Imaging equipment – preparation and use<br />

José Pires Jorge and Regis Lecoultre<br />

Quality control procedures that must be<br />

satisfactorily performed before imaging<br />

After acceptance testing, a QC protocol must<br />

be set up in each department and followed<br />

in accordance with national guidelines. The<br />

following routine quality control test schedule<br />

is typical:<br />

a) Daily energy peaking<br />

b) Daily flood uniformity tests<br />

c) Daily gamma camera sensitivity measurement<br />

d) Weekly linearity and resolution assessment<br />

e) Weekly centre-<strong>of</strong>-rotation calibration<br />

A routine quality control programme for a<br />

SPECT gamma camera includes quality control<br />

procedures appropriate to planar scintillation<br />

cameras (a–d) and specific SPECT quality controls<br />

(e). Further, more complex tests should be<br />

undertaken on a less frequent basis.<br />

Energy peaking<br />

This quality control procedure consists in<br />

“peaking” the gamma camera for relevant<br />

energies prior to obtaining flood images. It<br />

is mandatory that the energy peaking is undertaken<br />

on a daily basis and for each radionuclide<br />

used.<br />

1<br />

Checking the peaking is needed to ascertain<br />

that:<br />

•<br />

•<br />

•<br />

•<br />

The camera automatic peaking circuitry is<br />

working properly<br />

The shape <strong>of</strong> the spectrum is correct<br />

The energy peak appears at the correct<br />

energy<br />

There is no accidental contamination <strong>of</strong> the<br />

gamma camera<br />

It is recommended that the spectra obtained<br />

during peaking tests are recorded.<br />

Daily flood uniformity tests<br />

After a successful peaking test it is recommended<br />

that a uniformity test is performed<br />

on a daily basis. Flood fields are acquired and<br />

evaluation <strong>of</strong> camera uniformity can be made<br />

on a visual assessment. Quantitative parameters<br />

should also be computed regularly and<br />

recorded in order both to demonstrate sudden<br />

variations from normal and to alert the<br />

technologist to progressive deterioration in<br />

the equipment. On cameras that have interchangeable<br />

uniformity correction maps, it<br />

is vital that one is used that is for the correct<br />

nuclide, accurate and up to date.<br />

Daily gamma camera sensitivity<br />

measurement<br />

A practical means <strong>of</strong> measuring sensitivity is