IAP-Monitoring<br />

With the Air-Pouch System<br />

The Air-Pouch System consists of a<br />

hollow body connected to a pressure<br />

transducer by tubing. The pressure<br />

transducer, the electronic hardware,<br />

and the device for filling the Air-Pouch<br />

are integrated in the IAP-Monitor.<br />

For intraabdominal pressure measurement<br />

the Air-Pouch is placed in the<br />

stomach of the patient. The intraabdominal<br />

pressure is transmitted<br />

across the thin pouch wall to the air<br />

volume in the pouch and transformed<br />

into an electric signal by the pressure<br />

transducer.<br />

On the digital display the mean pressure<br />

and the amplitude of the pressure<br />

wave are shown. At the monitor<br />

output the pulsatile signal is available.<br />

The IAP-Monitor can be connected to<br />

all ICU bedside monitors through their<br />

pressure transducer inputs.<br />

A voltage output allows the connection<br />

to chart recorders.<br />

Through the RS 232 interface a computer<br />

can be connected to read out<br />

the pressure signal.<br />

The IAP-Monitor zeroes automatically<br />

once per hour. This automatic in-vivo<br />

zeroing is a unique feature of the Air-<br />

Pouch system.<br />

The catheters themselves do not contain<br />

electric or optic pressure transducers.<br />

That is the reason why they are<br />

very robust and not expensive.<br />

3


Monitors<br />

The Classic Choice with New Features<br />

The ICP-Monitor uses the Air-Pouch<br />

method for measuring intracranial<br />

pressure and intraabdominal pressure.<br />

It is compatible with the full range<br />

of Air-Pouch probes, with the CPP-<br />

Monitor and the Compliance-Monitor.<br />

The digital display indicates mean<br />

ICP or IAP, systolic and diastolic pressure.<br />

Additionally a mains power<br />

control light is visible.<br />

Two connectors in the back of the<br />

monitor take a connecting cable to a<br />

bedside monitor or chart recorder<br />

and a connecting cable to a CPP-Monitor,<br />

a Compliance-Monitor or a<br />

computer.<br />

The stainless steel enclosure is resistant<br />

to wear and tear and to influences<br />

from the environment.<br />

The ICP-Monitor with Batteries<br />

This ICP-Monitor has all the features<br />

of the classic choice. Additionally it is<br />

equipped with recharcheable batteries<br />

that allow more than three hours<br />

of independent operation. It is compatible<br />

with the full range of Air-<br />

Pouch probes, with the CPP-Monitor<br />

and the Compliance-Monitor.<br />

The digital display indicates mean<br />

ICP or IAP, systolic and diastolic pressure.<br />

Additionally a charge control light<br />

and a battery state indicator are<br />

visible.<br />

Two connectors in the back of the<br />

monitor take a connecting cable to a<br />

bedside monitor or chart recorder<br />

and a connecting cable to a CPP-Monitor,<br />

a Compliance-Monitor or a<br />

computer.<br />

The stainless steel enclosure is resistant<br />

to wear and tear and to influences<br />

from the environment.<br />

Description REF<br />

ICP-Monitor 230 V HDM 26.1/FV500<br />

ICP-Monitor 115 V HDM 26.1/FV503<br />

Technical Information<br />

Measurement range -50 to +100 mmHg<br />

Displays Mean ICP<br />

Systolic ICP<br />

Diastolic ICP<br />

Mains control<br />

Voltage output 1 V/50 mmHg<br />

Pressure monitor 5 µV/mmHg/V<br />

output<br />

RS 232 C interface 9600 Bd<br />

Accuracy class 2<br />

Safety class I, type BF<br />

Description REF<br />

ICP-Monitor with Batteries HDM 29.1<br />

(indicate voltage)<br />

Technical Information<br />

Measurement range -50 to +100 mmHg<br />

Displays Mean ICP<br />

Systolic ICP<br />

Diastolic ICP<br />

Battery state<br />

Charge control<br />

Running time More than 3 h<br />

on batteries<br />

Voltage output 1 V/50 mmHg<br />

Pressure monitor 5 µV/mmHg/V<br />

output<br />

RS 232 C interface 9600 Bd<br />

Accuracy class 2<br />

Safety class I, type BF


The IAP-Catheter<br />

In the tip region the<br />

IAP-Catheter is<br />

equipped with the Air-<br />

Pouch and with four<br />

holes.<br />

The IAP-Catheter is introduced like a<br />

naso-gastric tube. It is equipped with<br />

an air pouch at the tip. The catheter<br />

has one lumen that connects the airpouch<br />

with the IAP-Monitor and one<br />

lumen that takes the guide wire for<br />

introduction.<br />

After removal of the guide wire the<br />

second lumen can be used for<br />

ascertaining the position by<br />

auscultation. For this purpose the<br />

second lumen is equipped with holes<br />

in the tip region.<br />

For the second lumen a Luer-Lock<br />

connektor with cap is part of the<br />

package.<br />

For easy introduction the guide wire is<br />

spiralized.<br />

The IAP-Catheter is intended for single<br />

use. It is supplied sterile.<br />

TheLuer-Lock connektor<br />

is intended to connect<br />

to a syringe for<br />

ascertaining the position<br />

by auscultation.<br />

REF<br />

IAP-Catheter SND 32.1.11<br />

Technical Information<br />

Material Polyurethane<br />

Outer Diameter 3 mm<br />

Inner Diameter 1.6 mm<br />

Filling Volume 0.05 - 0.1 ml<br />

Diameter of Guide Wire 0.3 mm<br />

Total Length 2000 mm<br />

Length of Guide Wire 800 mm<br />

5


The CPP-Monitor<br />

The Versatile<br />

Monitoring Abdominal Perfusion<br />

Pressure<br />

Abdominal Perfusion Pressure (APP)<br />

is the difference of Arterial Blood<br />

Pressure (ABP) and Intraabdominal<br />

Pressure (IAP).<br />

The intensive therapy should aim at<br />

keeping the APP above 50 mmHg.<br />

The physiological APP is about<br />

80 mmHg.<br />

... with the CPP-Monitor<br />

The CPP-Monitor is originally intended<br />

for monitoring of Intracranial<br />

Pressure (ICP) and Cerebral Perfusion<br />

Pressure (CPP). Just as well it measures<br />

Arterial Blood Pressure and Intraabdominal<br />

Pressure, calculates the<br />

mean values, and shows them on the<br />

digital display. Abdominal Perfusion<br />

Pressure (APP) is calculated and also<br />

shown on the digital display.<br />

High and low alarm thresholds are<br />

programmable for ABP, IAP, and APP.<br />

With the built-in chart recorder Blood<br />

Pressure, Intraabdominal Pressure,<br />

and Perfusion Pressure can be recorded<br />

continuously. They can be recorded<br />

as momentary values or as<br />

mean values.<br />

Event recording allows automatic recording<br />

whenever one or more of the<br />

alarm thresholds are exceeded. Headers<br />

and trailers are additionally recorded<br />

in order to show the event<br />

completely.<br />

Besides the <strong>Spiegelberg</strong>: IAP-Monitor<br />

all other types of IAP transducers can<br />

be connected. All types of standard<br />

pressure transducers can be used to<br />

pick up Arterial Blood Pressure.<br />

The pressure signals are available at<br />

the outputs for graphic recorder and<br />

patient monitor. Using the RS 232 C<br />

interface the pressures can be fed<br />

into a computer.<br />

6<br />

A memory card can be inserted on<br />

the side of the CPP-Monitor. It can<br />

store 24 hours of signal. The stored<br />

pressures can be replayed on the recorder<br />

at different speeds.<br />

Description REF<br />

CPP-Monitor 230 V CPP 21.1-1/FV501<br />

CPP-Moniotr 115 V CPP 21.1-1/FV504<br />

Technical Information<br />

Pressure transducer inputs 5 µV/mmHg/V, CF Isolated<br />

Voltage input ICP 1 V/200 mmHg<br />

Voltage input ABP 1 V/400 mmHg<br />

RS 232 C interface 9600 Baud<br />

Measuring range ABP 200 mmHg<br />

Measuring range ICP 60 mmHg<br />

Measuring range CPP 200 mmHg<br />

Paper width 100 mm<br />

Recording speed 25 mm/s<br />

1 mm/s<br />

30 cm/h<br />

15 cm/h<br />

5 cm/h<br />

Monitor outputs 5 µV/mmHg/V, CF isolated<br />

Voltage outputs 1 V/50 mmHg<br />

Memory card 1 MB, 48 h signal,<br />

replay at 30 cm/h max.


Connecting Cables and Accessories<br />

Tell us what you want to connect, we supply the cable<br />

Description REF<br />

Interconnect Cables<br />

HDM 13.x, 26.x or HDM 29.x to KBL 21.005.00/FV600<br />

CPP 21.x or CMP 27.x<br />

HDM 26.x or HDM 29.x to CMP 27.x KBL 27.005.01/FV601<br />

and Drainage Valve<br />

Monitor Cables for HDM, CPP, and CMP<br />

Criticare KBL 13.037.01<br />

Datascope KBL 13.026.01<br />

Datex-Cardiocap KBL 13.007.00/FV608<br />

Digicare KBL 13.039.01<br />

Dixtal KBL 13.042.01<br />

Hellige 4./5. Generation KBL 13.003.00/FV609<br />

Hewlett Packard / Philips KBL 13.004.00/FV610<br />

Lohmeier / B.Braun, CombiCable KBL 13.027.01<br />

Marquette KBL 13.005.00/FV612<br />

Mindray KBL 13.038.01<br />

Nihon Kohden 5-pins KBL 13.029.01<br />

Nihon Kohden Lifescope KBL 13.035.01<br />

Ohmeda KBL 13.022.00/FV616<br />

Propaq/Mennen KBL 13.009.00/FV617<br />

Siemens / Draeger, 7-pins, ro<strong>und</strong> KBL 13.028.01<br />

Siemens / Draeger, 10-pins (Sirecust) KBL 13.002.00/FV620<br />

Siemens / Draeger, 16-pins, oval KBL 13.024.01<br />

Space-Labs KBL 13.006.00/FV622<br />

Transducer Cables for CPP-Monitor<br />

Abbott Transpac III, IV KBL 21.007.00/FV623<br />

Baxter Truewave KBL 21.014.01/FV657<br />

Braun-Melsungen KBL 21.002.00/FV624<br />

Gould Spectramed / Ohmeda / BD KBL 21.006.00/FV627<br />

Mallinckrodt / Utah KBL 21.004.00/FV628<br />

Medex LogiCal KBL 21.017.01/FV629<br />

Peter von Berg KBL 21.001.00/FV630<br />

Adapter for CPP-Monitor<br />

<strong>Spiegelberg</strong> ABP/ICP Two<br />

Channel Adapter KBL 21.009.00/FV631<br />

Description REF<br />

Transducer Cables for<br />

Compliance-Monitor ABP Input<br />

Abbott Transpac III, IV KBL 27.007.01/FV635<br />

Baxter Truewave KBL 27.014.01/FV636<br />

Braun-Melsungen KBL 27.002.01/FV637<br />

Gould Spectramed / Ohmeda / BD KBL 27.006.01/FV640<br />

Mallinckrodt / Utah KBL 27.004.01/FV641<br />

Medex LogiCal KBL 27.017.01/FV642<br />

Peter von Berg KBL 27.001.01/FV643<br />

Transducer Cables for<br />

Compliance-Monitor ICP Input<br />

Abbott Transpac III, IV KBL 27.107.01/FV644<br />

Baxter Truewave KBL 27.114.01/FV645<br />

Braun-Melsungen KBL 27.102.01/FV646<br />

Gould Spectramed / Ohmeda / BD KBL 27.106.01/FV649<br />

Mallinckrodt / Utah KBL 27.104.01/FV650<br />

Medex LogiCal KBL 27.117.01/FV651<br />

Peter von Berg KBL 27.101.01/FV652<br />

Cables for the Connection of HDM, CPP, CMP<br />

to Devices with Voltage Input<br />

Banana Plugs 4 mm KBL 13.031.00/FV653<br />

BNC Connector KBL 13.030.00/FV654<br />

Computer Cables RS 232 for HDM, CPP, and CMP<br />

IBM-AT, 9 pins, 1.5 m KBL 13.033.00/FV656<br />

IBM-AT, 9 pins, 25 cm KBL 13.033.01<br />

Accessories<br />

Memory Card 1 MB ZBH 21.403.00<br />

Memory Card 2 MB ZBH 21.402.00/FV603<br />

Pressure Transducer Peter von Berg VBM 21.100.00/FV607<br />

Drainage Valve for CMP 27.x ZBH 27.003.01/FV604<br />

Tube for Drainage Valve ZBH 27.001.01/FV605<br />

Thermal Paper Perforated VBM 21.002.00/FV606<br />

for Compliance-Monitor<br />

and CPP-Monitor, 4 Rolls


Manufacturer<br />

<strong>Spiegelberg</strong> (GmbH & Co.) KG<br />

Tempowerkring 4<br />

21079 Hamburg<br />

Germany<br />

Ph.: +49-40-790-178-0<br />

Fax: +49-40-790-178-10<br />

email: info@spiegelberg.de<br />

http://www.spiegelberg.de<br />

DIN EN ISO 13 485 certified<br />

0297<br />

© by <strong>Spiegelberg</strong> (GmbH & Co.) KG, 2007<br />

Prospekt IAP 2007 Eng V4.p65 2011-05-05

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!