Problem Likely Cause Solution<br />

Positive control has good<br />

call rates but samples are<br />

lower than expected.<br />

chapter 8 | Troubleshooting 217<br />

Genomic DNA not optimal. Ensure DNA samples are of high quality<br />

(i.e., run in a 1 to 2% gel and compare to<br />

Reference 103 DNA control). Use positive<br />

control sample as a reference guide for<br />

<strong>assay</strong> procedures. Prepare master mixes<br />

for samples and controls.<br />

MDR high, MCR low. Mixed or contaminated genomic<br />

DNA sample (page 197).<br />

Very low call rates Mixed up Nsp arrays and Sty sample,<br />

or vice versa.<br />

Use uncontaminated stock of DNA sample.<br />

Do Nsp and Sty work on separate days.