2 Operating safety<br />

2.1 Safety regulations<br />

The following instructions should be observed to ensure<br />

the safety of patients:<br />

• The temperature of the upper surface of the lamp housing<br />

might increase during operation.<br />

• Only a hospital grade 3-conductor electrical power supply<br />

cable must be used.<br />

• For <strong>USA</strong> and Canada<br />

Only a hospital grade 3-conductor electrical power supply<br />

cable must be used.<br />

Detachable Power Supply Cord Set, UL Listed, type SJE, SJT<br />

or SJO, 3-conductor, not smaller than 18 AWG.<br />

2.2 Transportation, storage and installation<br />

• As long as the equipment remains in its original packaging,<br />

it may be exposed to the following environmental/<br />

ambient conditions for transportation and storage purposes<br />

(opened or in the packaging) for a maximum of 15 weeks:<br />

•a range of ambient temperatures from -10 °C to +60 °C<br />

•a range of air pressures from 500 hPa to 1060 hPa<br />

•a relative humidity range from 10 % to 90 %, whereby<br />

the equipment should be allowed to adjust to the ambient<br />

temperature of the new location for several hours in the<br />

packaging (to avoid the forming of condensation).<br />

• Do not operate the equipment in humid locations. Ensure<br />

that no splashing, dripping or excessive surges of water<br />

occur near the equipment.<br />

• Always mount the slit lamp and the headrest on an<br />

electric insulated table top, like wood, plastic, etc. The table<br />

top must be fire resistant.<br />

2.3 Notes on usage<br />

• Only qualified and trained personnel should be permitted<br />

to operate the equipment.<br />

• The training of the operating personnel is the responsibility<br />

of the owner of the equipment.<br />

• It is hazardous to operate the equipment in locations<br />

where combustible gases or vapors could be present. The use<br />

of the equipment in the presence of volatile solvents such as<br />

alcohol, benzol or similar chemical products, as well as<br />

flammable anaesthetic agents is also hazardous.<br />

• The user should only open the slit lamp <strong>BQ</strong> 900 ® to<br />

change the light bulb. Any other intervention on the equipment<br />

can be hazardous to the safety of the examiner and the patient.<br />

• Before changing the light bulb: switch off the slit lamp<br />

<strong>BQ</strong> 900 ® , unplug and allow the light bulb to cool down for<br />

several minutes.<br />

• The equipment may only be operated with the listed<br />

accessories in accordance with the instruction manual or with<br />

the separate specific instruction manual for the accessories.<br />

• The rail covers (33) must be mounted to prevent from a<br />

tip over of the slit lamp.<br />

• In order to avoid possible eye damage, the sun should<br />

never be viewed through the ocular.<br />

1500.7001000.<strong>04050</strong><br />