IAP-Monitoring<br />
With the Air-Pouch System<br />
The Air-Pouch System consists of a<br />
hollow body connected to a pressure<br />
transducer by tubing. The pressure<br />
transducer, the electronic hardware,<br />
and the device for filling the Air-Pouch<br />
are integrated in the IAP-Monitor.<br />
For intraabdominal pressure measurement<br />
the Air-Pouch is placed in the<br />
stomach of the patient. The intraabdominal<br />
pressure is transmitted<br />
across the thin pouch wall to the air<br />
volume in the pouch and transformed<br />
into an electric signal by the pressure<br />
transducer.<br />
On the digital display the mean pressure<br />
and the amplitude of the pressure<br />
wave are shown. At the monitor<br />
output the pulsatile signal is available.<br />
The IAP-Monitor can be connected to<br />
all ICU bedside monitors through their<br />
pressure transducer inputs.<br />
A voltage output allows the connection<br />
to chart recorders.<br />
Through the RS 232 interface a computer<br />
can be connected to read out<br />
the pressure signal.<br />
The IAP-Monitor zeroes automatically<br />
once per hour. This automatic in-vivo<br />
zeroing is a unique feature of the Air-<br />
Pouch system.<br />
The catheters themselves do not contain<br />
electric or optic pressure transducers.<br />
That is the reason why they are<br />
very robust and not expensive.<br />
3
Monitors<br />
The Classic Choice with New Features<br />
The ICP-Monitor uses the Air-Pouch<br />
method for measuring intracranial<br />
pressure and intraabdominal pressure.<br />
It is compatible with the full range<br />
of Air-Pouch probes, with the CPP-<br />
Monitor and the Compliance-Monitor.<br />
The digital display indicates mean<br />
ICP or IAP, systolic and diastolic pressure.<br />
Additionally a mains power<br />
control light is visible.<br />
Two connectors in the back of the<br />
monitor take a connecting cable to a<br />
bedside monitor or chart recorder<br />
and a connecting cable to a CPP-Monitor,<br />
a Compliance-Monitor or a<br />
computer.<br />
The stainless steel enclosure is resistant<br />
to wear and tear and to influences<br />
from the environment.<br />
The ICP-Monitor with Batteries<br />
This ICP-Monitor has all the features<br />
of the classic choice. Additionally it is<br />
equipped with recharcheable batteries<br />
that allow more than three hours<br />
of independent operation. It is compatible<br />
with the full range of Air-<br />
Pouch probes, with the CPP-Monitor<br />
and the Compliance-Monitor.<br />
The digital display indicates mean<br />
ICP or IAP, systolic and diastolic pressure.<br />
Additionally a charge control light<br />
and a battery state indicator are<br />
visible.<br />
Two connectors in the back of the<br />
monitor take a connecting cable to a<br />
bedside monitor or chart recorder<br />
and a connecting cable to a CPP-Monitor,<br />
a Compliance-Monitor or a<br />
computer.<br />
The stainless steel enclosure is resistant<br />
to wear and tear and to influences<br />
from the environment.<br />
Description REF<br />
ICP-Monitor 230 V HDM 26.1/FV500<br />
ICP-Monitor 115 V HDM 26.1/FV503<br />
Technical Information<br />
Measurement range -50 to +100 mmHg<br />
Displays Mean ICP<br />
Systolic ICP<br />
Diastolic ICP<br />
Mains control<br />
Voltage output 1 V/50 mmHg<br />
Pressure monitor 5 µV/mmHg/V<br />
output<br />
RS 232 C interface 9600 Bd<br />
Accuracy class 2<br />
Safety class I, type BF<br />
Description REF<br />
ICP-Monitor with Batteries HDM 29.1<br />
(indicate voltage)<br />
Technical Information<br />
Measurement range -50 to +100 mmHg<br />
Displays Mean ICP<br />
Systolic ICP<br />
Diastolic ICP<br />
Battery state<br />
Charge control<br />
Running time More than 3 h<br />
on batteries<br />
Voltage output 1 V/50 mmHg<br />
Pressure monitor 5 µV/mmHg/V<br />
output<br />
RS 232 C interface 9600 Bd<br />
Accuracy class 2<br />
Safety class I, type BF
The IAP-Catheter<br />
In the tip region the<br />
IAP-Catheter is<br />
equipped with the Air-<br />
Pouch and with four<br />
holes.<br />
The IAP-Catheter is introduced like a<br />
naso-gastric tube. It is equipped with<br />
an air pouch at the tip. The catheter<br />
has one lumen that connects the airpouch<br />
with the IAP-Monitor and one<br />
lumen that takes the guide wire for<br />
introduction.<br />
After removal of the guide wire the<br />
second lumen can be used for<br />
ascertaining the position by<br />
auscultation. For this purpose the<br />
second lumen is equipped with holes<br />
in the tip region.<br />
For the second lumen a Luer-Lock<br />
connektor with cap is part of the<br />
package.<br />
For easy introduction the guide wire is<br />
spiralized.<br />
The IAP-Catheter is intended for single<br />
use. It is supplied sterile.<br />
TheLuer-Lock connektor<br />
is intended to connect<br />
to a syringe for<br />
ascertaining the position<br />
by auscultation.<br />
REF<br />
IAP-Catheter SND 32.1.11<br />
Technical Information<br />
Material Polyurethane<br />
Outer Diameter 3 mm<br />
Inner Diameter 1.6 mm<br />
Filling Volume 0.05 - 0.1 ml<br />
Diameter of Guide Wire 0.3 mm<br />
Total Length 2000 mm<br />
Length of Guide Wire 800 mm<br />
5
The CPP-Monitor<br />
The Versatile<br />
Monitoring Abdominal Perfusion<br />
Pressure<br />
Abdominal Perfusion Pressure (APP)<br />
is the difference of Arterial Blood<br />
Pressure (ABP) and Intraabdominal<br />
Pressure (IAP).<br />
The intensive therapy should aim at<br />
keeping the APP above 50 mmHg.<br />
The physiological APP is about<br />
80 mmHg.<br />
... with the CPP-Monitor<br />
The CPP-Monitor is originally intended<br />
for monitoring of Intracranial<br />
Pressure (ICP) and Cerebral Perfusion<br />
Pressure (CPP). Just as well it measures<br />
Arterial Blood Pressure and Intraabdominal<br />
Pressure, calculates the<br />
mean values, and shows them on the<br />
digital display. Abdominal Perfusion<br />
Pressure (APP) is calculated and also<br />
shown on the digital display.<br />
High and low alarm thresholds are<br />
programmable for ABP, IAP, and APP.<br />
With the built-in chart recorder Blood<br />
Pressure, Intraabdominal Pressure,<br />
and Perfusion Pressure can be recorded<br />
continuously. They can be recorded<br />
as momentary values or as<br />
mean values.<br />
Event recording allows automatic recording<br />
whenever one or more of the<br />
alarm thresholds are exceeded. Headers<br />
and trailers are additionally recorded<br />
in order to show the event<br />
completely.<br />
Besides the <strong>Spiegelberg</strong>: IAP-Monitor<br />
all other types of IAP transducers can<br />
be connected. All types of standard<br />
pressure transducers can be used to<br />
pick up Arterial Blood Pressure.<br />
The pressure signals are available at<br />
the outputs for graphic recorder and<br />
patient monitor. Using the RS 232 C<br />
interface the pressures can be fed<br />
into a computer.<br />
6<br />
A memory card can be inserted on<br />
the side of the CPP-Monitor. It can<br />
store 24 hours of signal. The stored<br />
pressures can be replayed on the recorder<br />
at different speeds.<br />
Description REF<br />
CPP-Monitor 230 V CPP 21.1-1/FV501<br />
CPP-Moniotr 115 V CPP 21.1-1/FV504<br />
Technical Information<br />
Pressure transducer inputs 5 µV/mmHg/V, CF Isolated<br />
Voltage input ICP 1 V/200 mmHg<br />
Voltage input ABP 1 V/400 mmHg<br />
RS 232 C interface 9600 Baud<br />
Measuring range ABP 200 mmHg<br />
Measuring range ICP 60 mmHg<br />
Measuring range CPP 200 mmHg<br />
Paper width 100 mm<br />
Recording speed 25 mm/s<br />
1 mm/s<br />
30 cm/h<br />
15 cm/h<br />
5 cm/h<br />
Monitor outputs 5 µV/mmHg/V, CF isolated<br />
Voltage outputs 1 V/50 mmHg<br />
Memory card 1 MB, 48 h signal,<br />
replay at 30 cm/h max.
Connecting Cables and Accessories<br />
Tell us what you want to connect, we supply the cable<br />
Description REF<br />
Interconnect Cables<br />
HDM 13.x, 26.x or HDM 29.x to KBL 21.005.00/FV600<br />
CPP 21.x or CMP 27.x<br />
HDM 26.x or HDM 29.x to CMP 27.x KBL 27.005.01/FV601<br />
and Drainage Valve<br />
Monitor Cables for HDM, CPP, and CMP<br />
Criticare KBL 13.037.01<br />
Datascope KBL 13.026.01<br />
Datex-Cardiocap KBL 13.007.00/FV608<br />
Digicare KBL 13.039.01<br />
Dixtal KBL 13.042.01<br />
Hellige 4./5. Generation KBL 13.003.00/FV609<br />
Hewlett Packard / Philips KBL 13.004.00/FV610<br />
Lohmeier / B.Braun, CombiCable KBL 13.027.01<br />
Marquette KBL 13.005.00/FV612<br />
Mindray KBL 13.038.01<br />
Nihon Kohden 5-pins KBL 13.029.01<br />
Nihon Kohden Lifescope KBL 13.035.01<br />
Ohmeda KBL 13.022.00/FV616<br />
Propaq/Mennen KBL 13.009.00/FV617<br />
Siemens / Draeger, 7-pins, ro<strong>und</strong> KBL 13.028.01<br />
Siemens / Draeger, 10-pins (Sirecust) KBL 13.002.00/FV620<br />
Siemens / Draeger, 16-pins, oval KBL 13.024.01<br />
Space-Labs KBL 13.006.00/FV622<br />
Transducer Cables for CPP-Monitor<br />
Abbott Transpac III, IV KBL 21.007.00/FV623<br />
Baxter Truewave KBL 21.014.01/FV657<br />
Braun-Melsungen KBL 21.002.00/FV624<br />
Gould Spectramed / Ohmeda / BD KBL 21.006.00/FV627<br />
Mallinckrodt / Utah KBL 21.004.00/FV628<br />
Medex LogiCal KBL 21.017.01/FV629<br />
Peter von Berg KBL 21.001.00/FV630<br />
Adapter for CPP-Monitor<br />
<strong>Spiegelberg</strong> ABP/ICP Two<br />
Channel Adapter KBL 21.009.00/FV631<br />
Description REF<br />
Transducer Cables for<br />
Compliance-Monitor ABP Input<br />
Abbott Transpac III, IV KBL 27.007.01/FV635<br />
Baxter Truewave KBL 27.014.01/FV636<br />
Braun-Melsungen KBL 27.002.01/FV637<br />
Gould Spectramed / Ohmeda / BD KBL 27.006.01/FV640<br />
Mallinckrodt / Utah KBL 27.004.01/FV641<br />
Medex LogiCal KBL 27.017.01/FV642<br />
Peter von Berg KBL 27.001.01/FV643<br />
Transducer Cables for<br />
Compliance-Monitor ICP Input<br />
Abbott Transpac III, IV KBL 27.107.01/FV644<br />
Baxter Truewave KBL 27.114.01/FV645<br />
Braun-Melsungen KBL 27.102.01/FV646<br />
Gould Spectramed / Ohmeda / BD KBL 27.106.01/FV649<br />
Mallinckrodt / Utah KBL 27.104.01/FV650<br />
Medex LogiCal KBL 27.117.01/FV651<br />
Peter von Berg KBL 27.101.01/FV652<br />
Cables for the Connection of HDM, CPP, CMP<br />
to Devices with Voltage Input<br />
Banana Plugs 4 mm KBL 13.031.00/FV653<br />
BNC Connector KBL 13.030.00/FV654<br />
Computer Cables RS 232 for HDM, CPP, and CMP<br />
IBM-AT, 9 pins, 1.5 m KBL 13.033.00/FV656<br />
IBM-AT, 9 pins, 25 cm KBL 13.033.01<br />
Accessories<br />
Memory Card 1 MB ZBH 21.403.00<br />
Memory Card 2 MB ZBH 21.402.00/FV603<br />
Pressure Transducer Peter von Berg VBM 21.100.00/FV607<br />
Drainage Valve for CMP 27.x ZBH 27.003.01/FV604<br />
Tube for Drainage Valve ZBH 27.001.01/FV605<br />
Thermal Paper Perforated VBM 21.002.00/FV606<br />
for Compliance-Monitor<br />
and CPP-Monitor, 4 Rolls
Manufacturer<br />
<strong>Spiegelberg</strong> (GmbH & Co.) KG<br />
Tempowerkring 4<br />
21079 Hamburg<br />
Germany<br />
Ph.: +49-40-790-178-0<br />
Fax: +49-40-790-178-10<br />
email: info@spiegelberg.de<br />
http://www.spiegelberg.de<br />
DIN EN ISO 13 485 certified<br />
0297<br />
© by <strong>Spiegelberg</strong> (GmbH & Co.) KG, 2007<br />
Prospekt IAP 2007 Eng V4.p65 2011-05-05