technical specifi cation<br />

measuring principle combined 180° absorption and 90°<br />

scattering<br />

resolution turbidity: 0.001 NTU/FTU<br />

UV254: 0.015 Abs/m<br />

accuracy turbidity: submersed: 0.05 NTU/<br />

FTU or +/- 5 %*<br />

in fl ow cell: 0.02 NTU/FTU or<br />

+/- 2.5 %*<br />

UV254: 0.06 Abs/m or +/- 15 %*<br />

(*whichever is greater)<br />

automatic compensation instrument dual-beam and 180° path<br />

precalibrated ex-works all parameters<br />

reference standard distilled water<br />

onboard memory 512 MB<br />

integration via con::cube<br />

con::lyte 1<br />

con::lyte 2<br />

con::lyte 4<br />

con::nect<br />

power supply 10 V - 18 V<br />

power consumption (typical) 60 mA @ 12V<br />

power consumption (max.) 130 mA @ 12V<br />

interface connection to s::<strong>can</strong><br />

terminals<br />

RS485, MODBUS<br />

cable length 7.5 m fi xed cable (-075) or<br />

plug connection (-000)<br />

housing material PEEK, POM-C<br />

weight (min.) approx. 440 g<br />

dimensions (diameter x length) 38.5 x 344,5 mm<br />

operating temperature 0 ... 45 °C<br />

storage temperature -20 ... 60 °C<br />

operating pressure -0,2 ... 6 bar<br />

installation / mounting submersed or in a fl ow cell<br />

<strong>can</strong> be mounted directly in a mains<br />

pipe / pressure pipe<br />

fl ow velocity 3 m/s (max.)<br />

automatic cleaning media: compressed air<br />

permissible pressure: 4 ... 6 bar<br />

cleaning interval: depending on<br />

application<br />

cleaning pressure 6 bar<br />

conformity - EMC EN 61326<br />

protection class (-000) IP67<br />

protection class (-075) IP68<br />

surface water<br />

typical concentration ranges for this application<br />

turbidity<br />

UV254<br />

part number<br />

[NTU/FTU]<br />

[Abs/m]<br />

i_s<strong>can</strong>_UV254 min. 0 0 Y4-3-035-p-000 / -075<br />

max. 800 60<br />

drinking water<br />

typical concentration ranges for this application<br />

turbidity<br />

UV254<br />

part number<br />

[NTU/FTU]<br />

[Abs/m]<br />

i_s<strong>can</strong>_UV2541 min. 0 0 Y4-3-035-p-000 / -075<br />

max. 800 60<br />

41<br />

© s::<strong>can</strong> Messtechnik GmbH<br />

Spectrometer<br />

Probes<br />

i::s<strong>can</strong><br />

i::s<strong>can</strong><br />

Ionselective<br />

Probes<br />

Physical Probes<br />

Terminals Software System<br />

Confi guration<br />

Monitoring<br />

Stations<br />

Spare Parts &<br />

Accessories<br />

Services &<br />

Solutions

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!