Caution when using the<br />

CW single probe 500mW<br />

The tip of the CW single probe 500mW can<br />

overheat with full power output and longer<br />

treatment times.<br />

As a result, you should not use this probe<br />

without an attachment or applicator.<br />

Otherwise, the probe tip can get too hot<br />

and may result in burns during contact<br />

with the skin.<br />

The CW single probe 500mW should not<br />

be operated at full power output for longer<br />

than 5 minutes. If you want to use it for longer,<br />

you should pause treatment for at least<br />

one minute every 5 minutes.<br />

Caution when using the pulsed<br />

<strong>probes</strong><br />

The maximum recommended frequency<br />

modulation with superpulsed <strong>probes</strong><br />

is 10,000 Hz. For therapeutic purposes we<br />

have enabled frequencies up to 40,000 Hz,<br />

for example, to allow more energy to be<br />

transferred in a shorter period. However,<br />

this results in an accelerated ageing of the<br />

laser diodes.<br />

The laser diodes should therefore be monitored<br />

more closely than usual, or in any<br />

case in the context of the annual safety<br />

check (see also the chapter Care and Maintenance<br />

on page 28)<br />

Caution when using the<br />

infrared <strong>probes</strong><br />

Every now and then you should test the<br />

functioning of your infrared <strong>probes</strong> with the<br />

help of the laser test diode on your control<br />

unit. Read the chapter on the <strong>Laser</strong><br />

Function Test in the user manual for your<br />

Physiolaser.<br />

22