12<br />

2.10. Final assembly<br />

2.10.1. Console<br />

If an additional instrument is to be connected to the binding<br />

posts or if any auxiliary equipment is to be plugged into<br />

the”AUX” power connector, make these connections at<br />

this time.<br />

2.11. Chair assembly<br />

2.11.1 The chair control switches located on the floor unit<br />

may control any Reliance® low voltage chair. Refer to the<br />

chair instructions for unpacking and assembling the chair.<br />

2.11.2. A chair control cable is provided to control the<br />

vertical movement of the chair base from the floor unit.<br />

One end of the cable is to be inserted into the black circular<br />

receptacle on the Outlet Plate Assembly. See Figure 5. The<br />

other end of the cable is to be inserted into the mating<br />

receptacle marked “CHAIR CONTROL CABLE” in the rear of<br />

your Reliance® chair.<br />

2.12. miscellaneous<br />

Figure 5<br />

2.12.1. When attaching an instrument to the slit lamp arm,<br />

note the thrust bearing which should be in place between<br />

the instrument and the arm.<br />

2.12.2. Proper balance of an instrument on the slit lamp<br />

arm is achieved by adding or removing counter-balance<br />

weights as shown in Figure 6. See Page 12 for a table of<br />

weight settings.<br />

2.12.3 Finally, plug the power cord into a wall receptacle<br />

and check all electrical functions as described in the<br />

Installation and Operating Instructions. If all functions are<br />

normal, reattach the base unit rear cover.<br />

2.12.4 To disconnect power to <strong>Instrument</strong> <strong>Stand</strong> unplug<br />

power cord from wall receptacle.<br />

nOTe: Outlet wires are hot.<br />

nOTe : Les fils de sortie sont chauds<br />

CauTiOn-“accessory equipment connected<br />

to the analog and interfaces must be certified<br />

according to the respective ieC standards<br />

(i.e. ieC 950 for data processing equipment<br />

and ieC 601-1 for medical equipment).<br />

Furthermore all configurations shall comply<br />

with the system standard ieC 601-1-1.<br />

everybody who connects additional equipment<br />

to the signal input part or output part<br />

configures a medical system, and is therefore<br />

responsible that the system complies with<br />

the requirements of ieC 601-1-1. if in doubt,<br />

consult the technical services department or<br />

your local representative.”<br />

equipment with siP/sOP connectors should<br />

either indicate the connected equipment<br />

comply with ieC 601-1 and/or ieC 601-<br />

1-1 harmonized national standard or the<br />

combination should be evaluated.<br />

aTTenTiOn- “les équipements branché sur<br />

le support ophtalmique doivent répondre aux<br />

normes ieC (i.e. ieC 950 pour le matériel<br />

informatique et le ieC 601-1 pour le matériel<br />

médical). en outre toutes les configurations<br />

devront étre conformes á la norme ieC<br />

601-1-1.<br />

Tout personne qui branché un équipement<br />

supplémentaire au systéme doit s’assurer<br />

que cet équipement répond aux normes<br />

médicales ieC 601-1-1 et en est responsable.<br />

en cas de doute, consultez le service<br />

technique de votre distributeur ou votre<br />

préposé technique local.”<br />

le matériel avec des connecteurs de type<br />

siP/sOP devront être conformes à la norme<br />

ieC 601-1 et/ou être conformes à la norme<br />

nationale harmonisée ieC 601-1-1.<br />

IN-<strong>7900</strong>