14.01.2013 Views

Analytical Chemistry Chemical Cytometry Quantitates Superoxide

Analytical Chemistry Chemical Cytometry Quantitates Superoxide

Analytical Chemistry Chemical Cytometry Quantitates Superoxide

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

contamination of the substances of interest (e.g., trace metals).<br />

When this strategy is used, the time needed to complete γ<br />

spectrometry before further analyses can be performed also<br />

needs to be taken into consideration.<br />

In many laboratories, the total 210 Pb concentration is determined<br />

by R spectrometry, through its daughter radionuclide<br />

210 Po in equilibrium. The technique is simple, rapid, reliable<br />

(recoveries >95%), and shows good precision. 10,11 Furthermore,<br />

although the technique is destructive, only ca. 250 mg of dry<br />

sediment is needed, and because laboratories usually have several<br />

R detectors (they are much cheaper than Ge detectors), sample<br />

throughput is much larger than with γ spectrometry. Typical<br />

counting times range from 1 to 7 days, depending on sample<br />

activity and desired uncertainty. An experimented analyst might<br />

produce a full total 210 Pb profile within a month after sample<br />

receipt (ca. 40 samples). When R spectrometry is used, the<br />

base 210 Pb can only be indirectly determined by assuming that<br />

the 226 Ra concentration is constant along the profile and<br />

therefore estimated as the average of concentrations in the<br />

profile bottom sections, as equilibrium should have been<br />

reached. 12 However, as 226 Ra may vary along the profile, this<br />

may lead to inaccurate 210 Pbex values. It is not uncommon that,<br />

once the total 210 Pb profile is known, γ spectrometry is carried<br />

out in selected sections, thus optimizing the use of this limiting<br />

resource, although because of sample size requirements this<br />

might be impracticable for many laboratories. As γ spectrometry<br />

is carried out on the untreated sample and 210 Po on a<br />

digestate, these techniques sometimes yield results which are<br />

not fully consistent.<br />

Liquid scintillation counting (LSC), mostly used for the<br />

determination of � emitters, 13,14 has also been extensively used<br />

to quantify R emitters, such as 226 Ra, in environmental samples<br />

(mainly waters). 15,16 Some of the reported methods include direct<br />

R LSC after water sample concentration, 17 226 Ra extraction from<br />

water samples with specific scintillation cocktails such as<br />

Radaex, 18 or generation of a radium-barium sulfate coprecipitate<br />

that is transformed into a soluble chloride or nitrate. 19,20<br />

Villa et al. 21 opted for this approach for sediment: the sediment<br />

is digested, and after the elimination of actinides as hydroxides,<br />

radium is recovered as Ra-Ba-SO4, dissolved in EDTA 0.2 M<br />

ammonia solution, and counted. However, most of these<br />

methods cannot be directly used with sediment digestates and/<br />

or are excessively resource-consuming.<br />

(10) Sanchez-Cabeza, J. A.; Masqué, P.; Ani-Ragolta, I. J. Radioanal. Nucl. Chem.<br />

1998, 227, 19–22.<br />

(11) Vesterbacka, P.; Ikaheimonen, T. K. Anal. Chim. Acta 2005, 545, 252–<br />

261.<br />

(12) Binford, M. W. J. Paleolimnol. 1990, 3, 253–268.<br />

(13) Pujol, L.; Sanchez-Cabeza, J. A. J. Radioanal. Nucl. Chem. 1999, 2, 391–<br />

398.<br />

(14) Liong Wee Kwong, L.; LaRosa, J. J.; Lee, S. H.; Povinec, P. P. J. Radioanal.<br />

Nucl. Chem. 2000, 248, 751–755.<br />

(15) Salonen, L. Sci. Total Environ. 1993, 130-131, 23–35.<br />

(16) Salonen, L.; Hukkanen, H. J. Radioanal. Nucl. Chem. 1997, 226, 67–74.<br />

(17) Sanchez-Cabeza, J. A.; Pujol, L. Analyst 1998, 123, 399–403.<br />

(18) Aupiais, J. Anal. Chim. Acta 2005, 532, 199–207.<br />

(19) Repinc, U.; Benedik, L. J. Radioanal. Nucl. Chem. 2002, 254, 181–185.<br />

(20) Galan-Lopez, M.; Martin-Sanchez, A.; Tosheva, Z.; Kies, A. In LSC 2005<br />

Advances in Liquid Scintillation Spectrometry; Chalupnik, S., Schoenhofer,<br />

F., Noakes, J., Eds.; Radiocarbon: Tucson, AZ, 2006; pp 165-170.<br />

(21) Villa, M.; Moreno, H. P.; Manjón, G. Radiat. Meas. 2005, 39, 543–550.<br />

6848 <strong>Analytical</strong> <strong>Chemistry</strong>, Vol. 82, No. 16, August 15, 2010<br />

After 210 Po analysis for R spectrometry, 226 Ra remains in<br />

solution. In this work, we propose, develop, and validate a new<br />

method to determine 226 Ra by 222 Rn emanation to a scintillation<br />

cocktail, which eliminates the need to perform any further<br />

purification, and counting with an ultralow background liquid<br />

scintillation system.<br />

EXPERIMENTAL SECTION<br />

Equipment. An ultralow background liquid scintillation system,<br />

Quantulus 1220 TM (Wallac, Turku, Finland), was used to<br />

carry out this work. In this system, background is reduced by<br />

an optimized combination of active and passive shields. A pulseshape<br />

analysis (PSA) circuit permits the discrimination of<br />

pulses produced by R and � radiation by comparing the area<br />

of the pulse tail after 50 ns from the start with its total area.<br />

Pulse-shape discrimination is accomplished using a software<br />

adjustable parameter (PSA parameter) which can vary between<br />

1 and 256. 16,22 Quenching (sample extinction) was quantified with<br />

the standard quenching parameter (SQP(E)) which is used to<br />

determine the counting efficiency for each sample through<br />

calibration curves. 23,24 Counting was performed with Wallac<br />

OptiScint HiSafe III, a diisopropyl naphthalene based aqueous<br />

immiscible cocktail, and low-diffusion PE counting vials (Packard<br />

BioScience).<br />

Counting Solutions. The tracer solutions were prepared by<br />

gravimetrically spiking 226 Ra/2 M HNO3 (NIST, SRM4967,<br />

U.S.A.) into known amounts of deionized water contained in<br />

20 mL low-diffusion PE counting vials. OptiScint HiSafe was<br />

then added to reach a total admixture volume of 20 mL. These<br />

were stored for 3 weeks in a dark temperature-controlled area<br />

to allow in-growth and equilibrium of the radioactive progenies.<br />

The background solutions, used for calibration purposes, were<br />

prepared with 10 mL of deionized water, acidified to match<br />

the standard solutions, to which 10 mL of the scintillation<br />

cocktail was added. In all cases, quenching was changed by<br />

adding different amounts of CCl 4, ranging from 0 to 200 µL.<br />

All reagents used in the experiments were of analytical grade<br />

(Fisher Scientific).<br />

RESULTS<br />

When counting a 226 Ra aqueous solution with an immiscible<br />

scintillant (such as OptiScint Hisafe), the R emitter 226 Ra decays<br />

to the R emitter 222 Rn (T1/2 ) 3.8332 ± 0.0008 days). Radon is<br />

highly soluble in oil-based scintillators and is selectively<br />

extracted in the cocktail, suffering some decay while this<br />

process takes place. Once solubilized in the organic phase,<br />

222 Rn decays to the R emitter 218 Po (T1/2 ) 3.094 ± 0.006 min),<br />

this mainly decays to the � emitter 214 Pb (T1/2 ) 26.8 ± 9 min),<br />

which decays to the � emitter 214 Bi (T1/2 ) 19.9 ± 0.4 min),<br />

which mainly decays to the R emitter 214 Po (T1/2 ) 162.3 ± 1.2<br />

µs), and this one to 210 Pb (T1/2 ) 22.23 ± 0.12 years). Therefore,<br />

in the scintillant, and after an appropriate equilibration time<br />

(usually set to about 3 weeks), the R emitters 222 Rn, 218 Pb, and<br />

214 Pb are in secular equilibrium (Figure 1). As the probability<br />

of these R decays is in all cases close to one, the maximum<br />

(22) Kaihola, L. J. Radioanal. Nucl. Chem. 2000, 243, 313–317.<br />

(23) Villa, M.; Manjon, G.; Garcia-Leon, M. Nucl. Instrum. Methods Phys. Res.,<br />

Sect. A 2003, 496 (2-3), 413–424.<br />

(24) Sanchez-Cabeza, J. A.; Pujol, L. Health Phys. 1995, 68 (5), 674–82.

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!