12 Operational Services<br />

European Ultraviolet Calibration Center (EUVC)<br />

Julian Gröbner and Gregor Hülsen<br />

The UV radiometers located at <strong>PMOD</strong>/<strong>WRC</strong> and the<br />

Weissfluhjoch measurement field worked reliably for all of<br />

20<strong>08</strong>, providing continuous UV measurements without significant<br />

interruptions. All radiometers were collected and<br />

re-calibrated during a two-week period in August at<br />

<strong>PMOD</strong>/<strong>WRC</strong>. The stability of the UV Radiometers was satisfactory,<br />

even though changes in some instruments were<br />

observed. This underlines the necessity of recalibrating UV<br />

broadband Radiometers on a yearly basis.<br />

In 20<strong>08</strong> the EUVC served as calibration center for an increasing<br />

number of customer UV radiometers. Furthermore,<br />

quality assurance audits were performed at three<br />

sites in Italy and Switzerland with the QASUME spectroradiometer.<br />

Results of these site audits can be found at the<br />

EUVC web-site:<br />

www.pmodwrc.ch/euvc/euvc.php?topic=qasume_audit.<br />

The 3 rd Regional Brewer Calibration Center-Europe (RBCC-<br />

E) campaign was hosted in 20<strong>08</strong> by the Lichtklimatisches<br />

Observatorium at Arosa, Switzerland. The reference spectroradiometer<br />

QASUME acted as the UV reference for the 7<br />

participating Brewer spectrophotometers whereas the<br />

Brewer #163 was calibrated against the RBCC-E total column<br />

ozone reference.<br />

A tutorial was held from July 7-11 at the EUVC on UV Radiometer<br />

operations. It provided on-site training and expert<br />

advice to station personnel to help them maintain and improve<br />

the required data quality of their measurement sites.<br />

One operator from Spain and two from Chile participated<br />

in this tutorial.<br />

Figure 1. On-Site Quality Assurance audits during the 3rd RBCC-E campaign<br />

held at the LKO at Arosa, Switzerland between the QASUME reference<br />

spectroradiometer and seven Brewer spectrophotometers. From<br />

front to back: Brewer#40, Brewer#72, Brewer#156, Brewer#17,<br />

Brewer#163, QASUME, Brewer#185, Brewer#64.