1724 1430 1382<br />

Ausschwingrotor 4-fach /<br />

Swing out rotor 4-times<br />

Gestell / frame<br />

1383 1348 1347 1384 1<strong>35</strong>1<br />

0765<br />

Gummieinlage<br />

/<br />

Rubber<br />

insert<br />

90°<br />

Röhrchen / tube<br />

Vacutainer Vacutainer Vacutainer 0509 0513 2078 0536<br />

VK 0534<br />

Chrombadröhrchen<br />

Chromium-bath<br />

tubes<br />

Skal. 10µl-300µl,<br />

15ml, 30ml<br />

VK 05<strong>35</strong><br />

Deckel (Gummi)<br />

Lid (rubber)<br />

Kapazität / capacity ml 4 – 7 4 - 7 8,5 - 10 15 50 1,5 2,0 30<br />

Maße / dimensions ∅ x L mm 13 x 100 16 x 75 16 x 100 17 x 120 29 x 115 11 x 38 44 x 105<br />

Anzahl p. Gestell/number p. frame 5 4 4 1 1 5 1<br />

Anzahl p. Rotor / number p. rotor 20 16 16 4 4 20 4<br />

Drehzahl / speed RPM 5000 5000 5000 5000 5000 5000 5000<br />

RZB / RCF 4388 4388 4388 4528 4528 4193 4332<br />

Radius / radius mm 157 157 157 162 162 150 155<br />

9 (97%) sec 16 16 16 16 16 16 16<br />

9 sec 19 19 19 19 19 19 19<br />

0 sec 232 232 232 232 232 232 232<br />

Temperatur / temperature °C 1) - 10 - 10 - 10 - 10 - 10 - 10 -10<br />

1724 1429<br />

Ausschwingrotor 4-fach /<br />

Swing out rotor 4-times<br />

Reduzierung / adapter<br />

Röhrchen / tube<br />

90°<br />

Kapazität / capacity ml 5 5 6<br />

Maße / dimensions ∅ x L mm 13,5 x 75 12 x 75 12 x 82<br />

Anzahl p. Gestell/number p. frame 19 19 19<br />

Anzahl p. Rotor / number p. rotor 76 76 76<br />

Drehzahl / speed RPM 5000 5000 5000<br />

RZB / RCF 4081 4081 4081<br />

Radius / radius mm 146 146 146<br />

9 (97%) sec 16 16 16<br />

9 sec 19 19 19<br />

0 sec 232 232 232<br />

Temperatur / temperature °C 1) - 10 - 10 - 10<br />

1) Tiefste erreichbare Temperatur bei maximaler Drehzahl und<br />

1 Stunde Laufzeit (nur bei Kühlzentrifuge)<br />

1) Lowest possible temperature during maximum speed and 1 hour<br />

running time (only with cooling centrifuges)<br />

43/56