152 GeneChip ® <strong>Mapping</strong> <strong>500K</strong> Assay Manual<br />

Shutting Down the Fluidics Station<br />

1. After removing a probe array from the probe array holder, the LCD<br />

window displays the message ENGAGE WASHBLOCK.<br />

2. For the FS-450, gently lift up the cartridge lever to engage, or<br />

close, the washblock.<br />

The fluidics station automatically performs a Cleanout procedure.<br />

The LCD window indicates the progress of the Cleanout<br />

procedure.<br />

3. When the fluidics station LCD window indicates REMOVE<br />

VIALS, the Cleanout procedure is complete.<br />

4. Remove the sample microcentrifuge vial(s) from the sample<br />

holder(s).<br />

5. If no other processing is to be performed, place wash lines into a<br />

bottle filled with deionized water.<br />

6. Choose Shutdown_450 for all modules from the drop-down<br />

Protocol list in the Fluidics Station dialog box. Click the Run<br />

button for all modules.<br />

The Shutdown protocol is critical to instrument reliability. Refer<br />

to the appropriate Fluidics Station User’s Guide for more<br />

information.<br />

7. After Shutdown protocol is complete, flip the ON/OFF switch of<br />

the fluidics station to the OFF position.<br />

8. Place buffer lines in a different bottle of deionized water than the<br />

one used for the shutdown protocol.<br />

To maintain the cleanliness of the fluidics station and obtain the<br />

highest quality image and data possible, a weekly bleach protocol<br />

is highly recommended.