Poster<br />

Topic: Biomaterials<br />

Abstract number: 25392<br />

WEAR BEHAVIOR OF AMC CERAMICS ON PE IN HIP PROSTHESES<br />

Bernard MASSON<br />

CeramTec, Vieille Toulouse (FRANCE)<br />

The introduction of ceramics in total hip arthroplasty contributed significantly to the<br />

wear reduction of polyethylene. This great benefit has been demonstrated by Dahl,<br />

showing that the wear of a 28mm alumina is considerably reduced compare to a<br />

CrCo ball head against Ultra High Molecular Weight Polyethylene (UHMWPE) after<br />

10 years. A first prospective, randomized study with a 15 years follow up compare<br />

the wear of polyethylene between alumina and metal ball head. A reduction of 44%<br />

penetration (linear wear) with the alumina-polyethylene bearing surface is<br />

demonstrated. In order to improve mechanical resistance, a new generation of<br />

alumina matrix composite (BIOLOX®delta) has been used in Orthopaedics since<br />

2001.The topic of this study is to demonstrate the excellent wear performance of the<br />

alumina ceramic composite against polyethylene, compared to alumina/PE in vivo.<br />

The BIOLOX®delta-PE bearing has been tested on a six station hip simulator. After 5<br />

million cycles, the wear rates calculated by linear interpolation were 13,52 mg per<br />

million cycles. When comparing the results for BIOLOX®forte on polyethylene with<br />

the same 28mm diameter and same testing parameter, we observed 26,57 mg/Mc.<br />

The BIOLOX®delta on UHMWPE bearing shows improved wear behavior with a<br />

much lower wear rate. Based on this results and the clinical performance of the<br />

alumina-UHMPE bearing from the literature, we can expect a further reduction of<br />

wear for the BIOLOX®delta on UHMWPE in vivo that will increase the survival rate of<br />

the total hip arthroplasty.<br />