14.01.2013 Views

Analytical Chemistry Chemical Cytometry Quantitates Superoxide

Analytical Chemistry Chemical Cytometry Quantitates Superoxide

Analytical Chemistry Chemical Cytometry Quantitates Superoxide

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

Figure 2. Influence of a surfactant (Tween 20) on microfluidic<br />

transport. The distance indicates that of the meniscus of the liquid<br />

column from the reservoir for the H2O2 solution in a straight flow<br />

channel of 500 µm × 75 µm in cross-section. Concentration of the<br />

surfactant: [, 0wt%;×, 0.005 wt %; 2, 0.01 wt %; b, 0.05 wt %; 9,<br />

0.1 wt %.<br />

channels. In addition, the movement occasionally became irregular,<br />

possibly due to the morphological or chemical nonuniformity<br />

of the channel. In the worst case, the column stopped<br />

midway in its journey and did not reach the lower compartment<br />

of the micropump. This problem was solved by adding a surfactant<br />

to the solution, which facilitated smooth movement. Figure 2<br />

shows the dependence of the movement of the solution on the<br />

concentration of the surfactant (Tween 20) that was added. The<br />

influence of the surfactant was dramatic, and the solution’s<br />

movement became smoother and faster with increasing concentration<br />

of surfactant. At concentrations higher than 0.01 wt %, the<br />

flow velocity almost leveled off. In the following experiments, the<br />

concentration was therefore fixed at 0.01 wt %.<br />

In biochemical analyses in microsystems, the length of time<br />

required for a reaction is often on the order of seconds or minutes.<br />

In view of this, an additional requirement in such cases is the<br />

presence of structures that can slow the flow velocity of solutions.<br />

For this reason, we then examined how the velocity of the column<br />

changed with changes in the width of the flow channel. In Figure<br />

3A, flow channels 2-4 are straight and have widths of 250 µm,<br />

500 µm, and 1.0 mm, respectively. With a change made only in<br />

the width, a marked difference in flow velocity was observed that<br />

demonstrated accelerated movement of liquid plugs in wider flow<br />

channels. The presence of compartments positioned along the flow<br />

channel exerts an additional similar influence. 34,35 Therefore,<br />

rectangular compartments with dimensions of 1.5 mm × 880 µm<br />

and 6.0 mm × 3.5 mm were attached to the 250 µm wide and 1.0<br />

mm wide flow channels (flow channels 1 and 5, respectively). A<br />

portion of the solution, however, also penetrated into the extending<br />

controlling flow channel while the solution filled each compartment.<br />

As a result, the movement of the column was not<br />

significantly different from the case in which there were no<br />

compartments. This result indicated that branched compartments<br />

are not effective for this purpose.<br />

We then tried a sequential arrangement. Figure 3B shows 500<br />

µm wide flow channels. Flow channel 1 is straight, and flow<br />

channels 2 and 3 have compartments of different sizes (2.0 mm ×<br />

2.0 mm and 3.5 mm × 3.5 mm, respectively). By locating the exit<br />

at an appropriate position in the compartment, the transport in<br />

the flow channel was resumed after the compartment was filled<br />

completely, and the effect of the structures was more significant<br />

than that in Figure 3A. There was a tendency for bubbles to<br />

remain in the corners of the square compartments. Although these<br />

bubbles were small and had no adverse effect on the transport of<br />

solutions, circular or elliptic compartments might be better, both<br />

to avoid this potential problem and to realize a more accurate<br />

adjustment of timing.<br />

The movement of the column could also be delayed by the<br />

introduction of constrictions. The width of the flow channels in<br />

Figure 3C is 500 µm. For flow channels 2 and 3, the constrictions<br />

were positioned near the inlet and had widths of 300 and 200 µm,<br />

respectively. The constrictions also had an effect, and the movement<br />

of the column was slowed with narrower constrictions.<br />

Although we used only simple delaying structures because of the<br />

limited space, microfluidic transport can be delayed further<br />

through the use of a more complicated network of flow channels. 22<br />

The movement of a solution in a flow channel with compartments<br />

and/or constrictions can be understood using numerical<br />

Figure 3. Movement of solutions in the flow channels with various delaying structures. (A) Effect of changing the width of the flow channel and<br />

attaching rectangular compartments on the sides. (B) Effect of adding rectangular compartments in series. (C) Effect of creating constrictions.<br />

The images were taken 1, 3, and 6 s (from left to right) after the introduction of the solution from the left. Scale bars correspond to 2 mm.<br />

<strong>Analytical</strong> <strong>Chemistry</strong>, Vol. 82, No. 16, August 15, 2010<br />

6873

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!