22<br />
SWitchpAck #2 (SW2)<br />
switch 1 through 3<br />
These switches are used to determine the type and voltage of<br />
the Indirect Ophthalmoscope bulb. See the Chart below.<br />
B UL B VOL T AG E A ND T Y P E -IND. OP HT HA L MOS C OP E<br />
S witch 1 2 3<br />
6.5V /18W, Incandescent OF F OF F OF F<br />
R S V D Incandescent ON OF F OF F<br />
R S V D Incandescent OF F ON OF F<br />
R S V D Incandescent ON ON OF F<br />
6V /10W Halogen OF F OF F ON<br />
6V /20W Halogen ON OF F ON<br />
12V /12W Halogen OF F ON ON<br />
R S V D Halogen ON ON ON<br />
switch 4<br />
• Turn Switch 4 ON if your system contains a indirect<br />
hanger (Figure 13) for an Indirect Ophthalmoscope and<br />
you want that hanger to automatically supply power to<br />
the instrument.<br />
note: Turn Switch 5 on SW1 OFF if turning Switch<br />
4 ON.<br />
note: Positionnez le contact 5 á la position<br />
arrêt sur l’interrupteur SW1 et positionnez le<br />
contact 4 EN position MARCHE.<br />
• Turn Switch 4 OFF if your system does not contain a hanger<br />
for an Indirect Ophthalmoscope. In this case, power is<br />
turned ON and OFF to the instrument via the B/P Switch<br />
on the front panel (See Switch 5 on SW1).<br />
switch 5<br />
• If you want your system to beep whenever a switch is<br />
pressed , set Switch 5 ON.<br />
• If you do not want to hear a beep whenever a switch is<br />
pressed, set Switch OFF.<br />
switch 6<br />
• Reserved<br />
switches 7 and 8<br />
These switches are used to determine the type and voltage<br />
of the Slit Lamp bulb. See chart to the above right.(DOES<br />
NOT APPLY TO INSTRUMENT STANDS)<br />
B UL B VOL T AG E A ND T Y P E -S L IT L A MP B UL B<br />
S witch 7 8<br />
6V /4.5A, Incandescent OF F OF F<br />
R S V D ON OF F<br />
6V /20W Halogen OF F ON<br />
12V /30W Halogen ON ON<br />
standby switch (sTbY)<br />
• OFF- Unit is deactivated, except for charging wells.<br />
• ON- Activates all unit functions<br />
Figure 14<br />
(F actory default)<br />
auxiliary switch (aux1)<br />
• Affects the room lights and is programmable.<br />
• See section 3.8.3 for programming instructions if you<br />
have a single-zone infrared receiver.<br />
• See Switch 7 of Switchpack #1 in the Programming<br />
Instructions section of this manual for details.<br />
IN-<strong>7900</strong>