ICP-Brochure - Spiegelberg
ICP-Brochure - Spiegelberg
ICP-Brochure - Spiegelberg
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
8<br />
Intraventricular Probe with Temperature Sensor<br />
In the tip of the probe a<br />
temperature sensor is<br />
located.<br />
The probe is outfitted<br />
with lettered marks.<br />
The Probe 3 XLT measures<br />
intraventricular pressure with an airpouch<br />
mounted in the tip region of a<br />
dual lumen probe. One lumen<br />
transmits the pressure to the Brain-<br />
Pressure Monitor. The second lumen<br />
is used for drainage of CSF. The<br />
measurement of pressure in the<br />
parenchyma is also possible. There is<br />
no interference of drainage and<br />
pressure measurement.<br />
As opposed to measurements via<br />
CSF coupled pressure transducers,<br />
<strong>ICP</strong> is still transmitted in the case of<br />
slit ventricles.<br />
Additionally a temperature sensor is<br />
located in the tip of the probe.<br />
Through an adapter cable the probe<br />
is connected directly to the<br />
temperature input of a patient<br />
monitor.<br />
The slotted suture<br />
clamp is used for<br />
fixation to the skin.<br />
Description REF<br />
Probe 3 XLT SND 13.1.13 XLT<br />
Technical Information<br />
The air tube connects<br />
the probe with the <strong>ICP</strong>-<br />
Monitor.<br />
The stylet is used for<br />
stiffening during<br />
insertion.<br />
Material Polyurethane<br />
Filling volume .05 - .1 ml<br />
Outer diameter 3 mm<br />
Inner diameter of 1.6 mm<br />
drainage lumen<br />
Length of dual lumen 130 mm<br />
part with radiopaque<br />
stripe<br />
Length of single lumen 150 mm<br />
part (drainage)<br />
Length of single lumen 1370 mm<br />
part (air system)<br />
Depth marks 50 - 100 mm<br />
Sensitivity of YSI 400<br />
temperature sensor<br />
Duration of use Short term,<br />
not more than 30 days<br />
Shelf life 3 years<br />
Double packed<br />
EO sterile<br />
For single use<br />
Latex free<br />
Adapter Cables for<br />
Temperature Input<br />
of Patient Monitor<br />
Description REF<br />
The adapter cable<br />
connects the probe to<br />
the patient monitor.<br />
The Luer-Lock connector<br />
connects the probe to<br />
the EVD kit.<br />
Adapter Cable with KBL 13.040.01<br />
6.35 mm Audio Plug<br />
Adapter Cable with KBL 13.041.01<br />
Draeger Connector<br />
Adapter Cable with KBL 13.043.01<br />
Philips/HP Connector