4 Operating the equipment<br />

• The height of the table must be selected to enable the patient<br />

to obtain a firm rest for the forehead and chin.<br />

• In order to ensure that only that part of the eye to be examined<br />

is illuminated and to avoid disturbing glare, the height of the<br />

light beam should be appropriately adjusted.<br />

Notice<br />

• Only use as much light intensity as required.<br />

• The light intensity should be reduced to a minimum on<br />

completion of each examination. This extends the life of the<br />

light bulb and protects the patient from severe glare at the<br />

beginning of an examination.<br />

• Those parts of the equipment coming into contact with<br />

patients should be cleaned with a dry cloth before each<br />

examination.<br />

4.1 Illumination unit<br />

Slit width<br />

The slit width is set with either of the control knobs (18). The<br />

division on the left hand knob enables any given slit width<br />

setting to be subsequently recovered; but provides no information<br />

on the actual width of the slit light.<br />

Slit length<br />

The slit length is set with the knurled control knob (22).<br />

Slit positioning<br />

The slit image can be positioned by turning the upper part of<br />

the illumination unit with the knurled control knob (22) from the<br />

vertical, through a range of any given oblique positions, to the<br />

horizontal.<br />

Examinations in indirect light<br />

The slit image can be moved away from the center of the visual<br />

field by loosening the centering screw (29). Tightening the<br />

screw will recenter the slit image in the visual field of the<br />

microscope.<br />

Binocular examinations (fundus)<br />

In order to ensure the possibility of conducting a binocular<br />

examination in optical cross section without difficulty from<br />

side angles between 3 and 10°:<br />

• the shorter mirror should be installed,<br />

• the illumination unit is rotated through 90° with the knurled<br />

control knob (22) and with the latch (30) in stages of 5°,<br />

• the illumination unit and the microscope are turned to the<br />

center position (0°).<br />

1500.7001000.<strong>04050</strong><br />

22<br />

29<br />

30<br />

18<br />

© HAAG-STREIT AG, CH-3098 Koeniz, Switzerland<br />

ACHTUNG<br />

Bitte das Kapitel "Gerätesicherheit" vor der<br />

Inbetriebnahme des Gerätes unbedingt<br />

durchlesen und beachten.<br />

ATTENTION<br />

Avant la mise en service de l'appareil consultez<br />

le chapitre "Sécurité de l'appareil".<br />

CAUTION<br />

It is imperative to read the chapter "Operating<br />

safety" before operating the equipment<br />

and to observe its precautions.<br />