18.12.2012 Views

Deutsche Tagung f ¨ur Forschung mit ... - SNI-Portal

Deutsche Tagung f ¨ur Forschung mit ... - SNI-Portal

Deutsche Tagung f ¨ur Forschung mit ... - SNI-Portal

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

Methoden und Instrumentierung Poster: Mi., 14:00–16:30 M-P42<br />

Innovative upgrading of the SANS instrument D11 at the ILL<br />

P. Lindner 1 , R. Schweins 1 , K. Lieutenant 1 , R. Gähler 1<br />

1 Institut Laue-Langevin, 6, rue Jules Horowitz, F-38042 Grenoble CEDEX 9, France<br />

Significant flux gains of 40 % - 140 % can be obtained at D11 for the whole range of collimation<br />

distances with a new guide design, recently confirmed by MC simulations [1].<br />

The 39 metres of straight glass guides with constant cross section 30 x 50 mm (width<br />

x height) will be replaced with a guide system with a diverging section, widening the<br />

guide width over the first metres by about 50 % and a focusing section over the last few<br />

metres. The changes are going to be realised in the period 2007-2008 and in conjunction<br />

with the ongoing Millennium Project “New Detector Tank” which concerns the<br />

development of a larger and faster SANS detector for D11 [2]. The new detector will<br />

be installed in a new detector tank with increased diameter and length, thus increasing<br />

the dynamic range and the resolution by 100 % each, allowing for about 6 times higher<br />

count-rates than at present and extending the range of momentum transfer Q both to<br />

higher Q as well as to lower Q, including the option of installing a multi-beam USANS.<br />

[1] K. Lieutenant, P. Lindner, R. Gähler (2006) manuscript in preparation<br />

[2] P. Lindner R. Schweins (2006) ILL Technical Report ILL06 LI 02T

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!