Probe Array Scan<br />

148 GeneChip ® <strong>Mapping</strong> <strong>500K</strong> Assay Manual<br />

The scanner is also controlled by GCOS Software 1.4. The probe array<br />

is scanned after the wash protocols are complete. Make sure the laser is<br />

warmed up prior to scanning by turning the GeneChip Scanner 3000<br />

7G laser on at least 10 minutes before use. If the probe array was stored<br />

at 4°C, allow to warm to room temperature before scanning.<br />

Refer to the GeneChip ® Operating Software User’s Guide (P/N<br />

701439), online help, and tutorials for more information on<br />

scanning.<br />

The scanner uses a laser and is equipped with a safety interlock<br />

system. Defeating the interlock system may result in exposure to<br />

hazardous laser light.<br />

Read and be familiar with the operation of the scanner before<br />

attempting to scan a probe array. Please refer to the GeneChip ®<br />

Scanner 3000 Quick Reference Card (P/N 08-0075).