<strong>Task</strong> <strong>Targeted</strong> <strong>Automation</strong>, TTA<br />
RSD Pro and RSA Pro<br />
<strong>The</strong> superior sorting and/or aliquoting solution<br />
for your laboratory<br />
RSD Pro RSA Pro
Enhance Your Lab<br />
<strong>Roche</strong>’s state-of-the-art automation allows<br />
small to large laboratories to realize customized<br />
sample handling solutions for improved<br />
workflow.<br />
For medium to very large laboratories, <strong>Task</strong><br />
<strong>Targeted</strong> <strong>Automation</strong> from <strong>Roche</strong> supports a<br />
high level of automation specific to the needs of<br />
your the laboratory, for example<br />
• Multiple tube types<br />
• Wide variety of sample carriers<br />
• Advanced camera system for tube identification<br />
• Selective decapping and recapping of tubes<br />
• Archiving of tubes.<br />
Fully scalable and extendable, <strong>Task</strong> <strong>Targeted</strong><br />
<strong>Automation</strong> can improve the efficiency of<br />
pre-analytical processing steps, while helping to<br />
reduce the operator’s exposure to potentially<br />
hazardous biological material and helps to eliminate<br />
repetitive stress injuries.<br />
*On-line centrifuge module is not yet available<br />
<strong>The</strong> Basic Systems<br />
Overview<br />
<strong>The</strong> RSD Pro and RSA Pro basic system delivers the following<br />
functions:<br />
• Administration of test requests downloaded from the<br />
laboratory IT system<br />
• Detection of tube type of each primary tube<br />
• Bi-directional host interface for data exchange<br />
• Distribution/sorting of primary tubes<br />
• Distribution/sorting of aliquot tubes (RSA Pro only)<br />
• Creation of one or more aliquot tubes (RSA Pro only)<br />
• Bar code labeling of the aliquot tubes (RSA Pro only)<br />
<strong>The</strong> RSD Pro and RSA Pro can be extended with the following<br />
options:<br />
• Recap unit (selective for primary/secondary)<br />
• Camera system<br />
• Online connected centrifuge*<br />
RSD Pro and RSA Pro Basic System<br />
Option: Recap Unit<br />
• for primary tubes<br />
or<br />
• for secondary tubes<br />
Option: Camera Extension<br />
• for tube type identification<br />
or<br />
• for tube type identification and sample<br />
volume recognition<br />
Option: On-line Centrifuge<br />
• Single or double centrifuge*
System Overview<br />
(RSA Pro only)<br />
Total Workflow Solutions from <strong>Roche</strong><br />
<strong>Roche</strong> offers powerful platforms and different levels of<br />
automation. Together with <strong>Roche</strong> Middleware complete<br />
workflow efficiency can be realized from order entry to<br />
sample archive.<br />
<strong>The</strong> primary benefits include:<br />
• Powerful sample flow management functions using a<br />
rule-based decision process from sample arrival to sample<br />
archive, sample tracking, and rerun/repeat/reflex test<br />
handling<br />
• <strong>The</strong> reduction of repetitive, error-prone and bio-hazardous<br />
pre- and post-analytical work steps<br />
• Extendable, flexible and scalable platforms. <strong>The</strong> <strong>Roche</strong><br />
process solutions are highly configurable and can be adapted<br />
to most customer requirements – including third party tubes<br />
and racks.<br />
Our Laboratory Process Analysis can determine the best<br />
solution for your lab. Specific needs may vary from the example<br />
listed.<br />
• Multiple workflow cycles to adapt to customer workloads
Functions and Features<br />
Input Area<br />
Tubes can be put into the input area in any order. Presorting<br />
according to tube diameter, tube height, cap color, open/closed<br />
is not required.<br />
<strong>The</strong> primary features of the input area are:<br />
• Large sample capacity for up to 600 tubes<br />
• Parallel processing of most of the established hematology,<br />
coagulation, serum, plasma and urine tubes<br />
• Continuous loading of sample tubes during routine operation<br />
• STAT handling capabilities<br />
• 4 configurable input drawers<br />
• <strong>The</strong> input drawers are freely definable with in total up to 30<br />
targets<br />
• <strong>The</strong> following <strong>Roche</strong> sample carriers are supported:<br />
– cobas ® 6000/cobas® 8000 tray with Hitachi 5 position rack<br />
– Elecsys 2010/cobas® 4000/COBAS Integra 800 tray with<br />
Hitachi 5 position rack<br />
– MODULAR ANALYTICS tray with Hitachi 5 position rack<br />
– COBAS Integra 400+ tray with 15 position rack<br />
In addition support is provided for:<br />
– Several neutral racks<br />
– Several Sysmex trays with 10 pos rack<br />
– Different centrifuge buckets<br />
– Other rack types (Generally RSD Pro and RSA Pro are able<br />
to handle non-<strong>Roche</strong> racks. Please contact your local <strong>Roche</strong><br />
representative for details).<br />
Decap Unit<br />
<strong>The</strong> decap unit removes the tube cap and turns the tube with<br />
the bar code label into a “good to read” position. <strong>The</strong> decap unit<br />
also pauses for a few milliseconds to prevent the buildup of<br />
aerosols.<br />
<strong>The</strong> following listed features are just a few examples of the RSD<br />
Pro and RSA Pro decap unit:<br />
• Different tube types (various closure types and tube dimensions)<br />
can automatically be decapped in one run<br />
• Tubes need not be pre-sorted<br />
• Support of selective decapping according to user defined rules<br />
Recap Unit<br />
<strong>The</strong> RSD Pro and RSA Pro can be equipped with a recapper.<br />
Selective, target specific recapping of plastic tubes based on user<br />
defined rules can be configured. For accurate tube sealing, the<br />
tube type information detected by the standard camera unit is<br />
sent to the recapper. Either the primary or aliquot tubes can be<br />
closed.<br />
<strong>The</strong> following features are also provided by the recap unit:<br />
• <strong>The</strong> sealing of the tubes with an aluminum foil<br />
• <strong>The</strong> support of various tube diameters (between 11.5 mm<br />
and 15.5 mm) and various tube heights (between 65.5 mm<br />
and 102.5 mm).
Aliquoter (RSA Pro only)<br />
<strong>The</strong> basic function of the aliquoter unit is the creation of secondary<br />
tubes based on user configurable rules.<br />
<strong>The</strong> primary features of the aliquoter include:<br />
• Secondary tubes are placed automatically into secondary<br />
tube storage<br />
• Bar code printer with label applicator offers the following:<br />
– A bar code printer with thermal transfer printing<br />
– Bar codes format is configurable to include type, length,<br />
and check digit.<br />
– Additional information such as other sample or patient<br />
information can be printed on the label in human readable<br />
characters<br />
– <strong>The</strong> printed bar code label is also placed on the secondary<br />
tube<br />
• Pipetting unit uses disposable tips and offers the following:<br />
– Customized based on each tube type and order request.<br />
<strong>The</strong> pipetting process is performed sample tube specific<br />
– Liquid level detection<br />
– <strong>The</strong> pipetting process is carry over free due to disposable<br />
tips<br />
Note: <strong>The</strong> disposable tip is not exchanged in the case of<br />
multiple sample uptake<br />
– Up to 28 secondary tubes can be pipetted from one<br />
primary tube<br />
– Up to 384 disposable tips on board in 4 trays<br />
– <strong>The</strong> pipetted volume is calculated based on the single test<br />
requests or based on fixed volume assigned to the sorting<br />
target<br />
– Clot detection is provided during the sample aspiration. If<br />
the tube is affected, an error handling process can be<br />
defined.<br />
*Sysmex products are not sold by <strong>Roche</strong>.<br />
Output (Sorting) Area<br />
All tubes (primary and secondary) are sorted into the output<br />
area. <strong>The</strong> sorting is done according to configurable sorting<br />
criteria.<br />
<strong>The</strong> primary features of the output area are:<br />
• Max sample capacity: up to 1200 tubes<br />
• Up to 8 configurable output drawers<br />
• In total up to 41 sorting targets can be defined<br />
• Parallel sample sorting, aliquoting and archiving can be<br />
performed (recursive workflow)<br />
• <strong>The</strong> system can be configured to turn the primary tube<br />
between 0 and 360 degrees before placing it into the target rack<br />
• Drawers (filled or partly filled) with racks can be changed<br />
while the system is running<br />
• <strong>The</strong> following <strong>Roche</strong> sample carriers are supported:<br />
– cobas® 6000/cobas® 8000 tray with Hitachi 5 position rack<br />
– Elecsys 2010/COBAS Integra 800/cobas 4000 tray with<br />
Hitachi 5 position rack<br />
– MODULAR ANALYTICS tray with Hitachi 5 position rack<br />
– COBAS Integra 400+ tray with 15 position rack<br />
Addition support is available for:<br />
– Several neutral racks<br />
– Several Sysmex* trays with 10 position rack<br />
– Different centrifuge buckets<br />
– Other rack types (Generally RSD Pro and RSA Pro are able<br />
to handle non <strong>Roche</strong> racks as well. Please contact your local<br />
<strong>Roche</strong> representative for details).
Standard Camera<br />
<strong>The</strong> RSD Pro and RSA Pro basic system is equipped with a CCD<br />
camera unit.<br />
<strong>The</strong> camera provides the following basic functions and features:<br />
• <strong>The</strong> cap color and tube dimensions are measured<br />
• Detection if the tube is closed or opened<br />
• <strong>The</strong> cap color is used to distinguish between tubes with the<br />
same tube dimensions<br />
• Processing of the different types of tubes (hematology,<br />
coagulation, serum, plasma and urine tubes) according to<br />
rules based on the received test requests<br />
• Up to 20 different tube types (various closure types, tube<br />
dimensions, cap colors) can be processed in one run<br />
• Positive verification between bar code with material<br />
identification and applied tube type. In the event of<br />
mismatch appropriate error handling procedures can be<br />
configured<br />
• Support of bar code alignment<br />
• Forwarding of the tube type information to the decap,<br />
aliquot and optional recap unit.<br />
Camera Options<br />
<strong>The</strong> following extensions are available:<br />
Tube Type Identification (TTI)<br />
<strong>The</strong> TTI option supports all functions of the basic camera system<br />
(e.g. the registration of cap color and tube dimensions, detection<br />
of whether the tube is open or closed, etc.). In addition, the<br />
recognized tube type (determined by the camera) is uploaded<br />
together with the sample query to the connected IT system.<br />
This information is a prerequisite to enable the IT system to<br />
download test requests (e.g. depending on the sample material).<br />
With tube type identification the cap color and tube dimensions<br />
(tube diameter and tube height) are detected. Up to 20 different<br />
tubes can be handled within one run. <strong>The</strong> tube type identification<br />
compares the measured tube values with the reference values<br />
of the currently known types. <strong>The</strong> tube type identification can be<br />
uploaded to the connected IT system.<br />
Tube Type Identification (TTI)<br />
Cap/no Cap<br />
Tube Dimension<br />
Upload Tube Type<br />
Identification<br />
Tube Type Identification<br />
QS1-Sample Quantity<br />
This camera option is used for the evaluation of the quantity of<br />
centrifuged serum samples. All functions of the “Tube Type<br />
Identification” extensions are supported. In addition, the<br />
following functions and features are delivered:<br />
• Sample volume detection for primary serum tubes<br />
• Upfront sorting rules in the event of insufficient sample<br />
volume<br />
• Use of non conductive tips<br />
QS1-Sample Quantity<br />
Cap/no Cap<br />
Tube Dimension<br />
Volume Calculation<br />
Upload Tube Type<br />
Identification<br />
Liquid Level<br />
Detection
Sample Quantity Detection<br />
A sample image of each processed tube is generated. This<br />
sample image is the basis for the volume calculation.<br />
Note: A minimum of 6.5 mm window for serum analysis must be<br />
available on all tubes for this feature to function.<br />
Sample Quantity Detection<br />
System Connectivity<br />
Host Interface<br />
<strong>The</strong> ASTM host interface of the RSD Pro and RSA Pro allows<br />
fast integration into the laboratory routine. <strong>The</strong> connection can<br />
be either serial or network.<br />
Remote Service and Support<br />
RSD Pro and RSA Pro can be serviced remotely. Configuration<br />
changes, instrument communication traces and software updates<br />
can be carried out via modem access.
Specifications<br />
Throughput: RSD Pro: Up to 1,000 primary tubes per hour<br />
RSA Pro: Up to 405 primary tubes per hour 1<br />
Sample<br />
Identification: Positive sample identification via bar code; able<br />
to handle most of the standard bar code<br />
formats such as Interleaved 2 of 5, Code 39,<br />
Codabar (NW 7), Code 128, others 2<br />
Centrifuge: One or two centrifuges – Hettice Rotanta 46<br />
RSC Robotic<br />
Sample Tube: RSD Pro: Accepts most 3, 5, 7 or 10 ml plastic<br />
tubes with an external tube diameter between<br />
11.5 and 15.5 mm and a total height between<br />
65.5 and 108 mm<br />
COBAS, COBAS INTEGRA, TTA, ELECSYS, LIFE NEEDS ANSWERS,<br />
MODULAR and MYLABONLINE are trademarks of <strong>Roche</strong>.<br />
© 2011 <strong>Roche</strong> <strong>Diagnostics</strong>. All Rights Reserved. 461-50923-0911<br />
<strong>Roche</strong> <strong>Diagnostics</strong><br />
9115 Hague Road<br />
Indianapolis, IN 46256<br />
1-800-981-8863<br />
www.mylabonline.com<br />
RSA Pro: Accepts most 3, 5, 7 or 10 ml plastic<br />
tubes Able to open sample tubes with<br />
hemogard, rubber stopper or screw cap type<br />
closure<br />
Sample Carrier: RD Hitachi MPA racks (5 positions) for input<br />
and output buffer. Other third party rack<br />
possible.<br />
Aliquot Tips: RSA Pro: Use of disposable pipette tips for<br />
secondary tube generation: 1100 µl volume<br />
conductive or 1000 µl volume non conductive<br />
384 tips on board in 4 tip racks<br />
Secondary<br />
Tube: Sarstedt secondary tube 13x75 mm<br />
Sarstedt false bottom 13x75 mm<br />
Greiner Bio-One false bottom 16 mm twin tubes<br />
1. <strong>The</strong> final throughput depends on different parameters as for example the aliquot<br />
volume or the number of aliquots per primary tube<br />
2. Other bar code types can be checked on request<br />
Dimensions: RSD Pro: Width: approx. 6.2 feet<br />
Depth (with touch screen): approx. 5.2 feet<br />
Height (incl. signal lamp): approx. 6.9 feet<br />
Weight: approx. 1918 lbs including centrifuge<br />
RSA Pro: Width: approx. 10.4 feet<br />
Depth (with touch screen): approx. 6.1 feet<br />
Height (incl. signal lamp): approx. 6.5 feet<br />
Weight: approx. 2425 lbs<br />
Operating<br />
Conditions: For indoor use up to 2000 m above sea level<br />
Operating<br />
Temperature: +15 °C to +30 °C (+59 °F to +86 °F)<br />
Air Conditions: 20–80 % relative humidity<br />
Compressed Air: D\Dry and oil free, min 6.0 bar (87.02 psi),<br />
max 8.0 bar (116.03 psi) Consumption approx.<br />
40.0 l (10.6 gals) per min<br />
Consumption approx. 66.0 l (17.4 gals) per min<br />
with recap unit<br />
Power: RSD Pro: 230/115 V, 50–60 Hz, 4950 VA<br />
RSA Pro: 230/115 V, 50–60 Hz, 1800 VA<br />
(with recap unit)<br />
Interface: ASTM protocol<br />
Network connection via TCP/IP or serial<br />
connection via RS 232