3.6.2<br />
• The round button under the “B/P” arrows is used to turn<br />
on and off the indirect hanger. The up and down arrows<br />
are used to control the output voltage (i.e. the intensity)<br />
of the indirect bulb.<br />
3.6.3<br />
• The round button under the “LAMP” arrows is used to turn<br />
on and off the overhead lamp. The up and down arrows<br />
are used to control the output voltage (i.e. the intensity)<br />
of the overhead lamp.<br />
3.7. Fuses<br />
�������<br />
Warning- DisCOnneCT FlOOr uniT FrOm<br />
main inPuT POWer beFOre PrOCeeDing<br />
WiTh eleCTriCal insPeCTiOns Or<br />
mainTenanCe.<br />
aVerTissemenT-DÉbranCheZ le<br />
suPPOrT OPhTalmiQue Du rÉseau<br />
ÉleCTriQue POur PrOCÉDer À une<br />
insPeCTiOns Ou mainTenanCe.<br />
CauTiOn- replace fuse(s) as marked. all<br />
fuses must be replaced with a fuse of the<br />
same size and rating. refer to the wiring<br />
diagrams on pages 56-59.<br />
l’aTTenTiOn- remplacement des fusible<br />
(s) tel qu’indiqué. Tous les fusibles doivent<br />
être remplacés avec un fusible de même<br />
taille et de même ampérage et/ou voltage.<br />
référez-vous aux diagrammes de câblage<br />
aux page 56-59.<br />
3.7.1. Floor units contain two fuses located inside the AC<br />
Input Module on the Outlet Plate Assembly.<br />
3.7.2. Consoles and Base Caps contain no additional<br />
fuses.<br />
3.7.3. Fuse replacement (aC input module)<br />
method for removing fuse drawer:<br />
1. The fuse drawer is located above the power cord.<br />
Remove the power cord.<br />
Place a small bladed screwdriver into appropriate slot<br />
(shown above) and push the flat spring towards the<br />
center of the power module until a click is heard. (Click<br />
may not be heard in all cases.) The drawer will come<br />
forward slightly.<br />
Once the flap has been pushed towards the center, the<br />
drawer will extend slightly beyond the power module.<br />
Remove the drawer by pulling it out by your fingers.<br />
CauTiOn- never install the drawer into the<br />
aC input module without fuses.<br />
aTTenTiOn- ne jamais introduire le<br />
module de support des fusibles sans avoir<br />
inséré les fusibles.<br />
IN-<strong>7900</strong> 17<br />
2.<br />
3.<br />
4.<br />
5.