chapter 6 | Fluidics Station Care and Maintenance 165<br />

Peristaltic Tubing Replacement<br />

Periodically the peristaltic tubing requires replacement because of<br />

wear, contamination, or in order to avoid salt buildup. Inspect the<br />

tubing, and if there is evidence of these conditions, follow the<br />

procedure outlined below.<br />

For systems in routine use, Affymetrix recommends monthly<br />

replacement of the tubing. To ensure proper performance, use only<br />

tubing available from Affymetrix. This tubing is manufactured to<br />

the required specifications to ensure proper fluid delivery and array<br />

performance. Additional tubing can be obtained by ordering from<br />

Affymetrix:<br />

Part Number Description Quantity<br />

400110 Tubing, Silicone<br />

Peristaltic, 8.5<br />

Wear gloves when changing tubing. Do not allow fluid from old<br />

tubing to spill onto surfaces.<br />

1. Open the module door (Figure 6.6).<br />