Topic: Arthroplasty - Hip<br />

Abstract number: 24504<br />

THE TITANIUM DUAL MOBILITY CUP, WITH MORE THAN TEN YEARS FOLLOW<br />

UP<br />

Remi PHILIPPOT 1 , Jean CAMILLERI 2 , Frederic FARIZON 1<br />

1 CHU Saint Etienne, Saint Etienne (FRANCE), 2 Hopital Nord, Saint Etienne<br />

(FRANCE)<br />

The concept of stainless steel dual mobility cups in total hip arthroplasty has<br />

demonstrated very low long-term instability rates and a 98% survival rate after 12<br />

years. We systematically implanted titanium alloy acetabular cups during a one year<br />

period. The purpose of our retrospective study was to report the 18-year clinical<br />

outcome data in a homogeneous and continuous series of 103 primary total hip<br />

replacements after implantation of a cementless titanium cup. All patients were<br />

implanted with NOVAE Ti (SERF®) cups made of titanium alloy combined with a<br />

retentive polyethylene liner and a 22.2 mm cobalt chrome prosthetic head. Mean<br />

patient age at the time of surgery was 53 years. All patients were clinically and<br />

radiographically evaluated. The overall 18-year actuarial cup survival rate with a 95%<br />

confidence interval was 87.4%. At last follow-up, there was no evidence of implant<br />

instability whereas acetabular aseptic loosening was reported in one case and high<br />

wear of the retentive liner in 9. The results of this investigation confirmed the longterm<br />

stability of dual-mobility implants. The main limitation of this system was early<br />

wear of the polyethylene liner in contact with the titanium metal back and reaction<br />

with third body along with loss of liner retentivity. In our study, titanium demonstrated<br />

favourable osteointegration properties but poor tribologic characteristics, therefore<br />

suggesting its interest at the bone-cup interface only.<br />