PROBLEMS AND SOLUTIONS<br />

Table 6.3<br />

Common error messages, their meanings, probable causes and solutions<br />

chapter 6 | Fluidics Station Care and Maintenance 171<br />

Error Message Problem Possible Cause Solution<br />

Missing Fluid Error Cartridge not filling<br />

completely with sample<br />

solution or buffer during<br />

initial stages of<br />

hybridization wash or<br />

staining protocol.<br />

Cartridge not filling<br />

completely with buffer<br />

during wash script<br />

System detects improper<br />

conditions while filling.<br />

Note where in protocol<br />

error occurred.<br />

Recovered less sample<br />

than initial input during<br />

Recover script.<br />

Possible holes in the<br />

septa of the cartridge.<br />

Sample or staining<br />

solution not in place<br />

properly.<br />

Insufficient volume of<br />

sample or staining<br />

solution (500 µL).<br />

Blocked sampling tube<br />

or line of the fluidics<br />

station.<br />

Failure of one of the<br />

fluidics sensors.<br />

Pump tubing stretched<br />

too tightly around the<br />

pump.<br />

Buffer bottle empty.<br />

Module not primed.<br />

Missing or insufficient<br />

stain or antibody in vial.<br />

Wash empty.<br />

Air bubbles in line.<br />

Leaks.<br />

Loose tubing<br />

attachments inside the<br />

fluidics station.<br />

Run Recover script, and then<br />

use another cartridge.<br />

Run Recover script. Make<br />

sure sample or stain vial is in<br />

the sample holder.<br />

Run Recover script. Add<br />

more sample solution to the<br />

sample vial.<br />

Run Recover script. Run the<br />

Clean or Prime script with<br />

fresh deionized (DI) water to<br />

flush out salt blockage.<br />

Call Affymetrix Technical<br />

Support for service.<br />

Loosen the tubing clamps,<br />

allow tubing to relax, close<br />

the clamps.<br />

Fill buffer bottles.<br />

Prime module.<br />

Identify if chip is filled<br />

If important to recover fluid<br />

in chip, then run Recovery<br />

script, followed by Resume<br />

function<br />

If not important to recover<br />

fluid in chip, run Resume<br />

function<br />

Call Affymetrix Technical<br />

Support for service.