The field properties listed in section 2.4.A5.1 have to be measured for all superconducting magnets.<br />

Besides this the dynamic behaviour of the magnetic field of all fast-ramped superconducting<br />

magnets during ramping must be measured. These so-called dynamic measurements will be carried<br />

out ramp by ramp, i.e. the coil will be placed at a certain angular position and the magnet will be<br />

ramped up and down and the induced signal will be measured. Then the coil will step to the next<br />

angular position and the measurement will be repeated, and so on. After the loop is finished, all<br />

multipoles can be computed from these data. In this case the power converter's reproducibility must<br />

be good enough to guarantee the necessary accuracy for this measurement procedure [114].<br />

Therefore the static reference coil of the mole (see Figure 2.4.158) is used to check the power<br />

converter's ramp-to-ramp reproducibility. As an alternative method a probe with many coils (in the<br />

order of 16) mounted on a cylindrical surface can be fabricated [115,116]. The different required<br />

systems are given in Table 2.4.46 and Table 2.4.47.<br />

Table 2.4.46: Magnetic measurement systems for the different machines for measuring the main dipoles.<br />

machine system main dimension comments<br />

<strong>Super</strong> FRS search coil curvature: 12.5 m separator magnets<br />

search coil curvature: 4.35 m Energy Buncher magnets<br />

SIS 100 mole coil radius 22.5 mm<br />

search coil curvature 52.5 m for cross checks of the<br />

integral length and angle<br />

single stretched<br />

only applicable if rectan-<br />

wire<br />

gular anticryostat will be<br />

available<br />

SIS 300 mole coil radius 30 mm<br />

search coil curvature 52.5 m for cross checks of the<br />

integral length<br />

Table 2.4.47: Magnetic measurement systems for the different machines for measuring the quadrupoles<br />

and higher order multipoles.<br />

machine system coil radius [mm] comments<br />

<strong>Super</strong> FRS adapted rotating<br />

coil<br />

175<br />

SIS 100<br />

single stretched<br />

wire<br />

mole<br />

single stretched<br />

wire<br />

coil radius 22.5<br />

SIS 300 mole coil radius 30<br />

single stretched<br />

wire<br />

for cross checks of angle,<br />

axis, integral gradient<br />

for cross checks of angle,<br />

axis, integral gradient<br />

for cross checks of angle,<br />

axis, integral gradient<br />

203