Digital Temperature Indicator DTI-1000 - Europascal GmbH
Digital Temperature Indicator DTI-1000 - Europascal GmbH
Digital Temperature Indicator DTI-1000 - Europascal GmbH
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
Pressure • <strong>Temperature</strong> • Flow<br />
europascal <strong>GmbH</strong> DKD – German Calibration Service *)<br />
<strong>Digital</strong> <strong>Temperature</strong> <strong>Indicator</strong><br />
<strong>DTI</strong>-<strong>1000</strong><br />
• Wide temperature range<br />
• Improve your accuracy: up to 0.005°C<br />
• „True-Ohm“ measurement<br />
• Wide selection of probes<br />
• Battery- and mains operation<br />
• RS232 communication<br />
Last update: 16.10.2009 *) pressure measures<br />
Kirchstraße 42 Phone: + 49 (0) 6182 / 78 3000 service@europascal.de<br />
63512 Hainburg Fax: + 49 (0) 6182 / 78 3005 http://www.europascal.de
Pressure • <strong>Temperature</strong> • Flow<br />
europascal <strong>GmbH</strong> DKD – German Calibration Service *)<br />
Specifications<br />
Input range:<br />
<strong>DTI</strong>-<strong>1000</strong> A (Pt100 / Pt25): 0 Ω - 360 Ω<br />
<strong>DTI</strong>-<strong>1000</strong> B (Pt25):<br />
Accuracy 12 month use:<br />
0 Ω – 95 Ω<br />
<strong>DTI</strong>-<strong>1000</strong> A: ± (6 ppm rdg + 1.4 mΩ<br />
<strong>DTI</strong>-<strong>1000</strong> B: ± (6 ppm rdg + 0.7 mΩ<br />
<strong>Temperature</strong> range: -200 °C to 750 °C<br />
<strong>Temperature</strong> coeffecient 0 °C – 50 °C: 0.8 ppm/ °C<br />
Input channels: 2<br />
Termination: Goldplated LEMO / 4 mm banana test sockets<br />
Resolution (user-selectable): 0.1 , 0.01 or 0.001 °C/°F/K/Ω<br />
Update rate (0.1 / 0.01 / 0.001°): 2 / 3 / 12 seconds<br />
Measuring current: 1 mA<br />
Dimensions (L x W x H): 225 mm x 135 mm x 195 mm<br />
Weight: 2.2 kg<br />
STS Standard Probe 100A<br />
<strong>Temperature</strong> range:<br />
Accuracy:<br />
-150 °C to 650 °C<br />
Hysteresis: 0.01 °C<br />
Long term stability at 0 °C: 0.014 °C<br />
Repeatability: 0.002 °C<br />
Minimum immersion depth:<br />
Operating conditions:<br />
70 mm<br />
(Probe, connection, and cable): Max. 70 °C<br />
Storage temperature: - 20 °C bis 60 °C<br />
Humidity: 0% bis 90% RH<br />
STS Standard Probe 102A<br />
<strong>Temperature</strong> range:<br />
Accuracy:<br />
-50 °C bis 155 °C<br />
Hysteresis: 0.01 °C<br />
Long term stability at 0 °C: 0.025 °C<br />
Repeatability: 0.002 °C<br />
Minimum immersion depth: 40 mm<br />
Operating conditions:<br />
(Probe, connection, and cable): 0 °C to 50 °C<br />
Humidity: 0% to 90% RH<br />
Last update: 16.10.2009 This document is subject to change without notice *) pressure measures<br />
Kirchstraße 42 Phone: + 49 (0) 6182 / 78 3000 service@europascal.de<br />
63512 Hainburg Fax: + 49 (0) 6182 / 78 3005 http://www.europascal.de