21 Displayed centrifugation data after switch-on.<br />

After switch-on the centrifugation data from program 1, or from the last program that was used, is displayed.<br />

If the rotor is at standstill, this can be set as follows:<br />

• Keep the PROG key held down for 8 seconds.<br />

After 8 seconds, ∗∗∗ Machine Menu ∗∗∗ appears in the display.<br />

• Press the PROG key as often as necessary until the ->Settings is displayed.<br />

• Press the START key. The SOUND / BELL = off/on is displayed.<br />

• Press the PROG key as often as necessary until Start program = Last/First is displayed.<br />

• Use the adjusting knob to set Last or First.<br />

Last = the program used last, First = program 1.<br />

• Press the START key to save the setting.<br />

Briefly, Store Settings ... is displayed as confirmation, followed by -> Settings.<br />

• Press the OPEN / STOP key once to leave the "Settings" menu or press the OPEN / STOP key twice to leave the<br />

"Machine Menu".<br />

22 Set temperature unit (only for centrifuge with refrigeration)<br />

The temperature can be entered in degrees Celsius (°C) or in degrees Fahrenheit (°F).<br />

When the rotor is at standstill, the temperature unit must be set as follows:<br />

• Keep the PROG key held down for 8 seconds.<br />

After 8 seconds, ∗∗∗ Machine Menu ∗∗∗ appears in the display.<br />

• Press the PROG key as often as necessary until the ->Settings is displayed.<br />

• Press the START key. The SOUND / BELL = off/on is displayed.<br />

• Press the PROG key as often as necessary until Temp Unit = Celsius/Fahrenheit is displayed.<br />

• Use the adjusting knob to set Celsius or Fahrenheit.<br />

• Press the START key to save the setting.<br />

Briefly, Store Settings ... is displayed as confirmation, followed by -> Settings.<br />

• Press the OPEN / STOP key once to leave the "Settings" menu or press the OPEN / STOP key twice to leave the<br />

"Machine Menu".<br />

23 Setting program locks<br />

The following program locks can be set:<br />

LOCK 1 LOCK 1 is displayed in the " " field.<br />

Programs can only be called up but not modified.<br />

LOCK 2 LOCK 2 is displayed in the " " field.<br />

No programs can be called up and modified.<br />

The centrifuge can be controlled via the interface (only for centrifuge with interface).<br />

LOCK 3 No status display.<br />

No program locking. Programs can be called up and modified.<br />

• Keep the PROG key held down for 8 seconds.<br />

After 8 seconds, ∗∗∗ Machine Menu ∗∗∗ appears in the display.<br />

• Press the PROG key as often as necessary until the ->Change LOCK is displayed.<br />

• Press the START key.The lock status is displayed.<br />

If no PIN has been entered, LOCK = 〈3〉 confirm by START, for example, will be displayed.<br />

If a PIN has been entered, LOCK = 3, for example, will be displayed.<br />

• Use the adjusting knob to set the desired lock status.<br />

If a PIN has been entered, PIN = ---- confirm by START will be shown. In this case, a valid PIN has to<br />

be entered with the adjusting knob and then the START button has to be pressed before the lock<br />

status can be set.<br />

• Press the START key to save the setting.<br />

Briefly, Store LOCK 2 ... is displayed as confirmation, followed by -> Change LOCK.<br />

• Press the OPEN / STOP key once to leave the "Change LOCK" menu or press the OPEN / STOP key twice to leave<br />

the "Machine Menu".<br />