14.01.2013 Views

Analytical Chemistry Chemical Cytometry Quantitates Superoxide

Analytical Chemistry Chemical Cytometry Quantitates Superoxide

Analytical Chemistry Chemical Cytometry Quantitates Superoxide

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

Figure 1. Schematic illustration of the setup showing the modification made in the GC-MS instrument. The two possible valve configurations<br />

are shown in the inset: Left, for structural identification by conventional GC-EI-MS analysis, and right, for absolute quantification using GC-<br />

Combustion-MS and postcolumn addition of 13 CO2. Adapted from reference. 15<br />

isotope ratio (12/13) in the tracer and Rn is the natural isotope<br />

ratio (13/12) in the sample, whereas Rb is the measured isotope<br />

ratio in the blend.<br />

Additionally, eq 1 can be used to quantify the mass flow of<br />

enriched carbon dioxide when an internal standard of known<br />

concentration is spiked to the sample.<br />

Head-Space SPME Procedure. Test solutions (∼10 ng/g) were<br />

prepared by diluting appropriately the BTEX standards in Milli<br />

Q water. Samples of ∼7 g were pipetted into 10 mL glass vials<br />

with open-top phenolic closures and PTFE/silicone septa (Supelco,<br />

Bellefonte, USA). Three fiber coatings, 100 µm polydimethylsiloxane<br />

(PDMS), 70/30 µm polydimethylsiloxane/divinylbenzene<br />

(PDMS/DVB), and 50/30 µm carboxen/polydimethylsiloxane/<br />

divinylbenzene (CAR/PDMS/DVB) were tested for comparative<br />

purposes. Samples were saturated in all cases with NaCl to<br />

improve extraction efficiency and stirred at 900 rpm at 30 °C<br />

during 30 min to allow equilibration between the liquid and gas<br />

phases. Then the fibers were inserted in the headspace and<br />

extracted during 15 min to ensure equilibrium conditions between<br />

the headspace and the fiber. After extraction, the fibers were<br />

removed and thermal desorption was carried out in the injection<br />

port of the GC at 250 °C.<br />

RESULTS AND DISCUSSION<br />

Online Continuous Measurement of the 12 C/ 13 C Isotope<br />

Ratios. The optimum conditions for the continuous measurement<br />

of 12 C/ 13 C isotope ratios were established by connecting directly<br />

the reservoir that contained the 13 CO2 spike to the mass<br />

spectrometer. 15 The mass spectrum obtained for the mass<br />

range 43-47 when such postcolumn flow was set at 1 mL/<br />

min is provided in the Supporting Information (Figure SI-1).<br />

The 13 C isotope enrichment in the postcolumn spike (∼97%)<br />

can be clearly seen from the spectrum. Resolution was<br />

optimized while still preserving a good sensitivity. The mass<br />

resolution finally obtained allowed baseline separation even for<br />

isotope ratios very far from 1. The experiments performed for<br />

the selection of the optimum integration time are given in the<br />

Supporting Information (Figure SI-2). An integration time of 70<br />

ms was finally selected considering both the precision in the ratio<br />

measurement and the number of points per chromatographic peak<br />

(∼15) to obtain well-defined peak profiles.<br />

Spike Enrichment and Stability of the 12 CO2/ 13 CO2<br />

Isotope Ratio. Again, to estimate the 13 C enrichment obtained<br />

after the preparation of the 13 CO2 spike, its reservoir was<br />

connected directly to the mass spectrometer. A constant flow<br />

of1mLmin -1 was set and the signals at m/z ratios 44 and 45<br />

were continuously monitored during 2 min for each independent<br />

isotope ratio measurement under the measurement<br />

conditions described in the procedures. The isotope ratio was<br />

calculated obtaining an average value of 0.034 ± 0.0003, which<br />

corresponds to a 13 C abundance of 96.70 ± 0.02% (n ) 5) (see<br />

the mass spectrum shown in the Figure SI-1). The slight<br />

difference between the experimentally obtained enrichment and<br />

that provided for the precursor Na2 13 CO3 by the supplier (99%),<br />

can be ascribed to small contaminations by natural abundance<br />

CO2 from the ambient air during the generation of the tracer<br />

<strong>Analytical</strong> <strong>Chemistry</strong>, Vol. 82, No. 16, August 15, 2010<br />

6865

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!