14.01.2013 Views

Analytical Chemistry Chemical Cytometry Quantitates Superoxide

Analytical Chemistry Chemical Cytometry Quantitates Superoxide

Analytical Chemistry Chemical Cytometry Quantitates Superoxide

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

Figure 3. Electrochemical detection experimental setup.<br />

ments, the microdialysis probe outlet will be connected to the<br />

glucose sample inlet of the device using a simple, low-deadvolume<br />

tubing connection.) For the comparison of system<br />

response to premixed glucose-GOx solutions and direct reaction<br />

of glucose and GOx in the channel, premixed solution was<br />

prepared using a 1:3 volume ratio of 5.6 mM glucose and 150<br />

U/mL GOx in a vial. After adding the two solutions, the vial<br />

was gently shaken 10 times and introduced in the reactor<br />

channel through both inlets approximately 5 min later after<br />

shaking. A two-position actuator switching valve (Valco Instrument<br />

Co., Inc.) was adapted to alternatively introduce buffer<br />

solution and glucose sample solution to measure background<br />

and sensing signals, respectively. Once the flow rate had been<br />

set each time, the valve was switched from buffer solution to<br />

sample solution. The same solution was used for measurements<br />

at each flow rate, and current measurements were made<br />

continuously as flow rate was varied. All experiments were<br />

carried out at room temperature, and data were saved directly<br />

on a computer. A schematic diagram of the electrochemical<br />

detection experimental setup is shown in Figure 3.<br />

3. RESULTS AND DISCUSSION<br />

3.1. Characterization of the Micromixers. A fabricated<br />

micromixer was characterized by capturing images at different<br />

distances from the Y-junction using a fluorescence microscope<br />

and determining the variation in fluorescence intensity across the<br />

width of the channel. The standard deviation (SD) was calculated<br />

using the following eq 3: 30<br />

SD ) � 1<br />

N<br />

N ∑ i)1<br />

(x i - x¯) 2<br />

where xi is the gray-scale intensity value of pixel i, and x¯ is the<br />

mean intensity value of pixels across the entire channel.<br />

Quantitative values of the SD of mixing efficiency varied<br />

between 0.5 for completely unmixed solutions (variation of<br />

intensity from 0 for KI solution to 100% fluorescence for<br />

fluorescein) and 0 for completely mixed solutions (intensity<br />

equal across the channel).<br />

Figure 4 shows the SD of intensity versus distance as a function<br />

of flow rate, comparing channels with slanted grooves and no<br />

(30) Xia, H. M.; Wan, S. Y. M.; Shu, C.; Chew, Y. T. Lab Chip 2005, 5, 748–<br />

755.<br />

6760 <strong>Analytical</strong> <strong>Chemistry</strong>, Vol. 82, No. 16, August 15, 2010<br />

(3)<br />

Figure 4. Comparison of microfluidic mixers having no grooves and<br />

slanted grooves as a function of distance from the Y-junction and at<br />

different flow rates (one device, n ) 3). The flow rate in each case is<br />

the total flow rate in the mixing channel. The flow rates of fluorescein<br />

and KI at the inlets are each half of the total flow rate.<br />

grooves. Flow rates that fall in the range required to achieve high<br />

glucose recoveries in microdialysis sampling of subcutaneous<br />

tissue were considered. The results showed that grooved channels<br />

have a higher mixing efficiency compared to channels without<br />

grooves. In fact, mixing in channels with slanted grooves is almost<br />

complete at a distance of 1 cm along the channel at total flow<br />

rates ranging from 0.4 to 2 µL/min (Reynolds number ) 0.08 at<br />

1 µL/min). This result confirms that grooves formed in the<br />

channel ceiling at an oblique angle with respect to flow direction<br />

enhance the mixing rate of two adjacent solution streams.<br />

Differences in measured standard deviations of fluorescence<br />

intensity at shorter distances along the channel are most likely<br />

due to the reproducibility of the experiment itself. The depth and<br />

width of the grooves were designed based on Stroock’s geometric<br />

parameters (relative groove height to channel height, R < 0.3;<br />

channel height, h E width of the channel, w) and have values of<br />

6-8 and 50 µm, respectively, as shown in Figure 2 (c). 27 These<br />

initial results of mixing characterization were used as important<br />

information for further mixing and reaction experiments.<br />

3.2. Cyclic Voltammograms of Three Different Types of<br />

Microfluidic Channel. Cyclic voltammograms were carried out<br />

to check the electrochemical properties of the microfluidic chip.<br />

A ferri/ferrocyanide couple provides an ideal electrochemical<br />

model for a preliminary chip test with integrated electrodes. The<br />

redox couple ferri/ferrocyanide reaction is as follows (eq 4):<br />

3- - 4-<br />

Fe(CN) 6 + e a Fe(CN)6 E° ) 0.356 V (4)<br />

First, the redox curve was examined with a stationary flow when<br />

1mM(K3Fe(CN)6) was introduced in the channel at a scan<br />

range of -0.3 to 0.7 V. The oxidation and reduction peaks were<br />

observed at 0.21 V (35 nA current) and 0.10 V (56 nA current)<br />

respectively, at a scan rate of 100 mV/s (data not shown). For<br />

cyclic voltammetry under continuous flow conditions, 1 mM<br />

Fe(CN)6 3- /1 mM Fe(CN)6 4- was prepared in 20 mM phosphatebuffered<br />

solution (pH 7.2) containing 0.1 M potassium nitrate<br />

as a supporting electrolyte. Experiments were performed at a

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!