Table 6.5 (Continued)<br />

Other Problems<br />

Sample needles do not properly<br />

enter vial.<br />

Bent sample needle<br />

User may be pressing the needle<br />

lever down too quickly or with too<br />

much force.<br />

chapter 6 | Fluidics Station Care and Maintenance 177<br />

Problem Possible Cause Solution<br />

Replace sample needle.<br />

Engage sample needle lever<br />

more slowly and/or with less<br />

force.