Table 6.4 (Continued)<br />

Common Error Messages<br />

Error Message Meaning<br />

“Missing Fluid Error”<br />

Examples:<br />

“Stage C” “WashA”<br />

“Sense/Threshold” “960/890”<br />

chapter 6 | Fluidics Station Care and Maintenance 175<br />

“Stage C” “WashA”<br />

“Sense/Threshold” “960/890”<br />

The Pump command completed its step<br />

count before the conductivity sensor<br />

determined that the cartridge contained<br />

a solution with conductivity below the<br />

set threshold value.<br />

The Missing Fluid Error (MFE) Display<br />

not only gives a visual notification of an<br />

error condition to the operator, but gives<br />

the operator information that enables<br />

him/her to determine the cause of the<br />

error. It does this by displaying<br />

information about the sensor value and<br />

the fluid that caused the error. It shows<br />

this internal information in a continuous<br />

loop until the machine is powered down<br />

or a script is started.<br />

For example:<br />

Missing Fluid<br />

Error for 4 seconds<br />

Stage A<br />

valvePos WashA for 4 seconds<br />

Sense/Threshold<br />

820/600 for 4 seconds