IDEXX Summary 14D<br />

Topic: Beta Trial Study report comparing <strong>Pseudalert*</strong> versus the ISO 16266 method<br />

for the <strong>de</strong>tection of Pseudomonas aeruginosa in 250 mL thermal pool and<br />

bottled water samples<br />

Rev 03252011<br />

i<strong>de</strong>xx.com<br />

Title: “Comparison of the performance of the IDEXX Pseudalert test against the ISO<br />

16266 method at recovering confirmed Pseudomonas aeruginosa from thermal<br />

pool and bottled water samples”<br />

Author: IDEXX Laboratories<br />

Date: March 2011<br />

Report Highlights:<br />

• Pseudalert was compared to the ISO 16266 method at the Nestlé Waters Quality<br />

Assurance Center in Vittel, France which regu<strong>la</strong>rly tests thermal pool and bottled<br />

water samples.<br />

• Data from the completed study showed:<br />

o Pseudalert had comparable <strong>de</strong>tection of P. aeruginosa versus the ISO 16266<br />

method from 250 mL thermal pool water samples naturally contaminated with<br />

bacteria<br />

o Pseudalert had comparable <strong>de</strong>tection of P. aeruginosa versus the ISO 16266<br />

method from 250 mL bottled water samples spiked with reference and wild<br />

strains.<br />

o Pseudalert was able to suppress the false <strong>de</strong>tection of a P. putida strain that<br />

was spiked into 250 mL bottled water samples at concentrations exceeding<br />

100 cfu/sample.<br />

• Pseudalert performed as well or better than the ISO 16266 method at <strong>de</strong>tecting<br />

P. aeruginosa in 250 mL thermal pool and bottled water samples<br />

*Pseudalert is a tra<strong>de</strong>mark or registered tra<strong>de</strong>mark of IDEXX Laboratories, Inc. or its affiliates in the United States and/or other<br />