Appendix A Installation of the equipment<br />
A.1 Mounting the microscope and the<br />
illumination unit<br />
Place the stereo microscope on the saddle bearing of the<br />
microscope arm. Move it forward against the stop piece and<br />
secure with the screw (13) appearing from obliquely below<br />
(use the Allen key 5 mm).<br />
Secure the breath shield (26) by means of the knurled screw<br />
(27) on the inner side of the bearer arm.<br />
Before attaching the illumination unit to the column of the<br />
instrument base, the set screw (53) in the hub of the bearer<br />
arm is to be sufficiently loosened so that its tip no longer<br />
protrudes into the screw hole.<br />
Grasp the illumination unit arm (52) as shown and place in<br />
position.<br />
Tighten the set screw (53) in the hub of the bearer arm with the<br />
small screwdriver.<br />
Small screwdriver<br />
HS-Part number 1001436<br />
Allen key 5 mm<br />
HS-Part number 1001602<br />
A.2 Mounting of headrest and power supply<br />
The corresponding HAAG-STREIT operating instructions<br />
should be observed!<br />