Q1 QuantiFast SYBR Green PCR<br />

Real-time PCR for SYBR ® Green detection (Template: DNA or cDNA), using QuantiFast SYBR Green PCR Kit<br />

from QIAGEN ®<br />

� Header Mc <strong>ep</strong> <strong>realplex</strong> Mc <strong>ep</strong> <strong>realplex</strong> S Comment<br />

Lid temperature 105 °C 105 °C<br />

TSP heated lid √ √<br />

Switch off lid at low Block Temp √ √<br />

<strong>Mastercycler</strong> <strong>ep</strong> any <strong>Mastercycler</strong> <strong>ep</strong> S<br />

Control Sim. Tube Sim. Tube<br />

Simulate <strong>Mastercycler</strong> gradient<br />

Impulse √<br />

� Program<br />

1. T = 95 °C 05:00<br />

2. T = 95 °C<br />

3. T = 60 °C<br />

4. Melting Curve<br />

00:10<br />

00:30<br />

35 x*<br />

Perform fluorescence data<br />

collection<br />

Q2 QuantiFast SYBR Green RT-PCR<br />

* Range 35 – 40 Cycles<br />

(d<strong>ep</strong>ends on the amount of<br />

template DNA)<br />

One-st<strong>ep</strong> real-time RT-PCR for SYBR ® Green detection (Template: RNA), using QuantiFast SYBR Green<br />

RT-PCR Kit from QIAGEN ®<br />

� Header Mc <strong>ep</strong> <strong>realplex</strong> Mc <strong>ep</strong> <strong>realplex</strong> S Comment<br />

Lid temperature 105 °C 105 °C<br />

TSP heated lid √ √<br />

Switch off lid at low Block Temp √ √<br />

<strong>Mastercycler</strong> <strong>ep</strong> any <strong>Mastercycler</strong> <strong>ep</strong> S<br />

Control Sim. Tube Sim. Tube<br />

Simulate <strong>Mastercycler</strong> gradient<br />

Impulse √<br />

� Program<br />

1. T = 50 °C 10:00<br />

2. T = 95 °C 05:00<br />

3. T = 95 °C<br />

4. T = 60 °C<br />

5. Melting Curve<br />

00:10<br />

00:30<br />

35 x*<br />

Perform fluorescence data<br />

collection<br />

* Range 35 – 40 Cycles<br />

(d<strong>ep</strong>ends on the amount of<br />

template DNA)


Q3 QuantiFast Probe PCR<br />

Real-time PCR for Singl<strong>ep</strong>lex Probe detection (Template: DNA or cDNA), using QuantiFast Probe PCR + ROX<br />

Vial Kit from QIAGEN ®<br />

� Header Mc <strong>ep</strong> <strong>realplex</strong> Mc <strong>ep</strong> <strong>realplex</strong> S Comment<br />

Lid temperature 105 °C 105 °C<br />

TSP heated lid √ √<br />

Switch off lid at low Block Temp √ √<br />

<strong>Mastercycler</strong> <strong>ep</strong> any <strong>Mastercycler</strong> <strong>ep</strong> S<br />

Control Sim. Tube Sim. Tube<br />

Simulate <strong>Mastercycler</strong> gradient<br />

Impuls √<br />

� Program<br />

1. T = 95 °C 03:00<br />

2. T = 95 °C<br />

3. T = 60 °C<br />

00:03<br />

00:30<br />

40 x*<br />

Perform fluorescence data<br />

collection<br />

* Range 35 – 40 Cycles<br />

(d<strong>ep</strong>ends on the amount of<br />

template DNA)<br />

Q4* QuantiFast Probe RT-PCR<br />

* available from August 2007<br />

Q5 QuantiTect Multiplex PCR Duplex<br />

Real-time PCR for Duplex Probe detection (Template: DNA or cDNA), using QuantiTect ® Multiplex PCR NoROX<br />

Kit from QIAGEN ®<br />

� Header Mc <strong>ep</strong> <strong>realplex</strong> Mc <strong>ep</strong> <strong>realplex</strong> S Comment<br />

Lid temperature 105 °C 105 °C<br />

TSP heated lid √ √<br />

Switch off lid at low Block Temp √ √<br />

<strong>Mastercycler</strong> <strong>ep</strong> any <strong>Mastercycler</strong> <strong>ep</strong> S<br />

Control Sim. Tube Sim. Tube<br />

Simulate <strong>Mastercycler</strong> gradient<br />

Impuls √<br />

� Program<br />

1. T = 95 °C 15:00<br />

2. T = 94 °C<br />

3. T = 60 °C<br />

01:00<br />

01:00<br />

40 x*<br />

Perform fluorescence data<br />

collection<br />

* Range 40 – 50 Cycles<br />

(d<strong>ep</strong>ends on the amount of<br />

template DNA)


Q6 QuantiTect Multiplex PCR Triplex<br />

Real-time PCR for Triplex Probe detection (Template: DNA or cDNA), using QuantiTect ® Multiplex PCR NoROX<br />

Kit from QIAGEN ®<br />

� Header Mc <strong>ep</strong> <strong>realplex</strong> Mc <strong>ep</strong> <strong>realplex</strong> S Comment<br />

Lid temperature 105 °C 105 °C<br />

TSP heated lid √ √<br />

Switch off lid at low Block Temp √ √<br />

<strong>Mastercycler</strong> <strong>ep</strong> any <strong>Mastercycler</strong> <strong>ep</strong> S<br />

Control Sim. Tube Sim. Tube<br />

Simulate <strong>Mastercycler</strong> gradient<br />

Impulse √<br />

� Program<br />

1. T = 95 °C 15:00<br />

2. T = 94 °C<br />

3. T = 60 °C<br />

01:00<br />

01:30<br />

40 x*<br />

Perform fluorescence data<br />

collection<br />

* Range 40 – 50 Cycles<br />

(d<strong>ep</strong>ends on the amount of<br />

template DNA)<br />

Q7 QuantiTect Multiplex RT-PCR Duplex<br />

One-st<strong>ep</strong> real-time RT-PCR for Duplex Probe detection (Template: RNA), using QuantiTect ® Multiplex RT-PCR<br />

NR Kit from QIAGEN ®<br />

� Header Mc <strong>ep</strong> <strong>realplex</strong> Mc <strong>ep</strong> <strong>realplex</strong> S Comment<br />

Lid temperature 105 °C 105 °C<br />

TSP heated lid √ √<br />

Switch off lid at low Block Temp √ √<br />

<strong>Mastercycler</strong> <strong>ep</strong> any <strong>Mastercycler</strong> <strong>ep</strong> S<br />

Control Sim. Tube Sim. Tube<br />

Simulate <strong>Mastercycler</strong> gradient<br />

Impulse √<br />

� Program<br />

1. T = 50 °C 20:00<br />

2. T = 95 °C 15:00<br />

3. T = 94 °C<br />

4. T = 60 °C<br />

00:45<br />

00:45<br />

40 x*<br />

Perform fluorescence data<br />

collection<br />

* Range 40 – 50 Cycles<br />

(d<strong>ep</strong>ends on the amount of<br />

template DNA)


Q8 QuantiTect Multiplex RT-PCR Triplex<br />

One-st<strong>ep</strong> real-time RT-PCR for Triplex Probe detection (Template: RNA), using QuantiTect ® Multiplex RT-PCR<br />

NR Kit from QIAGEN ®<br />

� Header Mc <strong>ep</strong> <strong>realplex</strong> Mc <strong>ep</strong> <strong>realplex</strong> S Comment<br />

Lid temperature 105 °C 105 °C<br />

TSP heated lid √ √<br />

Switch off lid at low Block Temp √ √<br />

<strong>Mastercycler</strong> <strong>ep</strong> any <strong>Mastercycler</strong> <strong>ep</strong> S<br />

Control Sim. Tube Sim. Tube<br />

Simulate <strong>Mastercycler</strong> gradient<br />

Impulse √<br />

� Program<br />

1. T = 50 °C 20:00<br />

2. T = 95 °C 15:00<br />

3. T = 94 °C<br />

4. T = 60 °C<br />

00:45<br />

01:15<br />

40 x*<br />

Perform fluorescence data<br />

collection<br />

* Range 40 – 50 Cycles<br />

(d<strong>ep</strong>ends on the amount of<br />

template DNA)<br />

Practice of the patented polymerase chain reaction (PCR) process requires a license. The <strong>Eppendorf</strong> [or appropriate trademark] Thermal Cycler is an Authorized Thermal Cycler and<br />

may be used with PCR licenses available from Applied Biosystems.<br />

Its use with Authorized Reagents also provides a limited PCR license in accordance with the label rights accompanying such reagents. This is a Licensed Real-Time Thermal Cycler<br />

under Applera's United States Patent No. 6,814,934 and corresponding claims in non-U.S. counterparts thereof, for use in research and for all other applied fields exc<strong>ep</strong>t human in<br />

vitro diagnostics. No right is conveyed expressly, by implication or by estoppel under any other patent claim.<br />

Trademarks: QIAGEN ® , QuantiFast, QuantiTect ® (QIAGEN Group); <strong>Mastercycler</strong> ® (<strong>Eppendorf</strong> AG); SYBR ® (Molecular Probes, Inc.).<br />

Practice of the patented polymerase chain reaction (PCR) process requires a license. The <strong>Mastercycler</strong> <strong>ep</strong> <strong>realplex</strong> Thermal Cycler is an Authorized Thermal Cycler and may be used<br />

with PCR licenses available from Applied Biosystems. Its use with Authorized Reagents also provides a limited PCR license in accordance with the label rights accompanying such<br />

reagents. This is a Licensed Real-Time Thermal Cycler under Applera's United States Patent No. 6,814,934 and corresponding claims in non-U.S. counterparts thereof, for use in<br />

research and for all other applied fields exc<strong>ep</strong>t human in vitro diagnostics. No right is conveyed expressly, by implication or by estoppel under any other patent claim.<br />

Purchase of QuantiFast SYBR Green Kits is accompanied by a limited, non-transferable immunity from suit to use it with detection by a dsDNA-binding dye as described in U.S.<br />

Patents Nos. 5,994,056 and 6,171,785 and corresponding patent claims outside the United States for the purchaser's own internal research. No real-time apparatus or system<br />

patent rights or any other patent rights, and no right to use this product for any other purpose are conveyed expressly, by implication or by estoppel.<br />

QuantiFast Probe Kits and QuantiTect Multiplex Kits are an Authorized 5' Nuclease Core Kit without Licensed Probe. Its purchase price includes a limited, non-transferable immunity<br />

from suit under certain patents owned by Roche Molecular Systems, Inc. or F. Hoffmann-La Roche Ltd, for using only this amount of the product in the practice of the 5' nuclease<br />

process solely for the purchaser's own internal research when used in conjunction with Licensed Probe. No right under any other patent claims (such as apparatus or system claims)<br />

and no right to use this product for any other purpose is hereby granted expressly, by implication or by estoppel. Further information on purchasing licenses may be obtained by<br />

contacting the Director of Licensing, Applied Biosystems, 850 Lincoln Centre Drive, Foster City, California 94404, USA.<br />

Purchase of the QuantiFast SYBR Green RT-PCR Kit, QuantiFast Probe RT-PCR +ROX Vial Kit, and QuantiTect Multiplex RT-PCR NR Kit is accompanied by a limited, nontransferable<br />

license under RT and Reverse Transcription-PCR patents owned by Roche Molecular Systems, Inc. and F. Hoffmann-La Roche Ltd to use it for the purchaser's own<br />

internal research. No real-time patent rights of any kind, no right under any other patent claims (such as apparatus or system claims), and no right to use this product for any other<br />

purpose is hereby granted expressly, by implication or by estoppel.<br />

1043239 01/2007 © 2007 QIAGEN, all rights reserved.

Hooray! Your file is uploaded and ready to be published.