14.01.2013 Views

Analytical Chemistry Chemical Cytometry Quantitates Superoxide

Analytical Chemistry Chemical Cytometry Quantitates Superoxide

Analytical Chemistry Chemical Cytometry Quantitates Superoxide

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

(pH 8.0) with cysteine and cystamine at concentrations of 6 and<br />

1 mM, respectively. Isoform-A was enriched by incubating the<br />

IgG2 under the same conditions but with the addition of 1.0 M<br />

guanidinium chloride (GuHCl) to the buffer. The samples were<br />

protected from light at 2-8 °C for approximately 48 h.<br />

The IgG2 mAb#3 was subjected to Cysf Ser mutagenesis at<br />

position 232 as described by Allen et al. 10 Site-directed mutagenesis<br />

was performed using the QuickChange XL site-directed mutagenesis<br />

kit (Stratagene, La Jolla, CA). The mutant was stably transfected into<br />

a serum-free suspension-adapted CHO cell line. 31 Following production,<br />

the mAb was purified by protein A affinity chromatography.<br />

Incorporation of the expected mutation was verified on the purified<br />

molecule by size exclusion chromatography (SEC) MS. 12<br />

For IMMS analysis, IgG samples were buffer exchanged and<br />

concentrated using Vivaspin 30 kDa molecular weight cutoff filters<br />

(GE Healthcare, Buckinghamshire, UK). For experiments performed<br />

under native conditions, the IgGs were diluted using 160 mM<br />

ammonium acetate (pH not adjusted) to a final working concentration<br />

of ∼3 µM.<br />

Mass spectrometry experiments were performed using a<br />

hybrid ion mobility quadrupole time-of-flight MS (Synapt, Waters<br />

Inc., Milford, MA) equipped with a nanoelectrospray ionization<br />

(ESI) source using metal coated borosilicate glass capillaries<br />

(nanoflow probe tips, long thin walled, Waters Corporation).<br />

Solution flow rates of ∼75 nL/min and an ESI capillary voltage of<br />

∼1.3 kV were used for all experiments. The source temperature<br />

was 50 °C, and the pressure of the vacuum/backing region was<br />

3.5 mbar. Each ion mobility mass spectrum was acquired from<br />

m/z 4000-8000 every 2 s; approximately 12 counts per scan were<br />

observed. The signal was typically averaged for approximately 10<br />

min. Gentle source conditions were used to minimize gas phase<br />

unfolding of the protein (sample cone: 40 V, trap voltage: 10 V,<br />

transfer lens: 12 V, bias: 25; cone gas: 30 L/Hr; P TRAP(Ar): 0.0175<br />

mbar). Nitrogen was used as the mobility carrier gas, and the<br />

following parameters were found to give optimal ion mobility<br />

separation (PIMS: 0.5 mbar, wave velocity: 300 m/s, wave height:<br />

9.8 V). The instrument was mass calibrated using a 50 µg/µL<br />

CsI solution. Waters’ raw data files were translated to Matlab<br />

binary files using software developed in-house. Data processing<br />

and plotting were performed in Matlab (The MathWorks Inc.<br />

Natick, MA) and Igor Pro (WaveMetrics Inc., Lake Oswego, OR).<br />

RESULTS AND DISCUSSION<br />

Antibodies belonging to the IgG2 subclass exist as a group of<br />

distinct isoforms with different disulfide connectivities between<br />

the Fab domain and the hinge region of the molecule. The goal<br />

of this study was to investigate the ability of IMMS to successfully<br />

separate these discrete IgG2 isoforms.<br />

A nano-ESI ion mobility mass spectrum for a solution of 3 µM<br />

mAb#1 (theoretical MW for the most abundant glycoform (G1F/<br />

G1F): 149821.50 Da) in 160 mM ammonium acetate is shown<br />

Figure 2a. A narrow charge state distribution centered on the 24+<br />

ion is observed. Ion mobility spectra for all MAbs analyzed were<br />

acquired with minimal acceleration voltages (sample cone: 40 V,<br />

trap: 10 V, transfer: 12 V). These gentle tuning conditions were<br />

found to provide optimal resolution in the ion mobility dimension<br />

(31) Rasmussen, B.; Davis, R.; Thomas, J.; Reddy, P. Cytotechnology 1998, 28,<br />

31–42.<br />

Figure 2. Ion mobility TOF mass spectra of (a) mAb#1 and (c) mAb#2.<br />

The normalized ion mobility intensities are graphed as contour plots with<br />

9 levels from 4% to 100% intensity. Extracted arrival time distributions<br />

for the 26+ charge states of (b) mAb#1 and (d) mAb#2.<br />

(vide infra). However, with these conditions, the mass spectral<br />

peaks are relatively broad making it impossible to observe the<br />

individual glycoforms of the IgG. The glycoforms for these IgG<br />

molecules can, however, be readily resolved using higher source<br />

voltages (data not shown). Robinson and co-workers have previously<br />

observed that optimal conditions for the mass and mobility<br />

dimensions are often not compatible on the Synapt instrument<br />

when analyzing large protein assemblies. 32<br />

Ion mobility separation of mAb#1 (Figure 2a) reveals two distinct<br />

conformer populations. For example, the mobility spectrum for the<br />

26+ charge state (Figure 2b) shows two distinct peaks with a 1.1<br />

ms difference in drift times (9.3 and 10.4 ms, respectively). Similar<br />

distributions are observed for each charge state. However, with the<br />

tuning conditions employed here, the best mobility resolution was<br />

achieved for the 27+, 26+, and 25+ charge states. To elucidate if<br />

the heterogeneity of the sugar chains on the antibody influences the<br />

observed gas-phase conformers, deglycosylated mAb#1 was analyzed.<br />

The same distribution of gas-phase conformers is revealed in the<br />

ion mobility mass spectrum of mAb#1 with the sugars released<br />

(Figure S1a, Supporting Information). This clearly demonstrates that<br />

these multiple conformers are not caused by sugar heterogeneity<br />

but instead are related to conformational differences in the protein<br />

chains. The arrival time distributions for each charge state virtually<br />

overlaps with that of the glycosylated protein. For example, the major<br />

peaks in the arrival time distribution plot of the 26+ charge state of<br />

deglycosylated mAb#1 are only shifted by -0.13 ms compared to<br />

(32) Ruotolo, B. T.; Benesch, J. L. P.; Sandercock, A. M.; Hyung, S.-J.; Robinson,<br />

C. V. Nat. Protoc. 2008, 3, 1139–1152.<br />

<strong>Analytical</strong> <strong>Chemistry</strong>, Vol. 82, No. 16, August 15, 2010<br />

6753

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!