22<br />

SWitchpAck #2 (SW2)<br />

switch 1 through 3<br />

These switches are used to determine the type and voltage of<br />

the Indirect Ophthalmoscope bulb. See the Chart below.<br />

B UL B VOL T AG E A ND T Y P E -IND. OP HT HA L MOS C OP E<br />

S witch 1 2 3<br />

6.5V /18W, Incandescent OF F OF F OF F<br />

R S V D Incandescent ON OF F OF F<br />

R S V D Incandescent OF F ON OF F<br />

R S V D Incandescent ON ON OF F<br />

6V /10W Halogen OF F OF F ON<br />

6V /20W Halogen ON OF F ON<br />

12V /12W Halogen OF F ON ON<br />

R S V D Halogen ON ON ON<br />

switch 4<br />

• Turn Switch 4 ON if your system contains a indirect<br />

hanger (Figure 13) for an Indirect Ophthalmoscope and<br />

you want that hanger to automatically supply power to<br />

the instrument.<br />

note: Turn Switch 5 on SW1 OFF if turning Switch<br />

4 ON.<br />

note: Positionnez le contact 5 á la position<br />

arrêt sur l’interrupteur SW1 et positionnez le<br />

contact 4 EN position MARCHE.<br />

• Turn Switch 4 OFF if your system does not contain a hanger<br />

for an Indirect Ophthalmoscope. In this case, power is<br />

turned ON and OFF to the instrument via the B/P Switch<br />

on the front panel (See Switch 5 on SW1).<br />

switch 5<br />

• If you want your system to beep whenever a switch is<br />

pressed , set Switch 5 ON.<br />

• If you do not want to hear a beep whenever a switch is<br />

pressed, set Switch OFF.<br />

switch 6<br />

• Reserved<br />

switches 7 and 8<br />

These switches are used to determine the type and voltage<br />

of the Slit Lamp bulb. See chart to the above right.(DOES<br />

NOT APPLY TO INSTRUMENT STANDS)<br />

B UL B VOL T AG E A ND T Y P E -S L IT L A MP B UL B<br />

S witch 7 8<br />

6V /4.5A, Incandescent OF F OF F<br />

R S V D ON OF F<br />

6V /20W Halogen OF F ON<br />

12V /30W Halogen ON ON<br />

standby switch (sTbY)<br />

• OFF- Unit is deactivated, except for charging wells.<br />

• ON- Activates all unit functions<br />

Figure 14<br />

(F actory default)<br />

auxiliary switch (aux1)<br />

• Affects the room lights and is programmable.<br />

• See section 3.8.3 for programming instructions if you<br />

have a single-zone infrared receiver.<br />

• See Switch 7 of Switchpack #1 in the Programming<br />

Instructions section of this manual for details.<br />

IN-<strong>7900</strong>