3 .4<br />

Follow the instructions for forward<br />

or reverse mode <strong>pipetting</strong><br />

The choice of operating mode can<br />

significantly affect the results of an<br />

analysis.<br />

<strong>Gilson</strong> pipettes are calibrated for forward mode<br />

<strong>pipetting</strong>. Manufacturers should list all pertinent<br />

modes of operation and specify for which mode<br />

the instrument is calibrated.<br />

The forward mode is the usual way of<br />

<strong>pipetting</strong> with an air-displacement pipette<br />

like Pipetman.<br />

See page 26.<br />

The forward mode for positive-displacement<br />

pipettes like Microman eliminates the purge<br />

stroke.<br />

See page 30.<br />

The reverse mode is only possible with<br />

air-displacement pipettes. It is used for<br />

solvents or slightly viscous liquids.<br />

See page 28.<br />

Wiping<br />

If necessary (viscous liquids such as cream), wipe the<br />

outside of the tip or the capillary with a clean medical<br />

wipe. Do not <strong>to</strong>uch the orifice. Choose a tissue which<br />

is resistant, lint-free, and inert <strong>to</strong> acids and solvents.<br />

Dispose of the tissue in a safe, hygienic manner.<br />

Important<br />

When working with high risk<br />

specimens, do not wipe<br />

the disposable part. Make sure fluid<br />

depth penetration does not exceed<br />

3 mm. Then <strong>to</strong>uch-off from<br />

the sidewall of the vessel.<br />

Learning more<br />

Pre-rinsing<br />

Greater uniformity and<br />

precision of dispensing are<br />

usually obtained by providing<br />

identical contact surfaces for<br />

all aliquots. This is achieved<br />

by pre-rinsing with the same<br />

liquid that is being dispensed.<br />

To pre-rinse, aspirate with the<br />

tip and then dispense back<br />

in<strong>to</strong> the original reservoir or<br />

<strong>to</strong> waste.<br />

Q<br />

AWhen should I pre-rinse<br />

the tip?<br />

Pre-rinsing should be<br />

performed<br />

• every time you change a tip<br />

• every time you increase the<br />

volume setting<br />

Don’t forget <strong>to</strong> pre-rinse<br />

• capillaries and pis<strong>to</strong>ns for<br />

Microman<br />

• DistriTips for Distriman<br />

Chapter 3<br />

25