160 GeneChip ® <strong>Mapping</strong> <strong>500K</strong> Assay Manual<br />

5. Open GeneChip ® Operating Software (GCOS), Microarray Suite,<br />

or the current version of the Affymetrix control software.<br />

Click Run → Fluidics... from the menu. Alternatively, click the<br />

down arrow on the Protocol list on the toolbar. The protocol<br />

window appears (Figure 6.3).<br />

Figure 6.3<br />

The Fluidics Station protocol window: select all modules.<br />

6. Choose the current bleach protocol (as of the writing of this<br />

manual, it is BLEACHv2_450) for each of the respective modules<br />

in the Protocol drop-down list. Select all four modules, 1 to 4, and<br />

click Run. The fluidics station will not start the bleach protocol<br />

until the needle lever is pressed down (Figure 6.4).<br />

Temperature will ramp up to 50°C.<br />

7. Follow the prompts on each LCD. Load empty 1.5 mL vials onto<br />

each module if not already done so.<br />

8. Press down on each of the needle levers to start the bleach<br />

protocol (Figure 6.4).