technical specifi cation<br />

measuring principle ISE<br />

measuring principle detail F: ionophore membrane<br />

reference electrode<br />

measuring range application 0.05 ... 1000 mg/l<br />

automatic compensation instrument temperature<br />

potential interfering ions OH- (at >pH8), Al3+, Ca2+, Fe3+,<br />

Si4+<br />

precalibrated ex-works all parameters<br />

integration via con::lyte 1<br />

con::lyte 2<br />

con::lyte 4<br />

con::nect<br />

con::cube<br />

power supply 10 ... 30 VDC<br />

power consumption (typical) 0.72 W<br />

interface connection to s::<strong>can</strong> sys plug, IP68, RS485, 12 VDC<br />

terminals<br />

cable length 7.5 m fi xed cable (-075) or<br />

plug connection (-000)<br />

53<br />

cable type PU jacket<br />

housing material stainless steel 1.4571, POM-C<br />

weight (min.) 2.7 kg<br />

dimensions (diameter x length) 60 x 326 mm<br />

operating temperature 0 ... 60 °C<br />

storage temperature 0 ... 60 °C<br />

installation / mounting submersed or in a fl ow cell<br />

process connection G 1 1/2" outside<br />

pH range 4.5 ... 7.5<br />

protection class IP68<br />

automatic cleaning media: compressed air<br />

permissible pressure: 2 ... 4 bar<br />

air volume: 3 ... 9 l per cleaning<br />

duration: 4 ... 12 sec. per cleaning<br />

cleaning interval: 30 … 120 min.,<br />

depending on application<br />

delay: 10 … 30 sec.<br />

drinking water<br />

typical concentration ranges for this application<br />

F<br />

temperature<br />

part number<br />

[mg/l]<br />

[°C]<br />

fl uor::lyser II min. 0.05 0 E-542-000 / -075<br />

max. 2 60<br />

© s::<strong>can</strong> Messtechnik GmbH<br />

Spectrometer<br />

Probes<br />

i::s<strong>can</strong> Ionselective<br />

Probes<br />

Physical Probes<br />

Terminals Software System<br />

Confi guration<br />

Monitoring<br />

Stations<br />

Spare Parts &<br />

Accessories<br />

Services &<br />

Solutions

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!