12.04.2014 Views

ROTO SILENTA 630 RS - Hettich Instruments

ROTO SILENTA 630 RS - Hettich Instruments

ROTO SILENTA 630 RS - Hettich Instruments

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

EN<br />

35 Faults<br />

If the fault cannot be eliminated with the help of the fault table, please inform Customer Service.<br />

Please specify the type of centrifuge and the serial number. Both numbers can be found on the name plate of the<br />

centrifuge.<br />

Perform a MAINS RESET:<br />

− Switch off the mains switch (switch position "0").<br />

− Wait at least 10 seconds and then switch on the mains switch again (switch position "Ι").<br />

Message / fault Cause Remedy<br />

No display --- No voltage.<br />

− Check supply voltage.<br />

Overvoltage protection tripped out (only for − Mains switch ON.<br />

types 5005-08).<br />

TACHO - ERROR 01 Faulty speedometer.<br />

− Open the cover.<br />

02 No rotor installed.<br />

Defective motor, frequency converter or<br />

− Switch off the mains switch<br />

(switch position "0").<br />

drive.<br />

− Wait at least 10 seconds.<br />

− Turn the rotor vigorously by hand.<br />

− Switch on the mains switch again<br />

(switch position "Ι"). The rotor<br />

must turn during switch-on.<br />

IMBALANCE /<br />

--- The rotor is unevenly loaded. − Open lid.<br />

UNWUCHT<br />

− Check the loading of the rotor,<br />

see chapter "Loading the rotor".<br />

− Repeat the centrifugation run.<br />

CONTROL - ERROR 04, Error in lid locking or lid closure.<br />

06 - 09<br />

N > MAX 05 Rotation too fast<br />

− Perform a MAINS RESET.<br />

N < MIN 13 Rotation too slow<br />

<strong>ROTO</strong>RCODE 10 Incorrect rotor coding<br />

MAINS INTERRUPT --- Power failure during the centrifugation run. − Open lid.<br />

(The centrifugation run was not finished.) − Push START button.<br />

− Repeat the centrifugation run if<br />

necessary.<br />

VE<strong>RS</strong>IONS-ERROR 12 Mismatch between electronic components<br />

SER I/O - ERROR 30 - 38 Error / defective interface<br />

° C * - ERROR 50 – 56, Error / defective cooling<br />

58<br />

LOCK - ERROR 57 Error / defective program locking<br />

− Perform a MAINS RESET.<br />

FU / CCI - ERROR 60 - 83 Error / defective motor control<br />

CONTROL-ERROR 26,<br />

90 – 95,<br />

97 - 99<br />

Error / defective control unit<br />

N > <strong>ROTO</strong>R-MAX 96 Speed in the selected program greater than<br />

the maximum speed of the rotor (n-max-<br />

Rotor).<br />

− Check the set speed.<br />

Reduce the set speed<br />

46/102

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!