14.01.2013 Views

Analytical Chemistry Chemical Cytometry Quantitates Superoxide

Analytical Chemistry Chemical Cytometry Quantitates Superoxide

Analytical Chemistry Chemical Cytometry Quantitates Superoxide

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

Table 1. Counting Properties and Detection Limits for a 226 Ra Standard Solution a<br />

window (channels) efficiency (%) background (cpm) b<br />

spectrometric point of view, a smaller window and higher R<br />

energy minimize the possibility of interferences with other<br />

emissions and should be favored when expected sample activity<br />

or desired analytical throughput are not compromised.<br />

DISCUSSION<br />

Although this technique is well-suited to easily determine 226 Ra<br />

in small samples from most environmental matrixes, the system<br />

has been designed and optimized for small sediment samples,<br />

subsequent to 210 Pb determination through 210 Po in equilibrium.<br />

Proposed Method. We recommend operating in batches of<br />

about 12 samples, where both a reference material and a blank<br />

are processed and measured at the same time. The solution from<br />

which the Po source was deposited (about 15 mL) is recovered<br />

and, in order to destroy ascorbic acid and its degradation<br />

compounds, heated under reflux with 5 mL of concentrated HNO3<br />

for half an hour or until a clear solution is obtained. This<br />

solution is then evaporated to incipient dryness, and small<br />

amounts of 0.5 M HCl are added to dissolve the residue. This<br />

operation is carried out two more times to ensure the elimination<br />

of HNO3, and the total volume is then adjusted to 10 mL<br />

in a 20 mL low-diffusion PE counting vial. Then 10 mL of<br />

OptiScint HiSafe cocktail is carefully added by avoiding disturbing<br />

the interface formed by the two immiscible liquids in the<br />

vial (Figure 7).<br />

The mixture is kept for 3 weeks in a dark temperaturecontrolled<br />

area, in order to wait for radioactive equilibrium and<br />

to minimize chemiluminescence, and is counted by liquid scintillation<br />

by using the following conditions:<br />

• Counting time: all samples of the batch are sequentially<br />

counted for 1 h each, during a minimum of 30 cycles.<br />

• Counting windows are approximately set to channels 750-900<br />

and 850-900. However, best results are obtained by summing<br />

all spectra and visually adjusting the windows manually to their<br />

optimum value.<br />

• Counting mode is set to discriminate R and other pulses,<br />

with PSA ) 145.<br />

Counting conditions and calibration are equipment-sensitive,<br />

and therefore, the quenching calibration must be performed for<br />

each instrument. We also recommend carrying out some tests<br />

with 226 Ra tracer solutions to confirm that the chosen PSA value<br />

is correct for each instrument.<br />

The counting information needed to calculate the activity are<br />

sample count rate, background count rate, and quenching parameter<br />

(such as SQP(E) in Quantulus 1220). From the quenching<br />

value the efficiency can be obtained (Figure 5), and the sample<br />

activity finally calculated.<br />

Low-Activity Test. We tested the linearity and the low-activity<br />

response of the detector in the chosen configuration by measuring<br />

a set of 226 Ra unquenched tracer solutions (0.5 M HCl) with<br />

Ld (cpm) MDA (Bq kg -1 ) c<br />

710-860 173 ± 12 0.0070 ± 0.0020 0.0076 0.29<br />

810-860 54 ± 4 0.0024 ± 0.0012 0.26 0.59<br />

a Activity ) 0.01 Bq. Quenched to SQP(E) ) 850. b Counting time of the background was 55 h. c MDA was calculated by assuming a typical<br />

sediment mass of 250 mg.<br />

Figure 7. Proposed procedure to measure 210 Po and 226 Ra in<br />

sediment samples.<br />

Table 2. Count Rates When Measuring 226 Ra<br />

Unquenched Standard Solutions by LSC<br />

sample activity (mBq) net count rate (cpm)<br />

0.017<br />

0.12<br />

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!