The value, parameters, conditions and compatibility of the FAT tests will be specified during<br />
Preliminary <strong>Design</strong> Review unless they are previously specified in the <strong>Technical</strong> Specification.<br />
As an example of a FAT with reference to the components is shown below:<br />
1. Cable and Coil Tests:<br />
• The most important dimensions of the coil are measured during the manufacturing<br />
process.<br />
• The CuNi tube for the cable must be certified for its tightness by the manufacturer.<br />
Otherwise a helium leakage test will be performed prior to the coil winding. After coil<br />
winding the coil will be tested for helium tightness.<br />
• Insulations tests (after coil winding, assembling with the yoke, assembling with the<br />
cryostat).<br />
• Continuity test after each steps involve instrumentation<br />
• The mass-flow through the cooling channel is measured with fixed input pressure > 1<br />
bar and output pressure ~ 1 bar.<br />
2. Yoke Tests:<br />
• The most important dimensions of the yoke are measured during the manufacturing<br />
process.<br />
• Measurement of magnetic permeability of the yoke material<br />
3. Cryostat Tests:<br />
• The most important dimensions of the cryostat are measured during the manufacturing<br />
process. After assembly of the cryostat the leakage is measured to assure that the required<br />
insulation vacuum can be achieved.<br />
• Geometry of the magnet inside the cryostat.<br />
4. Instrumentation and Sensor Tests:<br />
• The internal resistance and insulation versus ground is measured.<br />
• Voltage taps: a measurement of the contact to the coil and to the sensors is performed.<br />
5. Vacuum chamber Tests:<br />
The vacuum chamber test is described in related chapter.<br />
6. Magnet Tests:<br />
• Continuity Test of all type of taps (voltage, current, temperature sensors), resistance<br />
and inductance tests.<br />
• Insulation Test – min. twice the operation voltage plus 500V during 2 min. between coil<br />
and ground .<br />
• Shorts between wires in the coil – e.g. high voltage discharge test and comparison with<br />
the results of the manufacturing stage.<br />
• Vacuum Test of the cryostat – the results has to show a leak rate better than 10 -8 mbar/s.<br />
• Vacuum Test of the helium process lines – the results has to show a leak rate better than<br />
10 -8 mbar/s.<br />