collegare 10-10
configurazioni 2-19, 2-20
definire 2-18
Standard (procedura di valutazione)
parametro di valutazione 3-34, 3-53
riconoscimento del punto di equivalenza 3-28,
3-50
spiegazione 8-14
Statistica
modificare 6-22
includere nel calcolo 3-67
Stativo 1/2 (stativi di titolazione) 2-17
Stativo auto
definizione 2-17
serie di campioni 5-6, 5-15
Stativo di titolazione
mezzi ausiliari 2-17
montare 10-8
no. di comanda 11-7
selezionare 3-13
significato 7-8
Stativo esterno 2-17
Stato
campioni 4-7
comando a distanza 6-42
periferiche 2-18
Stirrer (agitatore) 6-8
Svolgimento dell’analisi (esempi)
calibrazione dell'elettrodo per pH 27
con l’autocampionatore 5-17
confronto: stativo di titol. 1/ST20A 5-16
determinazione del titolo (NaOH) 32
determinazione di CaCl 2 5-7
titolazione pH-Stat 5-10
Sync (funzione) 7-18
T (temperatura) 7-10
t (titolo) 3-72, 8-3
t(max)
funzione Misurare 3-18
funzione Temperatura 7-10
funzione Titolazione EP 3-39
funzione Titolazione EQP 3-25
funzione Titolazione EQP (Ipol/Upol) 3-49
t(min)
funzione Misurare 3-17
funzione Temperatura 7-10
funzione Titolazione EP 3-39
funzione Titolazione EQP 3-25
funzione Titolazione EQP (Ipol/Upol) 3-49
Tabella dei valori misurati
documentare 3-75, 6-24
visualizzare 5-7, 6-21
Tasti
delle istruzioni 9
di introduzione 10
freccia 7
Reset 7, 5-12
Tastiera
esterna 2-26
titolatore 7
Tastiera esterna
assegnazione dei tasti 10-15
collegare 10-15
definire 2-26
Temperatura
correttura 3-13, 3-69
funzione 7-10
introdurre 2-9, 4-6, 5-6
misurare 2-9, 3-13, 3-18, 3-20, 3-37, 3-62,
3-69, 7-10
misurare (funzione ausiliaria) 6-5
13-28 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Indice analitico
Tempo
di agitazione 3-16
di attesa 3-16
di mescolazione (funz. Titol. a due fasi) 7-14
di separazione (funz. Titol. a due fasi) 7-14
fisso (funz. Uscita comando a dist.) 7-16
funzione Agitare 3-16, 5-7
funzione Campione (condizionare) 3-15
introdurre 6-27
massimo (funz. Ingresso comando a dist.) 7-15
massimo (funzione Titolazione EP (Ipol/Upol))
3-61
selezionare il formato 6-26
TIME 5-4
Tendenza
funzione pH/mV-Stat 3-64
funzione Titolazione EP 3-42
funzione Titolazione EP (Ipol/Upol) 3-61
funzione Titolazione EQP 3-32
funzione Titolazione EQP (Ipol/Upol) 3-51
Terminale
assegnazione dei tasti 10-17
collegare 10-11
configurazione 10-16
configurazione (titolatore) 2-25
Testata del protocollo 6-27
TIME 5-4, 8-3
Titolante
aggiungere 2-6
cancellare 2-4
concentrazione 2-4, 8-3
memorizzare 2-6
modificare 2-4
Titolatore
accumulatore 6
aprire l’involucro 10-3
autodiagnosi 7
collegamenti 6, 10-10
comando a distanza 6-42
configurazione (titolatore) 6-26
documentazione 4
ID (identificazione) 6-27
orologio interno 6
pulire l’involucro 10-3
tastiera 7
vista anteriore 6
vista dorsale 6, 10-10
Titolazione
a punto d’equivalenza 3-20, 3-45
a punto finale 3-37, 3-56
EP (funzione) 3-37
EP (Ipol/Upol) (funzione) 3-56
EQP (funzione) 3-20
EQP (Ipol/Upol) (funzione) 3-45
manuale 6-14
manuale (Ipol/Upol) 6-16
Titolazione a due fasi
funzione 7-12
metodo speciale 7-12
Titolazione in auto-studio
funzione 3-43
resultati 5-9
Titolazione Karl Fischer
accessori 11-12
Titolo
funzione 3-11
metodi METTLER 19
metodi standard 19
Titolo (concentrazione)
controllare 2-5
determinazione (sol. di NaOH) 32
funzione 3-72
introdurre 2-4
spiegazione 30
trascrizione automatica 2-4
Trasferimento dei dati (Data Transfer) 6-36
Unidirezionale (modo di trasmissione) 2-21
Unità 3-67, 8-17, 8-18
Unità di misura (sonde) 2-8
Uscita del comando a distanza
attivare (funzione ausiliaria) 6-33
funzione 7-16
Uscite del comando a distanza
definire 7-5
mezzi ausiliari 7-5
Uscite TTL
attivare 6-33
avvolg. elettronico („wiring“) 10-14
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 13-29
Indice analitico
Valore(i) ausiliare(i)
funzione 3-73
introdurre 2-15
memorizzare 2-16
mezzi ausiliari 2-15
trascrizione automatica 2-15
Valore di semineutralizzazione 3-20
Valore di soglia
procedura di valutazione: asimmetrico 3-30
procedura di valutazione: minimo/massimo 3-30
procedura di valutazione: segmentato 3-31
procedura di valutazione: standard 3-30
Valore medio x
funzione Calcolo 3-67
funzione Titolo 3-72
funzione Valore ausiliario 3-73
Valutazione
funzione pH/mV-Stat 3-65
funzione Titolazione EQP 3-34
funzione Titolazione EQP (Ipol/Upol) 3-53
Varia (Misc. ...) 6-26
VDISP 3-19, 8-7
Velocità (agitatore)
definire 3-16
modificare 5-8, 6-8
Velocità della pompa (solventi) 2-27
Velocità di trasmissione
bilance 2-22
computer 2-23
stampante 2-20
terminale 2-25
VEND 8-4
VEQ 8-4
Versione del software 1-1
VEX 8-4
Vista dorsale del titolatore 10-9
Volume
introdurre 4-5, 5-6
selezionare 3-12
Volume della buretta
l’incremento più piccolo 3-24
selezionare 2-5
Volume fisso
avviso 5-5
selezionare 3-12
Volume massimo
funzione pH/mV-Stat 3-64
funzione Titolazione EP 3-42
funzione Titolazione EP (Ipol/Upol) 3-61
funzione Titolazione EQP 3-33
funzione Titolazione EQP (Ipol/Upol) 3-52
VP1/VP2 8-4
VSTAT 8-6
VT 3-65, 8-6
VT1/VT2 8-6
VTOT 8-6
x (valore medio) 8-7
z (numero di equivalenza) 3-13, 3-68, 8-3
ZERO (punto zero degli elettrodi) 8-3
13-30 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Contents
Page
ISO 9001 certificate for METTLER TOLEDO ...................................................................... 15-3
Certificado ISO 9001 para METTLER TOLEDO ................................................................. 15-3
Certificato ISO 9001 per la METTLER TOLEDO ................................................................ 15-3
Declaration of System Validation ................................................................................ 15-4
Déclaración de validación del sistema ........................................................................ 15-6
Dichiarazione di validazione del sistema .................................................................... 15-8
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 15-1
ISO 9001 certificate for METTLER TOLEDO
The Mettler-Toledo GmbH company, Greifensee, was examined and evaluated in 1991 by
the Swiss Association for Quality and Management Systems (SQS), and was awarded the
ISO 9001 certificate. This certifies that Mettler-Toledo GmbH, Greifensee, has a quality
management system that conforms with the international standards of the ISO 9000 series.
Repeat audits are carried out by the SQS at intervals to check that the quality management
system is operated in the proper manner and is continuously updated in relation to changes
brought about.
Certificado ISO 9001 para METTLER TOLEDO
La firma Mettler-Toledo GmbH, Greifensee, ha sido inspeccionada por la Asociación Suiza
para Sistemas de Calidad y Gestión (SQS), habiendo obtenido el certificado ISO 9001.
Esto acredita que Mettler-Toledo GmbH, Greifensee, dispone de un sistema de gestión de
calidad que cumple las normas internacionales (ISO serie 9000).
Con motivo de las inspecciones de repetibilidad por parte de la SQS, se comprueba
periódicamente si el sistema de gestión de calidad se manipula correctamente y se ajusta
de modo continuo.
Certificato ISO 9001 per la METTLER TOLEDO
Il sistema di garanzia della qualità della Società Mettler-Toledo GmbH, Greifensee, è
certificato ISO 9001 sin dal 1991 dall'Associazione Svizzera per Sistemi di Qualità e di
Gestione (SQS), e così fornisce la dimostrazione che il suo sistema Garanzia di Qualità
soddisfa i massimi requisiti.
Il sistema della garanzia della qualità Mettler-Toledo viene verificato periodicamente SQS,
dando così evidenza di un continuo aggiornamento e corretta gestione.
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 15-3
Declaration of System Validation
We herewith inform you that the products/systems:
DL50, DL53, DL55, DL58
including software and accessories were developed, tested and successfully validated according to the
international ISO9001:1994 based Life cycle rules of Mettler-Toledo GmbH, Analytical.
Life cycle checkpoint details were reviewed and approved by the Project Supervisory Group (PSG). The
products/systems were tested to meet functional and performance specifications and release criteria at
release to shipment. In order to support GLP and validation requirements, we will make the following
documents available to an authorized, governmental or regulatory agency for inspection.
• Product Guidelines
• Performance Specifications
• Documentation Plan
• Software Specifications
• Quality Plan
• Project Management System
• Test Plan
• Customer Requirements
• Review Reports
• Source Code
Mettler-Toledo GmbH, Analytical will maintain possession of all documents and their reproductions and
may require a non-disclosure agreement to be provided by those requiring access to these documents.
Schwerzenbach, Dr. Bernhard Grob Dr. Urs Spitz
January 1999 General Manager Manager Business Area
Business Unit Analytical Titration
15-4 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Product Life Cycle Model
Ideas
Project Study
Project Release
Project Start
Project End
Phase:
Basic Know how
Phase:
Feasibility
Phase:
Confirmation
Phase:
Readiness
Product Guidelines
Performance Specifications
Quality Plan
Documentation Plan
Project Plan
Review Report
Software Specifications
Test Plan
Review Report
Review Report
Review Report
Usage
Customer Complaints (CRIS)
Customer Requirements
Phase out
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 15-5
Déclaración de validación del sistema
Por la presente le informamos que los productos/sistemas
DL50, DL53, DL55, DL58
incluido el software y los accesorios, han sido desarrollados, probados y verificados con éxito de acuerdo
con las reglas sobre ciclo de vida de producto de Mettler-Toledo GmbH, Analytical. Estas reglas se basan
en la norma 9001:1994.
Los puntos de control de proyecto han sido comprobados y ratificados por el grupo de Control de Proyecto
(Project Supervisory Group o PSG). La comprobación de los productos/sistemas se ha realizado antes
de su entrega. Como apoyo a las exigencias GLP y de validación, ponemos el siguiente material a
disposición de las personas autorizadas para su examen:
• Imagen del producto
• Especificaciones
• Documentación del proyecto
• Especificaciones del software
• Plan de calidad
• Directiva Gestión de Proyecto
• Plan de ensayos
• Datos del servicio postventa y de deseos del cliente
• Protocolos de revisión
• Código fuente
Mettler-Toledo GmbH, Analytical, conservará la propiedad de todo el material y sus reproducciones y
llegará a un acuerdo de confidencialidad con quienes quieran examinar este material.
Schwerzenbach, Dr. Berbhard Grob Dr. Urs Spitz
Enero 1999 General Manager Manager Business Area
Business Unit Analytical Titration
15-6 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Modelo del ciclo de vida de un producto
Ideas
Estudio del proyecto
Encargo del proyecto
Comienzo del
proyecto
Fin del proyecto
Fase:
Conocimientos
básicos
Fase:
Viabilidad
Fase:
Verificación
Fase:
Preparación
Imagen del producto
Especificaciones
Plan de calidad
Documentación del proyecto
Plan del proyecto
Protocolo de revisión
Especificaciones del software
Plan de ensayos
Protocolo de revisión
Protocolo de revisión
Protocolo de revisión
Utilización
Servicio técnico (CRIS)
Deseos del cliente
Sustitución
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 15-7
Dichiarazione di validazione del sistema
Con la presente Vi informiamo che i prodotti/sistemi
DL50, DL53, DL55, DL58
inclusi software e accessori sono stati sviluppati, controllati e validati secondo le regole tecniche
relative al ciclo vita dei prodotti della Mettler-Toledo GmbH, Analytical. Queste regole si basano sulle
norme ISO 9001:1994.
I dettagli di controllo del progetto sono stati verificati ed approvati dall'organo di supervisione del
progetto (PSG: Project Supervisory Board). I prodotti/sistemi sono stati controllati e verificati prima
della fornitura. Per soddisfare le richieste di validazione e GLP i seguenti documenti sono messi a
disposizione per la visione da parte di personale autorizzato:
• Specifiche del prodotto
• Linee direttive
• Documentazione del progetto
• Specifiche del software
• Sistema di qualità
• Disposizioni per la gestione del progetto
• Piano dei test
• Dati del servizio clientela/esigenze dei clienti
• Rapporti di revisione
• Codice originale
La Mettler-Toledo GmbH, Analytical, rimarrà in possesso di tutti i documenti e di tutte le loro copie e
contrarrà un accordo di discrezione con coloro che desiderassero visionare tali documenti.
Schwerzenbach, Dr. Bernhard Grob Dr. Urs Spitz
Gennaio 1999 General Manger Manager Business Area
Business Unit Analytical
Titration
15-8 METTLER TOLEDO DL50/DL53/DL55/DL58 08/2004
Modello di ciclo vita di un prodotto
Idee
Studio del progetto
Realizzazione del progetto
Inizio del
progetto
Fine del progetto
Fase:
Conoscenze di
base
Fase:
Fattibilità
Fase:
Conferma
Fase:
Disponibilità
Specifiche del prodotto
Linee direttive
Sistema di qualità
Documentazione del progetto
Sistema del progetto
Rapporto di revisione
Specifiche del software
Piano dei test
Rapporto di revisione
Rapporto di revisione
Rapporto di revisione
Utilizzazione
Servizio clientela (CRIS)
Esigenze dei clienti
Sostituzione
08/2004 METTLER TOLEDO DL50/DL53/DL55/DL58 15-9
To protect your METTLER TOLEDO product’s future:
METTLER TOLEDO Service assures the quality, measuring accuracy and
preservation of value of all METTLER TOLEDO products for years to come.
Please send for full details about our attractive terms of service.
Thank you.
Printed on 100% chlorine-free paper, for the sake of our environment.
*P51709614*
Subject to technical changes and to the availability
of the accessories supplied with the instruments.
© Mettler-Toledo GmbH 1997, 1998, 1999, 2003, 2004 ME-51709614F Printed in Switzerland 0408/2.12
Mettler-Toledo GmbH, Analytical, Sonnenbergstrasse 74, CH-8603 Schwerzenbach,
Tel. ++41 1 806 77 11, Fax ++41 1 806 73 50, Internet: http://www.mt.com