Pressure • Temperature • Flow<br />
europascal <strong>GmbH</strong> DKD – German Calibration Service *)<br />
<strong>Dry</strong>-<strong>block</strong> <strong>calibrators</strong><br />
<strong>ATC</strong>-<strong>140</strong> A/B<br />
• Temperature range: -20 °C to <strong>140</strong> °C<br />
• Improved temperature homogeneity:<br />
0.05 °C (dry-<strong>block</strong>), 0.025 °C (liquid bath)<br />
• Unique dual-zone <strong>block</strong><br />
• Automatical control via PC (RS 232)<br />
• Easy to use<br />
Figure: <strong>ATC</strong>-<strong>140</strong>B<br />
Last update: 16.10.2009 *) pressure measures<br />
Kirchstraße 42 Phone: + 49 (0) 6182 / 78 3000 service@europascal.de<br />
63512 Hainburg Fax: + 49 (0) 6182 / 78 3005 http://www.europascal.de
Pressure • Temperature • Flow<br />
europascal <strong>GmbH</strong> DKD – German Calibration Service *)<br />
Specifications<br />
Temperature range:<br />
Maximum (<strong>Dry</strong>-<strong>block</strong>): <strong>140</strong> °C<br />
Minimum at ambient temperature 0 °C: -35 °C<br />
Minimum at ambient temperature 23 °C: -20 °C<br />
Minimum at ambient temperature 40 °C: -5 °C<br />
Maximum (Liquid bath): <strong>140</strong> °C<br />
Minimum at ambient temperature 0 °C: -33 °C<br />
Minimum at ambient temperature 23 °C: -18 °C<br />
Minimum at ambient temperature 40 °C: -3 °C<br />
Stability: ±0.02 °C<br />
Time to stability (approximate): 15 minutes<br />
Radial homogenity (difference between holes): 0.05 °C (dry-<strong>block</strong>), 0.025 °C (liquid bath)<br />
Accuracy (model B) with external sensor: ±0.04 °C<br />
Accuracy (model A+B) with internal sensor: ±0.2 °C (When used with the dry-<strong>block</strong> kit)<br />
±0.3 °C (When used with the liquid bath kit)<br />
Heating time: -23 to 23 °C 10 minutes<br />
23 to 100 °C 31 minutes<br />
100 to <strong>140</strong> °C 23 minutes<br />
Cooling time: <strong>140</strong> to 100 °C 7 minutes<br />
100 to 23 °C 27 minutes<br />
23 to 0° C 17 minutes<br />
0 to -15° C 35 minutes<br />
Immersion depth: 180 mm (dry-<strong>block</strong>)<br />
150 mm (liquid-bath)<br />
Resolution (user selectable): 1 °C or 0.1°C or 0.01°C<br />
Interface: RS232<br />
Instrument weight: 12.8 Kg<br />
Instrument dimensions (L x W x H): 352 x 156 x 360 mm<br />
Dual-zone heating <strong>block</strong><br />
The specialized <strong>block</strong> design increases the<br />
temperature homogeneity in the critical<br />
calibration zone. It also minimizes the need to<br />
insulate the sensor-under-test and makes it<br />
possible to calibrate liquid-filled and other<br />
mechanical sensors.<br />
The main, or lower, zone ensures optimum heat<br />
dissipation throughout the entire <strong>block</strong>. The<br />
secondary, or upper, zone compensates for the<br />
heat loss from the top of the <strong>block</strong> and from the<br />
sensor-under-test<br />