24.06.2014 Views

maenas (intertidal zone) and Segonzacia mesatlantica - Station ...

maenas (intertidal zone) and Segonzacia mesatlantica - Station ...

maenas (intertidal zone) and Segonzacia mesatlantica - Station ...

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

The Structural Analysis of Large Noncovalent Oxygen Binding Proteins Current Protein <strong>and</strong> Peptide Science, 2008, Vol. 9, No. 2 165<br />

below). Crossed immunoelectrophoresis suggested that a<br />

particular type of subunit, subunit beta, is necessary for dodecamer<br />

assembly in most species, but that a few groups<br />

have independently evolved the ability to produce dodecamers<br />

from the other subunit types [106].<br />

The quality of the MALLS analysis depends on the efficiency<br />

of the separation process: two badly-resolved molecular<br />

species will yield an average molecular mass with high<br />

polydispersity. In addition, a limitation of the MALLS<br />

method is that complexes with the same number of subunits<br />

but of slightly different types will differ only very slightly in<br />

mass. They cannot be resolved by SEC <strong>and</strong> thus MALLS<br />

will only give an average mass for all complexes of the same<br />

order. ESI-MS can provide further characterization in this<br />

case.<br />

ESI-MS has also been used recently for investigation of<br />

crustacean Hc. Noncovalent ESI-MS can resolve different<br />

molecular species in hemolymph without any prior separation.<br />

However, the sample preparation is a critical step in the<br />

ESI-MS procedure <strong>and</strong> Hc samples are desalted thoroughly<br />

in 10 mM ammonium acetate through at least 7 cycles of<br />

concentration/dilution on centrifugal filters before analysis;<br />

an incomplete desalting causes low signal-to-noise ratio.<br />

Under the adjusted conditions of noncovalent ESI-MS<br />

[62, 73] (see before), the different species are clearly visible<br />

as several Gaussian peak distributions along the m/z axis as<br />

presented in Fig. (3a). These distributions are well separated<br />

between different aggregation states (hexamers versus dodecamers);<br />

for a given aggregation state several close distributions<br />

are sometimes detected, corresponding to a slight mass<br />

difference. In Fig. (3a), two distributions are observed for<br />

hexamer (a,b) <strong>and</strong> dodecamer (c,d). As previously noted,<br />

such discrimination is not possible using the average mass<br />

determined by MALLS because of the very close mass <strong>and</strong><br />

coelution of these complexes. Thus ESI-MS is a complementary<br />

method enabling thorough investigation of the complex<br />

native masses.<br />

In addition to being a technique allowing high mass accuracy<br />

measurements, ESI-MS can also detect species present<br />

in low amount or hardly or not resolved by MALLS due to<br />

coelution of different complexes. In Fig. (3a) a higher aggregate<br />

is observed with a mass of 1 374 136 Da (18-meric<br />

form). Higher masses corresponding to 24-mers <strong>and</strong> even<br />

30-mers can be also be found [62]. It is known that aggregation<br />

can occur in the ESI when sample concentration is too<br />

high but it is unlikely that the assemblies observed here<br />

could be artifacts at least for the 18-mer as it can be observed<br />

by MALLS under low flow rate (0.25 ml/min, personal observation).<br />

However, the very low detected amount of these<br />

“unusual” complexes raises the question of their biological<br />

significance ; since no study by electron microscopy was<br />

performed on them, it is still possible that they arise from<br />

r<strong>and</strong>om post-translational cross-linking.<br />

For such high masses, charge state assignation may be<br />

difficult due to peak enlargement <strong>and</strong> the consecutive proximity<br />

of st<strong>and</strong>ard deviation values calculated for subsequent<br />

charge states [62, 74]. Even the use of deconvolution programs<br />

like MaxEnt cannot resolve the problem. The precision<br />

of the determined masses decreases with increasing<br />

masses but is nonetheless sufficient to unambiguously determine<br />

the number of subunit for each distribution as the<br />

uncertainty is inferior to the subunit average mass.<br />

Fig. (2). SEC-MALLS analysis obtained for a Rimicaris exoculata hemolymph sample.<br />

Analysis was performed with a Superose 6-C column. Optical density at 280nm (full curve) <strong>and</strong> estimated molecular weight (dots) are represented<br />

as a function of the eluted volume. Insert: cumulative molecular weight curve (reprinted from [105], with permission).<br />

86

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!