Appendix A Installation of the equipment<br />

A.1 Mounting the microscope and the<br />

illumination unit<br />

Place the stereo microscope on the saddle bearing of the<br />

microscope arm. Move it forward against the stop piece and<br />

secure with the screw (13) appearing from obliquely below<br />

(use the Allen key 5 mm).<br />

Secure the breath shield (26) by means of the knurled screw<br />

(27) on the inner side of the bearer arm.<br />

Before attaching the illumination unit to the column of the<br />

instrument base, the set screw (53) in the hub of the bearer<br />

arm is to be sufficiently loosened so that its tip no longer<br />

protrudes into the screw hole.<br />

Grasp the illumination unit arm (52) as shown and place in<br />

position.<br />

Tighten the set screw (53) in the hub of the bearer arm with the<br />

small screwdriver.<br />

Small screwdriver<br />

HS-Part number 1001436<br />

Allen key 5 mm<br />

HS-Part number 1001602<br />

A.2 Mounting of headrest and power supply<br />

The corresponding HAAG-STREIT operating instructions<br />

should be observed!<br />

1500.7001000.<strong>04050</strong><br />

13<br />

27<br />

26<br />

52<br />

53<br />

© HAAG-STREIT AG, CH-3098 Koeniz, Switzerland<br />

page 23 / 33