Warning-The eCliPse inFrareD rOOm<br />

lighTs COnTrOl sYsTem musT maTCh<br />

The reQuiremenTs OF The user. iF The<br />

sYsTem is nOT PrOPerlY COnFigureD<br />

beFOre seCOnDarY POWer TO The uniT is<br />

TurneD On Via The sTbY sWiTCh, Damage<br />

TO The sYsTem Or insTrumenT bulbs<br />

maY OCCur.<br />

aVerTissanT-le sYsTÈme De COnTrôle<br />

inFrarOuge D’ÉClairage ÉCliPse DOiT<br />

COrresPOnDre aux COnDiTiOns De<br />

l’uTilisaTeur. si le sYsTÈme n’esT<br />

Pas COrreCTemenT COnFigurÉ aVanT<br />

D’allumer le suPPOrT OPhTalmiQue, Par<br />

le COnTrôle “sTbY”, Des DOmmages au<br />

suPPOrT Ou aux amPOules ÉleCTriQues<br />

PeuVenT se PrODuire.<br />

ECLIPSE ROOM LIGHT SYSTEM<br />

Press Here<br />

to turn oFF<br />

indireCt<br />

BulB ratings<br />

slit lamP<br />

BulB ratings<br />

sW2-1<br />

sW2-2<br />

sW2-3<br />

Hanger<br />

BeeP on<br />

oPen<br />

sW2-7<br />

sW2-8<br />

Press Here<br />

to turn on<br />

oFF<br />

sW2<br />

REAR VIEW OF CONSOLE ASSEMBLY<br />

eCliPse inFrareD rOOm lighT COnTrOls<br />

PrOgramming guiDe<br />

Configuring the system is accomplished by setting dip<br />

switches found on the circuit board. The dip switches<br />

are labeled SW1 and SW2. The following paragraphs<br />

provide the necessary information required to configure<br />

the system.<br />

sWiTChPaCk #1 (sW1)<br />

switch 1<br />

• Turn Switch 1 OFF if your system contains a single-zone<br />

infrared receiver.<br />

• Turn Switch 1 ON if your system contains a dual-zone<br />

infrared receiver (Factory default setting).<br />

1 2 3 4 5 6 7 8<br />

oFF<br />

sW1<br />

1 2 3 4 5 6 7 8<br />

enaBle-on<br />

dual Zone<br />

Well #1<br />

Well #2<br />

Well #3<br />

Binding Post<br />

s/l arm<br />

auXiliary<br />

lamP<br />

switch 2<br />

• Turn Switch 2 ON to enable Well #1 to affect room lights<br />

and fixation lights.<br />

• Turn Switch 2 OFF to disable Well #1 from affecting room<br />

lights and fixation lights.<br />

switch 3<br />

• Turn Switch 3 ON to enable Well #2 to affect room lights<br />

and fixation lights.<br />

• Turn Switch 3 OFF to disable Well #2 from affecting room<br />

lights and fixation lights.<br />

switch 4<br />

• Turn Switch 4 ON to enable Well #3 to affect room lights<br />

and fixation lights.<br />

• Turn Switch 4 OFF to disable Well #3 from affecting room<br />

lights and fixation lights.<br />

switch 5<br />

• If power to the Indirect Ophthalmoscope is controlled by<br />

the B/P Switch on the front panel, turn Switch 5 ON if<br />

you want the B/P Switch to affect room lights.<br />

• Turn Switch 5 OFF if you do not want the B/P Switch to<br />

affect room lights.<br />

switch 6<br />

• Turn Switch 6 ON to enable Slit Lamp Arm Switch to affect<br />

room lights.<br />

• Turn Switch 6 OFF to disable Slit Lamp Arm Switch from<br />

affecting room lights.<br />

switch 7<br />

• Turn Switch 7 ON to enable the AUX1 Switch to affect<br />

room lights.<br />

• Turn Switch 7 OFF to disable the AUX1 Switch from<br />

affecting room lights.<br />

switch 8<br />

• Turn Switch 8 ON to enable the LAMP Switch to affect<br />

room lights.<br />

• Turn Switch 8 OFF to disable the LAMP Switch from<br />

affecting room lights.<br />

IN-<strong>7900</strong> 21