174 GeneChip ® <strong>Mapping</strong> <strong>500K</strong> Assay Manual<br />

MEANING OF ERROR MESSAGES<br />

The following lists some of the common error messages and what they<br />

mean (Table 6.4).<br />

Table 6.4<br />

Common Error Messages<br />

Error Message Meaning<br />

“Invalid Command” The script contains a command that can<br />

not be executed because its command<br />

code is either undefined or has a format<br />

error.<br />

“Improper Script” The first command of the script is not the<br />

required FS450 command.<br />

“Temperature Timeout” The Re-attempt command timed out<br />

before the set point temperature was<br />

reached.<br />

“Sensor Timeout” The Await Sensors command timed out<br />

before the anticipated sensor pattern<br />

was seen.<br />

“Valve not Homed” The Home command did not result in the<br />

valve reaching its HOME position.<br />

“Valve Motion Error” The Valve command did not result in the<br />

valve reaching its target valve position.<br />

“Valve out of Position” According to the outer valve encoder,<br />

the valve did not reach a valid position<br />

when it was last rotated.<br />

“Error while Filling” While filling the cartridge, the<br />

AwaitMotor command terminated<br />

because of the step count not the<br />

expected sensor pattern, and that the<br />

same error had occurred several times<br />

consecutively.<br />

“Error while Draining” While draining the cartridge, the<br />

AwaitMotor command terminated<br />

because the step count was not the<br />

expected sensor pattern, and the same<br />

error had occurred several times<br />

consecutively.