WiringIn order to prevent the cables from being torn apart when activating the motors, it is critical that youfollow the instructions below:1) Connect the cables of the bed end motors (head and foot end) directly to the control unit.2) Wire the cable of the bed end motor (foot) and the leg motor through the pipe pins of the backrest motor.3) Connect the mains cord to the 230V socket outlet (or the 24V cable and trafo on beds with external powersupply).4) Run the bed end motors to their top position.5) Place the motor cable of the foot end on the hook at the foot end.6) Run the backrest to its top position.7) Place the bed end motor (head) cables on the hooks at the head end.8) Fasten the locking cam to the control unit.GBWhen working with the foot end, note that there is a risk of getting squeezed since the adjustable leg section is notfirmly locked. Remember to remove the plug from the socket outlet when moving the bed.Wiring of the leg section motor and bed end motor (foot)through the backrest motor pipe pin.Wiring of the bed end motor (foot).Wiring at the lifting motor, head end.(On <strong>Medley</strong> with separate power supply) connect thewire from the power supply to the wire from the control unitand secure the connection with the enclosed wire lock11