Methoden und Instrumentierung Poster: Mi., 14:00–16:30 M-P30<br />

Triple Axis Spectrometer PUMA at FRM II: Characterisation, current<br />

Status and Perspectives<br />

Klaudia Hradil 1 , Harald Schneider 1 , Jürgen Neuhaus 2 , Götz Eckold 1<br />

1 IPC, Universität Göttingen – 2 ZWE FRM II, TU München<br />

The triple axis spectrometer PUMA with a view on the thermal source of FRM II is<br />

built up within a cooperation of the University of Göttingen and the Physics Department<br />

E13 of the Technical University of Munich. After the commissioning phase and<br />

first “friendly user” experiments, the instrument is now available to the user community<br />

within the proposal system of the FRM II. An overload factor of ∼2 already for<br />

the first proposal rounds reveals the high demand and acceptance of the instrument.<br />

The main characteristics of PUMA can be summarised as follows: high neutron flux<br />

on sample position; flexible and straightforward change of components as collimation,<br />

monochromators/analysers; Eularian cradle for flexible scattering plane adjustment;<br />

equipment for time resolved measurements; diverse and powerful sample environment.<br />

At present, PG(002) and Cu(220) monochromators and PG(002) as analyser, manufactured<br />

at IPC, are provided, covering an energy transfer range up to 80 meV. A multi<br />

analyser/detector option is under construction which allows survey measurements in<br />

(Q, ω) space and provides new features for time-resolved measurements of excitations.<br />

The high quality of the Cu(220) is displayed by the beam profiles at sample position<br />

(figure 1). Phonon measurements for larger energy transfers proved the high flux at<br />

sample position.<br />

Beside the standard sample environment of FRM II a closed-cycle cryostat (3-300K), a<br />

cryo furnace (3-800K), a high temperature furnace (-1300K) for the Eularian cradle and<br />

a high-pressure cell Paris-Edinburgh type VX3 (10 GPa) is available. In the context<br />

of this contribution, characteristics, current status of the instrument and a survey of<br />

the first measurements on PUMA will be presented.<br />

This project has been funded by the German Federal Ministry of Education and<br />

Research under contract no. 03EC6GO1<br />

Fig. 1: CCD picture of the beam<br />

profile, ki = 4.5 ˚A −1 . (left) horizontal/vertical<br />

focussed; (right) flat<br />

Cu(220) monochromator