maenas (intertidal zone) and Segonzacia mesatlantica - Station ...
maenas (intertidal zone) and Segonzacia mesatlantica - Station ...
maenas (intertidal zone) and Segonzacia mesatlantica - Station ...
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
The Structural Analysis of Large Noncovalent Oxygen Binding Proteins Current Protein <strong>and</strong> Peptide Science, 2008, Vol. 9, No. 2 165<br />
below). Crossed immunoelectrophoresis suggested that a<br />
particular type of subunit, subunit beta, is necessary for dodecamer<br />
assembly in most species, but that a few groups<br />
have independently evolved the ability to produce dodecamers<br />
from the other subunit types [106].<br />
The quality of the MALLS analysis depends on the efficiency<br />
of the separation process: two badly-resolved molecular<br />
species will yield an average molecular mass with high<br />
polydispersity. In addition, a limitation of the MALLS<br />
method is that complexes with the same number of subunits<br />
but of slightly different types will differ only very slightly in<br />
mass. They cannot be resolved by SEC <strong>and</strong> thus MALLS<br />
will only give an average mass for all complexes of the same<br />
order. ESI-MS can provide further characterization in this<br />
case.<br />
ESI-MS has also been used recently for investigation of<br />
crustacean Hc. Noncovalent ESI-MS can resolve different<br />
molecular species in hemolymph without any prior separation.<br />
However, the sample preparation is a critical step in the<br />
ESI-MS procedure <strong>and</strong> Hc samples are desalted thoroughly<br />
in 10 mM ammonium acetate through at least 7 cycles of<br />
concentration/dilution on centrifugal filters before analysis;<br />
an incomplete desalting causes low signal-to-noise ratio.<br />
Under the adjusted conditions of noncovalent ESI-MS<br />
[62, 73] (see before), the different species are clearly visible<br />
as several Gaussian peak distributions along the m/z axis as<br />
presented in Fig. (3a). These distributions are well separated<br />
between different aggregation states (hexamers versus dodecamers);<br />
for a given aggregation state several close distributions<br />
are sometimes detected, corresponding to a slight mass<br />
difference. In Fig. (3a), two distributions are observed for<br />
hexamer (a,b) <strong>and</strong> dodecamer (c,d). As previously noted,<br />
such discrimination is not possible using the average mass<br />
determined by MALLS because of the very close mass <strong>and</strong><br />
coelution of these complexes. Thus ESI-MS is a complementary<br />
method enabling thorough investigation of the complex<br />
native masses.<br />
In addition to being a technique allowing high mass accuracy<br />
measurements, ESI-MS can also detect species present<br />
in low amount or hardly or not resolved by MALLS due to<br />
coelution of different complexes. In Fig. (3a) a higher aggregate<br />
is observed with a mass of 1 374 136 Da (18-meric<br />
form). Higher masses corresponding to 24-mers <strong>and</strong> even<br />
30-mers can be also be found [62]. It is known that aggregation<br />
can occur in the ESI when sample concentration is too<br />
high but it is unlikely that the assemblies observed here<br />
could be artifacts at least for the 18-mer as it can be observed<br />
by MALLS under low flow rate (0.25 ml/min, personal observation).<br />
However, the very low detected amount of these<br />
“unusual” complexes raises the question of their biological<br />
significance ; since no study by electron microscopy was<br />
performed on them, it is still possible that they arise from<br />
r<strong>and</strong>om post-translational cross-linking.<br />
For such high masses, charge state assignation may be<br />
difficult due to peak enlargement <strong>and</strong> the consecutive proximity<br />
of st<strong>and</strong>ard deviation values calculated for subsequent<br />
charge states [62, 74]. Even the use of deconvolution programs<br />
like MaxEnt cannot resolve the problem. The precision<br />
of the determined masses decreases with increasing<br />
masses but is nonetheless sufficient to unambiguously determine<br />
the number of subunit for each distribution as the<br />
uncertainty is inferior to the subunit average mass.<br />
Fig. (2). SEC-MALLS analysis obtained for a Rimicaris exoculata hemolymph sample.<br />
Analysis was performed with a Superose 6-C column. Optical density at 280nm (full curve) <strong>and</strong> estimated molecular weight (dots) are represented<br />
as a function of the eluted volume. Insert: cumulative molecular weight curve (reprinted from [105], with permission).<br />
86